Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 93 - 93
1 Apr 2018
Todorov D Gueorguiev B Zderic I Stoffel K Richards G Lenz M Enchev D Baltov A
Full Access

Introduction. The incidence of distal femoral fractures in the geriatric population is growing and represents the second most common insufficiency fracture of the femur following fractures around the hip joint. Fixation of fractures in patients with poor bone stock and early mobilisation in feeble and polymorbide patients is challenging. Development of a fixation approach for augmentation of conventional LISS (less invasive stabilization system) plating may result in superior long-term clinical outcomes and enhance safe weight bearing. Objectives. The aim of this study was to investigate the biomechanical competence of two different techniques of augmented LISS plating for treatment of osteoporotic fractures of the distal femur in comparison to conventional LISS plating. Materials & methods. Unstable distal femoral fracture AO/OTA 33-A3 was set in artificial femora with low density simulating osteoporotic bone. Three study groups, consisting of 10 specimens each, were created for instrumentation with a 9-hole LISS plate, a LISS plate with an additional 3D-printed polyactide cylindrical intramedullary graft, as well as a LISS plate plus a medial 3.5mm LCP (locking compression plate) - double plating. All specimens were non-destructively tested under axial (20–150N) and torsional (0–4Nm) quasi-static loading. Each construct was tested with two different working length (WL) configurations (long and short) of the LISS plate. Relative movements between the most medial superior and inferior osteotomy aspects were investigated via three-dimensional motion tracking analysis. Results. Interfragmentary displacement along the femur axis (mm) under 150N axial loading was 2.03±0.23/1.65±0.27 for LISS with long/short WL, 0.18±0.06/0.18±0.04 for double plating with long/short WL, and 0.40±0.05/0.30±0.05 for LISS plus graft with long/short WL. Shear interfragmentary displacement (mm) under 4Nm torsional loading in internal rotation was 1.16±0.17/0.92±0.11 for LISS with long/short WL, 0.40±0.10/0.43±0.07 for double plating with long/short WL, and 1.09±0.13/0.82±0.11 for LISS plus graft with long/short WL. Double plating revealed significantly smaller longitudinal and shear displacement compared to the other two techniques for long and short WL, respectively (P≤0.010). In addition, LISS plus graft fixation was with significantly less longitudinal displacement in comparison to conventional LISS plating for long and short WL, respectively (P≤0.001). Long WL resulted in significantly higher longitudinal and shear displacement compared to short WL for LISS and LISS plus graft (P≤0.032), but not for double plating (P=1.000). Conclusion. Intramedullary grafting resulted in significantly increased fracture stability under axial loading in comparison to conventional LISS plating. However, it was not efficient enough to achieve comparable stability to double plating


Bone & Joint Research
Vol. 1, Issue 6 | Pages 118 - 124
1 Jun 2012
Grawe B Le T Williamson S Archdeacon A Zardiackas L

Objectives

We aimed to further evaluate the biomechanical characteristics of two locking screws versus three standard bicortical screws in synthetic models of normal and osteoporotic bone.

Methods

Synthetic tubular bone models representing normal bone density and osteoporotic bone density were used. Artificial fracture gaps of 1 cm were created in each specimen before fixation with one of two constructs: 1) two locking screws using a five-hole locking compression plate (LCP) plate; or 2) three non-locking screws with a seven-hole LCP plate across each side of the fracture gap. The stiffness, maximum displacement, mode of failure and number of cycles to failure were recorded under progressive cyclic torsional and eccentric axial loading.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 10 - 10
14 Nov 2024
Zderic I Kraus M Rossenberg LV Puls L Pastor T Gueorguiev B Richards G Pape HC Pastor T
Full Access

