Dual-taper implants provide surgeons with options to optimise patients' anatomy intraoperatively but are at risk of early revision due to adverse tissue reactions to corrosion debris. Risk factors for failure and linkage with symptoms however are not fully understood. We related retrieval findings to clinical and implant variables. This study involved 88 failed dual-taper implants with TMZF femoral stems and cobalt-chromium necks, revised for pain, elevated Co (median = 7.3μg/L) and Cr (median = 2.15μg/L) ions levels and fluid collection on MRI. Stem-neck surfaces were assessed for: 1) severity of corrosion using a published visual method and 2) severity of material lost and location of damage with a roundness-measuring machine. Five traces were taken on each round section of the taper surface at 45° increments to compute the relative depth of damage. The total area of these traces provided a measure of surface damage for comparative purposes. The stem-neck taper junctions were severely corroded; the deepest areas of damage were on the inferior-proximal and superior-distal part of the necks, compatible with cantilever bending. Elemental analysis revealed chromium rich deposits indicative of corrosion processes and metal transfer from the stem to the neck. There was a positive correlation between the severity of damage and time of implantation (p<0.0001). Co and Cr levels in the blood were also strongly correlated (p<0.0001, p=0.0002). No other implant or patient variables were linked. The stem-neck junction was severely corroded in all cases. The severity and location of the areas of surface damage did not link with implant or patient characteristics in this big cohort suggesting that the design and material combination is the predominant source of failure in these designs. Dual-taper hips are severely corroded at the stem-neck junction; this appears to be due to the use of a TMZF alloy stem paired with CoCrMo necks.
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
The rate and mode of
Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high
Aims. Femoral revision component subsidence has been identified as predicting
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Introduction. In the early 2000s hip resurfacing became an established bone conserving hip arthroplasty option particularly for the fit and active patient cohort. The performance of second-generation metal-on-metal bearings had led to the reintroduction of hip resurfacing. The Birmingham Hip resurfacing (BHR) was introduced in 1997. This was followed by a number of different designs of the hip resurfacing. The Durom hip resurfacing was introduced in 2001. These two designs had different metallurgical properties, design parameters particularly clearance and different implantation techniques. Data from joint registries show that both prosthesis perform well. Objectives. Our objective was to perform a retrospective survival analysis comparing the Birmingham to the Durom hip resurfacing and analyse the mode of failures of the cases revised. Methods. Data was collected prospectively but analysed retrospectively. The two cohorts comprised patients treated by two senior surgeons at different units. The follow up range was 1 to 14 years with a mean of 10 years. The end-point was revision for any cause. However this was further substratified. Results. The outcome of all patients was known. The two cohorts exhibited no significant difference in demographics. No failures in either cohort were attributed to adverse reaction to metal debris. Revision for any cause was analysed by plotting Kaplan-Meier Survival curves. The Durom cohort (n=273) had 5 deaths and 9 revisions. The Birmingham cohort (n=567) had 5 deaths and 22 revisions. The Kaplan-Meier survival curves for the two resurfacing designs were different. The Durom cohort demonstrated a concave curve with more
Total hip replacement in patients with advanced osteonecrosis of the femoral head is often complicated by early loosening of the femoral component. Recent evidence has suggested that abnormal bone extending into the proximal femur may be responsible for the
Background. The prevalence of immunocompromised patients undergoing Total Hip Arthroplasty (THA) is increasing worldwide as a consequence of advances in treatment. HIV is presenting in an older population group and concerns of higher rates of infection,
Pseudotumours have recently been reported in non-metal-on-metal total hip replacements (non-MoM THRs), however the magnitude and risk factors for this complication are unknown, as is the outcome of its treatment. 3340 primary THR undergoing revision for pseudotumour between 2008 and 2015 were identified in the National Joint Registry for England and Wales. 7.5% (n=249) of these pseudotumour revisions, had non-MoM bearing surfaces. The risk of revision for pseudotumour in non-MoM hips was 0.032% (249/789,397; 95% CI 0.028%–0.036%). The risk of pseudotumour revision was 2.35 times (95% CI 1.76–3.11) higher in ceramic-on-ceramic compared with hard-on-soft bearings, and 2.80 times (95% CI 1.74–4.36) higher in 36mm metal-on-polyethylene bearings compared to 28mm and 32mm metal-on-polyethylene bearings. The outcome of revision for pseudotumour non-MoM hips was studied in 185 hips. 13.5% (n=25) had re-revisions at a mean of 1.2 years (range 0.1–3.1 years). Infection (32%), dislocation/subluxation (24%), and aseptic loosening (24%) were the commonest indications for re-revision. The 4-year survival rate was 83.8% (95% CI=76.7%–88.9%). Multiple revision indications (Hazard Ratio (HR)=2.78; 95% CI=1.03–7.49) and incomplete revision procedures (HR=5.76; 95% CI=1.28–25.9) increased the risk of re-revision. Although the overall risk of revision for pseudotumour in non-MoM THRs is low, the risk is increasing and is significantly higher in ceramic-on-ceramic and large head metal-on-polyethylene THR. These revisions have a high
Introduction. Ceramic heads are used in hip revision surgery to mitigate corrosion concerns. Manufacturers recommend using a pristine titanium sleeve in conjunction with a well-fixed metal stem to prevent
Ceramic on ceramic (CoC) Total Hip Replacement revisions give specific problems and no consensus is made on the way to perform it. The aim of this study was to report the results of THA CoC revisions and to identify the specific concerns. Between April 2007 and May 2016, 50 consecutive revisions have been performed on average 2.8 years after the THA, in 46 patients (28 males, 18 females) at a mean age of 55 years. The main reasons for revision were loosening (34%), pain without loosening (20%), infection (12%), implant breakage (10%), and instability (8%). There were 41 implant revisions, five revisions without implant change (10%) and four revisions for head change (8%). Revisions were performed without bone reconstruction in 68% of hips and with a CoC couple in 83% of hips. Three patients have died and three were lost to follow up. There were six intraoperative femoral fractures, one early and two late non−recurrent dislocations. Eight hips (16%) had re−revision with implant replacement. At a mean follow-up of 5.5 years, the mean PMA, Harris and WOMAC scores were 14.8, 72 and 23. 80% of patients were satisfied of the procedure and 89% were ready to re-do the surgery if necessary. 16% of hips experienced a noise. There were no complete lucent lines, no osteolysis, and no implant migration. Wear was not measurable. The 6-year survival rate was 82 % and 90 % if the end-point was implant revision for any cause and for fixation failure, respectively. CoC THA failure mainly affects a young population and occurs
The
The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes. We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).Aims
Methods
With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases. A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant – Modular Endo-prosthetic Tumour System (METS).Aims
Methods
Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.Aims
Methods