Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_20 | Pages 3 - 3
12 Dec 2024
Bhalekar R Wells S Nargol M Nargol A Waller S Wildberg L Sidaginamale R Langton DJ
Full Access

Objectives

Unicompartmental knee replacement (UKR) is a conservative option for degenerative disease, with mobile (Oxford UKR) and fixed bearing (Physica ZUK) the most commonly used devices. The primary reasons for revising UKRs include disease progression (36.9%), loosening (31.7%), and pain (7.5%). Loosening typically occurs due to osteolysis caused by wear particles from the polyethylene (PE) inserts. There is limited published literature which has quantified volumetric material loss from the PE inserts of cemented fixed-bearing UKRs. This study aimed to quantify bearing wear and backside deformation of these PE components.

Design and Methods

At our national retrieval centre, we measured changes volumetric bearing wear and backside deformation of 31 explanted fixed-bearing Physica ZUK UKR PE inserts using a peer-reviewed methodology based on coordinate measuring machine analysis. These explants had been revised for any indication [Females (19) and Males (12)]. The location of the wear scars was identified and mapped. We compared the volumetric wear from the bearing surface with contemporary total knee replacement (TKR) inserts.


Bone & Joint Research
Vol. 8, Issue 6 | Pages 275 - 287
1 Jun 2019
Clement ND Bardgett M Merrie K Furtado S Bowman R Langton DJ Deehan DJ Holland J

Objectives

Our primary aim was to describe migration of the Exeter stem with a 32 mm head on highly crosslinked polyethylene and whether this is influenced by age. Our secondary aims were to assess functional outcome, satisfaction, activity, and bone mineral density (BMD) according to age.

Patients and Methods

A prospective cohort study was conducted. Patients were recruited into three age groups: less than 65 years (n = 65), 65 to 74 years (n = 68), and 75 years and older (n = 67). There were 200 patients enrolled in the study, of whom 115 were female and 85 were male, with a mean age of 69.9 years (sd 9.5, 42 to 92). They were assessed preoperatively, and at three, 12 and, 24 months postoperatively. Stem migration was assessed using Einzel-Bild-Röntgen-Analyse (EBRA). Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Harris Hip Score (HHS), Hip Disability and Osteoarthritis Outcome Score (HOOS), EuroQol-5 domains questionnaire (EQ-5D), short form-36 questionnaire (SF-36,) and patient satisfaction were used to assess outcome. The Lower Extremity Activity Scale (LEAS), Timed Up and Go (TUG) test, and activPAL monitor (energy expelled, time lying/standing/walking and step count) were used to assess activity. The BMD was assessed in Gruen and Charnley zones.


Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives

We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr.

Methods

For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood.


Bone & Joint Research
Vol. 7, Issue 6 | Pages 388 - 396
1 Jun 2018
Langton DJ Sidaginamale RP Joyce TJ Bowsher JG Holland JP Deehan D Nargol AVF Natu S

Objectives

We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL).

Patients and Methods

We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.


Bone & Joint Research
Vol. 6, Issue 2 | Pages 113 - 122
1 Feb 2017
Scholes SC Hunt BJ Richardson VM Langton DJ Smith E Joyce TJ

Objectives

The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.

Methods

In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.


Bone & Joint Research
Vol. 3, Issue 3 | Pages 60 - 68
1 Mar 2014
Langton DJ Sidaginamale RP Holland JP Deehan D Joyce TJ Nargol AVF Meek RD Lord JK

Objectives

Wear debris released from bearing surfaces has been shown to provoke negative immune responses in the recipient. Excessive wear has been linked to early failure of prostheses. Analysis using coordinate measuring machines (CMMs) can provide estimates of total volumetric material loss of explanted prostheses and can help to understand device failure. The accuracy of volumetric testing has been debated, with some investigators stating that only protocols involving hundreds of thousands of measurement points are sufficient. We looked to examine this assumption and to apply the findings to the clinical arena.

Methods

We examined the effects on the calculated material loss from a ceramic femoral head when different CMM scanning parameters were used. Calculated wear volumes were compared with gold standard gravimetric tests in a blinded study.