Introduction. The main postoperative complications in fixation of ulna shaft fractures are non-union and implant irritation using currently recommended 3.5-mm locking compression plates. An alternative approach using a combination of two smaller plates in orthogonal configuration has been proposed. The aim of this study was to compare the biomechanical properties of a single 3.5-mm locking compression plate versus double plating using one 2.5-mm and one 2.0-mm mandible plate in a human ulna shaft fracture model. Method. Eight pairs human ulnar specimens with a standardized 10-mm fracture gap were pairwise assigned for instrumentation with either a single 3.5-mm plate placed posteriorly, or for double plating using a 2.5-mm and a 2.0-mm mandible plate placed posteriorly under the flexor muscles and laterally under the extensor muscles. All constructs were initially non-destructively biomechanically tested in axial compression, torsion, and bending, which was followed by cyclic torsional loading to failure. Interfragmentary movements were monitored by means of optical motion tracking. Result. There were no significant differences between the two plating techniques for axial stiffness (p=0.335), torsional stiffness in supination (p=0.462), torsional stiffness in pronation (p=0.307), medio-lateral bending stiffness (p=0.522), and antero-posterior bending stiffness (p=0.143). During cyclic torsional loading over the first 3000 cycles, there were no significant differences between the two plating techniques for shear displacement across the fracture gap, p=0.324. The numbers of cycles until clinically relevant failure of 5° angular deformation were 1366±685 for double plating and 2024±958 for single plating, which was statistically non-significantly different, p>0.05. The constructs treated with both plating techniques failed due to bone breakage at the most distal screw. Conclusion. From a biomechanical perspective double plating of ulna shaft fractures using a 2.5-mm and a 2.0-mm locking mandible plate demonstrated equivalent fixation strength as conventional plating using a single 3.5-mm locking compression plate


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 472 - 472
1 Sep 2009
Buijze G Blankevoort L Kloen P
Full Access

New concepts in plate fixation have led to an evolution in plate design for olecranon fractures. The purpose of this study was. to compare the stiffness and strength of a contoured Locking Compression Plate (LCP) with a conventional plating method (one-third tubular plate) in a cadaveric comminuted olecranon fracture model with standardized osteotomy, and. to evaluate the LCP fixation method in a prospectively included group of patients with complex olecranon fractures using validated outcome scores. In the biomechanical study, five matched pairs of cadaveric elbows were randomly assigned for fixation by either LCP or a conventional plating method. Specimens were mounted to a custom-made testing bench and subjected to cyclic loading until failure occurred while measuring gapping at the osteotomy site. In the clinical study, twenty-one patients treated with LCP for complex olecranon fractures had a mean follow up of 20 months (3–39 months) and functional and patient rated outcome were evaluated. In the biomechanical study, there was no significant difference in fixation stiffness and strength between one third tubular plating and LCP (p > 0.05). In the clinical study, the mean time to union of the fracture was 6 months (2–28 months). According to the Mayo Elbow Performance Index (MEPI) most patients had a good or excellent outcome. No patients reported difficulty with activities of daily living. Physical capacity showed minimal loss of stability and strength. Six patients had their hardware removed. Technical ease of application and advantageous features of the LCP -such as unicortical screw fixation and improved holding power in osteopenic bone- make it a good alternative implant for comminuted olecranon fractures


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_16 | Pages 81 - 81
1 Nov 2018
Gueorguiev B
Full Access

Locking plates have led to important changes in bone fracture management, allowing flexible biological fracture fixation based on the principle of an internal fixator. The technique of locking plate fixation differs fundamentally from conventional plating and has its indications and limitations. Most of the typical locking plate failure patterns are related to basic technical errors, such as under-sizing of the implant, too short working length, and imperfect application of locking screws. After analysis of the fracture morphology and intrinsic stability following fracture reduction, a meticulous preoperative planning is mandatory under consideration of the principles of the internal fixator technique to avoid technical errors and inaccuracies leading to early implant failure


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_13 | Pages 89 - 89
1 Nov 2015
Ries M
Full Access

The Vancouver classification separates periprosthetic femur fractures after THA into three regions (A - trochanteric, B - around or just below the stem, and C - well below the stem), with fractures around or just below the stem further separated into those with a well-fixed (B1) or loose stem and good (B2) or poor (B3) bone stock. Trochanteric fractures may be associated with osteolysis and require treatment that addresses osteolysis as well as ORIF of displaced fractures. Fractures around a well-fixed stem can be treated with ORIF using cerclage or cable plating, while those around a loose stem require implant revision usually to a longer cementless tapered or distally porous coated cementless stem. Fractures around a loose stem with poor bone stock in which salvage of the proximal femur is not possible require replacement of the proximal femur with an allograft prosthetic composite or proximal femoral replacement. Fractures well below the stem can be treated with conventional plating methods. Periprosthetic acetabular fractures are rare and usually occur in the early post-operative period or late as a result of osteolysis or trauma. These can generally be separated into those with a stable acetabular component which can be treated non-operatively, and those with an unstable component often with discontinuity or posterior column instability which require complex acetabular reconstruction utilizing plating or revision to a cup-cage


