Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 26 - 26
1 Jan 2018
MacDonald S Howard J Goyal P Yuan X Lanting B Teeter M Naudie D McCalden R
Full Access

Lewinnek's safe zone recommendation to minimise dislocations was a target of 5–25° for anteversion angle and 30–50° for inclination angle. Subsequently, it was demonstrated that mal-positioning of the acetabular cup can also lead to edge loading, liner fracture, and greater conventional polyethylene wear. The purpose of this study was to measure the effect of acetabular cup position on highly crosslinked polyethylene wear in total hip arthroplasty (THA) at long-term follow-up. We identified all patients that underwent primary THA with a minimum of 10 years follow-up using an institutional database in London, Ontario, Canada. Patients with a single implant design consisting of a 28 mm cobalt chromium head and highly crosslinked polyethylene liner (ram extruded, GUR 1050, 100 kGy gamma irradiated, remelted, ethylene oxide sterilised) were selected for inclusion. In total, 85 hips from 79 recruited patients were analysed. Patients underwent a supine radiostereometric analysis (RSA) exam in which the x-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup anteversion angle, inclination angle, and 3D penetration rate (including wear and creep) were measured from the stereo radiograph pairs. At a mean follow-up of 13 years (range, 10–17 years) the mean penetration rate was 0.059 mm/year (95% CI: 0.045 to 0.073 mm/year). Mean anteversion angle was 18.2° (range, −14 to 40°) and mean inclination angle was 43.6° (range, 27 to 61°). With respect to the Lewinnek safe zone, 67% hips met the target for anteversion angle, 77% met the target for inclination angle, and 51% met the target for both. There was no correlation between anteversion angle and penetration rate (r = −0.14, p = 0.72) or between inclination angle and penetration rate (r = 0.11, p = 0.35). There was also no difference (p = 0.07) in penetration rate between hips located within the Lewinnek safe zone for both anteversion angle and inclination angle (mean 0.057 mm/year, 95% CI: 0.036 to 0.079 mm/year) and those outside the safe zone (mean 0.062 mm/year, 95% CI: 0.042 to 0.083 mm/year). Acetabular cup position had no effect on the wear rate of highly crosslinked polyethylene at long-term follow-up. Although care should still be taken to correctly position the acetabular cup for stability, highly crosslinked polyethylene is a forgiving bearing material that can withstand a wide range of cup positions without negatively impacting longevity due to wear


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 41 - 41
1 Oct 2018
Tatka J Brady AW Matta JM
Full Access

Introduction

Accurate acetabular position is an important goal during THA. It is also well known that accurate acetabular positioning is very frequently not achieved, even by experienced, high volume surgeons. Problems associated with cup malposition are: dislocation, accelerated poly wear, impingement, ceramic squeaking, metalosis. Murray et al described 3 methods of measurement and assessment of acetabular inclination and anteversion (I&A): anatomic, radiographic and operative. It is the hypothesis of the authors, that the differences and details of these 3 methods are poorly understood by many surgeons and this is contributory to inconsistent cup positioning. Additionally, the radiographic method, which is most commonly used for post op assessment and academic studies, contributes to misunderstanding and error. Modern computer guidance and software assessment of radiographs allows us to easily measure anatomic I&A which should be thought of as “true” I&A.

Methods

The mathematical criteria for radiographic measurement of anatomic I&A are defined as well as the mathematical relationships and discrepancies between anatomic and radiographic I&A for any given cup.

A = A n g l e o f a n t e v e r s i o n o f c u p I = A n g l e o f i n c l i n a t i o n o f c u p

E = Angle of ellipse major diameter to horizontal

E = Radiographic inclination

Sin A = H o r i z o n t a l w i d t h o f e l l i p s e L e n g t h o f e l l i p s e m a j o r d i a m e t e r Sin I = V e r t i c a l h e i g h t o f e l l i p s e L e n g t h o f e l l i p s e m a j o r d i a m e t e r

Tan I = Tan E / Cos A

Tan E = (Tan I) x (Cos A)


Bone & Joint Open
Vol. 3, Issue 6 | Pages 475 - 484
13 Jun 2022
Jang SJ Vigdorchik JM Windsor EW Schwarzkopf R Mayman DJ Sculco PK

Aims

Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error.

