The liner design is a key determinant of the constraint of a reverse total shoulder arthroplasty (rTSA). The aim of this study was to compare the degree of constraint of rTSA liners between different implant systems. An implant company’s independent 3D shoulder arthroplasty planning software (mediCAD 3D shoulder v. 7.0, module v. 2.1.84.173.43) was used to determine the jump height of standard and constrained liners of different sizes (radius of curvature) of all available companies. The obtained parameters were used to calculate the stability ratio (degree of constraint) and angle of coverage (degree of glenosphere coverage by liner) of the different systems. Measurements were independently performed by two raters, and intraclass correlation coefficients were calculated to perform a reliability analysis. Additionally, measurements were compared with parameters provided by the companies themselves, when available, to ensure validity of the software-derived measurements.Aims
Methods
Dual mobility (DM) is an established bearing option in Total Hip Arthroplasty (THA). The traditional mono-block DM designs have limited ability for additional fixation, whereas the modular DM designs allow additional screw fixation but limit internal diameter and have the potential to generate metal debris. We report the early results of a
Total knee arthroplasty is one of the most common surgeries. About 92% of all implanted knee endorposthesis in 2020 were manufactured from uncoated
The implantation of endoprosthesis is a routine procedure in orthopaedics. Endoprosthesis are mainly manufactured from ceramics, polymers, metals or metal alloys. To ensure longevity of the implants they should be as biocompatible as possible and ideally have antibacterial properties, to avoid periprosthetic joint infections (PJI). Various antibacterial implant materials have been proposed, but have so far only been used sporadically in patients. PJI is one of the main risk factors for revision surgeries. The aim of the study was to identify novel implant coatings that both exhibit antibacterial properties whilst having optimal biocompatibility. Six different novel implant coatings and surface modifications (EBM TiAl6V4, strontium, TiCuN, TiNbN, gentamicin phosphate (GP), gentamicin phosphate+cationic polymer (GP+CP)) were compared to standard CoCrMo-alloy. The coatings were further characterized with regard to the surface roughness. E. coli and S. capitis were cultured on the modified surfaces to investigate the antibacterial properties. To quantify bacterial proliferation the optical density (OD) was measured and viability was determined using colony forming units (CFU). Murine bone marrow derived macrophages (BMMs) were cultured on the surfaces and differentiated into osteoblasts to quantify the mineralisation using the alizarin red assay. All novel coatings showed reduced bacterial proliferation and viability compared to standard CoCrMo-alloy. A significant reduction was observed for GP and GP+CP coated samples compared to
Abstract. Aims. Ceramic coatings in total knee arthroplasty have been introduced with the aim of reducing wear and consequently improving implant survivorship. We studied both cobalt-chrome-molybdenum and ceramic-coated components of the same implant design from a single centre to identify if the ceramic coating conferred any benefit at mid-term review. Patients and Methods. We identified 1641 Columbus TKAs (Aesculap AG, Tüttlingen, Germany) from a prospectively collected arthroplasty database. 983 were traditional
Acetabular edge-loading was a cause of increased wear rates in metal-on-metal hip arthroplasties, ultimately contributing to their failure. Although such wear patterns have been regularly reported in retrieval analyses, this study aimed to determine their in vivo location and investigate their relationship with acetabular component positioning. 3D CT imaging was combined with a recently validated method of mapping bearing surface wear in retrieved hip implants. The asymmetrical stabilizing fins of Birmingham hip replacements (BHRs) allowed the co-registration of their acetabular wear maps and their computational models, segmented from CT scans. The in vivo location of edge-wear was measured within a standardized coordinate system, defined using the anterior pelvic plane.Aims
Methods
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
Ti-6Al-4V is the most common alloy used for orthopaedic implants. Its popularity is due to low density, superior corrosion resistance, good osseointegration and lower elastic modulus when compared to other commonly used alloys such as
Introduction. Total hip replacement failure due to fretting-corrosion remains a clinical concern. We recently described that damage within
Little is known about the relationship between head-neck corrosion and its effect on the periprosthetic tissues and distant organs of patients hosting well-functioning devices. The purpose of this study was to investigate in postmortem retrieved specimens the degree and type of taper damage, and the corresponding histologic responses in periprosthetic tissues and distant organs. Fifty postmortem THRs (34 primaries, 16 revisions) retrieved after 0.5 to 26 years were analyzed. Forty-three implants had a
Aims. This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys. Methods. Murine peritoneal macrophages were cultured and placed on stainless steel,
