Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 29 - 29
19 Aug 2024
Kayani B Konan S Tahmassebi J Giebaly D Haddad FS
Full Access

The direct superior approach (DSA) is a modification of the posterior approach (PA) that preserves the iliotibial band and short external rotators except for the piriformis or conjoined tendon during total hip arthroplasty (THA). The objective of this study was to compare postoperative pain, early functional rehabilitation, functional outcomes, implant positioning, implant migration, and complications in patients undergoing the DSA versus PA for THA. This study included 80 patients with symptomatic hip arthritis undergoing primary THA. Patients were prospectively randomised to receive either the DSA or PA for THA, surgery was undertaken using identical implant designs in both groups, and all patients received a standardized postoperative rehabilitation programme. Predefined study outcomes were recorded by blinded observers at regular intervals for two-years after THA. Radiosteriometric analysis (RSA) was used to assess implant migration. There were no statistical differences between the DSA and PA in postoperative pain scores (p=0.312), opiate analgesia consumption (p=0.067), and time to hospital discharge (p=0.416). At two years follow-up, both groups had comparable Oxford hip scores (p=0.476); Harris hip scores (p=0.293); Hip disability and osteoarthritis outcome scores (p=0.543); University of California at Los Angeles scores (p=0.609); Western Ontario and McMaster Universities Arthritis Index (p=0.833); and European Quality of Life questionnaire with 5 dimensions scores (p=0.418). Radiographic analysis revealed no difference between the two treatment groups for overall accuracy of acetabular cup positioning (p=0.687) and femoral stem alignment (p=0.564). RSA revealed no difference in femoral component migration (p=0.145) between the groups at two years follow-up. There were no differences between patients undergoing the DSA versus PA for THA with respect to postoperative pain scores, functional rehabilitation, patient-reported outcome measurements, accuracy of implant positioning, and implant migration at two years follow-up. Both treatment groups had excellent outcomes that remained comparable at all follow-up intervals


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 105 - 105
10 Feb 2023
Xu J Veltman W Chai Y Walter W
Full Access

Navigation in total hip arthroplasty has been shown to improve acetabular positioning and can decrease the incidence of mal-positioned acetabular components. The aim of this study was to assess two surgical guidance systems by comparing intra-operative measurements of acetabular component inclination and anteversion with a post-operative CT scan. We prospectively collected intra-operative navigation data from 102 hips receiving conventional THA or hip resurfacing arthroplasty through either a direct anterior or posterior approach. Two guidance systems were used simultaneously: an inertial navigation system (INS) and optical navigation system (ONS). Acetabular component anteversion and inclination was measured on a post-operative CT. The average age of the patients was 64 years (range: 24-92) and average BMI was 27 kg/m. 2. (range 19-38). 52% had hip surgery through an anterior approach. 98% of the INS measurements and 88% of the ONS measurements were within 10° of the CT measurements. The mean (and standard deviation) of the absolute difference between the post-operative CT and the intra-operative measurements for inclination and anteversion were 3.0° (2.8) and 4.5° (3.2) respectively for the ONS, along with 2.1° (2.3) and 2.4° (2.1) respectively for the INS. There was significantly lower mean absolute difference to CT for the INS when compared to ONS in both anteversion (p<0.001) and inclination (p=0.02). Both types of navigation produced reliable and reproducible acetabular cup positioning. It is important that patient-specific planning and navigation are used together to ensure that surgeons are targeting the optimal acetabular cup position. This assistance with cup positioning can provide benefits over free-hand techniques, especially in patients with an altered acetabular structure or extensive acetabular bone loss. In conclusion, both ONS and INS allowed for adequate acetabular positioning as measured on postoperative CT, and thus provide reliable intraoperative feedback for optimal acetabular component placement


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 34 - 34
10 Feb 2023
Farey J Chai Y Xu J Sadeghpour A Marsden-Jones D Baker N Vigdorchik J Walter W
Full Access

