Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 424 - 424
1 Sep 2009
Jeffs R Weston-Simons JS Twyman R
Full Access

Study Purposes: A retrospective study to assess patient’s height as a predictor for sizing the femoral component of the Phase 3 Oxford Unicompartmental Knee replacement.

Methods and Results: 118 patients post oxford uni-compartmental knee replacement were identified from the elective orthopaedic unit. Their post-operative radiographs were reviewed by 3 independent orthopaedic surgeons according to the Oxford Group criteria, measuring the femoral component fit. Patients were allocated to two cohorts: patients with ill-fitting prostheses and well fitting prosthesis.

Patient’s height and size of femoral component used at time of surgery was recorded. The results showed a positive correlation between the patient’s height and a well fitting femoral prosthesis.

Conclusions: Current templates with the Oxford Phase Three system may not provide accurate guide to the correct femoral prosthesis size for a patient. Recent studies have highlighted the wide range of magnifications used in computerised X-ray systems leading to variance with templating. Other studies have shown large inter-observer variability when sizing the femoral component. A variable as simple as patient height may offer a more accurate guide for choosing the size of the femoral component.

We have demonstrated that the height of a patient can be a used to guide and assist in the sizing of the femoral component of the Oxford uni-compartmental knee replacement. This study may also have implications for the sizing of other prostheses currently in use.