Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims. This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. Methods. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth. Results. Overall, 36 mm heads had lower corrosion onset load (p = 0.009) and change in open circuit potential (OCP) during simulated gait with (p = 0.006) and without joint movement (p = 0.004). Discontinuing gait’s joint movement decreased corrosion currents (p = 0.042); however, wear testing showed no significant effect of joint movement on taper damage. In addition, 36 mm heads had greater corrosion area (p = 0.050), but no significant difference was found for maximum linear wear depth (p = 0.155). Conclusion. Larger heads are more susceptible to taper corrosion; however, not due to frictional torque as hypothesized. An alternative hypothesis of taper flexural rigidity differential is proposed. Further studies are necessary to investigate the clinical significance and underlying mechanism of this finding. Cite this article: Bone Jt Open 2021;2(11):1004–1016


Bone & Joint Open
Vol. 2, Issue 8 | Pages 611 - 617
10 Aug 2021
Kubik JF Bornes TD Klinger CE Dyke JP Helfet DL

Aims. Surgical treatment of young femoral neck fractures often requires an open approach to achieve an anatomical reduction. The application of a calcar plate has recently been described to aid in femoral neck fracture reduction and to augment fixation. However, application of a plate may potentially compromise the regional vascularity of the femoral head and neck. The purpose of this study was to investigate the effect of calcar femoral neck plating on the vascularity of the femoral head and neck. Methods. A Hueter approach and capsulotomy were performed bilaterally in six cadaveric hips. In the experimental group, a one-third tubular plate was secured to the inferomedial femoral neck at 6:00 on the clockface. The contralateral hip served as a control with surgical approach and capsulotomy without fixation. Pre- and post-contrast MRI was then performed to quantify signal intensity in the femoral head and neck. Qualitative assessment of the terminal arterial branches to the femoral head, specifically the inferior retinacular artery (IRA), was also performed. Results. Quantitative MRI revealed a mean reduction of 1.8% (SD 3.1%) of arterial contribution in the femoral head and a mean reduction of 7.1% (SD 10.6%) in the femoral neck in the plating group compared to non-plated controls. Based on femoral head quadrant analysis, the largest mean decrease in arterial contribution was in the inferomedial quadrant (4.0%, SD 6.6%). No significant differences were found between control and experimental hips for any femoral neck or femoral head regions. The inferior retinaculum of Weitbrecht (containing the IRA) was directly visualized in six of 12 specimens. Qualitative MRI assessment confirmed IRA integrity in all specimens. Conclusion. Calcar femoral neck plating at the 6:00 position on the clockface resulted in minimal decrease in femoral head and neck vascularity, and therefore it may be considered as an adjunct to laterally-based fixation for reduction and fixation of femoral neck fractures, especially in younger patients. Cite this article: Bone Jt Open 2021;2(8):611–617


Bone & Joint Research
Vol. 13, Issue 4 | Pages 184 - 192
18 Apr 2024
Morita A Iida Y Inaba Y Tezuka T Kobayashi N Choe H Ike H Kawakami E

Aims

This study was designed to develop a model for predicting bone mineral density (BMD) loss of the femur after total hip arthroplasty (THA) using artificial intelligence (AI), and to identify factors that influence the prediction. Additionally, we virtually examined the efficacy of administration of bisphosphonate for cases with severe BMD loss based on the predictive model.

Methods

The study included 538 joints that underwent primary THA. The patients were divided into groups using unsupervised time series clustering for five-year BMD loss of Gruen zone 7 postoperatively, and a machine-learning model to predict the BMD loss was developed. Additionally, the predictor for BMD loss was extracted using SHapley Additive exPlanations (SHAP). The patient-specific efficacy of bisphosphonate, which is the most important categorical predictor for BMD loss, was examined by calculating the change in predictive probability when hypothetically switching between the inclusion and exclusion of bisphosphonate.