Imageless computer navigation systems have the potential to improve acetabular cup position in total hip arthroplasty (THA), thereby reducing the risk of revision surgery. This study aimed to evaluate the accuracy of three alternate registration planes in the supine surgical position generated using imageless navigation for patients undergoing THA via the direct anterior approach (DAA). Fifty-one participants who underwent a primary THA for osteoarthritis were assessed in the supine position using both optical and inertial sensor imageless navigation systems. Three registration planes were recorded: the anterior pelvic plane (APP) method, the anterior superior iliac spines (ASIS) functional method, and the Table Tilt (TT) functional method. Post-operative acetabular cup position was assessed using CT scans and converted to radiographic inclination and anteversion. Two repeated measures analysis of variance (ANOVA) and Bland-Altman plots were used to assess errors and agreement of the final cup position. For inclination, the mean absolute error was lower using the TT functional method (2.4°±1.7°) than the ASIS functional method (2.8°±1.7°, ρ = .17), and the ASIS anatomic method (3.7°±2.1, ρ < .001). For anteversion, the mean absolute error was significantly lower for the TT functional method (2.4°±1.8°) than the ASIS functional method (3.9°±3.2°, ρ = .005), and the ASIS anatomic method (9.1°±6.2°, ρ < .001). All measurements were within ± 10° for the TT method, but not the ASIS functional or APP methods. A functional registration plane is preferable to an anatomic reference plane to measure intra-operative acetabular cup inclination and anteversion accurately. Accuracy may be further improved by registering patient location using their position on the operating table rather than anatomic landmarks, particularly if a tighter target window of ± 5° is desired


Abstract. Optimal acetabular component position in Total Hip Arthroplasty is vital for avoiding complications such as dislocation and impingement, Transverse acetabular ligament (TAL) have been shown to be a reliable landmark to guide optimum acetabular cup position. Reports of iliopsoas impingement caused by acetabular components exist. The Psoas fossa (PF) is not a well-regarded landmark for Component positioning. Our aim was to assess the relationship of the TAL and PF in relation to Acetabular Component positioning. A total of 22 cadavers were implanted on 4 occasions with the an uncemented acetabular component. Measurements were taken between the inner edge of TAL and the base of the acetabular component and the distance between the lower end of the PF and the most medial end of TAL. The distance between the edge of the acetabular component and TAL was a mean of 1.6cm (range 1.4–18cm). The distance between the medial end of TAL and the lowest part of PF was a mean of 1.cm (range 1,3–1.8cm) It was evident that the edge of PF was not aligned with TAL. Optimal acetabular component position is vital to the longevity and outcome following THA. TAL provides a landmark to guide acetabular component position. However we feel the PF is a better landmark to allow appropriate positioning of the acetabular component inside edge of the acetabulum inside the bone without exposure of the component rim and thus preventing iliopsoas impingement at the psoas notch and resultant groin pain


Bone & Joint Open
Vol. 3, Issue 6 | Pages 475 - 484
13 Jun 2022
Jang SJ Vigdorchik JM Windsor EW Schwarzkopf R Mayman DJ Sculco PK

Aims

Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error.

Methods

A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (Δsacral slope(SS)stand-sit > 30°), or stiff (SSstand-sit < 10°) spinopelvic mobility contributed to increased error rates.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 9 - 9
1 Nov 2021
Farey J Chai Y Xu J Sadegpour A Jones DM Baker N Vigdorchik J Walter W
Full Access

Imageless computer navigation systems in total hip arthroplasty (THA) improve acetabular cup position, thereby reducing the risk of revision surgery for all causes as well as dislocation. We aimed to evaluate the registration accuracy of 3 alternate registration planes. A prospective, observational study was conducted with 45 THA in the supine position using two imageless navigation systems and 3 registration planes. Patient position was registered sequentially using an optical system (Stryker OrthoMap) and an inertial sensor-based system (Navbit Sprint) with 3 planes of reference: (Plane 1) an anatomical plane using the anterior superior iliac spines (ASISs) and the pubic symphysis; (Plane 2) a functional plane parallel to the line between the ASISs and the table plane; and, (Plane 3) a functional plane that was perpendicular to the gravity vector and aligned with the longitudinal axis of the patient. The 3 measurements of acetabular cup inclination and anteversion were compared with the measurements from postoperative computed tomography (CT) scans. For inclination, the mean absolute error was significantly lower for Plane 3 (1.80°) than for Plane 2 (2.74°), p = .038 and was lower for both functional planes than for the anatomical plane (3.75°), p < .001. For anteversion, the mean absolute error was significantly lower for Plane 3 (2.00°) than for Plane 2 (3.69°), p = .004 and was lower for both functional planes than for the anatomical plane (8.58°), p < .001. Patient registration using functional planes more accurately measured the acetabular cup position than registration using anatomic planes


