To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications? A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.Aims
Methods
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
Angular proximal femoral deformities increase the technical complexity of primary total hip arthroplasties (THAs). The goals were to determine the long-term implant survivorship, risk factors, complications, and clinical outcomes of contemporary primary THAs in this difficult cohort. Our institutional total joint registry was used to identify 119 primary THAs performed in 109 patients with an angular proximal femoral deformity between 1997 and 2017. The deformity was related to previous femoral osteotomy in 85%, and developmental or metabolic disorders in 15%. 53% had a predominantly varus angular deformity. The mean age was 44 years, mean BMI was 29 kg/m. 2. , and 59% were female. An uncemented metaphyseal fixation stem was used in 30%, an uncemented diaphyseal fixation stem in 28%, an uncemented modular body stem with metaphyseal fixation sleeve in 24%, and a cemented stem in 18%. Simultaneous corrective femoral osteotomy was performed in 18%. Kaplan-Meier survivorships and Harris hip scores were reported. Mean follow-up was 8 years. The 10-year survivorships free of
We previously reported the five to ten-year results of the Birmingham Hip Resurfacing (BHR) implant. The purpose of this study was to evaluate the survivorship, radiographic results, and clinical outcomes of the BHR at long-term follow-up. We retrospectively reviewed 250 patients from the original cohort of 324 BHRs performed from 2006 to 2013 who met contemporary BHR indications. Of these, 4 patients died and 4 withdrew. From the 242 patients, 224 patients (93%) were available for analysis. Modified Harris hip score (mHHS) and University of California, Los Angeles (UCLA) scores were collected and compared to a matched total hip arthroplasty (THA) cohort. Mean follow-up was 14 years. Survivorship free of aseptic revision was 97.4% and survivorship free of any revision was 96.0% at 15 years. Revisions included 3 periprosthetic joint infections, 2 for elevated metal ions and symptomatic pseudotumor, 2 for aseptic
Although data on uncemented short stems are available, studies on cemented short-stemmed THAs are limited. These cemented short stems may have inferior long-term outcomes and higher femoral component fracture rates. Hence, we examined the long-term follow-up of cemented short Exeter stems used in primary THA. Within the Exeter stem range, 7 stems have a stem length of 125 mm or less. These stems are often used in small patients, in young patients with a narrow femoral canal or patients with anatomical abnormalities. Based on our local database, we included 394 consecutive cemented stems used in primary THA (n=333 patients) with a stem length ≤125 mm implanted in our tertiary referral center between 1993 and December 2021. We used the Dutch Arthroplasty Registry (LROI) to complete and cross-check the data. Kaplan-Meier survival analyses were performed to determine 20-year survival rates with stem revision for any reason, for septic loosening, for aseptic loosening and for femoral component fracture as endpoints. The proportion of male patients was 21% (n=83). Median age at surgery was 42 years (interquartile range: 30–55). The main indication for primary THA was childhood hip diseases (51%). The 20-year stem survival rate of the short stem was 85.4% (95% CI: 73.9–92.0) for revision for any reason and 96.2% (95%CI: 90.5–98.5) for revision for septic loosening. No stems were revised for aseptic
Aims. After failed acetabular fractures, total hip arthroplasty (THA) is a challenging procedure and considered the gold standard treatment. The complexity of the procedure depends on the fracture pattern and the initial fracture management. This study’s primary aim was to evaluate patient-reported outcome measures (PROMs) for patients who underwent delayed uncemented acetabular THA after acetabular fractures. The secondary aims were to assess the radiological outcome and the incidence of the associated complications in those patients. Methods. A total of 40 patients underwent cementless acetabular THA following failed treatment of acetabular fractures. The postoperative clinical and radiological outcomes were evaluated for all the cohort. Results. The median (interquartile range (IQR)) Oxford Hip Score (OHS) improved significantly from 9.5 (7 to 11.5), (95% confidence interval (CI) (8 to 10.6)) to 40 (39 to 44), (95% CI (40 to 43)) postoperatively at the latest follow-up (p < 0.001). It was worth noting that the initial acetabular fracture type (simple vs complex), previous acetabular treatment (ORIF vs conservative), fracture union, and restoration of anatomical centre of rotation (COR) did not affect the final OHS. The reconstructed centre of rotation (COR) was restored in 29 (72.5%) patients. The mean abduction angle in whom acetabular fractures were managed conservatively was statistically significantly higher than the surgically treated patients 42.6° (SD 7.4) vs 38° (SD 5.6)) (p = 0.032). We did not have any case of acetabular or