Bone & Joint Research
Vol. 5, Issue 11 | Pages 531 - 537
1 Nov 2016
Burgo FJ Mengelle DE Ozols A Fernandez C Autorino CM

Objectives. Studies reporting specifically on squeaking in total hip arthroplasty have focused on cementless, and not on hybrid, fixation. We hypothesised that the cement mantle of the femur might have a damping effect on the sound transmitted through the metal stem. The objective of this study was to test the effect of cement on sound propagation along different stem designs and under different fixation conditions. Methods. An in vitro model for sound detection, composed of a mechanical suspension structure and a sound-registering electronic assembly, was designed. A pulse of sound in the audible range was propagated along bare stems and stems implanted in cadaveric bone femurs with and without cement. Two stems of different alloy and geometry were compared. Results. The magnitudes of the maximum amplitudes of the bare stem were in the range of 10.8 V to 11.8 V, whereas the amplitudes for the same stems with a cement mantle in a cadaveric bone decreased to 0.3 V to 0.7 V, implying a pulse-attenuation efficiency of greater than 97%. The same magnitude is close to 40% when the comparison is made against stems implanted in cadaveric bone femurs without cement. Conclusion. The in vitro model presented here has shown that the cement had a remarkable effect on sound attenuation and a strong energy absorption in cement mantle and bone. The visco-elastic properties of cement can contribute to the dissipation of vibro-acoustic energy, thus preventing hip prostheses from squeaking. This could explain, at least in part, the lack of reports of squeaking when hybrid fixation is used. Cite this article: F. J. Burgo, D. E. Mengelle, A. Ozols, C. Fernandez, C. M. Autorino. The damping effect of cement as a potential mitigation factor of squeaking in ceramic-on-ceramic total hip arthroplasty. Bone Joint Res 2016;5:531–537. DOI: 10.1302/2046-3758.511.BJR-2016-0058.R1


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 90 - 101
1 Jan 2020
Davis ET Pagkalos J Kopjar B

Aims. The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Methods. Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad). Results. A total of 5,329 THAs were revised, 1,290 of which were due to aseptic loosening. Total radiation dose, face asymmetry, and stabilization treatments were found to significantly affect implant survival. G1 had the highest revision risk for any reason and for aseptic loosening and G3 and G4 the lowest. Compared with G1, the adjusted hazard ratio for G2 was 0.74 (95% confidence interval (CI) 0.64 to 0.86), G3 was 0.36 (95% CI 0.30 to 0.43), and G4 was 0.38 (95% CI 0.31 to 0.47). The cumulative incidence of revision for aseptic loosening at 12 years was 0.52 and 0.54 per 100 THAs for G3 and G4, respectively, compared with 1.95 per 100 THAs in G1. Asymmetrical liners had a lower revision risk due to aseptic loosening and reasons other than aseptic loosening compared with symmetric (flat) liners. In G3 and G4, stabilization with vitamin E and heating above melting point performed best. Conclusion. Polyethylene liners with a total radiation dose of ≥ 5 Mrad, an asymmetrical liner face, and stabilization with heating above the melting point demonstrate best survival. Cite this article: Bone Joint J 2020;102-B(1):90–101


Bone & Joint Open
Vol. 1, Issue 8 | Pages 500 - 507
18 Aug 2020
Cheruvu MS Bhachu DS Mulrain J Resool S Cool P Ford DJ Singh RA

Aims

Our rural orthopaedic service has undergone service restructure during the COVID-19 pandemic in order to sustain hip fracture care. All adult trauma care has been centralised to the Royal Shrewsbury Hospital for assessment and medical input, before transferring those requiring operative intervention to the Robert Jones and Agnes Hunt Orthopaedic Hospital. We aim to review the impact of COVID-19 on hip fracture workload and service changes upon management of hip fractures.

Methods

We reviewed our prospectively maintained trust database and National Hip Fracture Database records for the months of March and April between the years 2016 and 2020. Our assessment included fracture pattern (intrascapular vs extracapsular hip fracture), treatment intervention, length of stay and mortality.


