Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Foot & Ankle

RISK OF SAPHENOUS NERVE INJURY DURING SYNDESMOTIC STABILIZATION WITH THE TIGHTROPE TECHNIQUE

The British Orthopaedic Foot & Ankle Society (BOFAS) Annual Congress 2019, Nottingham, England, 13–15 November 2019.



Abstract

Background

The use of a knotless TightRope for the stabilisation of a syndesmotic injury is a well-recognised mode of fixation. It has been described that the device can be inserted using a “closed” technique. This presents a risk of saphenous nerve entrapment and post-operative pain.

Aim

We aimed to establish the actual risk of injury to the Saphenous Nerve using a “closed” technique for the insertion of a TightRope.

Method

20 TightRopes were inserted into Fresh Frozen Cadavers. This was done using the senior authors preferred technique of divergent tightropes with the distal implant directed slightly anterior to the fibula-tibia axis and the proximal implant slightly posterior in order to simulate the greatest risk to the nerve. This was done under image Intensifier guidance to simulate an intraoperative environment. The medial side of the distal tibia was then dissected to directly record and measure the relationship of the TightRope to the Saphenous Nerve. Measurements were taken using digital calipers from the centre of the button on the medial side of the TightRope to the centre of the nerve at the point of closest proximity.

Results

12 TightRopes were found to exit posterior to the nerve, 7 anterior and 1 penetrated through the centre of the nerve. The mean distance from the centre of the button to the nerve was 6.99mm (range 0.72–14.52mm, standard deviation 4.33mm). In 9 of the 20 TightRopes, the nerve was found to be less than 5mm away.

Conclusion

Our findings demonstrated that the risks of damaging or indeed entrapping the Saphenous nerve were high, and therefore we would advocate an open incision on the medial side with judicious exploration to ensure there is no damage to the medial neurological structures.