The Bone & Joint Journal
Vol. 107-B, Issue 1 | Pages 58 - 64
1 Jan 2025
Carender CN Bedard NA Fruth KM Taunton MJ Pagnano MW Abdel MP

Aims

The purpose of this study was to directly compare the Modular Dual Mobility (MDM) Mobile Bearing Hip System (Stryker, USA) and large femoral heads (LFHs) in revision total hip arthroplasties (THAs) at mid-term follow-up, with specific emphasis on survival free of re-revision for dislocation, any re-revision, dislocation, and the risk of metal-related complications.

Methods

We identified 299 revision THAs performed at a single tertiary care academic institution from March 2011 to July 2014. Aseptic loosening of the acetabular component (n = 65), dislocation (n = 58), and reimplantation as part of a two-stage exchange protocol (n = 57) were the most common reasons for index revision. MDM constructs were used in 123 cases, and LFHs were used in 176 cases. Mean age was 66 years (28 to 93), mean BMI was 31 kg/m2 (18 to 55), and 45% (n = 136) were female. Mean follow-up was seven years (2 to 12).


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 13 - 13
19 Aug 2024
DeBenedetti A Weintraub MT Valle CJD Jacobs JJ Nam D
Full Access

The purpose of this randomized controlled trial was to evaluate serum metal levels in patients undergoing total hip arthroplasty (THA) with a conventional compared to a modular dual-mobility bearing. Patients undergoing primary THA for osteoarthritis were randomized to receive either a modular dual-mobility or conventional polyethylene bearing. All patients received the same titanium acetabular and femoral component and a ceramic femoral head. Serum metal levels were drawn pre-operatively then annually for a minimum of two years postoperatively. An a priori power analysis determined that 40 patients (20 per cohort) were needed to identify a clinically relevant difference in serum cobalt of 0.35 ng/ml (ppb) at 90% power. Forty-six patients were randomized to a modular dual-mobility (n=25) or conventional bearing (n=21) with 40 at a minimum follow-up of two years. No differences in serum cobalt (mean 0.14 ppb [range, 0.075–0.29] vs. 0.20 ppb [range, 0.075–0.57], p=0.39) or chromium levels (mean 0.14 ppb [range, 0.05–0.50] vs. 0.12 ppb [range, 0.05–0.35], p=0.65) were identified between the modular dual-mobility and conventional cohorts, respectively. There was no statistically significant difference in serum Co or Cr at two years postoperatively in subjects implanted with a ceramic head and this particular dual mobility bearing in comparison to a ceramic head and a conventional acetabular component. While modest expected elevations in serum Co and Cr were observed in the dual mobility group, in no case did the Co level exceed the laboratory reference range nor the threshold of one part per billion that has been associated with adverse local tissue reactions to mechanically-assisted crevice corrosion


Traditional mechanical debridement can only remove visibly infected tissue and is unable to completely clear all the biofilm that hides within muscle crevices and nerves. This study aims to determine the results of single-stage revision using noncontact low frequency ultrasonic debridement in treating chronic periprosthetic joint infections (PJI). A prospective study of consecutive patients requiring single-stage revision for chronic PJI was performed since August 2021. After mechanical debridement, an 8‑mm handheld non‑contact low‑frequency ultrasound probe was used for ultrasonic debridement at a frequency of (25±5) kHz and power of 90% for 5 minutes. Each ultrasound lasted 10 seconds with 3‑seconds intervals. The probe was repeatedly sonicated among all soft tissue and bsingle interface. The distal femoral canal and the posterior capsule of the knee were fully sonicated with a special right‑angle probe. Chemical debridement was then performed to irrigation the whole operative area. Recurrence of infection, culture results and number of colonies 24 hours after ultrasonic debridement were recorded. A total of 45 patients (25 hips and 20 knees) were included and 43 of them (95.6%) were free of infection at a mean follow-up time of 29 months (24 to 33). There were no intraoperative complications related to ultrasonic debridement (neurovascular and muscle injury, poor wound healing and fat liquefaction). The culture‑positive rate of wound liquid before ultrasonic debridement was 40.0% (18/45), which significantly increased to 75.6% (34/45) after ultrasonic debridement (P=0.001). The median number of colonies 24 hours after ultrasonic debridement was 2372 CFU/ml (310 to 4340 CFU/ml), which was significantly higher than that before debridement (307 CFU/ml; 10 to 980 CFU/ml) (P=0.000). Single-stage revision with non‑contact low‑frequency ultrasonic debridement can fully expose bacteria within biofilm, increase the efficacy of chemical debridement and lead to a favorable short‑term outcome without related complications