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 69 - 69
1 Nov 2021
Pastor T Zderic I Richards G Gueorguiev B Knobe M
Full Access

Introduction and Objective. Distal femoral fractures are commonly treated with a straight plate fixed to the lateral aspects of both proximal and distal fragments. However, the lateral approach may not always be desirable due to persisting soft-tissue or additional vascular injury necessitating a medial approach. These problems may be overcome by pre-contouring the plate in helically shaped fashion, allowing its distal part to be fixed to the medial aspect of the femoral condyle. The objective of this study was to investigate the biomechanical competence of medial femoral helical plating versus conventional straight lateral plating in an artificial distal femoral fracture model. Materials and Methods. Twelve left artificial femora were instrumented with a 15-hole Locking Compression Plate – Distal Femur (LCP-DF) plate, using either conventional lateral plating technique with the plate left non-contoured, or the medial helical plating technique by pre-contouring the plate to a 180° helical shape and fixing its distal end to the medial femoral condyle (n=6). An unstable extraarticular distal femoral fracture was subsequently simulated by means of an osteotomy gap. All specimens were tested under quasi-static and progressively increasing cyclic axial und torsional loading until failure. Interfragmentary movements were monitored by means of optical motion tracking. Results. Initial axial stiffness was significantly higher for helical (185.6±50.1 N/mm) versus straight (56.0±14.4) plating, p<0.01. However, initial torsional stiffness in internal and external rotation remained not significantly different between the two fixation techniques (helical plating:1.59±0.17 Nm/° and 1.52±0.13 Nm/°; straight plating: 1.50±0.12 Nm/° and 1.43±0.13Nm/°), p≥0.21. Helical plating was associated with significantly higher initial interfragmentary movements under 500 N static compression compared to straight plating in terms of flexion (2.76±1.02° versus 0.87±0.77°) and shear displacement under 6 Nm static rotation in internal (1.23±0.28° versus 0.40±0.42°) and external (1.21±0.40° versus 0.57±0.33°) rotation, p≤0.01. In addition, helical plating demonstrated significantly lower initial varus/valgus deformation than straight plating (4.08±1.49° versus 6.60±0.47°), p<0.01. Within the first 10000 cycles of dynamic loading, helical plating revealed significantly bigger flexural movements and significantly lower varus/valgus deformation versus straight plating, p=0.02. No significant differences were observed between the two fixation techniques in terms of axial and shear displacement, p≥0.76. Cycles to failure was significantly higher for helical plating (13752±1518) compared to straight plating (9727±836), p<0.01. Conclusions. Although helical plating using a pre-contoured LCP-DF was associated with higher shear and flexion movements, it demonstrated improved initial axial stability and resistance against varus/valgus deformation compared to straight lateral plating. Moreover, helical plate constructs demonstrated significantly improved endurance to failure, which may be attributed to the less progressively increasing lever bending moment arm inherent to this novel fixation technique. From a biomechanical perspective, helical plating may be considered as a valid alternative fixation technique to standard straight lateral plating of unstable distal femoral fractures


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 73 - 73
1 Sep 2012
Littlechild J Keating J Kahn K
Full Access

The outcome of 77 high energy tibial plateau fractures treated by locking or conventional plating was reviewed. The aim of the study was to determine if there was any advantage of locking plates in reducing the complication rates associated with fixation of these injuries. All patients had a high energy injury pattern (medial or bicondylar plateau fractures). There were 32 locked plates and 45 non-locking plates used. Compartment syndrome complicated 5 patients (16%) in the locked plate group and 3 (7%) in the non-locked group (p = 0.198). Superficial infection occurred in 4 (13%) patients with locked plates and 7 (16%) patients with non-locked plates. Thromboembolic complications occurred in 3 (7%) patients treated with non-locked plates. There were no thrombembolic complications in the locked plate group (p = 0.135). Overall, malunion of the plateau occurred in 10 (22%) patients treated with non-locked plates compared to 7 (22%) patients who received locked plates. This was due to residual malreduction in 4 (13%) patients in the locked plate group and 6 (13%) patients in the non-locked plate group at the time of surgery. In the remaining cases loss of reduction after fixation occurred in 4 (9%) patients who received non-locked plates and in 3 (9%) patients who were treated with locked plates. No statistically significant difference was noted in the treatment outcomes of patients managed with locked plates or non-locked plates, regardless of fracture severity. We concluded that there is no definite advantage associated with the use of locked plating for high energy tibial plateau fractures


