Aim
The current antibiotic treatment of periprosthetic joint infection (PJI) is optimized by measuring concentrations in plasma. However, it remains unclear whether effective concentrations of the antibiotics are reached at the site of PJI. Nonetheless, adequate target site concentrations are important to achieve effective eradication of the micro-organism. In order to determine the efficacy of cefuroxime and flucloxacillin in synovial fluid, synovial tissue and bone tissue in relation to the minimal inhibitory concentration (MIC) of the pathogen causing the PJI, we perform a pharmacokinetic/pharmacodynamic (PK/PD) study. Therefore, we aimed to develop validated analytical methods for analysis of cefuroxime and flucloxacillin in synovial fluid, synovial tissue and bone tissue.
Method
Blank samples of synovial fluid, synovial tissue and bone tissue were obtained by orthopedic surgeons during surgery. For validation the samples of each matrix were spiked with both cefuroxime and flucloxacillin. Synovial tissue and bone tissue was pulverized with a mikro-dismembrator. Samples were kept frozen at −20°C until analysis. After a sample preparation quantification of cefuroxime and flucloxacillin in each matrix was performed on the ultra-performance convergence chromatography-tandem mass spectrometry (UPC2-MS/MS). Stable-isotope-labeled meropenem-d6 served as internal standard. The linearity, limits of quantification, accuracy and precision and carry-over were determined for all methods separately. The methods were validated according to the European Medicine Agency (EMA) and Food and Drug Administration (FDA) guidelines on bioanalytical method validation.
Results
These methods were successfully validated for cefuroxime and flucloxacillin quantification in all matrices according to the EMA and FDA guidelines. The limits of quantification were adequate to cover potential cefuroxime and flucloxacillin concentration in synovial fluid, synovial tissue and bone tissue as described in literature, with a range of 1–100mg/L for synovial fluid and 1–20 µg/g for synovial tissue and bone tissue (r >0.995). Accuracy and within-run precision were validated according to acceptance values (RSD <15%). Carry over was less than 20%. Matrix effects and recovery were investigated for synovial fluid. The results were within the range of 80–120%
Conclusions
The results of the validation fall within the limits of quantification according to the EMA and FDA guidelines. Therefore, these methods can be applied during a PK/PD study to discover the exposure of antibiotics in synovial fluid, synovial tissue and bone tissue at the site of infection in patients with a PJI.