Abstract
Introduction and Objective
In the elderly population, chronic rotator cuff tears are often associated with high re-rupture rates after surgical tendon refixation. Implant materials, especially in combination with additives are supposed to positively influence healing outcome. Furthermore, adequate mechanical properties are crucial. In order to realize degradable implants with high specific surface area, polycaprolactone (PCL) was chosen as basic material and processed by electrospinning to achieve a high surface area for growth factor implementation and subsequent cell attachment.
Materials and Methods
PCL (Mn approx. 80,000 g/mol) was used to generate fibre mats by electrospinning (relative collector velocity 8 m/s; flow rate of 4 ml/h). Mechanical analysis was performed according to EN ISO 527–2:2012 with test specimen 1BA (5 mm in diameter). Maximum force at failure (Fmax) as well as stiffness were evaluated. For preclinical in vivo testing, a coating with CS-g-PCL was performed to increase cellular adhesion and biological integration. Native and TGF-ß3 loaded mats were examined in a chronic rat tendon defect model with dissection of the M. infraspinatus, four week latency and following refixation at the humerus with different PCL-fibre mats (approval Nr. 33.12–42502–04–15/2015). After 8 weeks, rats were finalized and tendon-bone insertions were analyzed biomechanically and via histological methods.
Results
Electrospun PCL-fibre mats (n = 6) showed maximum forces of 2.19 ± 0.8 N and a stiffness of 0.38 ± 0.12 N/mm. Native rat infraspinatus tendons showed Fmax values of 28.4 ± 7.2 N and a stiffness of 11.8 ± 4.9 N/mm. After implantation, Fmax of the implant-tendon-regenerate was significantly lower in CS-g-PCL - fibre mat groups compared to native control tendons (mean 52 % of native tendon value). Functionalization with TGF-ß3 led to increased Fmax (78 % of the native tendon value). However, differences were not statistically significant. Histological evaluation revealed no differences between non loaded and TGF-ß3 loaded mats. The implants were strongly disintegrated. Granulation tissue and a high number of foreign body giant cells were present.
Conclusions
Although mechanical properties of fabricated mats were low, loading of the fibre mats influenced the biomechanical outcome of refixed tendons, presumably due to their high potential for binding biological active substances like TGF-ß3. However, in ongoing studies these cell reactions, especially regarding polarization of macrophages and foreign body cells need to be characterized. This research project has been supported by the German Research Foundation “Graded Implants FOR 2180 – tendon- and bone junctions” WE 4262/6-2 and parts were published in J Tissue Eng Regen Med. 2020 Jan;14(1):186–197. doi: 10.1002/term.2985