Aims. The aim of this study was to compare the design of the generic
OptiStem XTR femoral stem with the established Exeter femoral stem. Materials and Methods. We obtained five boxed, as manufactured, implants of both designs
at random (ten in total). Two examiners were blinded to the implant
design and independently measured the mass, volume, trunnion surface
topography, trunnion roughness, trunnion cone angle, Caput-Collum-Diaphyseal
(CCD) angle, femoral
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).Aims
Methods
Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.Aims
Methods
Complex joint fractures of the lower extremity are often accompanied by soft-tissue swelling and are associated with prolonged hospitalization and soft-tissue complications. The aim of the study was to evaluate the effect of vascular impulse technology (VIT) on soft-tissue conditioning in comparison with conventional elevation. A total of 100 patients were included in this prospective, randomized, controlled monocentre study allocated to the three subgroups of dislocated ankle fracture (n = 40), pilon fracture (n = 20), and intra-articular calcaneal fracture (n = 40). Patients were randomized to the two study groups in a 1:1 ratio. The effectiveness of VIT (intervention) compared with elevation (control) was analyzed separately for the whole study population and for the three subgroups. The primary endpoint was the time from admission until operability (in days).Aims
Methods
In this review, we discuss the evidence for patients returning to sport after hip arthroplasty. This includes the choices regarding level of sporting activity and revision or complications, the type of implant, fixation and techniques of implantation, and how these choices relate to health economics. It is apparent that despite its success over six decades, hip arthroplasty has now evolved to accommodate and support ever-increasing patient demands and may therefore face new challenges. Cite this article:
The aim of this randomized trial was to compare the functional outcome of two different surgical approaches to the hip in patients with a femoral neck fracture treated with a hemiarthroplasty. A total of 150 patients who were treated between February 2014 and July 2017 were included. Patients were allocated to undergo hemiarthroplasty using either an anterolateral or a direct lateral approach, and were followed for 12 months. The mean age of the patients was 81 years (69 to 90), and 109 were women (73%). Functional outcome measures, assessed by a physiotherapist blinded to allocation, and patient-reported outcome measures (PROMs) were collected postoperatively at three and 12 months.Aims
Patients and Methods
The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years. We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score.Aims
Patients and Methods
This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation.Aims
Patients and Methods
Controversy about the use of an anatomical total shoulder arthroplasty
(aTSA) in young arthritic patients relates to which is the ideal
form of fixation for the glenoid component: cemented or cementless.
This study aimed to evaluate implant survival of aTSA when used
in patients aged < 60 years with primary glenohumeral osteoarthritis (OA),
and to compare the survival of cemented all-polyethylene and cementless
metal-backed glenoid components. A total of 69 consecutive aTSAs were performed in 67 patients
aged < 60 years with primary glenohumeral OA. Their mean age
at the time of surgery was 54 years (35 to 60). Of these aTSAs,
46 were undertaken using a cemented polyethylene component and 23
were undertaken using a cementless metal-backed component. The age, gender,
preoperative function, mobility, premorbid glenoid erosion, and
length of follow-up were comparable in the two groups. The patients
were reviewed clinically and radiographically at a mean of 10.3
years (5 to 12, Aims
Materials and Methods
The aim of this study was to compare the cost-effectiveness of
treatment with an osseointegrated percutaneous (OI-) prosthesis
and a socket-suspended (S-) prosthesis for patients with a transfemoral
amputation. A Markov model was developed to estimate the medical costs and
changes in quality-adjusted life-years (QALYs) attributable to treatment
of unilateral transfemoral amputation over a projected period of
20 years from a healthcare perspective. Data were collected alongside
a prospective clinical study of 51 patients followed for two years.Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
Total knee arthroplasty (TKA) is a major orthopaedic
intervention. The length of a patient's stay has been progressively
reduced with the introduction of enhanced recovery protocols: day-case
surgery has become the ultimate challenge. This narrative review shows the potential limitations of day-case
TKA. These constraints may be social, linked to patient’s comorbidities,
or due to surgery-related adverse events (e.g. pain, post-operative
nausea and vomiting, etc.). Using patient stratification, tailored surgical techniques and
multimodal opioid-sparing analgesia, day-case TKA might be achievable
in a limited group of patients. The younger, male patient without
comorbidities and with an excellent social network around him might
be a candidate. Demographic changes, effective recovery programmes and less invasive
surgical techniques such as unicondylar knee arthroplasty, may increase
the size of the group of potential day-case patients. The cost reduction achieved by day-case TKA needs to be balanced
against any increase in morbidity and mortality and the cost of
advanced follow-up at a distance with new technology. These factors
need to be evaluated before adopting this ultimate ‘fast-track’
approach. Cite this article:
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article: