A modular femoral head–neck junction has practical
advantages in total hip replacement. Taper fretting and corrosion
have so far been an infrequent cause of revision. The role of design
and manufacturing variables continues to be debated. Over the past
decade several changes in technology and clinical practice might
result in an increase in clinically significant taper fretting and
corrosion. Those factors include an increased usage of large diameter
(36 mm) heads, reduced femoral neck and taper dimensions, greater
variability in taper assembly with smaller incision surgery, and
higher taper stresses due to increased patient weight and/or physical
activity. Additional studies are needed to determine the role of
taper assembly compared with design, manufacturing and other implant
variables. Cite this article:
Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of
Aims. Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for
I believe ceramic-on-polyethylene should be used in all patients undergoing THA. I believe the issues that one must look at include wear and osteolysis, bearing fracture and in 2018 corrosion/adverse local tissue reaction (ALTR). If one looks at these aspects it is clear ceramic-on-polyethylene is the bearing of choice. In the literature, there is a paucity of studies comparing metal-on-polyethylene with ceramic-on-polyethylene total hips. The data suggests no real difference in survivorship but less wear. However, most studies are not comparative studies and are underpowered to see these differences. The only data that is powered to see these differences is registry data. In first decade survivorship of metal-on- polyethylene is similar to ceramic-on-polyethylene. However, in the second decade the advantage is clearly with ceramic-on-polyethylene. However, in 2018, the major issue is corrosion and subsequent
Background. Adverse local tissue reactions (ALTR) in metal-on-polyethylene (MoP) total hip arthroplasty (THA) with head-neck taper corrosion is likely to be multifactorial involving implant and patient factors. However, there is a paucity of clinical data on implant parameters as predisposing factors in MoP head-neck taper corrosion. The aim of this study was to identify any potential implant factors associated with failed MoP THA due to head-neck taper corrosion. Methods. A total of 67 MoP THA patients in two groups was investigated: 1)
High failure rates have been associated with large diameter metal-on-metal total hip replacements (MoM THR). However there is limited literature describing the outcomes following the revision of MoM THR for adverse local tissue reaction (ALTR). A total of 98 large diameter MoM THRs underwent revision for
The differential diagnosis of the painful total
hip arthroplasty (resurfacing or total hip) includes infection,
failure of fixation (loosening), tendinitis, bursitis, synovitis,
adverse local tissue reaction (ALTR) to cobalt-chromium alloys,
and non-hip issues, such as spinal disorders, hernia, gynecologic,
and other pelvic pain. Assuming that the hip is the source of pain,
the first level question is prosthetic or non-prosthetic pain generator?
The second level prosthetic question is septic or aseptic? The third
level question (aseptic hips) is well-fixed or loose?
Introduction. Modular dual-mobility constructs reduce the risk of dislocation after total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chrome (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the mid-term serum Co and Cr levels after modular dual-mobilities used in revision and complex primary THAs, and specifically included patients revised for
Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing total hip arthroplasty (THA) has been recognized as a cause of adverse local tissue reactions (ALTR). Serum metal levels have been advocated as a tool for the diagnosis of
Introduction. Failure of metal on metal (MOM) total hip arthroplasty (THA) and resurfacing arthroplasty (HRA) due to development of adverse local tissue reaction (ALTRs) is a significant problem. The prevalence of ALTRs in asymptomatic MOM arthroplasty patients is highly variable. The purpose of this prospective, longitudinal study was to: 1) determine MRI
Introduction:. To date, there have been few reports of the results of revisions for failed metal-on-metal hip arthroplasties (MoM HA's). These series have included relatively modest numbers, and classification of the severity of adverse local tissue reaction (ALTR) has been under-reported. In this study, early outcomes and complications are analyzed as a function of pre-operative MRI grade and intra-operative
Increased failure rates due to metallic wear particle-associated adverse local tissue reactions (ALTR) is a significant clinical problem in resurfacing and total hip arthroplasty. Histological analysis and particle characterization are important elements for understanding the biological mechanisms of the reaction and different histological subtypes may have unique needs for longitudinal clinical follow-up and complication rates after revision arthroplasty. Consecutive patients (N=285 cases) presenting with