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 324 - 324
1 Mar 2004
Bajwa A Green S Finn P Elfick A Gregg P Port A
Full Access

Aims: Does PHILOS (Proximal Humeral Internal Locking system) construct provides better þxation than Clover leaf plate and T-plate in a simulated 2-part fracture of proximal humerus, in an osteoporotic bone model?Materials and Methods: Biomechanical laboratory study. Third generation composite Humerus model was used, with short e-glass epoxy þbres forming cortex and polyurethane cancellous core. Low density polyure-thane core (1.2gm/cc) was used to simulate an osteoporotic model. Osteotomy at surgical neck of humerus was carried out to create 2-part fracture of proximal humerus. Samples were randomised to receive one of the implants. Following þxation samples were placed in a custom made jig to þx proximal and distal ends without interfering with implants and osteotomy site. All samples were subjected to cyclical torque, Torque to failure, Cyclical compression and Compression loading to failure. Results were entered in a database. Results: PHILOS provided signiþcantly better þxation in ÔTorque to failureñ experiment. PHILOS construct shows less plastic deformation in cyclical torque and cyclical compression. Locking screws did not Ôback offñ in any of the experiments involving PHILOS construct, however ordinary screws did back off both in Ôtoque and compressionñ testing. Conclusions: PHILOS construct provides better stability in Torque and compression as compared to conventional plating devices, in an osteoporotic bone model


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 395 - 395
1 Oct 2006
Nanda R Ahmad M Bajwa A Green S Gregg P Port A
Full Access

Aim: To compare of strength of constructs using the newer and part specific nail systems: Polarus and European Humeral Nail with that using PHILOS and Conventional plate systems in a simulated 2-part fracture of proximal humerus, in an osteoporotic bone model. Materials and Methods: A Biomechanical laboratory study was undertaken. Third generation composite Humerus model was used, with short e-glass epoxy fibres forming cortex and polyurethane cancellous core. Low-density polyurethane core (1.2gm/cc) was used to simulate an osteoporotic model. Osteotomy at surgical neck of humerus was carried out to create 2-part fracture of proximal humerus. Samples were fixed using one of the implants- the Polarus nail, the European Humeral Nail, PHILOS Plate, Clover Leaf Plate or T-Plate. Following fixation samples were placed in a custom made jig to fix proximal and distal ends without interfering with implants and osteotomy site. All samples were subjected to cyclical torque, torque to failure, cyclical compression and Compression loading to failure. Results: The two Nail systems that are specifically designed for fixation of proximal humerus fracture provided significantly better fixation in all the test modalities. PHILOS construct shows less plastic deformation in cyclical torque and cyclical compression when compared to the other plates but the 2 nail systems were far superior. Locking screws did not ‘back off’ in any of the experiments involving the Polarus, European Humeral Nail and PHILOS construct, however ordinary screws used with the conventional plates did back off both in ‘torque and compression’ testing. Conclusions: Polarus and European Humeral Nail constructs provide better stability in torque and compression as compared to PHILOS, which in turn is a more stable construct in comparison to conventional plating devices


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 236 - 236
1 May 2009
Foster MP Papp S Poitras P
Full Access