Abstract. Background. Optimal acetabular component position in Total Hip Arthroplasty is vital for avoiding complications such as dislocation, impingement, abductor muscle strength and range of motion. Transverse acetabular ligament (TAL) and posterior labrum have been shown to be a reliable landmark to guide optimum acetabular cup position. There have been reports of iliopsoas impingement caused by both cemented and uncemented acetabular components. Acetabular component mal-positioning and oversizing of acetabular component are associated with iliopsoas impingement. The Psoas fossa (PF) is not a well-regarded landmark to help with Acetabular Component positioning. Our aim was to assess the relationship of the TAL and PF in relation to Acetabular Component positioning. Methods. A total of 12 cadavers were implanted with the an uncemented acetabular component, their position was initially aligned to TAL. Following optimal seating of the acetabular component the distance of the rim of the shell from the PF was noted. The Acetabular component was then repositioned inside the PF to prevent exposure of the rim of the Acetabular component. This study was performed at Smith & Nephew wet lab in Watford. Results. Out of the twelve acetabular components that were implanted parallel to the TAL, all had the acetabular rim very close or outside to the psoas notch with a potential to cause iliopsoas impingement. Alteration of the acetabular component position was necessary in all cadavers to inside the PF to prevent iliopsoas impingement. It was evident that the edge of PF was not aligned with TAL. Conclusion. Optimal acetabular component position is vital to the longevity and outcome following THA. TAL provides a landmark to guide acetabular component position. We feel the PF is a better landmark to allow appropriate positioning of the acetabular component inside bone without exposure of the component rim and thus preventing iliopsoas impingement at the psoas notch. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 29 - 29
1 Feb 2020
Abe I Shirai C
Full Access

Background. Accurate acetabular cup positioning is considered to be essential to prevent postoperative dislocation and improve the long-term outcome of total hip arthroplasty (THA). Recently various devices such as navigation systems and patient-specific guides have been used to ensure the accuracy of acetabular cup positioning. Objectives. The present study evaluated the usefulness of CT-based three-dimensional THA preoperative planning for acetabular cup positioning. Methods. This study included 120 hips aged mean 68.3 years, who underwent primary THA using CT-based THA preoperative planning software ZedHip® (LEXI, Tokyo Japan) and postoperative CT imaging (Fig.1). The surgical approach adopted the modified Watson-Jones approach in the lateral decubitus position and Trident HA acetabular cups were used for all cases. Preoperatively the optimum cup size and position in the acetabular were decided using the ZedHip® software, taking into consideration femoral anteversion and to achieve the maximum range of motion in dynamic motion simulation. Radiographic inclination (RI) was selected in the range between 40°∼45° and radiographic anteversion (RA) in the range between 5°∼25°. Three-dimensional planning images of the cup positioning were obtained from the ZedHip® software, and the distances between the edge of the implant and anatomical landmarks such as the edge of the anterior or superior acetabular wall were measured on the three-dimensional images and recorded (Fig.2). Intraoperatively, the RI and RA were confirmed by reference to these distances and the acetabular cup was inserted. Relative positional information of the implant was extracted from postoperative CT imaging using the ZedHip® software and used to reproduce the position of the implant on preoperative CT imaging with the software image matching function. The difference between the preoperative planning and the actual implant position was measured to assess the accuracy of acetabular cup positioning using the ZedHip® software. Results. Actual cup size corresponded with that of preoperative planning in 95% of cases (114 hips). Postoperative mean RI was 42.3° ± 4.2° (95% confidence interval (CI), 41.5° ∼ 43.0°) and mean RA was 16.1° ± 5.9° (95%CI, 15.0° ∼ 17.1°). Deviation from the target RI was 4.2° ± 3.7° (95%CI, 3.5° ∼ 4.9°) and deviation from the target RA was 4.0° ± 3.6° (95%CI, 3.4° ∼ 4.7°). Overall 116 hips (96.7%) were within the RI safe zone (30° ∼ 50°) and 108 hips (90.0%) were within the RA safe zone (5° ∼ 25°), and 105 hips (87.5%) were within both the RI and RA safe zones (Fig.3). Mean cup shift from preoperative planning was 0.0mm ± 3.0mm to the cranial side in the cranio-caudal direction, 2.1mm ± 3.0mm to the anterior side in the antero-posterior direction, and 1.7mm ± 2.1mm to the lateral side in the medio-lateral direction. Conclusion. The accuracy of acetabular cup positioning using our method of CT-based three-dimensional THA preoperative planning was slightly inferior to reported values for CT-based navigation, but obviously superior to those without navigation and similar to those with portable navigation. CT-based three-dimensional THA preoperative planning is effective for acetabular cup positioning, and has better cost performance than expensive CT-based navigation. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 93 - 93
1 Feb 2020
Cipparone N Robinson M Chen J Muir J Shah R
Full Access

