Introduction: To enable navigated-assisted orthopedic surgery, a reference frame must be rigidly fixed to a stable bony structure. This may create technical obstacles and wound complications. Instead, we propose to attach the reference frame to the fracture table.
Methods: The study population consisted of 10 patients who underwent fixation of subcapital femoral neck fracture with three cannulated screws, using fluoroscopy-based navigation. Step 1 – the patient was positioned on a fracture table and the reference frame was attached to the iliac crest. Three guide wires were inserted under fluoroscopy-based navigation. 2 – New fluoroscopic images were acquired. 3 – Navigated drill guide placed over each guide wire to record final navigated drill guide position – these images include actual guide wire positions and the trajectories of the navigated drill guide. Navigation accuracy was validated, measuring translational and angular deviations of the virtual trajectory from the implant on the same fluoroscopic image in anteroposterior and lateral views. 4 – The reference frame was removed from the iliac crest and attached to the fracture table. Step 3 was then repeated.
Results: The translational deviation of the virtual trajectory from the inserted guide wire when the reference frame was attached to the iliac crest was not statistically significant from the deviation when it was attached to the fracture table. Angular differences were also not statistically significant.
Conclusions: In our experience, attaching the reference frame to the fracture table instead of to the iliac crest allows for similar accuracy of the navigation process with the possible benefit of reducing patient morbidity.
The abstracts were prepared by Orah Naor, IOA Co-ordinator and Secretary. Correspondence should be addressed to Israel Orthopaedic Association, PO Box 7845, Haifa 31074, Israel.