Vol. 95-B, Issue SUPP_34 | Pages 343 - 343
1 Dec 2013
Hayashi S Fujishiro T Hashimoto S Kanzaki N Nishiyama T Kurosaka M
Full Access

Introduction:

Implant dislocations are often caused by implant or bone impingement, and less impingement is critical to prevent dislocations. Several reports demonstrated that greater femoral offset delayed bony impingement and led to an improved range of motion (ROM) after THA. Therefore, an increase in the femoral offset may improve ROM and decrease implant dislocation. The aim of this study was to clarify the effect of the femoral offset in avoiding component or bony impingement after total hip arthroplasty (THA).

Methods:

Seventy-eight patients underwent THA with a Pinnacle cup and Summit stem (DePuy). Intraoperative kinematic analysis was performed with a navigation system, which was used to obtain intraoperative range of motion (ROM) measurements during trial insertion of stems of 2 different offset lengths with the same head size. Further, ROM was also measured after actual component insertion.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 174 - 174
1 Mar 2013
Fujishiro T Nishiyama T Hayashi S Kanzaki N Hashimoto S Shibanuma N Kurosaka M
Full Access

Background

The cementless acetabular component fixed with several screws is one of the most widely used approaches in THA. These screws rely on contact pressure and the resultant friction between the screw head and the cup to control translation and angulation of the prosthesis. However, intraoperative change of the acetabular component alignment during screw fixation should be hardly detected. Acetabular component alignment can be assessed using computer-assisted navigation systems with realtime adjustments for component position. The purpose of the current study was to evaluate intraoperative change of acetabular component alignment during screw fixation using navigation system.

Patients and Methods

Primary THAs were performed in 74 hips using CT based fluoroscopic matching navigation system (VectorVision, BrainLAB). The patients were 18 men and 56 women with a mean age of 64.4 years (range, 47–78 years) at operation. Intraoperative acetabular component inclination and anteversion were measured at the time of press-fit, and after screw fixation using the cup verification function in the system. Mean of the absolute difference between at the time of press-fit and after screw fixation was evaluated as intraoperative change of acetabular component. We measured the distance from the center of the femoral head to the inter-teardrop line as a horizontal and vertical reference on the postoperative radiograph. The number of screws was also investigated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 54 - 54
1 Sep 2012
Fujishiro T Nishiyama T Hayashi S Kanzaki N Takebe K Kurosaka M
Full Access

Background

Total hip arthroplasty for Crowe type IV developmental dysplasia of the hip is a technically demanding procedure. Restoration of the anatomical hip center frequently requires limb lengthening in excess of 4 cm and increases the risk of neurologic traction injury. However, it can be difficult to predict potential leg length change, especially in total hip arthroplasty for Crowe type IV developmental hip dysplasia. The purpose of the present study was to better define features that might aid in the preoperative prediction of leg length change in THAs with subtrochanteric femoral shortening osteotomy for Crowe type IV developmental dysplasia of the hip.

Patients and Methods

Primary total hip arthroplasties with subtrochanteric femoral shortening osteotomy were performed in 70 hips for the treatment of Crowe type IV developmental hip dysplasia. The patients were subdivided into two groups with or without iliofemoral osteoarthritis. Leg length change after surgery was measured radiographically by subtracting the amount of resection of the femur from the amount of distraction of the greater trochanter. Preoperative passive hip motion was retrospectively reviewed from medical records and defined as either higher or lower motion groups.