Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Wear of high-density polyethylene on bone and cartilage



Download PDF

Abstract

Wear of high-density polyethylene on bone and cartilage has resulted in a large volume of plastic particles being shed into the two knees and two hips studied. The giant-cell foreign-body reaction of the synovium may not be sufficient to cope with the amount of debris presented and the destruction of the endosteal bone in one hip, caused by the wear particles and movement of the prosthesis, has made revision impossible. Articulation of high-density polyethylene against bone or cartilage either by design or by the failure of alignment of the component must be avoided.

For access options please click here