The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.Aims
Methods
A conventional arthroscopic capsuloligamentous repair is a reliable surgical solution in most patients with scapholunate instability. However, this repair does not seem to be sufficient for more advanced injuries. The aim of this study was to evaluate the functional results of a wide arthroscopic dorsal capsuloligamentous repair (WADCLR) in the management of severe scapholunate instability. This was a prospective single-centre study undertaken between March 2019 and May 2021. The primary outcome was the evaluation of the reduction of the radiological deformity and the functional outcomes after WADCLR. A secondary outcome was the evaluation of the effectiveness of this technique in patients with the most severe instability (European Wrist Arthroscopy Society (EWAS) stage 5). The patients were reviewed postoperatively at three, six, and 12 months.Aims
Methods
As many as 25% to 40% of unicompartmental knee
replacement (UKR) revisions are performed for pain, a possible cause
of which is proximal tibial strain. The aim of this study was to
examine the effect of UKR implant design and material on cortical
and cancellous proximal tibial strain in a synthetic bone model.
Composite Sawbone tibiae were implanted with cemented UKR components
of different designs, either all-polyethylene or metal-backed. The tibiae
were subsequently loaded in 500 N increments to 2500 N, unloading
between increments. Cortical surface strain was measured using a
digital image correlation technique. Cancellous damage was measured
using acoustic emission, an engineering technique that detects sonic
waves (‘hits’) produced when damage occurs in material. Anteromedial cortical surface strain showed significant differences
between implants at 1500 N and 2500 N in the proximal 10 mm only
(p <
0.001), with relative strain shielding in metal-backed implants.
Acoustic emission showed significant differences in cancellous bone
damage between implants at all loads (p = 0.001). All-polyethylene implants
displayed 16.6 times the total number of cumulative acoustic emission
hits as controls. All-polyethylene implants also displayed more
hits than controls at all loads (p <
0.001), more than metal-backed
implants at loads ≥ 1500 N (p <
0.001), and greater acoustic
emission activity on unloading than controls (p = 0.01), reflecting
a lack of implant stiffness. All-polyethylene implants were associated
with a significant increase in damage at the microscopic level compared
with metal-backed implants, even at low loads. All-polyethylene
implants should be used with caution in patients who are likely
to impose large loads across their knee joint. Cite this article:
Between 1990 and 2001, 24 children aged between 15 months and 11 years presented with late orthopaedic sequelae after meningococcal septicaemia. The median time to presentation was 32 months (12 to 119) after the acute phase of the disease. The reasons for referral included angular deformity, limb-length discrepancy, joint contracture and problems with prosthetic fitting. Angular deformity with or without limb-length discrepancy was the most common presentation. Partial growth arrest was the cause of the angular deformity. Multiple growth-plate involvement occurred in 14 children. The lower limbs were affected much more often than the upper. Twenty-three children underwent operations for realignment of the mechanical axis and limb-length equalisation. In 15 patients with angular deformity around the knee the deformity recurred. As a result we recommend performing a realignment procedure with epiphysiodesis of the remaining growth plate when correcting angular deformities.
We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants. This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.