Aims. A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Methods. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package. Results. In the Durasul group, no patients underwent revision due to loosening or showed radiological evidence of osteolysis. In the Sulene group, four patients (four hips) were revised due to
Hip resurfacing arthroplasty (HRA) is being offered as a viable alternative to total hip arthroplasty (THA). For very young patients (under 30 years old), THA achieves fair results but with high revision and complication rates. This retrospective study was designed to evaluate the results of HRA performed in patients under 30 years old with a long follow-up. During the inclusion period (2007–2021), we collected in a single operator database, all HRA performed in patients under 30 with a minimum follow up of 2 years. 103 HRA in 93 patients (77 male / 16 females; a mean age at operation 27.7 years (18 to 29.9)) were included. The two mean preoperative diagnoses were osteoarthritis in 54 HRA (52.5%) and DDH or hip dislocation in 19.4% (20 cases). No patient was lost to follow-up and 3 patients (5 HRA) deceased. There were 2 revisions (one for
The increasing number of total hip arthroplasty (THA) used in young patients will inevitably lead to more revision procedures at younger ages, especially since the outcome of primary THA in young patients is already inferior compared to older patients. However, these data are lacking in literature. The aim of this study was to determine the survival of both acetabular and femoral components placed during primary and revision hip arthroplasty in patients under 55 years using Dutch Arthroplasty Register (LROI) data. All primary THA registered in the LROI between 2007–2018 in patients under 55 years were selected (n=25,682). Subsequent cup- and stem revision procedures were included. Kaplan-Meier survival analyses were used to estimate the survival probability of primary and revised cup- and stem components. Mean follow-up of primary cups and stems was 5.8 years (SD 3.2) and 5.9 years (SD 3.2), respectively. In total, 659 cup revision procedures and 532 stem revision procedures were registered. Most common reason for cup revision was acetabular loosening (n=163), most common reason for stem revision was
Introduction. The Opera acetabular component (Smith & Nephew Ltd) shares many same design features with the Ogee (Depuy Int.). Differences are a malleable flange, instrumentation designed to improve cement pressurisation and the specifics of the UHMW-poly used. Methods. Data was collected prospectively on 419 consecutive replacements (390 patients), performed between March 2000 and February 2005. A posterior approach was used, and the acetabulum was prepared with multiple key holes. Palacos-R bone cement containing Gentamicin was used with a cement pressuriser. Results. There were 243 Operas (221 patients) and 176 Ogees (169 patients). There were 262 in female patients (62.5%) and age at surgery averaged 68.3yrs (25–92). Average follow-up of surviving patients was 185 months (165 – 225). Acetabular abduction angle in both groups averaged 46° (25–65). Opera group: 27 (11.1%) acetabular components loosened aseptically, all associated with rapid wear. 16 (6.6%) acetabular components were revised (15 (6.2%) - aseptic loosening and 1 late infection. At 10 years 85% were graded Hodgkinson 0 or 1. Six