Acetabular cup positioning remains a real challenge and component malpositioning after total hip arthroplasty (THA) can lead to increased rates of dislocation and wear. It is a common cause for revision THA. A novel 3D imageless mini-optical navigation system was used during THA to provide accurate, intraoperative, real-time, and non-fluoroscopic data including component positioning to the surgeon. This retrospective comparative single surgeon and single approach study examined acetabular component positioning between traditional mini-posterolateral THA and mini-posterolateral THA using the 3D mini-optical navigation system. A retrospective chart review was conducted of 157 consecutive (78 3D mini-optical navigation and 79 traditional non-navigation methods) THAs performed by the senior author using a mini-posterolateral approach at an ambulatory surgery center and hospital setting. Two independent reviewers analyzed postoperative radiographs in a standardized fashion to measure acetabular component positioning. Demographic, clinical, surgical, and radiographic data were analyzed. These groups were found to have no statistical difference in age, gender, and BMI (Table I). There was no difference between groups in acetabular components in the Lewinnek safe zone, 31.2% vs 26.6% (p = 0.53). Cup anteversion within the safe zone did not differ, 35.1% vs 40.5% (p = 0.48); while cup inclination within the safe zone differed, with more in the navigation group, 77.9% vs 51.9% (p < 0.01). Change in leg length was significantly different with the navigation group's leg length at 1.9 ± 6.3, less than the traditional at 5.4 ± 7.0 (p < 0.01). There was no difference in mean change in offset between groups (4.5 ± 5.9 vs 6.2 ± 7.9, p = 0.12); navigation, traditional) (Table II). The 3D mini-optical navigation group did have significantly longer operative time (98.4 ± 17.5 vs 89.3 ± 15.5 p < 0.01). Use of the novel 3D Mini-optical Navigation System significantly improved cup inclination compared to traditional methods while increasing operative time. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 44 - 44
1 Feb 2020
Mays R Benson J Muir J White P Meftah M
Full Access

Proper positioning of the acetabular cup deters dislocation after total hip arthroplasty (THA) and is therefore a key focus for orthopedic surgeons. The concept of a safe zone for acetabular component placement was first characterized by Lewinnek et al. and furthered by Callanan et al. The safe zone concept remains widely utilized and accepted in contemporary THA practice; however, components positioned in this safe zone still dislocate. This study sought to characterize current mass trends in cup position identified across a large study sample of THA procedures completed by multiple surgeons. This retrospective, observational study reviewed acetabular cup position in 1,236 patients who underwent THA using computer-assisted navigation (CAS) between July 2015 and November 2017. Outcomes included acetabular cup position (inclination and anteversion) measurements derived from the surgical navigation device and surgical approach. The overall mean cup position of all recorded cases was 21.8° (±7.7°, 95% CI = 6.7°, 36.9°) of anteversion and 40.9° (±6.5°, 95% CI = 28.1°, 53.7°) of inclination (Table 1). For both anteversion and inclination, 65.5% (809/1236) of acetabular cup components were within the Lewinnek safe zone and 58.4% (722/1236) were within the Callanan safe zone. Acetabular cups were placed a mean of 6.8° of anteversion (posterior/lateral approach: 7.0°, anterior approach: 5.6°) higher than the Lewinnek and Callanan safe zones whereas inclination was positioned 0.9° higher than the reported Lewinnek safe zone and 3.4° higher than the Callanan safe zone (Figure 1,2). Our data shows that while the majority of acetabular cups were placed within the traditional safe zones, the mean anteversion orientation is considerably higher than those suggested by the Lewinnek and Callanan safe zones. The implications of this observation warrant further investigation. For any figures or tables, please contact the authors directly


