Relative motion at the modular head-neck junction of hip prostheses can lead to severe surface damage through mechanically-assisted corrosion. One factor affecting the mechanical performance of modular junctions is the frictional resistance of the mating surfaces to relative motion. Low friction increasing forces normal to the head-neck interface, leading to a lower threshold for slipping during weight-bearing. Conversely, a high friction coefficient is expected to limit interface stresses but may also allow uncoupling of the interface in service. This study was performed to examine this trade-off using finite element models of the modular head-neck junction A finite element model (FEM) of the trunnion/ head assembly of a total hip prosthesis was initially created and experimentally validated. CAD models of a stem trunnion (taper size: 12/14mm) and a prosthetic femoral head (diameter: 28mm) were discretized into elements for finite element analysis (FEA). The trunnion (Ti6Al4V) was modelled with a hexahedral mesh (33,648 elements) and the femoral head (CoCrMo) with a tetrahedral mesh (51,182 elements). A friction-based sliding contact interface was defined between the mating surfaces. The model was loaded in 2 stages: (i) an assembly load of 4000N applied along the trunnion axis, and (ii) 500N applied along the trunnion axis in combination with a torque of 10Nm. A linear static solution was set up using Siemens NX-Nastran solver. Multiple simulations were executed by modulating the frictional coefficient at the taper-bore interface from 0.05 to 0.15 in increments of 0.01, the coefficient of 0.1 serving as the control case (Swaminathan and Gilbert, 2012).Introduction
Methods
Current implant designs and materials provide a high grade of quality and safety, but aseptic implant loosening is still the main reason for total hip revision. Highly cross-linked polyethylene (HX-PE) is used successfully in total hip replacements (THR) since several years. The good wear properties lead to a reduction of wear debris and may contribute to a longer survival time of the THRs. Furthermore, thin HX-PE liner allows the use of larger femoral heads associated with a decreased risk of dislocation and an improved range of motion. However, the cross-linking process is associated with a loss of mechanical properties of the polyethylene material which compromise the use of thin HX-PE liner in terms of high stress situations. The aim of the present study was the experimental wear analysis of HX-PE liner under steep acetabular cup position. Furthermore, a finite element analysis (FEA) was performed in order to calculate the stress within the HX-PE material in case of steep cup position under physiological loading. Experimental wear testing was performed for 5 Mio load cycles, using highly cross-linked polyethylene (HX-PE) acetabular liner combined with 44 mm ceramic femoral heads at a standard position of the acetabular cup (30° inclination) according to ISO 14242 as well as at 60° cup inclination. The wall thickness of the HX-PE liner was 3.8 mm. A hip wear simulator, according to ISO 14242 (EndoLab GmbH, Rosenheim, Germany), was used and wear was determined gravimetrically. Moreover, finite element models of the THR system at standard and steep cup position was created by Abaqus/CAE (Dessault Systemes Providence, USA). Using the finite element software Abaqus (Dessault Systemes Providence, USA) the total hip implants were physiologically loaded with maximum force of the gait cycle (3.0 kN). Thereby, the stresses within the HX-PE material were analysed. The average gravimetrical wear rates of the HX-PE liners at standard implant position (30°) and 60° cup inclination showed small wear amounts of 3.15 ± 0.32 mg and 1.92 ± 1.00 mg per million cycles, respectively. The FEA revealed a clear increase of stresses at the HX-PE liner with respect to steep cup position (von Mises stress of 8.78 MPa) compared to ISO standard implant position (von Mises stress of 5.70 MPa). The wear simulator tests could not demonstrate significant differences of gravimetrical wear amount of HX-PE liners under steep hip cup position compared to standard implant position. The small contact surface between the femoral head and the SX-PE liner during the wear testing may lead to the low wear rate of the misaligned acetabluar cup. Moreover, the FEA showed that the effect of a misaligned acetabular cup on the stresses within the polyethylene liner can be critical. Although an increase of wear could not be detected a steeper acetabular cup position using thin HX-PE liners should be avoided due to higher stresses preventing implant failure in clinical application.