Bone & Joint Research
Vol. 2, Issue 5 | Pages 84 - 95
1 May 2013
Sidaginamale RP Joyce TJ Lord JK Jefferson R Blain PG Nargol AVF Langton DJ

Objectives

The aims of this piece of work were to: 1) record the background concentrations of blood chromium (Cr) and cobalt (Co) concentrations in a large group of subjects; 2) to compare blood/serum Cr and Co concentrations with retrieved metal-on-metal (MoM) hip resurfacings; 3) to examine the distribution of Co and Cr in the serum and whole blood of patients with MoM hip arthroplasties; and 4) to further understand the partitioning of metal ions between the serum and whole blood fractions.

Methods

A total of 3042 blood samples donated to the local transfusion centre were analysed to record Co and Cr concentrations. Also, 91 hip resurfacing devices from patients who had given pre-revision blood/serum samples for metal ion analysis underwent volumetric wear assessment using a coordinate measuring machine. Linear regression analysis was carried out and receiver operating characteristic curves were constructed to assess the reliability of metal ions to identify abnormally wearing implants. The relationship between serum and whole blood concentrations of Cr and Co in 1048 patients was analysed using Bland-Altman charts. This relationship was further investigated in an in vitro study during which human blood was spiked with trivalent and hexavalent Cr, the serum then separated and the fractions analysed.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 56 - 63
1 Apr 2012
Langton DJ Sidaginamale R Lord JK Nargol AVF Joyce TJ

Objectives

An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer.

Methods

Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular bearing surfaces and also the surfaces of the taper junctions. The relationship between taper wear rates and a number of variables, including bearing diameter and orientation of the acetabular component, was investigated.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 66 - 66
1 Jan 2011
Joyce TJ Langton DJ Nargol AVF
Full Access

Introduction: One of the latest groups of replacement hip joints are known as hip resurfacings and they consist of a relatively large diameter femoral head articulating within a thin acetabular cup. Many of these devices show good short to medium term clinical results. However there are concerns over such implants including fracture of the femur and possible wear debris related reactions. Much valuable data can be learnt from explanted prostheses which have ‘failed’ and then been removed from patients. As hip resurfacing prostheses have only recently been introduced, there are relatively few such retrieval studies.

Methods and materials: Nineteen femoral and acetabular components from metal-on-metal hip resurfacing prostheses were obtained at revision operations. There were eight patients who had femoral fractures and the remainder experienced worsening groin pain and a characteristic sterile effusion. There were eleven head components and four pairs of matching heads and cups. Each of these was examined using a Zeiss TSK Rond-com60A roundness measuring machine and a Mitutoyo LEGEX co-ordinate measuring machine (CMM). Out of roundness measurements were taken on three planes for each acetabular and femoral component. The CMM was used to obtain 12 traces at 30° intervals for each acetabular and femoral component, allowing areas of localised wear to be identified and the maximum wear depth to be quantified.

Results: The maximum out of roundness values for the nineteen components ranged from 1.8 to 91.8 microns. A similar range of values was obtained from the CMM results. From the paired components, out of roundness was greater in the head than in the cup. All eight femoral heads which were retrieved after fracture of the femur showed out of roundness of less than 5 microns.

Discussion: Both out of roundness measurements and those from the CMM provided information about the wear of the implants. For a new component, a typical out of roundness value would be no greater than 5 microns. Therefore, from the out of roundness values it was seen that the eight ‘fracture’ components showed minimal distortion or wear after removal. In contrast the remaining components, which had a minimum out of roundness of 15 microns, showed much greater wear, thus suggesting that the groin pain was associated with relatively large volumes of wear debris. CMM scans helped to identify localised areas of wear and maximum wear depths. Values in the range of < 2 microns to 164 microns have been reported previously and show good agreement with the findings of this study. Retrieved components which had been implanted at high angles of inclination and anteversion tended to show the greatest wear, implying that correct positioning in vivo is crucial to the longevity of hip resurfacing prostheses.