Introduction. Robotic-assisted hip arthroplasty helps acetabular preparation and implantation with the assistance of a robotic arm. A computed tomography (CT)-based navigation system is also helpful for acetabular preparation and implantation, however, there is no report to compare these methods. The purpose of this study is to compare the acetabular cup position between the assistance of the robotic arm and the CT-based navigation system in total hip arthroplasty for patients with osteoarthritis secondary to developmental dysplasia of the hip. Methods. We studied 31 hips of 28 patients who underwent the robotic-assisted hip arthroplasty (MAKO group) between August 2018 and March 2019 and 119 hips of 112 patients who received THA under CT-based navigation (CT-navi group) between September 2015 and November 2018. The preoperative diagnosis of all patients was osteoarthritis secondary to developmental dysplasia of the hip. They received the same cementless cup (Trident, Stryker). Robotic-assisted hip arthroplasty were performed by four surgeons while THA under CT-based navigation were performed by single senior surgeon. Target angle was 40 degree of radiological cup inclination (RI) and 15 degree of radiological cup anteversion (RA) in all patients. Propensity score matching was used to match the patients by gender, age, weight, height, BMI, and surgical approach in the two groups and 30 patients in each group were included in this study. Postoperative cup position was assessed using postoperative anterior-posterior pelvic radiograph by the Lewinnek's methods. The differences between target and postoperative cup position were investigated. Results. The acetabular cup position of all cases in both Mako and CT-navi group within Lewinnek's safe zone (RI: 40±10 degree; RA: 15±10 degree) in group were within this zone. Three was no significant difference of RI between Mako and CT-navi group (40.0 ± 2.1 degree vs 39.7± 3.6 degree). RA was 15.0 ± 1.2 degree and 17.0 ± 1.9 degree in MAKO group and in CT-navi group, respectively, with significant difference (p<0.001). The differences of RA between target and postoperative angle were smaller in MAKO group than CT-navi group (0.60± 1.05 degree vs 2.34± 1.40 degree, p<0.001). The difference or RI in MAKO group was smaller than in CT-navi, however, there was no significance between them (1.67± 1.27 degree vs 2.39± 2.68 degree, p=0.197). Conclusions. Both the assistance of the robotic arm and the CT-based navigation system were helpful to achieve the acetabular cup implantation, however, MAKO system achieved more accurate acetabular cup implantation than CT-based navigation system in total hip arthroplasty for the patients with OA secondary to DDH. Longer follow-up is necessary to investigate the clinical outcome


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_8 | Pages 96 - 96
1 May 2019
Su E
Full Access

Acetabular implant position is important for the stability, function, and long-term wear properties of a total hip arthroplasty (THA). Prior studies of acetabular implant positioning have demonstrated a high percentage of outliers, even in experienced hip surgeons, when conventional instruments are used. Computer navigation is an attractive tool for use in (THA, as it has been shown to improve the precision of acetabular component placement and reduce the incidence of outliers. However, computer navigation with imageless, large-console systems is costly and often interrupts the surgeon's workflow, and thus, has not been widely adopted. Another method to improve acetabular component positioning during THA is the use of fluoroscopy with the direct anterior approach. Studies have demonstrated that the supine position of the patient during surgery facilitates the use of fluoroscopic guidance, thus improving acetabular component position. A handheld, accelerometer based navigation unit for use in total hip replacement has recently become available to assist the surgeon in positioning the acetabular component during anterior approach THA, potentially reducing the need for intraoperative fluoroscopic studies. We sought to compare the radiographic results of direct anterior THA performed with conventional instrumentation vs. handheld navigation to determine the accuracy of the navigation unit, and to see whether or not there was a reduction in the fluoroscopic time used during surgery. Furthermore, we timed the use of the navigation unit to see whether or not it required a substantial addition to surgical time. Our results demonstrate that a handheld navigation unit used during anterior approach THA had no difference with regard to acetabular cup positioning when compared to fluoroscopically assisted THA, but led to a reduction in the use of intraoperative fluoroscopy time


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 138 - 138
1 Apr 2019
Harold R Delagrammaticas D Stover M Manning DW
Full Access