Background. Surgical reconstruction of the anterior cruciate ligament is a common practice to treat the disability or chronic instability of the knee. Several factors associated with success or failure of the ACL reconstruction, including surgical technique and graft material and graft tension. We aimed to show how we can optimize the graft properties and achieve better post surgical outcomes during ACL reconstruction using 3-dimensional computational
Experimental knee simulators for component evaluation or An existing finite element model of the KKS was modified to extend the capability, and improve the fidelity, of the computational model beyond the experimental setup. An actuator to allow anterior-posterior (A-P) motion of the hip was included and used to prescribe relative hip-ankle A-P kinematics during the simulations. The quadriceps muscle, which in the experimental simulator consisted of a single quadriceps bundle with a point-to-point line of action, was divided into four heads of the quadriceps with physiological muscle paths. The hamstrings muscle, which was not present in the experiment, was represented by point-to-point actuators in four bundles. A flexible control system was developed which allowed control of the quadriceps and hamstrings actuators to match a knee flexion profile, similar to actuation of the experimental KKS, but also allowed control of the compressive tibiofemoral (TF) joint force, medial-lateral (M-L) load distribution, internal-external (I-E) torque and A-P load at the joint. A series of sensors, measuring all six load components on the medial and lateral compartments of the tibial insert, as well as knee flexion angle, were incorporated into the simulation. Instantaneous measurements from the sensors were fed to a control system, implemented within an Abaqus/Explicit user subroutine (Figure 1). The controller was used to drive actuators in the FE model to match target
The wear particles released from the polyethylene (PE) tibial insert of modular total knee replacements (TKRs) have been shown to cause wear particle induced osteolysis, which may necessitate revision surgery [1]. Wear occurs at the backside surface of the PE insert of modular TKRs, resulting from the relative movement between the PE insert and the tibial tray [2]. Wear particles generated from the backside surface of the PE insert have been shown to be smaller in size than those originating from the articular surface [1], and may therefore have increased biological activity and osteolytic potential [3-4]. The ability to predict backside micromotion and contact pressure by
Surgical education of fracture fixation biomechanics relies mainly on simplified illustrations to distill the essence of the underlying principles. These mostly consist of textbook drawings or hands-on exercises during courses, both with unique advantages such as broad availability and haptics, respectively. Computer simulations are suited to bridge these two approaches; however, the validity of such simulations must be guaranteed to teach the correct aspects. Therefore, the aim of this study was to validate
Osteosynthesis aims to maintain fracture reduction until bone healing occurs, which is not achieved in case of mechanical fixation failure. One form of failure is plastic plate bending due to overloading, occurring in up to 17% of midshaft fracture cases and often necessitating reoperation. This study aimed to replicate in-vivo conditions in a cadaveric experiment and to validate a
Introduction. Achieving an appropriate primary stability after implantation is a prerequisite for the long-term viability of a dental implant. Virtual testing of the bone-implant construct can be performed with
Despite past advances of implant technologies, complication rates of fixations remain high at challenging sites such as the proximal humerus [1]. These may not only be owed to the implant itself but also to dissatisfactory surgical execution of fracture reduction and implant positioning. Therefore, the aim of this study was to quantify the instrumentation accuracy of a highly standardised and guided procedure and its influence on the biomechanical outcome and predicted failure risk. Preoperative planning of osteotomies creating an unstable 3-part fracture and fixation with a locking plate was performed based on CT scans of eight pairs of low-density proximal humerus samples from elderly female donors (85.2±5.4 years). 3D-printed subject-specific guides were used to osteotomise and instrument the samples according to the pre-OP plan. Instrumentation accuracies in terms of screw lengths and orientations were evaluated by comparing post-OP CT scans with the pre-OP plan. The fixation constructs were biomechanically tested until cyclic cut-out failure [2]. Failure risks of the planned and the post-OP configurations were predicted using a validated sample-specific
Introduction. Metallic resurfacing systems have been widely used until pseudotumors and ALTR have been clinically found and related to excessive wear of these metal-on-metal hip systems. Hence, surgeons widely abandoned the use of resurfacing systems. Meanwhile, there is a ceramic on ceramic (CoC) resurfacing system (Embody, London, UK) made of zirconia toughened alumina (BIOLOX. ®. delta, CeramTec, Plochingen, Germany) in a clinical safety study. Even though conventional CoC hip systems are known for their excellent wear behavior, it has to be ensured that intraoperative and in-vivo deformations of the ceramic acetabular cup do not infringe the proper functionality of the system. The method of determining the minimum clearance of such a system will be presented here. Materials and Methods. Combined experimental and numerical results were used to determine the deformation of the ceramic shell. In a cadaver lab, the resulting deformations after impaction of generic metal shells have been measured, see e.g. [1] for the method of measurement. The maximum deformation has been chosen for further calculation. Additionally, the stiffness of both generic metal and ceramic shells has been measured using ISO 7206–12. The deformation of the ceramic shells were then calculated by the equation. where u. c. and u. m. are the deformations of the ceramic and the metal shell, respectively, and K. m. and K. c. are the respective stiffnesses. Additionally, in a
