Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 27 - 27
23 Apr 2024
Howard A Harwood P Benton A Merrel C Culmer P Bolton W Stewart T
Full Access

Introduction

Ilizarov fixators are reliant on tensioned fine wires for stability. The tension in the wires is generated using specific tensioning devices. Loss of wire tension over time may lead of loss a stability and complications. A series of in vitro experiments were undertaken to explore wire tensioner accuracy, the impact of fixation bolt torque and initial tension on loss of tension in ilizarov constructs under static and dynamic loads.

Materials & Methods

Medical grade materials were applied to a synthetic bone analogue using surgical instruments in all experiments. Bolt torque was fixed at 6, 10 or 14 Nm using a torque limiting wrench. Wire tension was assessed using a strain measurement bridge. Wires were tensioned to 90, 110 and 130kg as measured by a commercial dynamometric tensioner. Static and dynamic testing was undertaken using an instron testing machine. Cyclical loads from 50–750N were applied for 5000 cycles.