INTRODUCTION. Corrosion of modular tapers is increasingly recognized as a source of adverse tissue reaction (ALTR) and revision surgery in total hip arthroplasty (THA). The incidence of corrosion and rate of revision for
Aims. Advances in surgical technique and implant design may influence the incidence and mechanism of failure resulting in revision total hip arthroplasty (rTHA). The purpose of the current study was to characterize aetiologies requiring rTHA, and to determine whether temporal changes existed in these aetiologies over a ten-year period. Methods. All rTHAs performed at a single institution from 2009 to 2019 were identified. Demographic information and mode of implant failure was obtained for all patients. Data for rTHA were stratified into two time periods to assess for temporal changes: 2009 to 2013, and 2014 to 2019. Operative reports, radiological imaging, and current procedural terminology (CPT) codes were cross-checked to ensure the accurate classification of revision aetiology for each patient. Results. In all, 2,924 patients with a mean age of 64.6 years (17 to 96) were identified. There were 1,563 (53.5%) female patients, and the majority of patients were Caucasian (n = 2,362, 80.8%). The three most frequent rTHA aetiologies were infection (27.2%), aseptic loosening (25.2%), and wear (15.2%). The frequency of rTHA for adverse local tissue reaction (ALTR) was significantly greater from 2014 to 2019 (4.7% vs 10.0%; p < 0.001), while the frequency of aseptic loosening was significantly greater from 2009 to 2013 (28.6% vs 21.9%; p < 0.001). Conclusion. Periprosthetic joint infection was the most common cause for rTHA in the current cohort of patients. Complications associated with
Introduction. Modular neck (MN) components in total hip arthroplasty (THA) offer intraoperative flexibility, but adverse local tissue reactions (ALTR) due to tribocorrosion at modular junctions are a potential complication of such designs. Serum ion levels and metal artifact reduction sequence (MARS) MRI are used to assess
Adverse local tissue reactions (ALTR), such as so-called pseudotumours associated with metal-metal bearings, can also occur secondary to corrosion products from modular tapers where at least one side is composed of cobalt alloy. In 1988, Svensson et al. reported a fulminant soft-tissue pseudotumour following a cementless, metal-on-polyethylene total hip. This case had all of the features of
Introduction. Taper corrosion at modular junctions can cause a spectrum of adverse local tissue reactions (ALTR) in the periprosthetic soft tissues in patients who have undergone total hip arthroplasty (THA). Because these reactions are usually painful, taper corrosion has become part of the differential diagnosis of hip pain following THA. However these destructive lesions may not always cause pain, and can occasionally result in other atypical presentations. The purpose of this study is to describe a cohort of patients presenting with late and recurrent instability following THA due to underlying
Introduction. Neck-stem corrosion has been associated with Adverse Local Tissue Reaction (ALTR) in dual-taper femoral stems. Several diagnostic tests, of varying specificity and sensitivity, are used to identify
Introduction. Complication and revision rates have shown to be high for all metal-on-metal (MoM) bearings, especially for the ASR Hip System (ASR hip resurfacing arthroplasty (HRA) and ASR XL total hip arthroplasty (THA)). This prompted the global recall of the ASR Hip System in 2010. Many studies have previously explored the association between female gender and revision surgery MoM HRA implants; yet less research has been dedicated to exploring this relationship in MoM THA. The first purpose of this study was to assess the associations between gender and implant survival, as well as adverse local tissue reaction (ALTR), in patients with MoM THA. Secondly, we sought to report the differences between genders in metal ion levels and patient reported outcome measures (PROMs) in patients with MoM THA. Methods. The study population consisted of 729 ASR XL THA patients (820 hips) enrolled from September 2012 to June 2015 in a multicenter follow-up study at a mean of 6.4 (3–11) years from index surgery. The mean age at the time of index surgery was 60 (22–95) years and 338 were women (46%). All patients enrolled had complete patient and surgical demographic information, blood metal ion levels and PROMs obtained within 6 months, and a valid AP pelvis radiograph dating a maximum of 2 years prior to consent. Blood metal ion levels and PROMs were then obtained annually after enrollment. A sub-set of patients from a single center had annual metal artifact reduction sequence (MARS) MRI performed and were analyzed for the presence of moderate-to-severe
Background. Reasons for revision of metal-on-metal hip resurfacing arthroplasty (MoMHRA) have evolved with improving surgical experience and techniques. Early revisions were often due to fracture of the femoral neck while later revisions are associated with loosening and/or adverse local tissue reactions (ALTR) to wear debris. In some studies, revisions of MoMHRA with