Background. Osteoporotic fracture fixation in the proximal humerus remains a critical challenge. While the biomechanical benefits of screw augmentation with bone cement are established, minimising the cement volume may help control any risk of extravasation and reduce surgical procedure time. Previous experimental studies suggest that it may be sufficient to only augment the screws at the sites of the lowest bone quality. However, adequately testing this hypothesis in vitro is not feasible. Methods. This study systematically evaluated the 64 possible strategies for augmenting six screws in the humeral head through
Introduction. Pedicle screw loosening in posterior instrumentation of thoracolumbar spine occurs up to 60% in osteoporotic patients. These complications may be alleviated using more flexible implant materials and novel designs that could be optimized with reliable computational modeling. This study aimed to develop and validate non-linear homogenized
Introduction. In total hip arthroplasty, press-fit anchorage is one of the most common fixation methods for acetabular cups and mostly ensures sufficient primary stability. Nevertheless, implants may fail due to aseptic loosening over time, especially when the surrounding bone is affected by stress-shielding. The use of acetabular cups made of isoelastic materials might help to avoid stress-shielding and osteolysis. The aim of the present numerical study was to determine whether a modular acetabular cup with a shell made of polyetheretherketone (PEEK) may be an alternative to conventional titanium shells (Ti6Al4V). For this purpose, a 3D finite element analysis was performed, in which the implantation of modular acetabular cups into an artificial bone stock using shells made of either PEEK or Ti6Al4V, was simulated with respect to stresses and deformations within the implants. Methods. The implantation of a modular cup, consisting of a shell made of PEEK or Ti6Al4V and an insert made of either ceramic or polyethylene (PE), into a bone cavity made of polyurethane foam (20 pcf), was analysed by 3D
Morsellised cortico-cancellous bone (MCB) is used extensively in impaction grafting procedures, such as the filling of cavitory defects on the femoral and acetabular sides during hip arthroplasty. Several experimental studies have attempted to describe the mechanical behaviour of MCB in compression and shear, and it has been found that it’s properties can be improved by washing and rigorous impaction at the time of surgery. However their focus has not been on the development of constitutive models that can be used in computational simulation. The results of serial confined compaction tests are presented and used to develop constitutive models describing the non-linear elasto-plastic behaviour of MCB, as well as its time dependent visco-elastic behaviour. It is found that the elastic modulus, E of MCB increases linearly with applied pressure, p, with E achieving a value of around 30 MPa at a pressure of around 1 MPa. The plastic behaviour of MCB can be described using a Drucker Prager Cap yield criterion, capable of describing yielding of the graft in shear and compression. The time dependent visco-elastic behaviour of MCB can be accurately modelled using a spring and dashpot model that can be numerically expressed using a fourth order Prony series. The role of impaction in reducing subsequent plastic deformation was also investigated. The developed relationships allow the constitutive modelling of MCB in
There is a critical need for safe innovation in total joint replacements to address the demands of an ageing yet increasingly active population. The development of robust implant designs requires consideration of uncertainties including patient related factors such as bone morphology but also activity related loads and the variability in the surgical procedure itself. Here we present an integrated framework considering these sources of variability and its application to assess the performance of the femoral component of a total hip replacement (THR). The framework offers four key features. To consider variability in bone properties, an automated workflow for establishing statistical shape and intensity models (SSIM) was developed. Here, the inherent relationship between shape and bone density is captured and new meshes of the target bone structures are generated with specific morphology and density distributions. The second key feature is a virtual implantation capability including implant positioning, and bone resection. Implant positioning is performed using automatically identified bone features and flexibly defined rules reflecting surgical variability. Bone resection is performed according to manufacturer guidelines. Virtual implantation then occurs through Boolean operations to remove bone elements contained within the implant's volume. The third feature is the automatic application of loads at muscle attachment points or on the joint contact surfaces defined on the SSIM. The magnitude and orientation of the forces are derived from models of similar morphology for a range of activities from a database of musculoskeletal (MS) loads. The connection to this MS loading model allows the intricate link between morphology and muscle forces to be captured. Importantly, this model of the internal forces provides access to the spectrum of loading conditions across a patient population rather than just typical or average values. The final feature is an environment that allows