Background. Supine positioning during direct anterior approach total hip arthroplasty (DAA THA) facilitates use of fluoroscopy, which has been shown to improve acetabular component positioning on plane radiograph. This study aims to compare 2- dimensional intraoperative radiographic measurements of acetabular component position with RadLink to postoperative 3- dimensional SterEOS measurements. Methods. Intraoperative fluoroscopy and RadLink (El Segundo, CA) were used to measure acetabular cup position intraoperatively in 48 patients undergoing DAA THA. Cup position was measured on 6-week postoperative standing EOS images using 3D SterEOS software and compared to RadLink findings using Student's t-test. Safe-zone outliers were identified. We evaluated for measurement difference of > +/− 5 degrees. Results. RadLink acetabular cup abduction measurement (mean 43.0°) was not significantly different than 3D SterEOS in the anatomic plane (mean 42.6°, p = 0.50) or in the functional plane (mean 42.7°, p = 0.61) (Fig. 1–2). RadLink acetabular cup anteversion measurement (mean 17.9°) was significantly different than 3D SterEOS in both the anatomic plane (mean 20.6°, p = 0.022) and the functional plane (mean 21.2°, p = 0.002) (Fig. 3–4). RadLink identified two cups outside of the safe-zone. However, SterEOS identified 12 (anatomic plane) and 10 (functional plane) outside of the safe-zone (Fig. 5–7). In the functional plane, 58% of anteversion and 92% of abduction RadLink measurements were within +/− 5° of 3D SterEOS. Conclusion. Intraoperative fluoroscopic RadLink acetabular anteversion measurements are significantly different than 3D SterEOS measurements, while abduction measurements are similar. Significantly more acetabular cups are placed outside of the safe- zone when evaluated with 3D SterEOS versus RadLink


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 41 - 41
1 Oct 2018
Tatka J Brady AW Matta JM
Full Access

Introduction

Accurate acetabular position is an important goal during THA. It is also well known that accurate acetabular positioning is very frequently not achieved, even by experienced, high volume surgeons. Problems associated with cup malposition are: dislocation, accelerated poly wear, impingement, ceramic squeaking, metalosis. Murray et al described 3 methods of measurement and assessment of acetabular inclination and anteversion (I&A): anatomic, radiographic and operative. It is the hypothesis of the authors, that the differences and details of these 3 methods are poorly understood by many surgeons and this is contributory to inconsistent cup positioning. Additionally, the radiographic method, which is most commonly used for post op assessment and academic studies, contributes to misunderstanding and error. Modern computer guidance and software assessment of radiographs allows us to easily measure anatomic I&A which should be thought of as “true” I&A.

Methods

The mathematical criteria for radiographic measurement of anatomic I&A are defined as well as the mathematical relationships and discrepancies between anatomic and radiographic I&A for any given cup.

A = A n g l e o f a n t e v e r s i o n o f c u p I = A n g l e o f i n c l i n a t i o n o f c u p

E = Angle of ellipse major diameter to horizontal

E = Radiographic inclination

Sin A = H o r i z o n t a l w i d t h o f e l l i p s e L e n g t h o f e l l i p s e m a j o r d i a m e t e r Sin I = V e r t i c a l h e i g h t o f e l l i p s e L e n g t h o f e l l i p s e m a j o r d i a m e t e r

Tan I = Tan E / Cos A

Tan E = (Tan I) x (Cos A)


Robotic assisted surgery aims to reduce surgical errors in implant positioning and better restore native hip biomechanics compared to conventional techniques for total hip arthroplasty (THA). The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual THA versus robotic-arm assisted THA. Secondary objectives were to determine differences between these treatment techniques for THA in achieving the planned combined offset, cup inclination, cup version, and leg-length correction. This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. All operative procedures were undertaken by a single surgeon using the minimally-invasive posterior approach. Two independent blinded observers recoded all radiological outcomes of interest using plain radiographs. Patients in both treatment groups were well-matched for age, gender, body mass index, laterality of surgery, and ASA scores. Interclass correlation coefficient was 0.92 (95% CI: 0.84 – 0.95) for intra-observer agreement and 0.88 (95% CI: 0.82–0.94) for inter-observer agreement in all study outcomes. Robotic THA was associated with improved accuracy in restoring the native horizontal (p<0.001) and vertical (p<0.001) centres of rotation, and improved preservation of the patient's native combined offset (P<0.001) compared to conventional THA. Robotic THA improved accuracy in positioning of the acetabular cup within the combined safe zones of inclination and anteversion described by Lewinnek et al (p=0.02) and Callanan et al (p=0.01) compared to conventional THA (figures 1–2). There was no difference between the two treatment groups in achieving the planned leg-length correction (p=0.10). Robotic-arm assisted THA was associated with improved accuracy in restoring the native centre of rotation, better preservation of the combined offset, and more precise acetabular cup positioning within the safe zones of inclination and anteversion compared to conventional manual THA. Robotic-arm assisted THA enables improved preservation of native hip biomechanics compared to conventional manual THA. For any figures or tables, please contact authors directly: . fsh@fareshaddad.net