Introduction. Metal ion and particle release, particularly cobalt, has become an important subject in total hip arthroplasty, as it has shown to induce metal hypersensitivity, adverse local tissue reactions and systemic ion related diseases. The purpose of the following study was compare the ion release barrier function of a zirconium nitride (ZrN) multilayer coated hip stem for cemented use, designed for patients with metal ion hypersensitivity, against its uncoated version in a test configuration simulating the worst case scenario of a severely debonded hip stem. The ZrN multilayer coating is applied on a
Introduction. Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure. There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference and may be influenced by assembly method. Objective. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability. Methods. A two-dimensional, axisymmetric model representative of a
Wear and corrosion debris generated from total hip replacements (THR) can cause adverse local tissue reactions (ALTR) or osteolysis, often leading to premature implant failure. The tissue response can be best characterized by histopathological analysis, which accurately determines the presence of cell types, but is limited in the characterization of biochemical changes (e.g. protein conformation alteration). Fourier transform infrared micro-spectroscopy imaging (FTIRI) enables rapid analysis of the chemical structure of biological tissue with a high spatial resolution, and minimal additional sample preparation. The data provides the most information through multivariate method carried out by hierarchical clustering analysis (HCA). It is the goal of this study to demonstrate the beneficial use of this multivariate approach in providing pathologist with biochemical information from cellular and subcellular organization within joint capsule tissue retrieved from THR patients. Joint capsule tissue from 2 retrieved THRs was studied. Case 1: a metal-on-polyethylene THR, and Case 2: a dual modular metal-on-metal THR. Prior to FTIRI analysis, tissue samples were formalin-fixed paraffin-embedded and 5μm thick microtome sectioned samples were prepared and mounted on BaF. 2. discs and deparaffinized. FTIRI data were collected using high-definition transmission mode (pixel size: ∼1.1 μm. 2. ). Hyperspectral images were exported to CytoSpec V2.0.06 for processing and reconstruction into pseudo-color maps based on cluster assignments. Case 1 exhibited a strong presence of lymphocytes and macrophages (Fig. 1a). Since the process of taking second derivatives reduces the half width of the spectral peaks, it increases the sensitivity toward detecting shoulders or second peaks that may not be apparent in the raw spectra (Fig. 1b). Thus, areas occupied by lymphocytes and macrophages can be easily distinguished providing a fast tissue screening method. Here, HCA was able to distinguish macrophages and lymphocytes based on the infrared response, even in areas where both occurred intermixed. (Fig. 1c) The tissue in direct proximity to cells had a slightly altered collagenous structure. Case 1 also exhibited multiple glassy, green particles which can typically observed around THRs that underwent taper corrosion (Fig. 2a). HCA image was able to visualize and distinguish large CrPO. 4. particles, embedded within fibrin exudate rich areas, collagenous tissue without inflammatory cells, and a nearby area with a strong macrophage presence and some finer CrPO. 4. particles (Fig. 2d). Moreover, this method can not only locate macrophages, but distinguish particle-laden macrophages depending the type of particles within the cells. In Case 2 (Fig. 3a), clustering results (Fig. 3 b&c) are consistent with the fact that different particle types are associated with MoM bearing surface wear (Co rich particles), corrosion of the
Introduction. Modular junctions in total hip replacement (THR) have been a primary source of fretting and corrosion which can lead to implant failure. Fretting is a result of unintended micromotion between the femoral head and stem tapers and is suspected to result after improper taper seating during assembly. Two design factors known to influence in-vitro taper assembly mechanics are relative taper alignment—mismatch angle—and the surface finish—micro-grooves. However, these factors have not been systematically evaluated together. Objective. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of taper mismatch angle and taper surface finish—smooth and rough—on the modular junction mechanics during assembly. Methods. A two-dimensional, axisymmetric model of a
Introduction. Metal alloys have been commonly used for surgical applications due to their suitable mechanical characteristics and relatively good biocompatibility. However, direct cellular corrosion of orthopaedic implants remains a controversial topic and is still not fully understood. This study aims to examine a possible aspect of this corrosion mechanism by determining if macrophages can attach and directly affect the surfaces of 316L stainless steel, Ti6Al4V, and
Introduction. Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure. 1. There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey. 2. found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference strength. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference strength and may be influenced by assembly method. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability. Methods. A two-dimensional, axisymmetric FEA model representative of a
