Advertisement for orthosearch.org.uk
Results 1 - 20 of 39
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 49 - 49
19 Aug 2024
Beverland D Wong ML Robinson M Cassidy R Lamb J Diamond O
Full Access

A: - determine the incidence of intraoperative periprosthetic femoral fracture (PFF)

B: - determine the incidence of, and the reoperation rate for postoperative PFF

When using either CC or PTS femoral components.

Retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All PFFs were identified using regional radiograph archiving and electronic care systems.

Of the 11,018 THAs 4,952 were CC and 6,066 were PTS. Between groups, age, sex, and BMI did not differ.

A: - 55 (0.5%) had an intraoperative PFF. 44 CC and 11 PTS (p<0.001). 3 patients in each group had a femoral shaft fracture, remaining fractures were either the calcar (20 CC and 2 group) or the greater trochanter (11 CC and 6 PTS).

B: - 91(0.8%) sustained a postoperative PFF. Of those 15 were managed conservatively, 15 were revised and 61 (80.3%) had an ORIF. The CC group had both a lower overall rate of postoperative PFFs (0.7% (36/4,952) vs 0.9% (55/6,066); p = 0.341), and a lower rate of return to theatre (0.4% (22/4,952) vs 0.9% (54/6,066); p = 0.005). 1.3% of male PTS (36/2,674) had a reoperation compared to 0.3% of male CC (7/2,121) (p<0.001).

With regard to stem fracture there were none in the Corail group and 5 in the Exeter group. Of these 2 were sub trunnion and 3 were basal neck.

A: - There were significantly more intraoperative PFFs with CC 44 (0.8%) than PTS 11 (0.2%). However, the majority of fractures were either of the calcar or greater trochanter with no impact on early recovery or one year Oxford scores.

B: - Male PTS were five times more likely to have a reoperation for postoperative PFF. Females had the same incidence of reoperation with either component type.

There were 5 stem fractures in the Exeter group and none in the Corail.

These results represent robust estimates, which are likely to be more accurate than revision only studies typically generated from registry data.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 20 - 20
1 May 2019
Lamb J King S van Duren B West R Pandit H
Full Access

Background

Method of fixation in THA is a contentious issue, with proponents of either technique citing improved implant survival and outcomes. Current comparisons rely on insufficiently powered studies with short-term follow up or larger poorly controlled registry studies. Patient factors are considered a key variable contributing to the risk of implant failure. One way to overcome this confounder is to compare the survival of cementless and cemented THAs patients who have undergone bilateral THAs with cemented hip on one side and cementless hip on the other. We compared stem survival of patients who have bilateral THA with one cemented stem in one hip and a cementless stem in the contralateral hip in the National Joint Registry.

Methods

UK National Joint Registry is the largest registry of its kind in the world. This study included 2934 patients with 5868 THAs who underwent bilateral THAs s between 2003 and 2016. These patients had undergone bilateral sequential THAs within 3 years of each other: cemented THA on one side and cementless on the other, Patients had identical pre-operative American Society of Anaesthesiologists group for both THAs and same indication for surgery. Implant survival was compared using Cox regression with an endpoint of stem revision.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 755 - 761
1 Jun 2015
Flatøy B Röhrl SM Rydinge J Dahl J Diep LM Nordsletten L

Cemented femoral stems with force closed fixation designs have shown good clinical results despite high early subsidence. A new triple-tapered stem in this category (C-stem AMT) was introduced in 2005. This study compares this new stem with an established stem of similar design (Exeter) in terms of migration (as measured using radiostereometric analysis), peri-prosthetic bone remodelling (measured using dual energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. . A total of 70 patients (70 hips) with a mean age of 66 years (53 to 78) were followed for two years. Owing to missing data of miscellaneous reasons, the final analysis represents data from 51 (RSA) and 65 (DXA) patients. Both stems showed a typical pattern of migration: Subsidence and retroversion that primarily occurred during the first three months. C-stem AMT subsided less during the first three months (p = 0.01), before stabilising at a subsidence rate similar to the Exeter stem from years one to two. The rate of migration into retroversion was slightly higher for C-stem AMT during the second year (p = 0.03). Whilst there were slight differences in movement patterns between the stems, the C-stem AMT exhibits good early clinical outcomes and displays a pattern of migration and bone remodelling that predicts good clinical performance. Cite this article: Bone Joint J 2015;97-B:755–61


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 33 - 37
1 Jul 2021
Ennin KA Elsharkawy KA Dasgupta S Emerson RH

Aims

To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications?