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_10 | Pages 55 - 55
1 Jun 2018
Su E
Full Access

Acetabular implant position is important for the stability, function, and long-term wear properties of a total hip arthroplasty (THA). Prior studies of acetabular implant positioning have demonstrated a high percentage of outliers, even in experienced hip surgeons, when conventional instruments are used. Computer navigation is an attractive tool for use in THA, as it has been shown to improve the precision of acetabular component placement and reduce the incidence of outliers. However, computer navigation with imageless, large-console systems is costly and often interrupts the surgeon's workflow, and thus, has not been widely adopted. Another method to improve acetabular component positioning during THA is the use of fluoroscopy with the direct anterior approach. Studies have demonstrated that the supine position of the patient during surgery facilitates the use of fluoroscopic guidance, thus improving acetabular component position. A handheld, accelerometer-based navigation unit for use in total hip replacement has recently become available to assist the surgeon in positioning the acetabular component during anterior approach THA, potentially reducing the need for intra-operative fluoroscopic studies. We sought to compare the radiographic results of direct anterior THA performed with conventional instrumentation vs. handheld navigation to determine the accuracy of the navigation unit, and to see whether or not there was a reduction in the fluoroscopic time used during surgery. Furthermore, we timed the use of the navigation unit to see whether or not it required a substantial addition to surgical time. Our results demonstrate that a handheld navigation unit used during anterior approach THA had no difference with regard to acetabular cup positioning when compared to fluoroscopically assisted THA, but led to a reduction in the use of intra-operative fluoroscopy time


Bone & Joint 360
Vol. 7, Issue 2 | Pages 12 - 15
1 Apr 2018


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 63 - 63
1 Apr 2018
Shon W Han S Lee D Kim H
Full Access

Background. Recent clinical studies have suggested that systemic metal ion levels are significantly elevated at midterm follow-up after ceramic-on-metal (COM) bearing. However, it is not clear whether there is a correlation between patient- and surgical-related factors including the lifestyle and elevated levels of serum metal ions following COM total hip arthroplsty (THA). Material and Methods. Two hundred and one patients (234 hips) including 121 COM patients (140 hips) and 80 non-COM patients (94 hips) were enrolled in accordance with the inclusion criteria. The patients were divided into three groups based on the type of surgical bearings used. The Harris Hip Score (HHS), University of California, Los Angeles (UCLA) activity scale score, and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score were measured, and radiographs were obtained for the analysis. Serum metal ion levels of cobalt and chromium were measured using a high-resolution inductively coupled plasma mass spectrometry. Patient- and surgical-related factors were analyzed to determine which group of patients is at a high risk of metal ion-related problems. Results. Significantly higher serum levels of Co and Cr were detected in the serum of the COM THA group (Co: 1.86±4.0 µg/L, range: 0.30 to 34.20 µg/L; Cr: 1.81±2.87 µg/L, range: 0.10 to 27.80 µg/L) than in the serum of the non-COM THA group (Co: 0.27±0.14 µg/L, range: 0.15 to 0.90 µg/L; Cr: 0.19±0.25 µg/L, range; 0.10 to 2.30 µg/L) (p<0.001). The HHS in the COM group was significantly better than that in the non-COM group (p=0.013). The total ROM of the THAs was significantly greater in the 36-mm COM THA group (272.7°, range: 200°–345°) than in the non-COM group (248.5°, range: 135°–300°) (p<0.001). No radiolucency, osteolysis, or loosening was found during the follow-up radiographic examination. The serum Co levels of patients who achieved the squatting position were significantly higher than those of patients who could not squat (Co: p=0.033; Cr: p=0.074). The serum Co and Cr levels of patients who achieved the kneeling position were significantly higher than those of patients who could not kneel (Co: p=0.049; Cr: p=0.031). There was no significant difference between the two groups in the cross-legged sitting position. The metal ion levels of the COM THA group correlated with the total ROM (Co: p=0.0293; Cr: p=0.0399), and those of the patients who were capable of squatting and kneeling were significantly higher than those of the patients who were unable (p<0.05). However, age, BMI, acetabular cup position and patient activity did not show significant correlations with the serum metal ion levels. Conclusions. Patients who underwent a 36-mm COM THA had good clinical outcomes with an excellent hip function at the short and midterm follow-up intervals. However, high levels of metal ions were detected in the serum of COM THA patients. We found that COM THA patients who were capable of greater ROMs, squatting, and kneeling are at risk of metal ion-related problems


