Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_I | Pages - 26
1 Mar 2002
Boisrenoult P Bricteux S Beaufils P Hardy P
Full Access

Purpose of the study: We compared in vitro the efficacy of screw-plate fixation versus double screw fixation on a model of type 2 Schatzker fracture of the lateral tibial plateau.

Materials and methods: Ten screw-plate fixations using a lateral prebent plate and 10 double-screw fixations (6.5 mm screws) were made on 10 pairs of non-embalmed cadaver knees after simulation of type 2 Schatzker fractures. The strength of each fixation was tested with a compression device. Criteria indicating failure were displacements greater than 2 mm of one or more fracture lines. The force applied at rupture and the stiffness of each type of fixation were compared. Wilcoxon’s test was used for statistical analysis.

Results: Force at rupture and stiffness of the fixation were similar for the two types of fixation. There was no statistical difference (p > 0.05) between the screw-plate and the double-screw fixations.

Discussion: Our findings on a model of type 2 Schatzker fractures are in agreement with previous data obtained by other authors working on models of type 1 Schatsker fractures. The biomechanical stability of the double-screw fixation is as good as that obtained with screw-plate fixation for the treatment of fractures of the lateral tibial plateau.