Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

THE EFFECT OF A HYDROXYAPATITE-COATED, GROOVED FEMORAL COLLAR ON RADIOGRAPHIC LOOSENING OF DISTAL FEMORAL ENDOPROSTHESES



Abstract

Purpose: The purpose of this study was to determine whether the low rate of mechanical loosening of the SMILES rotating hinge distal femoral endoprosthesis relates to the hydroxyapatite (HA)-coated, grooved collar of the femoral component.

Methods and results: A database was used to identify two groups of cases of primary distal femoral replacement with a custom-designed and manufactured SMILES endoprosthesis at our unit; those with the collared femoral component (“collar group”), and those without a collar (“non-collar group”). From these two groups, patients were pair-matched for age and length of bone resection. We performed a retrospective review of serial biplanar pairs of radiographs of each patient, assessing radiolucent lines and extracortical bone pedicle. 11 matched pairs were identified, (14 females, 8 males), with a mean age of 36 years, (range 16–66). The pathology was primary bone tumour in 20 cases, (17 malignant, 3 benign), and metastatic disease in 2 cases. Mean length of follow-up was 85 months, (range 27–122). Radiolucent line score (RLS) progression over time was significantly lower in the collar group, (0.01 vs 0.73, p=0.001) (fig. 1 & 2), as was the mean final RLS, (2.72 vs 7.81, p=0.02). Mean RLS per radiographic quadrant was 0.56 in cases in which a bony pedicle was ingrown onto the prosthesis, (exclusively in the collared-group), 2.41 in cases in which the pedicle was not ingrown, (most prevalent in the non-collared group), and 1.02 in those cases without any pedicle formation, (ANOVA analysis, p=0.0002).

Conclusion: This study demonstrates that the HA-coated, grooved collar significantly reduces the progression of radiolucent lines, and consequently the overall RLS, explaining the reduced rate of mechanical loosening of the SMILES prosthesis. A bony pedicle that does not incorporate onto the prosthesis surface may be associated with an increase in radiographic loosening.

The abstracts were prepared by Mr Roger Tillman. Correspondence should be addressed to BOOS at the Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PN