This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
Decreased head-neck ratio diameter and component malposition in total hip arthroplasty are factors known to result in impingement, increased rates of dislocation, wear and failure. In addition to these complications, impingement of the femoral neck on the acetabular component of a hip resurfacing may result in femoral neck fracture and loosening of the acetabular component. Little is known regarding the optimum femoral and acetabular hip resurfacing component position to avoid impingement. In the first part of this study we analysed the radiographic component position of 131 consecutive hip resurfacings. In the second part the effect of three component variables on the range of motion to impingement were analysed using a dry bone model:. Inclination of the acetabular cup. Version of the acetabular cup. Femoral head-neck diameter ratio. The mean