Abstract
Recent case reports have described V40 taper failure with clinically relevant adverse reaction to metal debris (ARMD). The real incidence of V40 taper damage and potential consequences are currently unclear, however. Aim of this study is therefore, to evaluate the long-term incidence of pseudotumors in a consecutive series of THA with V40 taper and identify potential influencing factors.
From 2006 to 2007 a total number of 120 patients (127 hips) received either an uncemented (Accolade©) or cemented hip stem (ABGII©), both with V40 taper (Stryker© Mahwah, New Jersey, USA). They all were combined with 36 mm Vitallium (CoCrMo) heads and uncemented cups (Trident©) with XLPE inlays. 11,2 +/− 0,5 years post-op 82 patients with 87 hips (mean age 74 years, 58 % female) underwent clinical (PROMs) and radiographic evaluation. In 71 patients (75 hips) MARS- MRI of the hip was performed. 38 patients were lost to follow-up. In 81 patients (86 hips) chrome and cobalt levels were determined.
MRI-investigation revealed 20 pseudotumors (26%) and 18 of them had a diameter of >2cm. Patients with pseudotumors had significant higher median cobalt ion levels compared to those without (2,85 μg/l vs. 1,32 μg/l; p=0,022) and a significant correlation between pseudotumors and cobalt levels was found. Radiographic osteolysis was associated with pseudotumors as well (p= 0,014). Neither approach, BMI, gender, age, type of stem, head length, inclination nor heterotope ossification showed a significant correlation to pseudotumor occurrence.
Due to the high incidence of local ARMD in in asymptomatic patients with V40 taper and metal heads we recommend regular post-operative follow-up investigations including routine metal ion screening and consecutive MRI investigation upon elevation.