Fretting corrosion of taper junctions is long known and of great concern, because of metal ion and particle release and their related adverse local and systemic effects on the human body (1–3). Orthopedic taper junctions are often comprised of CoCr29Mo6/TiAl6V4 pairings. Beside others the imprinting of the TiAlV-machining marks into the CoCrMo-taper is of clinical interest (4, 5). Thus, the multifactorial details and their interdependencies on the macro-, micro, and nanoscale are still a matter of research (6). This contribution presents the mechanisms of imprinting found in an in-vitro fretting corrosion test. The worn surfaces, the lubricant as well as its remains were analyzed after test and the findings brought into relation to the characteristic wear sub-mechanisms. The fretting tests were conducted by means of a cylinder-on-pin set-up. All details about the test and the sequence of analyses can be found in (7, 8). A marked tribofilm of C-rich organic matter and oxidized wear particles of both bodies was generated at the TiAlV/
Introduction. Corrosion at modular junctions of total hip replacements has been identified as a potential threat to implant longevity, resulting in efforts to determine appropriate countermeasures. Visual scoring and volumetric material loss measurements have been useful tools to elucidate various clinical and design factors associated with corrosion damage. However, corrosion involves electron exchange that results in chemical changes to biomedical alloys, and electrochemical assessment may therefore be a more appropriate approach to understand the phenomenon. The purpose of this pilot study was to electrochemically distinguish the severity of corrosion in retrieved femoral heads. A secondary goal was to identify the potential of electrochemical impedance spectroscopy (EIS) as a method to identify different forms of corrosion damage. Methods. Twenty femoral heads were identified from a larger study of total hip replacements, obtained as part of an ongoing multi- center IRB-approved retrieval program. Using a previously established 4-point scoring method, components were binned by taper damage: 10 components were identified as having severe damage, 7 with moderate damage and 2 with mild damage. One (1) unimplanted control was included to represent minimal corrosion damage. All components were then characterized using electrochemical impedance spectroscopy under the frequency domain: a 10 mV sinusoidal voltage, ranging from 20 kHz to 2 mHz, was applied to the taper of a femoral head (working electrode) filled with a 1M solution of PBS, a platinum counter electrode and a chlorided silver reference electrode. Absolute impedance at 2 mHz (|Z. 0.002. |), and max phase angle (θ) were assessed relative to taper damage severity. After least-squares fitting of the EIS data to a Randles circuit with a constant phase element, circuit elements: polarization resistance (Rp), CPE-capacitance, and CPE-exponent were also evaluated. The seven (7) most severely corroded components were further examined with scanning electron microscopy to identify corrosion modes. For all statistical analyses, significance was determined at alpha=0.05. Results. Taper damage was strongly correlated with both |Z. 0.002. | (ρ = −0.857, p<0.001) and CPE-capacitance (ρ=0.913, p<0.001). Taper damage was moderately associated with max phase angle (ρ= −0.483, p=0.031), CPE-exponent (ρ= −0.653, p=0.002) and Rp (ρ=0.556, p=0.011). Log-log plots of the strongest predictors of taper damage (|Z. 0.002. | and CPE- capacitance) identified some clustering among severely corroded components. SEM analysis identified evidence of grain/phase boundary corrosion on four components, all with log CPE-capacitance ≥ −4.4. Discussion. The results of this pilot study highlight that electrochemical impedance spectroscopy is useful in determining corrosion severity in retrieved femoral heads, and may also identify intergranular corrosion attack. For an undamaged taper, the self- passivating behavior of
Recent clinical data suggest improvement in the fixation of tibia trays for total knee arthroplasty when the trays are additive manufactured with highly porous bone ingrowth structures. Currently, press-fit TKA is less common than press-fit THA. This is partly because the loads on the relatively flat, porous, bony apposition area of a tibial tray are more demanding than those same porous materials surrounding a hip stem. Even the most advanced additive manufactured (AM) highly porous structures have bone ingrowth limitations clinically as aseptic loosening still remains more common in press-fit TKA vs. THA implants. Osseointegration and antibacterial properties have been shown in vitro and in vivo to improve when implants have modified surfaces that have biomimetic nanostructures designed to mimic and interact with biological structures on the nano-scale. Pre-clinical evaluations show that TiO. 2. nanotubes (TNT), produced by anodization, on Ti6Al4V surfaces positively enhance the rate at which osseointegration occurs and TNT nano-texturization enhances the antibacterial properties of the implant surface. 2. In this in vivo sheep study, identical Direct Metal laser Sintered (DMLS) highly porous Ti6Al4V specimens with and without TNT surface treatment are compared to sintered bead specimens with plasma sprayed hydroxyapatite-coated surface treatment. Identical DMLS specimens made from