Methods

A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1339 - 1344
1 Aug 2021
Jain S Mohrir G Townsend O Lamb JN Palan J Aderinto J Pandit H

Aims

This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems.

Methods

Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 466 - 475
8 Jul 2021
Jain S Lamb J Townsend O Scott CEH Kendrick B Middleton R Jones SA Board T West R Pandit H

Aims

This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty.

Methods

Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and stem fixation philosophy (polished taper-slip (PTS) vs composite beam (CB)). The effect of multiple variables was estimated using multinomial logistic regression to estimate odds ratios (ORs) with 95% confidence intervals (CIs). Surgical treatment (revision vs fixation) was compared by UCS grade and AO/OTA type.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 51 - 52
1 Nov 2013
Murray DW

There is a North Atlantic divide, with cementless femoral stems being used more frequently in the USA and cemented stems being used more frequently in many countries in Europe. This is primarily because different cemented stems have been used on different sides of the Atlantic and the results of the cemented stems in the US have often been poor, whereas the results of the stems used in Europe have been good. In the National registers in Europe, cemented stems have tended to achieve better results than cementless.

Cite this article: Bone Joint J 2013;95-B, Supple A:51–2.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001).

The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs.


Background. The acknowledged benefit of the direct anterior (DA) approach is early functional return. Most surgeons in the U.S. use cementless femoral replacement given the negative track record of some cemented designs. However, delayed osseointegration of a femoral stem typically seen in older patients with poor bone quality will delay recovery, diminishing the benefits of the DA approach. Registry studies have shown a higher revision rate and complications in this patient population leading to a renewed interest in cemented fixation. Questions posed. To achieve the functional benefits of the DA approach and the fixation benefits of cemented replacement, this study combined the 2 techniques posing the following questions:1) Does the limited access of the DA approach adversely affect the cement technique? 2) Does such a cementing technique reduce the incidence of cementless complications?. Methods. A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016–2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 in the cementless group and 76 in the cemented group, 70% females. Femoral complications were compared between the 2 groups using the T-test. Results. The cementless group had a higher rate of femoral complications (8 versus 0, P=0.011). There were 2 loose stems and 6 fractures, all requiring revision. Fractures occurred about 14.5 days and loosening about 10 months postoperatively. Conclusion. A higher rate of complications occurred with cementless stems. Cemented stems are protective in patients above 70 and can be safely done through the DA approach


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 64 - 64
2 May 2024
Lamb J West R Relton S Wilkinson M Pandit H
Full Access

Current estimates of periprosthetic fracture risk associated with femoral implants is mostly limited to revision only estimates and does not accurately represent stem performance. The aim of this study was to estimate the risk of surgically treated post-operative periprosthetic femoral fracture (POPFF) more accurately associated with frequently used femoral implants used for total hip arthroplasty (THA). A cohort study of patients who underwent primary THA in England between January 1, 2004, and December 31, 2020. Periprosthetic fractures were identified from prospectively collected revision records and national procedure coding records. Survival modelling was used to estimate POPFF incidence rates, adjusting for potential confounders. Subgroup analyses were performed for patients over 70 years, with non-osteoarthritic indications, and neck of femur fracture. POPFF occurred in 0.6% (5100/809,832) of cases during a median (IQR) follow up of 6.5 (3.9 to 9.6) years. The majority of POPFF were treated with fixation after implantation of a cemented stem. Adjusted patient time incidence rates (PTIR) for POPFF varied by stem design, regardless of cement fixation. Cemented composite beam stems (CB stems) demonstrated the lowest risk of POPFF. Collared cementless stems had an equivalent or lower rate of POPFF versus the current gold standard polished taper slip cemented stem. POPFF account for a quarter of all revisions following primary THA. Cemented CB stems are associated with the lowest POPFF risk. Stem design is strongly associated with POPFF risk, regardless of the presence of cement. Surgeons, policymakers, and patients should consider these findings when recommending femoral implants in those most at risk of POPFF


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 435 - 441
1 May 2024
Angelomenos V Mohaddes M Kärrholm J Malchau H Shareghi B Itayem R