Aims. Adult patients with history of childhood infection pose a surgical challenge for total hip arthroplasty (THA) due to distorted bony anatomy, soft-tissue contractures, risk of reinfection, and relatively younger age. Therefore, the purpose of the present study was to determine clinical outcome, reinfection rate, and complications in patients with septic sequelae after THA. Methods. A retrospective analysis was conducted of 91 cementless THAs (57 male and 34 female) performed between 2008 and 2017 in patients who had history of hip infection during childhood. Clinical outcome was measured using Harris Hip Score (HHS) and Modified Merle d’Aubigne and Postel (MAP) score, and quality of life (QOL) using 12-Item Short Form Health Survey Questionnaire (SF-12) components: Physical Component Score (PCS) and Mental Component Score (MCS); limb length discrepancy (LLD) and radiological assessment of the prosthesis was performed at the latest follow-up. Reinfection and revision surgery after THA for any reason was documented. Results. There was significant improvement in HHS, Modified Merle d’Aubigne Postel hip score, and QOL index SF 12-PCS and MCS (p < 0.001) and there was no case of reinfection reported during the follow-up. The minimum follow-up for the study was three years with a mean of 6.5 (SD 2.3; 3 to 12). LLD decreased from a mean of 3.3 cm (SD 1) to 0.9 cm (SD 0.8) during follow-up. One patient required revision surgery for
Aims. We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. Methods. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason. Results. During this period of study 144 patients (157 hips) died (69.78%). Two patients were lost to follow-up, leaving 61 patients (66 hips) available for review (29.33%). Of the 225 hips, three underwent revision for femoral failure with osteolysis. One underwent femoral component revision for treatment of a periprosthetic fracture. Eight underwent revision of the acetabular component only for loosening. Two hips had both components revised, when components were found to be loose at time of revision for acetabular loosening, though no radiological femoral osteolysis. Two patients underwent revision for infection. Using
Aims. Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA. Methods. A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified. Results. The AOANJRR reported two revisions: one due to infection, and the second due to
Total hip joint replacement (THJR) for high riding congenital hip dislocation (CDH) is often performed in young patients, and presents unique problems with acetabular cup placement and leg length inequality. A database and the NZ Joint Registry were used to identify 76 hips in 57 patients with a diagnosis of CDH who underwent THJR in the Wellington region between 1994 and 2015. Records and radiographs of 46 hips in 36 patients classified pre-operatively as Crowe II, III or IV were reviewed. Surgical technique used a direct lateral approach, the uncemented acetabular component was located in the anatomic hip center and a primary femoral stem was used in all but one hip. Whether a step-cut sub-trochanteric femoral osteotomy was performed depended on degree of correction, tension on the sciatic nerve, and restoration of leg length. For the 36 patients classified as Crowe II or higher, the average age at operation was 44 years (26 – 66), female:male ratio was 4.5:1 and follow-up averaged 10 years (2 – 22.3). Of the 15 hips classified as Crowe IV, 10 required a step-cut sub-trochanteric femoral osteotomy to shorten the femur, but 5 were lengthened without undo tension on the sciatic nerve. Nine Crowe IV hips received a conventional proximally coated tapered primary femoral component. Oxford hip scores for 76% of patients was excellent (> 41/48), and 24% had good scores (34 – 41). All femoral osteotomies healed. Five hips have been revised, one at 2 years for
Introduction. The BIRMINGHAM HIP. ◊. Resurfacing is a metal-on-metal (MOM) hip implant system approved by the US FDA in 2006. The approval required a multicenter, prospective, post-approval study (PAS). Our purpose is to report the current minimum 10-year results. Methods. 253 patients (280 hips) had surgery between October 2006 and December 2009 at one of 5 sites. We report revisions, survivorship, EQ-5D, Harris Hip Score (HHS), radiographic findings, and metal levels including cobalt (Co) and chromium (Cr). The mean age at surgery was 51 years, 74% male, BMI 28, osteoarthritis 95%. 243 (87%) of hips have known outcome or 10-year minimum follow-up (fup). Prior to 10 years, 5 patients died, 20 hips were revised, and 37 hips did not complete 10-year fup. Results. The 10-year component survivorship is 92.9% (95% CI 89.8–96.1%). The reasons for revision were femoral neck fracture (3),