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 89 - 96
1 Mar 2024
Heckmann ND Chung BC Liu KC Chen XT Lovro LR Kistler NM White E Christ AB Longjohn DB Oakes DA Lieberman JR

Aims

Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.

Methods

A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 41 - 41
23 Jun 2023
Hernigou P
Full Access

The purpose was to determine the lifetime risk of re-operation due to specific complications related to dual mobility using re-operation as a competing risk, excluding loosening, periprosthetic fracture, and infection. 1503 mono-block dual mobility total hip arthroplasty (DM-THAs). Defining the re-operation when anesthesia (for dislocation) and revision when the implant changed. Surgery (801 for primary, 702 for revision with 201 for recurrent dislocation and 501 for loosening) performed between 1990 and 2020 in average 81-year-old (range 50–102) patients, with 522 living patients at 10 years follow-up. During the first month, outer dislocation (60 cases; 4%) was the cause re-operation (1% among primary and 6 % among revisions). Twenty-four intra-prosthetic dislocations (IPD) were an iatrogenic consequence of a failed closed reduction (reduction maneuver dissociating the inner head) with 1.6% revision. Between 1 month and 1 year, 22 new outer dislocations, while 25 of the 60 “first month” dislocations had recurrent dislocation. Fifteen other IPDs as iatrogenic consequences were observed. At one year, the cumulative revision was 3% (49 of 82 dislocations). Between 1- 10-year FU, 132 other dislocations, and 45 other revisions for dislocations were observed. Corrosion was another cause of revision (37 cases): between the cobalt-chromium shell and the femoral neck (23 hips), or 14 crevice corrosion between the trunnion and the metal head (trunnion damage). In summary, at 10-year: dislocation first cause of re-operation (214 anesthesia, 14%), while among 131 revisions (8.9 %) the 55 iatrogenic intra-prosthetic dislocations were the first revision cause before 39 recurrent dislocations and 37 corrosions. The 522 patients followed ten years or more had a 15% risk revision due to DM specific complications during their lifetime and 10% more risk associated with loosening (6%), periprosthetic fracture (2%) and infection (2%)


Bone & Joint Research
Vol. 12, Issue 3 | Pages 155 - 164
1 Mar 2023
McCarty CP Nazif MA Sangiorgio SN Ebramzadeh E Park S

Aims

Taper corrosion has been widely reported to be problematic for modular total hip arthroplasty implants. A simple and systematic method to evaluate taper damage with sufficient resolution is needed. We introduce a semiquantitative grading system for modular femoral tapers to characterize taper corrosion damage.

Methods

After examining a unique collection of retrieved cobalt-chromium (CoCr) taper sleeves (n = 465) using the widely-used Goldberg system, we developed an expanded six-point visual grading system intended to characterize the severity, visible material loss, and absence of direct component contact due to corrosion. Female taper sleeve damage was evaluated by three blinded observers using the Goldberg scoring system and the expanded system. A subset (n = 85) was then re-evaluated following destructive cleaning, using both scoring systems. Material loss for this subset was quantified using metrology and correlated with both scoring systems.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims

Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors.

Methods

Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 41 - 41
1 Apr 2022
Holleyman R Petheram T Reed M Burton P Malviya A
Full Access