Aims. Refobacin Bone Cement R and Palacos R + G bone cement were introduced to replace the original cement Refobacin Palacos R in 2005. Both cements were assumed to behave in a biomechanically similar fashion to the original cement. The primary aim of this study was to compare the migration of a polished triple-tapered femoral stem fixed with either Refobacin Bone Cement R or Palacos R + G bone cement. Repeated radiostereometric analysis was used to measure migration of the femoral head centre. The secondary aims were evaluation of cement mantle, stem positioning, and patient-reported outcome measures. Methods. Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected. Results. Median distal migration (y-axis) at two years for the Refobacin-Optivac system was -0.79 mm (-2.01 to -0.09), for the Refobacin-Optipac system was -0.75 mm (-2.16 to 0.20), and for the Palacos-Optivac system was -1.01 mm (-4.31 to -0.29). No statistically significant differences were found between the groups. Secondary outcomes did not differ statistically between the groups at the two-year follow-up. Conclusion. At two years, we found no significant differences in distal migration or clinical outcomes between the three groups. Our data indicate that Refobacin Bone Cement R and Palacos R + G are comparable in terms of stable fixation and early clinical outcomes. Cite this article: Bone Joint J 2024;106-B(5):435–441


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 37 - 37
1 Nov 2015
Kazi H Whitehouse S Timperley A
Full Access

Introduction. Cemented stems have given good results with no difference in outcome when comparing taper slip and composite beam designs in short-term randomised trials. We aimed to ascertain differences in outcome between the different cemented design philosophies using a large-scale registry analysis. Materials and Methods. A retrospective cohort study of National Joint Registry of England and Wales (NJREW) was conducted. The study population included all primary total hip operations performed in the UK from 1 April 2003 to 31 September 2012. All cemented stems were identified and categorised as taper slip (polished) or composite beam (non-polished). Results. Non-polished stems accounted for 16.6% of all primary cemented hip replacements with polished stems making up the remainder. Mean age for non-polished stems was 73.0 years and 71.8 for the polished group (p<0.001) suggesting selection bias. There was a statistically significant increased chance of revision in the non-polished group compared with the polished group (p<0.001). Reasons for revision by fixation type:. Reason for revision. Total. Polished n=253667. Non-polished n=50661. N. %. n. %. Aseptic loosening stem. 532. 300. 0.12. 232. 0.46. Infection. 965. 735. 0.29. 230. 0.45. Stem lysis. 149. 99. 0.04. 50. 0.10. Cup lysis. 156. 125. 0.05. 31. 0.06. Malalignment stem. 114. 86. 0.03. 28. 0.06. The risk of aseptic loosening, stem lysis, malalignment and infection were several magnitudes higher for non-polished stems than for polished varieties (p<0.05). Discussion. Theoretical concerns are reflected in clinical practice with an increased incidence of revision for non-polished stems for all indications recorded. Conclusions. Incidence of revision is higher in cemented non-polished stems compared with polished designs


Bone & Joint Open
Vol. 4, Issue 5 | Pages 370 - 377
19 May 2023
Comeau-Gauthier M Bzovsky S Axelrod D Poolman RW Frihagen F Bhandari M Schemitsch E Sprague S

Aims

Using data from the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial, we sought to determine if a difference in functional outcomes exists between monopolar and bipolar hemiarthroplasty (HA).

Methods

This study is a secondary analysis of patients aged 50 years or older with a displaced femoral neck fracture who were enrolled in the HEALTH trial and underwent monopolar and bipolar HA. Scores from the Western Ontario and McMaster University Arthritis Index (WOMAC) and 12-Item Short Form Health Survey (SF-12) Physical Component Summary (PCS) and (MCS) were compared between the two HA groups using a propensity score-weighted analysis.


Aims

The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless polished taper slip stem (PTS) for management of an intracapsular hip fracture.

Methods

This retrospective study included hip fracture patients aged 50 years and older treated with Exeter (PTS) bipolar hemiarthroplasty between 2019 and 2022. Patient demographics, place of domicile, fracture type, delirium status, American Society of Anesthesiologists (ASA) grade, length of stay, and mortality were collected. Reoperation and mortality were recorded up to a median follow-up of 29.5 months (interquartile range 12 to 51.4). Cox regression was performed to evaluate independent risk factors associated with reoperation and mortality.


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 476 - 485
8 Jul 2021
Scheerlinck T De Winter E Sas A Kolk S Van Gompel G Vandemeulebroucke J

Aims

Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre.

Methods

Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre.


Bone & Joint Research
Vol. 11, Issue 3 | Pages 180 - 188
1 Mar 2022
Rajpura A Asle SG Ait Si Selmi T Board T

Aims

Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane.

Methods

Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 486 - 491
1 Mar 2021
Di Martino A Bordini B Ancarani C Viceconti M Faldini C

Aims

Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017).

Methods

A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 7 | Pages 974 - 978
1 Sep 2004
Thomas SR Shukla D Latham PD

Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem debonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted. We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims

There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA.

Methods

We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality.