Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Content I can access

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_III | Pages 494 - 494
1 Aug 2008
McGowan J Nicol S Kumar CS
Full Access

Purpose: The purpose of the study was to compare the maximum compression force of three different 6.5 mm cancellous screws commonly used in hindfoot fusions.

Materials and methods: Screw 1 was a solid core standard fragment partially threaded cancellous screw (Smith and Nephew).

Screw 2 was a titanium cannulated screw with a medium thread pitch (Asnis III, Stryker).

Screw 3 was also a titanium cannulated screw with a large core diameter but with a small thread pitch (Ace, DePuy).

Four different densities of polyurethane foams were used simulating cancellous bone and the compression

Results: Screw 3 had the highest compression force in tests with the low density foams (p< 0.05) and screw 1 performed better in higher density foams (p< 0.05). In medium density foams, both screws 1 and 3 showed significantly more compression than screw 2 (p< 0.05).

Clinical relevance: The results indicate that the 6.5 mm standard fragment non-cannulated cancellous screw may provide more compression in a normal density bone whereas in an osteoporotic bone a cannulated titanium screw may be preferred for producing better compression during arthrodesis.