Bone & Joint Research
Vol. 10, Issue 6 | Pages 354 - 362
1 Jun 2021
Luo Y Zhao X Yang Z Yeersheng R Kang P

Aims

The purpose of this study was to examine the efficacy and safety of carbazochrome sodium sulfonate (CSS) combined with tranexamic acid (TXA) on blood loss and inflammatory responses after primary total hip arthroplasty (THA), and to investigate the influence of different administration methods of CSS on perioperative blood loss during THA.

Methods

This study is a randomized controlled trial involving 200 patients undergoing primary unilateral THA. A total of 200 patients treated with intravenous TXA were randomly assigned to group A (combined intravenous and topical CSS), group B (topical CSS), group C (intravenous CSS), or group D (placebo).


Bone & Joint Research
Vol. 2, Issue 9 | Pages 200 - 205
1 Sep 2013
Amarasekera HW Campbell PC Parsons N Achten J Masters J Griffin DR Costa ML

Objectives . We aimed to determine the effect of surgical approach on the histology of the femoral head following resurfacing of the hip. Methods. We performed a histological assessment of the bone under the femoral component taken from retrieval specimens of patients having revision surgery following resurfacing of the hip. We compared the number of empty lacunae in specimens from patients who had originally had a posterior surgical approach with the number in patients having alternative surgical approaches. Results. We found a statistically significant increase in the percentage of empty lacunae in retrieval specimens from patients who had the posterior approach compared with other surgical approaches (p < 0.001). . Conclusions. This indicates that the vascular compromise that occurs during the posterior surgical approach does have long-term effects on the bone of the femoral head, even if it does not cause overt avascular necrosis. Cite this article: Bone Joint Res 2013;2:200–5


Bone & Joint Research
Vol. 5, Issue 9 | Pages 362 - 369
1 Sep 2016
Oba M Inaba Y Kobayashi N Ike H Tezuka T Saito T

Objectives. In total hip arthroplasty (THA), the cementless, tapered-wedge stem design contributes to achieving initial stability and providing optimal load transfer in the proximal femur. However, loading conditions on the femur following THA are also influenced by femoral structure. Therefore, we determined the effects of tapered-wedge stems on the load distribution of the femur using subject-specific finite element models of femurs with various canal shapes. Patients and Methods. We studied 20 femurs, including seven champagne flute-type femurs, five stovepipe-type femurs, and eight intermediate-type femurs, in patients who had undergone cementless THA using the Accolade TMZF stem at our institution. Subject–specific finite element (FE) models of pre- and post-operative femurs with stems were constructed and used to perform FE analyses (FEAs) to simulate single-leg stance. FEA predictions were compared with changes in bone mineral density (BMD) measured for each patient during the first post-operative year. Results. Stovepipe models implanted with large-size stems had significantly lower equivalent stress on the proximal-medial area of the femur compared with champagne-flute and intermediate models, with a significant loss of BMD in the corresponding area at one year post-operatively. Conclusions. The stovepipe femurs required a large-size stem to obtain an optimal fit of the stem. The FEA result and post-operative BMD change of the femur suggest that the combination of a large-size Accolade TMZF stem and stovepipe femur may be associated with proximal stress shielding. Cite this article: M. Oba, Y. Inaba, N. Kobayashi, H. Ike, T. Tezuka, T. Saito. Effect of femoral canal shape on mechanical stress distribution and adaptive bone remodelling around a cementless tapered-wedge stem. Bone Joint Res 2016;5:362–369. DOI: 10.1302/2046-3758.59.2000525


Aims

Surgical treatment of hip fracture is challenging; the bone is porotic and fixation failure can be catastrophic. Novel implants are available which may yield superior clinical outcomes. This study compared the clinical effectiveness of the novel X-Bolt Hip System (XHS) with the sliding hip screw (SHS) for the treatment of fragility hip fractures.

Methods

We conducted a multicentre, superiority, randomized controlled trial. Patients aged 60 years and older with a trochanteric hip fracture were recruited in ten acute UK NHS hospitals. Participants were randomly allocated to fixation of their fracture with XHS or SHS. A total of 1,128 participants were randomized with 564 participants allocated to each group. Participants and outcome assessors were blind to treatment allocation. The primary outcome was the EuroQol five-dimension five-level health status (EQ-5D-5L) utility at four months. The minimum clinically important difference in utility was pre-specified at 0.075. Secondary outcomes were EQ-5D-5L utility at 12 months, mortality, residential status, mobility, revision surgery, and radiological measures.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 758 - 765
12 Oct 2023
Wagener N Löchel J Hipfl C Perka C Hardt S Leopold VJ

Aims. Psychological status may be an important predictor of outcome after periacetabular osteotomy (PAO). The aim of this study was to investigate the influence of psychological distress on postoperative health-related quality of life, joint function, self-assessed pain, and sports ability in patients undergoing PAO. Methods. In all, 202 consecutive patients who underwent PAO for developmental dysplasia of the hip (DDH) at our institution from 2015 to 2017 were included and followed up at 63 months (SD 10) postoperatively. Of these, 101 with complete data sets entered final analysis. Patients were assessed by questionnaire. Psychological status was measured by Brief Symptom Inventory (BSI-18), health-related quality of life was raised with 36-Item Short Form Survey (SF-36), hip functionality was measured by the short version 0f the International Hip Outcome Tool (iHOT-12), Subjective Hip Value (SHV), and Hip Disability and Outcome Score (HOS). Surgery satisfaction and pain were assessed. Dependent variables (endpoints) were postoperative quality of life (SF-36, HOS quality of life (QoL)), joint function (iHOT-12, SHV, HOS), patient satisfaction, and pain. Psychological distress was assessed by the Global Severity Index (GSI), somatization (BSI Soma), depression (BSI Depr), and anxiety (BSI Anx). Influence of psychological status was assessed by means of univariate and multiple multivariate regression analysis. Results. In multiple multivariate regression, postoperative GSI, BSI Soma, and BSI Depr had a negative effect on postoperative SF-36 (e -2.07, -3.05, and -2.67, respectively; p < 0.001), iHOT-12 (e -1.35 and -4.65, respectively; p < 0.001), SHV (e -1.20 and -2.71, respectively; p < 0.001), HOS QoL (e -2.09 and -4.79, respectively; p < 0.001), HOS Function (e -1.00 and -3.94, respectively; p < 0.001), and HOS Sport (e -1.44 and -5.29, respectively; p < 0.001), and had an effect on postoperative pain (e 0.13 and 0.37, respectively; p < 0.001). Conclusion. Psychological distress, depression, and somatization disorders affect health-related quality of life, perceived joint function, and sports ability. Pain perception is significantly increased by somatization. However, patient satisfaction with surgery is not affected. Cite this article: Bone Jt Open 2023;4(10):758–765


Bone & Joint Open
Vol. 3, Issue 10 | Pages 795 - 803
12 Oct 2022
Liechti EF Attinger MC Hecker A Kuonen K Michel A Klenke FM

