This was a randomised controlled trial studying
the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised
that it would have similar early clinical results and micromovement
to those of a standard-length tapered Synergy metaphyseal fixation
stem. Using radiostereometric analysis (RSA) we compared the two
stems in 43 patients. A short metaphyseal fixation stem was used
in 22 patients and a Synergy stem in 21 patients. No difference
was found in the clinical outcomes pre- or post-operatively between
groups. RSA showed no significant differences two years post-operatively
in mean micromovement between the two stems (except for varus/valgus tilt
at p = 0.05) (subsidence 0.94 mm (. sd. 1.71) vs 0.32
mm (. sd. 0.45), p = 0.66; rotation 0.96° (. sd. 1.49) vs 1.41°
(. sd. 2.95), p = 0.88; and total migration 1.09 mm (. sd. 1.74) vs 0.73
mm (. sd. 0.72), p = 0.51). A few stems (four SMF and three Synergy)
had initial migration >
1.0 mm but stabilised by three to six months,
with the exception of one SMF stem which required revision three
years post-operatively. For most stems, total micromovement was
very low at two years (subsidence <
0.5 mm, rotation <
1.0°,
total migration <
0.5 mm), which was consistent with osseous ingrowth.
The small sample makes it difficult to confirm the universal applicability
of or elucidate the potential contraindications to the use of this
particular new design of stem. Cite this article: Bone Joint J 2015; 97-B:595–602