Biomechanical stability is important for fracture healing. With standard plate and screw constructs, longer plates with screws well spaced, near and far from the fracture site, are biomechanically superior. Newer locked plates have been shown to be superior to conventional plating for difficult fractures. The ideal screw configuration for fixation with locked plates has yet to be addressed. This study investigates the effects of screw position on construct stiffness as well as strain in both the plate and bone during fixation of a diaphyseal comminuted fracture using a locking plate with bicortical fixation. A composite cylinder (Sawbones) was machined to produce two models:. (a) comminuted model (4mm gap) and. (b) whole model (no gap) to simulate the remodelling phase. Five strain gauges were mounted to the bone models and one between the center holes of the locking plate. Four different configurations of screw number and position were evaluated using a twelve-hole locking plate (Smith & Nephew Perilock). Plate holes were numbered on each side of the gap from one to six. Screw configuration 654321, 621, 654 and 321 were tested in four-point bending on an MTS 858 Mini-Bionix. Force (N) and displacement (mm) as well as strain readings were recorded at 10 Hz. Plate strain in the gap model did not vary significantly for the different configurations. Construct stiffness of the 654 model (all screws far from gap) showed a 30% decrease in stiffness as compared to other screw configurations (p< 0.001). In the whole bone model, the maximal bone strain was outside the farthest screw from the center of the plate (stress shielding) and bone strain at the fracture site in 654 was significantly higher than in 621 (p< 0.001). Results showed that three screw fixation produced similar construct stiffness to a six screw construct when well spaced. Three screws placed far from the fracture gap (654) as compared to three screws evenly spaced (621) showed decreased stability in the comminuted model but resulted in increased bone strain at the fracture site in the whole bone model. All configurations produced similar plate strain


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 183 - 183
1 Mar 2006
Rubel I Corcoran A
Full Access

Introduction: Since the introduction of periarticular locking plates (PLP) open reduction and internal fixation of periarticular fractures has gained popularity. Although initial trials have shown encouraging results, no studies to date has focused on its use for metaphyseal fractures. The purpose of this study is to report on the performance of PLP for fixation of periarticular fractures. Material and Methods: 49 with at least one year follow up were included in this prospective review. All fractures involved the metaphyseal area and 39 had an intraarticular extension as well. Fixation was performed by a combined locking-regular screws technique. The parameters included in the analysis were fracture displacement, type of callus formation, healing of the fracture, screw pull-out, screw breakage, plate breakage, stress rising and stress shielding with subsequent bone loss. Results: At one year follow up the results showed that: 1 fracture had lost reduction in the early post op; 2 locking screws had backed up despite being initially locked to the plate; 1 screw broke, 1 plate broke. There were no problems with stress rising or stress fractures at the end of the plate. The most impressive finding was the high rate of stress shielding with subsequent bone loss within the range delimited by the locking screws (27 %). Callus formation: 18 % had no visible callus on plain radiographs despite being clinically healed. 62% had very little callus. 15 % had moderate callus and 5% had robust callus formation. Discussion and conclusion: The use of PLP appears to have some unique characteristics, different than conventional plating.: 80 % of the fractures healed with no or very little callus. The use of locking screws in the shaft portion of the fixation may have been the cause for the high rate of stress shielding, which typically occurred within the locking screws in our series. Since the review of our data we are no longer using locking screws in the shaft portion of the fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 2 | Pages 253 - 258
1 Mar 2001
Bhaskar AR Roberts JA

Unstable fractures of the forearm in children present problems in management and in the indications for operative treatment. In children, unlike adults, the fractures nearly always unite, and up to 10° of angulation is usually considered to be acceptable. If surgical intervention is required the usual practice in the UK is to plate both bones as in an adult. We studied, retrospectively, 32 unstable fractures of the forearm in children treated by compression plating. Group A (20 children) had conventional plating of both forearm bones and group B (12 children) had plating of the ulna only. The mean age was 11 years in both groups and 23 (71%) of the fractures were in the midshaft. In group B an acceptable position of the radius was regarded as less than 10° of angulation in both anteroposterior (AP) and lateral planes, and with the bone ends hitched. This was achieved by closed means in all except two cases, which were therefore included in group A. Union was achieved in all patients, the mean time being 9.8 weeks in group A and 11.5 weeks in B. After a mean interval of at least 12 months, 14 children in group A and nine in group B had their fixation devices removed. We analysed the results after the initial operation in all 32 children. The 23 who had the plate removed were assessed at final review. The results were graded on the ability to undertake physical activities and an objective assessment of loss of rotation of the forearm. In group A, complications were noted in eight patients (40%) after fixation and in six (42%) in relation to removal of the radial plate. No complications occurred in group B. The final range of movement and radiological appearance were compared in the two groups. There was a greater loss of pronation than supination in both. There was, however, no limitation of function in any patient and no difference in the degree of rotational loss between the two groups. The mean radiological angulation in both was less than 10° in both AP and lateral views, which was consistent with satisfactory function. The final outcome for 23 patients was excellent or good in 12 of 14 (90%) in group A, despite the complications, and in eight of nine in group B (90%). If reduction and fixation of the fracture of the ulna alone restores acceptable alignment of the radius in unstable fractures of the forearm, operation on the radius can be avoided