Aims. Traditionally, total hip arthroplasty (THA) templating has been performed on anteroposterior (AP) pelvis radiographs. Recently, additional AP hip radiographs have been recommended for accurate measurement of the femoral offset (FO). To verify this claim, this study aimed to establish quantitative data of the measurement error of the FO in relation to leg position and X-ray source position using a newly developed geometric model and clinical data. Methods. We analyzed the FOs measured on AP hip and pelvis radiographs in a prospective consecutive series of 55 patients undergoing unilateral primary THA for hip osteoarthritis. To determine sample size, a power analysis was performed. Patients’ position and X-ray beam setting followed a standardized protocol to achieve reproducible projections. All images were calibrated with the KingMark calibration system. In addition, a geometric model was created to evaluate both the effects of leg position (rotation and abduction/adduction) and the effects of X-ray source position on FO measurement. Results. The mean FOs measured on AP hip and pelvis radiographs were 38.0 mm (SD 6.4) and 36.6 mm (SD 6.3) (p < 0.001), respectively. Radiological view had a smaller effect on FO measurement than inaccurate leg positioning. The model showed a non-linear relationship between projected FO and femoral neck orientation; at 30° external neck rotation (with reference to the detector plane), a true FO of 40 mm was underestimated by up to 20% (7.8 mm). With a neutral to mild external neck rotation (≤ 15°), the underestimation was less than 7% (2.7 mm). The effect of abduction and adduction was negligible. Conclusion. For routine THA templating, an AP pelvis radiograph remains the gold standard. Only patients with femoral neck malrotation > 15° on the AP pelvis view, e.g. due to external rotation contracture, should receive further imaging. Options include an additional AP hip view with elevation of the entire affected hip to align the femoral neck more parallel to the detector, or a CT scan in more severe cases. Cite this article: Bone Jt Open 2022;3(10):795–803


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 331 - 340
1 Mar 2022
Strahl A Kazim MA Kattwinkel N Hauskeller W Moritz S Arlt S Niemeier A

Aims. The aim of this study was to determine whether total hip arthroplasty (THA) for chronic hip pain due to unilateral primary osteoarthritis (OA) has a beneficial effect on cognitive performance. Methods. A prospective cohort study was conducted with 101 patients with end-stage hip OA scheduled for THA (mean age 67.4 years (SD 9.5), 51.5% female (n = 52)). Patients were assessed at baseline as well as after three and months. Primary outcome was cognitive performance measured by d2 Test of Attention at six months, Trail Making Test (TMT), FAS-test, Rivermead Behavioural Memory Test (RBMT; story recall subtest), and Rey-Osterrieth Complex Figure Test (ROCF). The improvement of cognitive performance was analyzed using repeated measures analysis of variance. Results. At six months, there was significant improvement in attention, working speed and concentration (d2-test; p < 0.001), visual construction and visual memory (ROCF; p < 0.001), semantic memory (FAS-test; p = 0.009), verbal episodic memory (RBMT; immediate recall p = 0.023, delayed recall p = 0.026), as well as pain (p < 0.001) with small to large effect sizes. Attention, concentration, and visual as well as verbal episodic memory improved significantly with medium effect sizes over η. 2. partial. = 0.06. In these cognitive domains the within-group difference exceeded the minimum clinically important difference. Conclusion. THA is associated with clinically relevant postoperative improvement in the cognitive functions of attention, concentration, and memory. These data support the concept of a broad interaction of arthroplasty with central nervous system function. Cite this article: Bone Joint J 2022;104-B(3):331–340


The Bone & Joint Journal
Vol. 106-B, Issue 7 | Pages 656 - 661
1 Jul 2024
Bolbocean C Hattab Z O'Neill S Costa ML

Aims. Cemented hemiarthroplasty is an effective form of treatment for most patients with an intracapsular fracture of the hip. However, it remains unclear whether there are subgroups of patients who may benefit from the alternative operation of a modern uncemented hemiarthroplasty – the aim of this study was to investigate this issue. Knowledge about the heterogeneity of treatment effects is important for surgeons in order to target operations towards specific subgroups who would benefit the most. Methods. We used causal forest analysis to compare subgroup- and individual-level treatment effects between cemented and modern uncemented hemiarthroplasty in patients aged > 60 years with an intracapsular fracture of the hip, using data from the World Hip Trauma Evaluation 5 (WHiTE 5) multicentre randomized clinical trial. EuroQol five-dimension index scores were used to measure health-related quality of life at one, four, and 12 months postoperatively. Results. Our analysis revealed a complex landscape of responses to the use of a cemented hemiarthroplasty in the 12 months after surgery. There was heterogeneity of effects with regard to baseline characteristics, including age, pre-injury health status, and lifestyle factors such as alcohol consumption. This heterogeneity was greater at the one-month mark than at subsequent follow-up timepoints, with particular regard to subgroups based on age. However, for all subgroups, the effect estimates for quality of life lay within the confidence intervals derived from the analysis of all patients. Conclusion. The use of a cemented hemiarthroplasty is expected to increase health-related quality of life compared with modern uncemented hemiarthroplasty for all subgroups of patients aged > 60 years with a displaced intracapsular fracture of the hip. Cite this article: Bone Joint J 2024;106-B(7):656–661