Methods

A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (Δsacral slope(SS)stand-sit > 30°), or stiff (SSstand-sit < 10°) spinopelvic mobility contributed to increased error rates.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 9 - 9
1 Nov 2021
Farey J Chai Y Xu J Sadegpour A Jones DM Baker N Vigdorchik J Walter W
Full Access

Imageless computer navigation systems in total hip arthroplasty (THA) improve acetabular cup position, thereby reducing the risk of revision surgery for all causes as well as dislocation. We aimed to evaluate the registration accuracy of 3 alternate registration planes. A prospective, observational study was conducted with 45 THA in the supine position using two imageless navigation systems and 3 registration planes. Patient position was registered sequentially using an optical system (Stryker OrthoMap) and an inertial sensor-based system (Navbit Sprint) with 3 planes of reference: (Plane 1) an anatomical plane using the anterior superior iliac spines (ASISs) and the pubic symphysis; (Plane 2) a functional plane parallel to the line between the ASISs and the table plane; and, (Plane 3) a functional plane that was perpendicular to the gravity vector and aligned with the longitudinal axis of the patient. The 3 measurements of acetabular cup inclination and anteversion were compared with the measurements from postoperative computed tomography (CT) scans. For inclination, the mean absolute error was significantly lower for Plane 3 (1.80°) than for Plane 2 (2.74°), p = .038 and was lower for both functional planes than for the anatomical plane (3.75°), p < .001. For anteversion, the mean absolute error was significantly lower for Plane 3 (2.00°) than for Plane 2 (3.69°), p = .004 and was lower for both functional planes than for the anatomical plane (8.58°), p < .001. Patient registration using functional planes more accurately measured the acetabular cup position than registration using anatomic planes


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 29 - 29
19 Aug 2024
Kayani B Konan S Tahmassebi J Giebaly D Haddad FS
Full Access

The direct superior approach (DSA) is a modification of the posterior approach (PA) that preserves the iliotibial band and short external rotators except for the piriformis or conjoined tendon during total hip arthroplasty (THA). The objective of this study was to compare postoperative pain, early functional rehabilitation, functional outcomes, implant positioning, implant migration, and complications in patients undergoing the DSA versus PA for THA. This study included 80 patients with symptomatic hip arthritis undergoing primary THA. Patients were prospectively randomised to receive either the DSA or PA for THA, surgery was undertaken using identical implant designs in both groups, and all patients received a standardized postoperative rehabilitation programme. Predefined study outcomes were recorded by blinded observers at regular intervals for two-years after THA. Radiosteriometric analysis (RSA) was used to assess implant migration. There were no statistical differences between the DSA and PA in postoperative pain scores (p=0.312), opiate analgesia consumption (p=0.067), and time to hospital discharge (p=0.416). At two years follow-up, both groups had comparable Oxford hip scores (p=0.476); Harris hip scores (p=0.293); Hip disability and osteoarthritis outcome scores (p=0.543); University of California at Los Angeles scores (p=0.609); Western Ontario and McMaster Universities Arthritis Index (p=0.833); and European Quality of Life questionnaire with 5 dimensions scores (p=0.418). Radiographic analysis revealed no difference between the two treatment groups for overall accuracy of acetabular cup positioning (p=0.687) and femoral stem alignment (p=0.564). RSA revealed no difference in femoral component migration (p=0.145) between the groups at two years follow-up. There were no differences between patients undergoing the DSA versus PA for THA with respect to postoperative pain scores, functional rehabilitation, patient-reported outcome measurements, accuracy of implant positioning, and implant migration at two years follow-up. Both treatment groups had excellent outcomes that remained comparable at all follow-up intervals


Robotic assisted surgery aims to reduce surgical errors in implant positioning and better restore native hip biomechanics compared to conventional techniques for total hip arthroplasty (THA). The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual THA versus robotic-arm assisted THA. Secondary objectives were to determine differences between these treatment techniques for THA in achieving the planned combined offset, cup inclination, cup version, and leg-length correction. This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. All operative procedures were undertaken by a single surgeon using the minimally-invasive posterior approach. Two independent blinded observers recoded all radiological outcomes of interest using plain radiographs. Patients in both treatment groups were well-matched for age, gender, body mass index, laterality of surgery, and ASA scores. Interclass correlation coefficient was 0.92 (95% CI: 0.84 – 0.95) for intra-observer agreement and 0.88 (95% CI: 0.82–0.94) for inter-observer agreement in all study outcomes. Robotic THA was associated with improved accuracy in restoring the native horizontal (p<0.001) and vertical (p<0.001) centres of rotation, and improved preservation of the patient's native combined offset (P<0.001) compared to conventional THA. Robotic THA improved accuracy in positioning of the acetabular cup within the combined safe zones of inclination and anteversion described by Lewinnek et al (p=0.02) and Callanan et al (p=0.01) compared to conventional THA (figures 1–2). There was no difference between the two treatment groups in achieving the planned leg-length correction (p=0.10). Robotic-arm assisted THA was associated with improved accuracy in restoring the native centre of rotation, better preservation of the combined offset, and more precise acetabular cup positioning within the safe zones of inclination and anteversion compared to conventional manual THA. Robotic-arm assisted THA enables improved preservation of native hip biomechanics compared to conventional manual THA. For any figures or tables, please contact authors directly: . fsh@fareshaddad.net