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1447 - 1458
1 Nov 2019
Chatziagorou G Lindahl H Kärrholm J

Aims

We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis.

Patients and Methods

The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1385 - 1391
1 Oct 2019
Nicholson JA Gribbin H Clement ND Robinson CM

Aims

The primary aim of this study was to determine if delayed clavicular fixation results in a greater risk of operative complications and revision surgery.

Patients and Methods

A retrospective case series was undertaken of all displaced clavicular fractures that underwent plate fixation over a ten-year period (2007 to 2017). Patient demographics, time to surgery, complications, and mode of failure were collected. Logistic regression was used to identify independent risk factors contributing towards operative complications. Receiver operating characteristic (ROC) curve analysis was used to determine if a potential ‘safe window’ exists from injury to delayed surgery. Propensity score matching was used to construct a case control study for comparison of risk.


The Bone & Joint Journal
Vol. 97-B, Issue 10 | Pages 1423 - 1427
1 Oct 2015
Rand BCC Penn-Barwell JG Wenke JC

Systemic antibiotics reduce infection in open fractures. Local delivery of antibiotics can provide higher doses to wounds without toxic systemic effects. This study investigated the effect on infection of combining systemic with local antibiotics via polymethylmethacrylate (PMMA) beads or gel delivery.

An established Staphylococcus aureus contaminated fracture model in rats was used. Wounds were debrided and irrigated six hours after contamination and animals assigned to one of three groups, all of which received systemic antibiotics. One group had local delivery via antibiotic gel, another PMMA beads and the control group received no local antibiotics. After two weeks, bacterial levels were quantified.

Combined local and systemic antibiotics were superior to systemic antibiotics alone at reducing the quantity of bacteria recoverable from each group (p = 0.002 for gel; p = 0.032 for beads). There was no difference in the bacterial counts between bead and gel delivery (p = 0.62).

These results suggest that local antibiotics augment the antimicrobial effect of systemic antibiotics. Although no significant difference was found between vehicles, gel delivery offers technical advantages with its biodegradable nature, ability to conform to wound shape and to deliver increased doses. Further study is required to see if the gel delivery system has a clinical role.

Cite this article: Bone Joint J 2015;97-B:1423–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 527 - 534
1 Apr 2010
Streubel PN Gardner MJ Morshed S Collinge CA Gallagher B Ricci WM

It is unclear whether there is a limit to the amount of distal bone required to support fixation of supracondylar periprosthetic femoral fractures. This retrospective multicentre study evaluated lateral locked plating of periprosthetic supracondylar femoral fractures and compared the results according to extension of the fracture distal with the proximal border of the femoral prosthetic component.

Between 1999 and 2008, 89 patients underwent lateral locked plating of a supracondylar periprosthetic femoral fracture, of whom 61 patients with a mean age of 72 years (42 to 96) comprising 53 women, were available after a minimum follow-up of six months or until fracture healing. Patients were grouped into those with fractures located proximally (28) and those with fractures that extended distal to the proximal border of the femoral component (33).

Delayed healing and nonunion occurred respectively in five (18%) and three (11%) of more proximal fractures, and in two (6%) and five (15%) of the fractures with distal extension (p = 0.23 for delayed healing; p = 0.72 for nonunion, Fisher’s exact test). Four construct failures (14%) occurred in more proximal fractures, and three (9%) in fractures with distal extension (p = 0.51). Of the two deep infections that occurred in each group, one resolved after surgical debridement and antibiotics, and one progressed to a nonunion.

Extreme distal periprosthetic supracondylar fractures of the femur are not a contra-indication to lateral locked plating. These fractures can be managed with internal fixation, with predictable results, similar to those seen in more proximal fractures.