Bone & Joint Research
Vol. 11, Issue 5 | Pages 317 - 326
23 May 2022
Edwards TC Guest B Garner A Logishetty K Liddle AD Cobb JP

Aims. This study investigates the use of the metabolic equivalent of task (MET) score in a young hip arthroplasty population, and its ability to capture additional benefit beyond the ceiling effect of conventional patient-reported outcome measures. Methods. From our electronic database of 751 hip arthroplasty procedures, 221 patients were included. Patients were excluded if they had revision surgery, an alternative hip procedure, or incomplete data either preoperatively or at one-year follow-up. Included patients had a mean age of 59.4 years (SD 11.3) and 54.3% were male, incorporating 117 primary total hip and 104 hip resurfacing arthroplasty operations. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and the MET were recorded preoperatively and at one-year follow-up. The distribution was examined reporting the presence of ceiling and floor effects. Validity was assessed correlating the MET with the other scores using Spearman’s rank correlation coefficient and determining responsiveness. A subgroup of 93 patients scoring 48/48 on the OHS were analyzed by age, sex, BMI, and preoperative MET using the other metrics to determine if differences could be established despite scoring identically on the OHS. Results. Postoperatively the OHS and EQ-5D demonstrate considerable negatively skewed distributions with ceiling effects of 41.6% and 53.8%, respectively. The MET was normally distributed postoperatively with no relevant ceiling effect. Weak-to-moderate significant correlations were found between the MET and the other two metrics. In the 48/48 subgroup, no differences were found comparing groups with the EQ-5D, however significantly higher mean MET scores were demonstrated for patients aged < 60 years (12.7 (SD 4.7) vs 10.6 (SD 2.4), p = 0.008), male patients (12.5 (SD 4.5) vs 10.8 (SD 2.8), p = 0.024), and those with preoperative MET scores > 6 (12.6 (SD 4.2) vs 11.0 (SD 3.3), p = 0.040). Conclusion. The MET is normally distributed in patients following hip arthroplasty, recording levels of activity which are undetectable using the OHS. Cite this article: Bone Joint Res 2022;11(5):317–326


Bone & Joint Open
Vol. 4, Issue 8 | Pages 628 - 635
22 Aug 2023
Hedlundh U Karlsson J Sernert N Haag L Movin T Papadogiannakis N Kartus J

Aims. A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision THAs without infection performed using the same lateral approach. Methods. A group of eight patients revised due to a PJI with a previous lateral approach was compared with a group of 21 revised THAs without infection, performed using the same approach. The primary variables of the study were the fibril diameter, as seen in transmission electron microscopy (TEM), and the total degeneration score (TDS), as seen under the light microscope. An analysis of bacteriology, classification of infection, and antibiotic treatment was also performed. Results. Biopsy samples from the GMED from infected patients revealed a larger fibril diameter than control patients, as seen in the TEM (p < 0.001). Uninfected patients were slightly older and had their revisions performed significantly later than the infected patients. Histologically, samples from infected patients revealed significantly more vascularity (p < 0.001), the presence of glycosaminoglycans (p < 0.001), and a higher TDS (p = 0.003) than the control patients. The majority of patients had staphylococcal infections of various species. Conclusion. More histological degeneration in the GMED was found in patients undergoing THA revision surgery due to PJI than in patients undergoing THA revision surgery due to other reasons. Cite this article: Bone Jt Open 2023;4(8):628–635


Bone & Joint Open
Vol. 3, Issue 4 | Pages 340 - 347
22 Apr 2022
Winkler T Costa ML Ofir R Parolini O Geissler S Volk H Eder C

