Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 172 - 172
1 Mar 2008
Yagihashi K Nishimura I Ishida T Ito H Tanino H Nakamura T Matsuno T Mitamura Y
Full Access

Prosthetic impingement after THA is to different for the angle and shape of the implant. Purpose of this study is examine the range of motion(ROM) on a computer when angle and shape of the implant are changed.

The 3D implant models were created on a computer. The angle was measured in the flexion, extension, adduction direction byevery 0.1 degrees. There are three kinds of acetabular abduction angle, two kinds of acetabular anteversion angle and two kinds of femoral anteversion angle. There are three kinds of the radius of neck and the neck shaft angle. All 324 patterns of the above model were measured.

When the radius of neck decreased, the ROM increased in all cases. When the neck shaft angle decreased, the ROM increased by almost all cases. When the acetabular anteversion angle increased, the ROM of flexion direction increased and adduction direction decreased, and as for the extension direction, all the factors had influenced the change in the ROM. When the acetabular angle increased, the ROM of the extension direction increased and the flexion directions decreased. As for adduction direction, femoral anteversion angle, acetabular anteversion angles, and the radius of neck had influenced the ROM. When the femoral anteversion angle increased, the ROM of flexion direction increased and extension, adduction direction decreased.

The clinical ROM is affected by the impingement of non-implant and the strain of the soft tissue. Therefore, It’ s considered that the clinical ROM is smaller than the ROM which was investigated in this study in many cases. When the radius of neck and the neck shaft angle decrease, the increase of the ROM expected. However the radius of the neck should not be decreased too much to avoid the decrease of the neck strength.