Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 81 - 81
19 Aug 2024
Angelomenos V Shareghi B Itayem R Mohaddes M
Full Access

Early micromotion of hip implants measured with radiostereometric analysis (RSA) is a predictor for late aseptic loosening. Computed Tomography Radiostereometric Analysis (CT-RSA) can be used to determine implant micro-movements using low-dose CT scans. CT-RSA enables a non-invasive measurement of implants. We evaluated the precision of CT-RSA in measuring early stem migration. Standard marker-based RSA was used as reference. We hypothesised that CT-RSA can be used as an alternative to RSA in assessing implant micromotions. We included 31 patients undergoing Total Hip Arthroplasty (THA). Distal femoral stem migration at 1 year was measured with both RSA and CT-RSA. Comparison of the two methods was performed with paired-analysis and Bland-Altman plots. Furthermore, the inter- and intraobserver reliability of the CT-RSA method was evaluated. No statistical difference was found between RSA and CTMA measurements. The Bland-Altman plots showed good agreement between marker-based RSA and CT-RSA. The intra- and interobserver reliability of the CT-RSA method was found to be excellent (≥0.992). CT-RSA is comparable to marker-based RSA in measuring distal femoral stem migration. CTMA can be used as an alternative method to detect early implant migration