Aims. The aim of the HIPGEN consortium is to develop the first cell therapy product for hip fracture patients using PLacental-eXpanded (PLX-PAD) stromal cells. Methods. HIPGEN is a multicentre, multinational, randomized, double-blind, placebo-controlled trial. A total of 240 patients aged 60 to 90 years with low-energy femoral neck fractures (FNF) will be allocated to two arms and receive an intramuscular injection of either 150 × 10. 6. PLX-PAD cells or placebo into the medial gluteal muscle after direct lateral implantation of total or hemi hip arthroplasty. Patients will be followed for two years. The primary endpoint is the Short Physical Performance Battery (SPPB) at week 26. Secondary and exploratory endpoints include morphological parameters (lean body mass), functional parameters (abduction and handgrip strength, symmetry in gait, weightbearing), all-cause mortality rate and patient-reported outcome measures (Lower Limb Measure, EuroQol five-dimension questionnaire). Immunological biomarker and in vitro studies will be performed to analyze the PLX-PAD mechanism of action. A sample size of 240 subjects was calculated providing 88% power for the detection of a 1 SPPB point treatment effect for a two-sided test with an α level of 5%. Conclusion. The HIPGEN study assesses the efficacy, safety, and tolerability of intramuscular PLX-PAD administration for the treatment of muscle injury following arthroplasty for hip fracture. It is the first phase III study to investigate the effect of an allogeneic cell therapy on improved mobilization after hip fracture, an aspect which is in sore need of addressing for the improvement in standard of care treatment for patients with FNF. Cite this article: Bone Jt Open 2022;3(4):340–347


Bone & Joint Open
Vol. 5, Issue 9 | Pages 776 - 784
19 Sep 2024
Gao J Chai N Wang T Han Z Chen J Lin G Wu Y Bi L

Aims. In order to release the contracture band completely without damaging normal tissues (such as the sciatic nerve) in the surgical treatment of gluteal muscle contracture (GMC), we tried to display the relationship between normal tissue and contracture bands by magnetic resonance neurography (MRN) images, and to predesign a minimally invasive surgery based on the MRN images in advance. Methods. A total of 30 patients (60 hips) were included in this study. MRN scans of the pelvis were performed before surgery. The contracture band shape and external rotation angle (ERA) of the proximal femur were also analyzed. Then, the minimally invasive GMC releasing surgery was performed based on the images and measurements, and during the operation, incision lengths, surgery duration, intraoperative bleeding, and complications were recorded; the time of the first postoperative off-bed activity was also recorded. Furthermore, the patients’ clinical functions were evaluated by means of Hip Outcome Score (HOS) and Ye et al’s objective assessments, respectively. Results. The contracture bands exhibited three typical types of shape – feather-like, striped, and mixed shapes – in MR images. Guided by MRN images, we designed minimally invasive approaches directed to each hip. These approaches resulted in a shortened incision length in each hip (0.3 cm (SD 0.1)), shorter surgery duration (25.3 minutes (SD 5.8)), less intraoperative bleeding (8.0 ml (SD 3.6)), and shorter time between the end of the operation and the patient’s first off-bed activity (17.2 hours (SD 2.0)) in each patient. Meanwhile, no serious postoperative complications occurred in all patients. The mean HOS-Sports subscale of patients increased from 71.0 (SD 5.3) to 94.83 (SD 4.24) at six months postoperatively (p < 0.001). The follow-up outcomes from all patients were “good” and “excellent”, based on objective assessments. Conclusion. Preoperative MRN analysis can be used to facilitate the determination of the relationship between contracture band and normal tissues. The minimally invasive surgical design via MRN can avoid nerve damage and improve the release effect. Cite this article: Bone Jt Open 2024;5(9):776–784


Bone & Joint Open
Vol. 5, Issue 2 | Pages 123 - 131
12 Feb 2024
Chen B Duckworth AD Farrow L Xu YJ Clement ND

