Abstract
Abstract
Objective
To estimate the effect of calcar collar separation on the likelihood of calcar collar contact during in vitro periprosthetic fracture.
Methods
Three groups of six composite femurs were implanted with a collared cementless femoral stem. Neck resection was increased between groups (group 1 = normal, group 2 = 3mm additional, group 3 = 6mm additional), to simulate failure to obtain calcar collar contact. Prior to each trial, the distances between anterior (ACC) and posterior (PCC) collar and the calcar were measured. Periprosthetic fractures of the femur were simulated using a previously published technique. High speed video recording identified when collar to calcar contact (CCC) occurred. The ACC and PCC were compared between trials where the CCC was and was not achieved. Regression estimated the odds of failing to achieve CCC for a given ACC or PCC.
Results
CCC was achieved prior to fracture in all cases in group one, 50% in group two and 0% in group three. The median (range) ACC for those trials where CCC was achieved was 0.40 (0.00, 3.37) mm versus 6.15 (3.06 to 6.88) mm, where CCC was not achieved (p <0.01). The median (range) PCC for those trials where CCC was achieved was 0.85 (0.00 to 3.71) mm versus 5.97 (2.23 to 7.46) mm, where CCC was not achieved (p <0.01). Binomial logistic regression estimated risk of failure to obtain CCC increased 3.8 fold (95% confidence interval 1.6 to 30.2, p <0.05) for each millimetre of PCC.
Conclusions
Increased separation between collar and calcar reduced the likelihood of calcar collar contact during a simulated periprosthetic fracture of the femur. Surgeons should aim to achieve a calcar-collar distance of less than 1mm following implantation to ensure calcar collar contact during periprosthetic femoral fracture and to reduce the risk of fracture.
Declaration of Interest
(a) fully declare any financial or other potential conflict of interest