DENSITY CALIBRATION IN COMPUTED RADIOGRAPHY



Abstract

An uncomplicated, quantitative method of determining density from X-rays would be of extreme value to clinicians. In this study we perform a thorough assessment of applying a step wedge to grey level calibration method to X-rays obtained using Computed Radiography (CR).

An Aluminium step wedge of ten, 5mm-thick steps was X-rayed with a Fuji CR system together with a knee phantom (3M) at various energy and Fuji processing settings. Automatic detection of the steps by means of the Hough transform was used to assess optimum CR settings. Background variation due to the anode Heel effect was evaluated by acquiring an “empty field” X-ray at different energy settings and with copper filtering. The effects of beam hardening were considered with a custom-made phantom which was also used to assess correcting for soft tissue and bone thickness.

X-rays taken at higher energy settings and with wider windowing imaged the widest number of steps (nine) and gave the best accuracy in modelling the step thickness to grey level relationship. Fitting a straight line to the log of the net grey levels gives an excellent model of the data (R2 = 0.99). X-rays of copper sheeting show that automatic histogram analysis is performed by the Fuji CR system, which can have unpredictable effects on aluminium thickness to grey level relationship. Background variation in the anode-cathode direction due to the Heel effect was corrected with a 1D exponential model (R2 = 0.99), allowing position-independent measurements to be obtained. Correcting for bone thickness, soft tissue and beam hardening further improves measurement quality.

Use of step wedge calibration to provide quantitative information on plain X-rays without altering their clinical quality is possible using digital radiography. However, a thorough assessment of the entire X-ray process is necessary to achieve accurate and comparable information.

Correspondence should be addressed to EORS Secretariat Mag. Gerlinde M. Jahn, c/o Vienna Medical Academy, Alserstrasse 4, 1090 Vienna, Austria. Fax: +43-1-4078274. Email: eors@medacad.org