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 26 - 26
1 Jan 2018
MacDonald S Howard J Goyal P Yuan X Lanting B Teeter M Naudie D McCalden R
Full Access

Lewinnek's safe zone recommendation to minimise dislocations was a target of 5–25° for anteversion angle and 30–50° for inclination angle. Subsequently, it was demonstrated that mal-positioning of the acetabular cup can also lead to edge loading, liner fracture, and greater conventional polyethylene wear. The purpose of this study was to measure the effect of acetabular cup position on highly crosslinked polyethylene wear in total hip arthroplasty (THA) at long-term follow-up. We identified all patients that underwent primary THA with a minimum of 10 years follow-up using an institutional database in London, Ontario, Canada. Patients with a single implant design consisting of a 28 mm cobalt chromium head and highly crosslinked polyethylene liner (ram extruded, GUR 1050, 100 kGy gamma irradiated, remelted, ethylene oxide sterilised) were selected for inclusion. In total, 85 hips from 79 recruited patients were analysed. Patients underwent a supine radiostereometric analysis (RSA) exam in which the x-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup anteversion angle, inclination angle, and 3D penetration rate (including wear and creep) were measured from the stereo radiograph pairs. At a mean follow-up of 13 years (range, 10–17 years) the mean penetration rate was 0.059 mm/year (95% CI: 0.045 to 0.073 mm/year). Mean anteversion angle was 18.2° (range, −14 to 40°) and mean inclination angle was 43.6° (range, 27 to 61°). With respect to the Lewinnek safe zone, 67% hips met the target for anteversion angle, 77% met the target for inclination angle, and 51% met the target for both. There was no correlation between anteversion angle and penetration rate (r = −0.14, p = 0.72) or between inclination angle and penetration rate (r = 0.11, p = 0.35). There was also no difference (p = 0.07) in penetration rate between hips located within the Lewinnek safe zone for both anteversion angle and inclination angle (mean 0.057 mm/year, 95% CI: 0.036 to 0.079 mm/year) and those outside the safe zone (mean 0.062 mm/year, 95% CI: 0.042 to 0.083 mm/year). Acetabular cup position had no effect on the wear rate of highly crosslinked polyethylene at long-term follow-up. Although care should still be taken to correctly position the acetabular cup for stability, highly crosslinked polyethylene is a forgiving bearing material that can withstand a wide range of cup positions without negatively impacting longevity due to wear


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_15 | Pages 111 - 111
1 Aug 2017
Su E
Full Access

Acetabular implant position is important for the stability, function, and long-term wear properties of a total hip arthroplasty (THA). Prior studies of acetabular implant positioning have demonstrated a high percentage of outliers, even for experienced hip surgeons, when conventional instruments are used. Computer navigation is an attractive tool for use in THA, as it has been shown to improve the precision of acetabular component placement and reduce the incidence of outliers. However, computer navigation with imageless, large-console systems is costly and often interrupts the surgeon's workflow, and thus has not been widely adopted. Another method to improve acetabular component positioning during THA is the use of fluoroscopy with the direct anterior approach. Studies have demonstrated that the supine position of the patient during surgery facilitates the use of fluoroscopic guidance, thus improving acetabular component position. A handheld, accelerometer based navigation unit for use in total hip replacement has recently become available to assist the surgeon in positioning the acetabular component during anterior approach THA, potentially reducing the need for intra-operative fluoroscopic studies. We sought to compare the radiographic results of direct anterior THA performed with conventional instrumentation vs. handheld navigation to determine the accuracy of the navigation unit, and to see whether or not there was a reduction in the fluoroscopic time used during surgery. Furthermore, we timed the use of the navigation unit to see whether or not it required a substantial addition to surgical time. Our results demonstrate that a handheld navigation unit used during anterior approach THA had no difference with regard to acetabular cup positioning when compared to fluoroscopically assisted THA, but led to a reduction in the use of intra-operative fluoroscopy time