Modular femoral stems offer surgeons great flexibility in biomechanical configuration during total hip replacement (THR) however introduce a taper-trunnion articulation known to be a source of additional wear debris through crevice, fretting and galvanic corrosion with mixed material combinations. This study aimed to investigate the influence of the trunnion bearing surface combination on the revision rate following primary total hip replacement (THR). All patients who underwent THR using an Exeter V40 cemented stainless steel stem and monobloc cemented polyethylene acetabular component (uncemented cups excluded to standardise the acetabular bearing surface and fixation) between January 2003 and December 2019 were extracted from the National Joint Registry for England, Wales, Northern Ireland, and the Isle of Man. The primary exposure was the head substrate used corresponding to the trunnion bearing. Time-to-event was determined by duration of implantation from primary surgery to revision with cases censored at death or end of available follow-up. Multivariable Cox proportional hazard models were used to identify predictors of all cause revision, adjusted for age, sex, American Association of Anaesthesiologists (ASA) grade, body mass index, surgical indication (osteoarthritis or other), and femoral head size. 229,870 THR were identified (66% female, mean age 73.4 years (SD 9.1) with the majority (91%) performed for osteoarthritis of which 4,598 were revised. Mean time from primary to revision or censoring was 6.8 years (SD 4.0). Multivariable modelling showed CoCr/SS trunnions were associated with a significantly higher risk of revision (hazard ratio (HR) 1.31 (95%CI 1.15 to 1.48, p<0.0001) as compared to SS/SS (reference). Both Alumina/SS (HR 0.74 (0.65 to 0.84), p<0.0001) and Zirconia/SS (HR 0.61 (0.49 to 0.74), p<0.0001) were associated with a significantly lower risk of revision. Ceramic heads on an Exeter stem were associated with significantly improved survivorship compared to metal heads in primary THR. CoCr/SS trunnion articulations had the poorest survivorship which may be contributed to by trunnionosis


Bone & Joint 360
Vol. 11, Issue 2 | Pages 15 - 18
1 Apr 2022


Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims

This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.

Methods

In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.


Bone & Joint 360
Vol. 10, Issue 5 | Pages 15 - 18
1 Oct 2021


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims. This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported. Methods. This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed. Results. Adjacent to the nail’s telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail’s telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion. Conclusion. Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425–436


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims. Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. Methods. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy. Results. Fretting was present on 88% (53/60) of the retrieved liners, and corrosion was present on 97% (58/60). Fretting was most often found on the lip of the taper at the transition between the lip and the dome regions. Macrophages and particles reflecting an innate inflammatory reaction to corrosion debris were noted in six of the 48 cases for which periprosthetic tissues were examined, and all were associated with retrieved components that had high corrosion scores. Conclusion. Our results show that corrosion occurs at the interface between MDM liners and shells and that it can be associated with reactions in the local tissues, suggesting continued concern that this problem may become clinically important with longer-term use of these implants. Cite this article: Bone Joint J 2021;103-B(7):1238–1246


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims

Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA.

Methods

This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m2, and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims

The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components.

Methods

Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 33 - 40
1 Jul 2020
Gustafson JA Pourzal R Levine BR Jacobs JJ Lundberg HJ

Aims

The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions.

Methods

Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims

We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice.

Methods

Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims

This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber.

Methods

For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 116 - 121
1 Jul 2020
Heise G Black CM Smith R Morrow BR Mihalko WM

Aims

This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys.

Methods

Murine peritoneal macrophages were cultured and placed on stainless steel, CoCrMo, and Ti6Al4V discs into a 96-well plate. Cells were activated with interferon gamma and lipopolysaccharide. Macrophages on stainless steel discs produced significantly more nitric oxide (NO) compared to their control counterparts after eight to ten days and remained elevated for the duration of the experiment.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 138 - 144
1 Jun 2020
Heckmann ND Nahhas CR Yang J Della Valle CJ Yi PH Culvern CN Gerlinger TL Nam D

Aims

In patients with a “dry” aspiration during the investigation of prosthetic joint infection (PJI), saline lavage is commonly used to obtain a sample for analysis. The aim of this study was to investigate prospectively the impact of saline lavage on synovial fluid analysis in revision arthroplasty.

Methods

Patients undergoing revision hip (THA) or knee arthroplasty (TKA) for any septic or aseptic indication were enrolled. Intraoperatively, prior to arthrotomy, the maximum amount of fluid possible was aspirated to simulate a dry tap (pre-lavage) followed by the injection with 20 ml of normal saline and re-aspiration (post-lavage). Pre- and post-lavage synovial white blood cell (WBC) count, percent polymorphonuclear cells (%PMN), and cultures were compared.