Aims. This study aimed to determine whether lateral femoral wall thickness (LWT) < 20.5 mm was associated with increased revision risk of intertrochanteric fracture (ITF) of the hip following sliding hip screw (SHS) fixation when the medial calcar was intact. Additionally, the study assessed the association between LWT and patient mortality. Methods. This retrospective study included ITF patients aged 50 years and over treated with SHS fixation between 2019 and 2021 at a major trauma centre. Demographic information, fracture type, delirium status, American Society of Anesthesiologists grade, and length of stay were collected. LWT and tip apex distance were measured. Revision surgery and mortality were recorded at a mean follow-up of 19.5 months (1.6 to 48). Cox regression was performed to evaluate independent risk factors associated with revision surgery and mortality. Results. The cohort consisted of 890 patients with a mean age of 82 years (SD 10.2). Mean LWT was 27.0 mm (SD 8.6), and there were 213 patients (23.9%) with LWT < 20.5 mm. Overall, 20 patients (2.2%) underwent a revision surgery following SHS fixation. Adjusting for covariates, LWT < 20.5 mm was not independently associated with an increased revision or mortality risk. However, factors that were significantly more prevalent in LWT < 20.5 mm group, which included residence in care home (hazard ratio (HR) 1.84; p < 0.001) or hospital (HR 1.65; p = 0.005), and delirium (HR 1.32; p = 0.026), were independently associated with an increased mortality risk. The only independent factor associated with increased risk of revision was older age (HR 1.07; p = 0.030). Conclusion. LWT was not associated with risk of revision surgery in patients with an ITF fixed with a SHS when the calcar was intact, after adjusting for the independent effect of age. Although LWT < 20.5 mm was not an independent risk factor for mortality, patients with LWT < 20.5 mm were more likely to be from care home or hospital and have delirium on admission, which were associated with a higher mortality rate. Cite this article: Bone Jt Open 2024;5(2):123–131


Bone & Joint Open
Vol. 4, Issue 3 | Pages 198 - 204
16 Mar 2023
Ramsay N Close JCT Harris IA Harvey LA

Aims. Cementing in arthroplasty for hip fracture is associated with improved postoperative function, but may have an increased risk of early mortality compared to uncemented fixation. Quantifying this mortality risk is important in providing safe patient care. This study investigated the association between cement use in arthroplasty and mortality at 30 days and one year in patients aged 50 years and over with hip fracture. Methods. This retrospective cohort study used linked data from the Australian Hip Fracture Registry and the National Death Index. Descriptive analysis and Kaplan-Meier survival curves tested the unadjusted association of mortality between cemented and uncemented procedures. Multilevel logistic regression, adjusted for covariates, tested the association between cement use and 30-day mortality following arthroplasty. Given the known institutional variation in preference for cemented fixation, an instrumental variable analysis was also performed to minimize the effect of unknown confounders. Adjusted Cox modelling analyzed the association between cement use and mortality at 30 days and one year following surgery. Results. The 30-day mortality was 6.9% for cemented and 4.9% for uncemented groups (p = 0.003). Cement use was significantly associated with 30-day mortality in the Kaplan-Meier survival curve (p = 0.003). After adjusting for covariates, no significant association between cement use and 30-day mortality was shown in the adjusted multilevel logistic regression (odd rati0 (OR) 1.1, 95% confidence interval (CI) 0.9 to 1.5; p = 0.366), or in the instrumental variable analysis (OR 1.0, 95% CI 0.9 to 1.0, p=0.524). There was no significant between-group difference in mortality within 30days (hazard ratio (HR) 0.9, 95% CI 0.7to 1.1; p = 0.355) or one year (HR 0.9 95% CI 0.8 to 1.1; p = 0.328) in the Cox modelling. Conclusion. No statistically significant difference in patient mortality with cement use in arthroplasty was demonstrated in this population, once adjusted for covariates. This study concludes that cementing in arthroplasty for hip fracture is a safe means of surgical fixation. Cite this article: Bone Jt Open 2023;4(3):198–204