Adverse spinopelvic mobility (SPM) has been shown to increase risk of dislocation of primary total hip arthroplasty (THA). In patients undergoing THA, prevalence of adverse SPM has been shown to be as high as 41%. Stiff lumbar spine, large posterior standing pelvic tilt and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Dislocation rates for dual mobility articulations have been reported to be 0% to 1.1%. The aim of this study was to determine the early survivorship from the Australian National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a dual mobility articulation. A multicentre study was performed using data from 229 patients undergoing primary THA, enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameters had a dual mobility articulation inserted at the time of their surgery. Average age was 76 (22 to 93) years and 63% were female. At a mean of 2.1 (1 – 3.3) years post-op, the AOANJRR was analysed for follow-up. Reasons for revision and types of revision were identified. The AOANJRR reported two revisions. One due to infection and the second due to
The aim of this study was to compare open reduction and internal fixation (ORIF) with revision surgery for the surgical management of Unified Classification System (UCS) type B periprosthetic femoral fractures around cemented polished taper-slip femoral components following primary total hip arthroplasty (THA). Data were collected for patients admitted to five UK centres. The primary outcome measure was the two-year reoperation rate. Secondary outcomes were time to surgery, transfusion requirements, critical care requirements, length of stay, two-year local complication rates, six-month systemic complication rates, and mortality rates. Comparisons were made by the form of treatment (ORIF vs revision) and UCS type (B1 vs B2/B3). Kaplan-Meier survival analysis was performed with two-year reoperation for any reason as the endpoint.Aims
Methods
Introduction. Cementation of a new liner into an existing well-fixed acetabular component is common during revision total hip arthroplasties (THAs) for many indications, but most commonly for lack of a modern compatible crosslinked polyethylene liner. However, little is known about the long-term durability of this strategy. The purpose of this study was to evaluate the long-term implant survivorship, risk of complications, clinical outcomes, and radiographic results of cementing a new highly cross-linked polyethylene (HXLPE) liner into a well-fixed acetabular component. Methods. We retrospectively identified 326 revision THAs where a non-constrained HXLPE liner was cemented into a well-fixed acetabular component. Mean age at revision THA was 63 years, with 50% being female. The most common indications for revision THA were wear and osteolysis (49%), aseptic
Introduction. Many surgeons are reluctant to use a constrained liner at the time of acetabular component revision given concerns this might result in early acetabular component loosening. We hypothesized that with appropriate initial implant stabilization of highly porous acetabular components with supplemental screw fixation, constrained liners could be safely used at the time of acetabular revision. Methods. We retrospectively identified 148 revision total hip arthroplasties (THAs) where a constrained liner of one design was cemented into a newly placed highly porous acetabular component fixed with supplemental screws (mean 5 screws). Mean age at revision THA was 69 years, with 68% being female. The most common indications for revision were two-stage re-implantation (33%), recurrent dislocation (30%), and aseptic loosening (22% acetabular; 9% acetabular/femoral component). Mean follow-up was 8 years. Results. There were no failures at the bone-implant interface, and there were no revisions for aseptic loosening of the acetabular component. Furthermore, all acetabular components were bone ingrown on radiographic analysis. The 10-year survivorships free from any acetabular revision and free from any reoperation were 75% and 67%, respectively. Overall, 33 hips (22%) required revision or reoperation for infection/wound complications (n=12), dislocation (n=11), periprosthetic femur fracture (n=4),
The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented tapered femoral component in primary total hip arthroplasty (THA). The original cohort consisted of 145 consecutive THAs performed by a single surgeon in 138 patients. A total of 37 patients (40 hips) survived a minimum of 30 years, and are the focus of this review. The femoral component used in all cases was a first-generation Taperloc with a non-modular 28 mm femoral head. Clinical follow-up at a minimum of 30 years was obtained on every living patient. Radiological follow-up at 30 years was obtained on all but four.Aims
Methods