Total Hip Replacement (THR) is one of the most successful operations in all of medicine, however surgeons just rely on their experience and expertise when choosing between cemented or cementless stem, without having any quantitative guidelines. The aim of this project is to provide clinicians with some tools to support in their decision making. A novel method based on bone mineral density (BMD) measurements and assessments was developed 1) to estimate the periprosthetic fracture risk (FR) while press-fitting cementless stem; 2) to evaluate post-operative bone remodeling in terms of BMD changes after primary THR. Data for 5 out of over 70 patients (already involved in a previous study. 1. ) that underwent primary THA in Iceland were selected for developing novel methods to assess intra-operative FR and bone mineral density (BMD) changes after the operation. For each patient three CT images were acquired (Philips Brilliance 64 Spiral-CT, 120 kVp, slice thickness: 1 mm, slice increment: 0.5 mm): pre-op, 24 hours and 1 year post-operative. Pre-op CT scan was used to create 3D finite element model (Materialise Mimics) of the proximal femur. The material properties were assigned based on Hounsfield Units. Different strategies were analyzed for simulating the press-fitting operation, developing what has already been done in prior study. 1. In the
Introduction: Mechanical complications following lumbar fixation are due to the combination of various factors related to morphology, pathology, and surgery. The aim of this study was to provide a patient-specific Finite Element Model of the lumbar spine for the simulation of surgical strategies, and to use it as a predictive tool aiming to detect and reduce preoperatively the risks of mechanical complications. Materials &
Methods: A pre-existing 3D personalized FEM of the lumbar spine was used. Posterior implants and main degenerative pathologies were also modelled. After in vitro validation based on 24 specimens and 4 different instrumentations, the model was used to simulate real cases. Applied loads were based on patient characteristics (weight, imbalance). Simulation results included mechanical stresses in the discs and within the implants. Clinical consistency of the simulations was tested through the gathering of clinical data for 66 patients instrumented with lumbo-sacral rigid screw-rod systems. Two subsets were considered: “mechanical successes” (53), and “mechanical failures” (13, including 11 screw breakage and 2 screw loosening). Blind comparison was then performed between these observed clinical outcomes and numerical simulations results. Results &
Discussion: Among the 66 patients, simulation results highlighted specific behaviours for 9 patients for which mechanical loads on implants were significantly higher. All of these 9 patients were actual “mechanical failures”. None of the actual “mechanical successes” were associated with “abnormal” simulation results. Conclusion: This is the first time
Introduction. As population grows older, and patients receive primary joint replacements at younger age, more and more patients receive a total hip prosthesis nowadays. Ten-year failure rates of revision hip replacements are estimated at 25.6%. The acetabular component is involved in over 58% of those failures. From the second revision on, the pelvic bone stock is significantly reduced and any standard device proves inadequate in the long term [Villanueva et al. 2008]. To deal with these challenges, a custom approach could prove valuable [Deboer et al. 2007]. Materials and methods. A new and innovative CT-based methodology allows creating a biomechanically justified and defect-filling personalized implant for acetabular revision surgery [Figure 1]. Bone defects are filled with patient-specific porous structures, while thin porous layers at the implant-bone interface facilitate long-term fixation. Pre-operative planning of screw positions and lengths according to patient-specific bone quality allow for optimal fixation and accurate transfer to surgery using jigs. Implant cup orientation is anatomically analyzed for required inclination and anteversion angles. The implant is patient-specifically analyzed for mechanical integrity and interaction with the bone based upon fully individualized muscle modeling and
In recent years 3D preoperative planning has become increasingly popular with orthopaedic surgeons. One technique that has shown to be successful in transferring this preoperative plan to the operating room is based on surgical templates that guide various surgical instruments. Such a patient-specific template is designed using both the 3D reconstructed anatomy and the preoperative plan and is then typically produced via additive manufacturing technology. The combination of a preoperative plan and a surgical template has the potential to result in a more accurate procedure than an unguided one, when the following three criteria are met: the template needs to achieve a stable fit on the surgical field, it needs to fit in a unique position, and the surgeon needs to be able to determine the correct, planned position during the surgery. When the template fails one of these conditions, it can be used incorrectly. Consequently the process could result in an inaccurate outcome. This research focuses on modelling the stability of a surgical template on bone. The relationship between the contact surface of the template and the resulting stability is investigated with a focus on methods to quantify the template stability. The model calculates a quality score on the designed contact surface, which reflects the likelihood of positioning the template on the bone in a stable position. The model used in this study has been experimentally validated to verify its ability to provide a reliable indication of the template stability. This was analysed using finite element analysis where multiple templates and support models with different contact surface shapes were created. The application of forces and moments in varying directions was simulated. Stability is then defined as the ability of a template to resist an applied force or moment. The displacements of the templates were computed and analysed. The results show a minimal displacement of less than 0.01 mm and a maximal displacement larger than 10 mm. The former is considered to be a very stable template design; the latter to be very unstable and hence, would result in an insecure contact. The geometry of the contact surface had a clear influence on the template stability. Overall, the coverage of curvature variations improved the stability of the template. The displacements of the different
In biomechanical