Abstract
While Oxidized Zirconium (OxZr) femoral heads matched with highly cross-linked polyethylene (XLPE) have demonstrated the lowest rate of revision compared to other bearing couples in the Australian National Joint Registry, it has been postulated that these results may, in part, be due to the fact that a single company offers this bearing option with a limited combination of femoral and acetabular prostheses. The purpose of this study was to assess clinical and radiographic outcomes in a matched cohort of total hip replacements (THR) utilizing an identical cementless femoral stem and acetabular component with either an Oxidized Zirconium (OxZr) or Cobalt-Chrome (CoCr) femoral heads at a minimum of 10 years follow-up.
We reviewed our institutional database to identify all patients whom underwent a THR with a single cementless femoral stem, acetabular component, XLPE liner and OxZr femoral head with a minimum of 10 years of follow-up. These were then matched to patients who underwent a THR with identical prosthesis combinations with CoCr femoral head by gender, age and BMI. All patients were prospectively evaluated with WOMAC, SF-12 and Harris Hip Score (HHS) preoperatively and postoperatively at 6 weeks, 3 months, 1 and 2 years and every 2 years thereafter. Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years for both cohorts. Paired analysis was performed to determine if differences exist in patient reported outcomes.
There were 208 OxZr THRs identified which were matched with 208 CoCr THRs. There was no difference in average age (OxZr, 54.58 years, CoCr, 54.75 years), gender (OxZr 47.6% female, CoCr 47.6% female), and average body max index (OxZr, 31.36 kg/m2, CoCr, 31.12 kg/m2) between the two cohorts. There were no significant differences preoperatively in any of the outcome scores between the two groups (WOMAC (p=0.449), SF-12 (p=0.379), HHS(p=0.3718)). Both the SF12 (p=0.446) and the WOMAC (p=0.278) were similar between the two groups, however the OxZr THR cohort had slightly better HHS compared to the CoCr THR cohort (92.6 vs. 89.7, p=0.039). With revision for any reason as the end point, there was no significant difference in 10 years survivorship between groups (OxZr 98.5%, CoCr 96.6%, p=0.08). Similarly, aseptic revisions demonstrated comparable survivorship rates at 10 year between the OxZr (99.5%) and CoCr groups (97.6%)(p=0.15).
Both THR cohorts demonstrated outstanding survivorship and improvement in patient reported outcomes. The only difference was a slightly better HHS score for the OxZr cohort which may represent selection bias, where OxZr implants were perhaps implanted in more active patients. Implant survivorship was excellent and not dissimilar for both the OxZr and CoCr groups at 10 years. Therefore, with respect to implant longevity at the end of the first decade, there appears to be no clear advantage of OxZr heads compared to CoCr heads when paired with XLPE for patients with similar demographics. Further follow-up into the second and third decade may be required to demonstrate if a difference does exist.