Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 75 - 75
1 Dec 2020
Burkhard B Schopper C Ciric D Mischler D Gueorguiev B Varga P
Full Access

Proximal humerus fractures (PHF) are the third most common fractures in the elderly. Treatment of complex PHF has remained challenging with mechanical failure rates ranging up to 35% even when state-of-the-art locked plates are used. Secondary (post-operative) screw perforation through the articular surface of the humeral head is the most frequent mechanical failure mode, with rates up to 23%. Besides other known risk factors, such as non-anatomical reduction and lack of medial cortical support, in-adverse intraoperative perforation of the articular surfaces during pilot hole drilling (overdrilling) may increase the risk of secondary screw perforation. Overdrilling often occurs during surgical treatment of osteoporotic PHF due to minimal tactile feedback; however, the awareness in the surgical community is low and the consequences on the fixation stability have remained unproved. Therefore, the aim of this study was to evaluate biomechanically whether overdrilling would increase the risk of cyclic screw perforation failure in unstable PHF.

A highly unstable malreduced 3-part fracture was simulated by osteotomizing 9 pairs of fresh-frozen human cadaveric proximal humeri from elderly donors (73.7 ± 13.0 ys, f/m: 3/6). The fragments were fixed with a locking plate (PHILOS, DePuy Synthes, Switzerland) using six proximal screws, with their lengths selected to ensure 6 mm tip-to-joint distance. The pairs were randomized into two treatment groups, one with all pilot holes accurately predrilled (APD) and another one with the boreholes of the two calcar screws overdrilled (COD). The constructs were tested under progressively increasing cyclic loading to failure at 4 Hz using a previously developed setup and protocol. Starting from 50 N, the peak load was increased by 0.05 N/cycle. The event of initial screw loosening was defined by the abrupt increase of the displacement at valley load, following its initial linear behavior. Perforation failure was defined by the first screw penetrating the joint surface, touching the artificial glenoid component and stopping the test via electrical contact.

Bone mineral density (range: 63.8 – 196.2 mgHA/cm3) was not significantly different between the groups. Initial screw loosening occurred at a significantly lower number of cycles in the COD group (10,310 ± 3,575) compared to the APD group (12,409 ± 4,569), p = 0.006. Number of cycles to screw perforation was significantly lower for the COD versus APD specimens (20,173 ± 5,851 and 24,311 ± 6,318, respectively), p = 0.019. Failure mode was varus collapse combined with lateral-inferior translation of the humeral head. The first screw perforating the articular surface was one of the calcar screws in all but one specimen.

Besides risk factors such as fracture complexity and osteoporosis, inadequate surgical technique is a crucial contributor to high failure rates in locked plating of complex PHF. This study shows for the first time that overdrilling of pilot holes can significantly increase the risk of secondary screw perforation. Study limitations include the fracture model and loading method. While the findings require clinical corroboration, raising the awareness of the surgical community towards this largely neglected risk source, together with development of devices to avoid overdrilling, are expected to help improve the treatment outcomes.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 62 - 62
1 Apr 2017
Inzana J Münch C Varga P Hofmann-Fliri L Südkamp N Windolf M
Full Access

Background

Osteoporotic fracture fixation in the proximal humerus remains a critical challenge. While the biomechanical benefits of screw augmentation with bone cement are established, minimising the cement volume may help control any risk of extravasation and reduce surgical procedure time. Previous experimental studies suggest that it may be sufficient to only augment the screws at the sites of the lowest bone quality. However, adequately testing this hypothesis in vitro is not feasible.

Methods

This study systematically evaluated the 64 possible strategies for augmenting six screws in the humeral head through finite element simulations to determine the relative biomechanical benefits of each augmentation strategy. Two subjects with varying levels of local bone mineral density were each modeled with a 2-part and 3-part fracture that was stabilised with a PHILOS plate. The biomechanical fixation was evaluated under physiological loads (muscle and joint reaction forces) that correspond to three different motions: 45 degrees abduction, 45 degrees abduction with 45 degrees internal rotation, and 45 degrees flexion.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XI | Pages 11 - 11
1 Apr 2012
Brownson N Rymaszewski L Elliott J
Full Access

The aim of management of an adult distal humeral fracture is to restore mobility, stability and pain-free elbow function. Good results are usually achieved in the majority of fractures treated with ORIF, but the management of comminuted fractures in elderly, frail patients with osteoporotic bone remains controversial. The literature focuses on elbow replacement if stable internal fixation cannot be achieved, with “bag-of-bones” management now rarely discussed eg. key-note paper - 10 successful cases reported by Brown RF & Morgan RG in 1971 (JBJS 53-B(3):425-428). We present the experience in two units in which conservative management has been actively adopted in selected cases by consultants with a subspecialty interest in the elbow.

All patients over the age of 60 with distal humeral fractures (2007 – 2009) who had been treated conservatively were reviewed clinically and radiologically. Duration of follow-up and outcome, including the Oxford and quick DASH scores, were recorded, with the fractures classified using the AO system.

There were 25 patients, 19 female and 6 male. 19/25 patients have been successfully treated conservatively with a mean Range Of Movement: Extension/Flexion: 45/125, Pronation/Supination 74/70. Only 5 underwent subsequent total elbow replacement and one delayed ORIF. There is a significant complication rate following surgical treatment with ORIF or elbow replacement in elderly, frail patients, including infection, painful non-union and/or stiffness. We believe that there is a role for initial conservative treatment in selected patients with low, displaced, comminuted humeral fractures in osteoporotic bone. Initial early mobilisation as pain allows can give good functional results without the risks of operation. It does not preclude future surgery if conservative treatment fails, but this is not required in the majority of cases.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 103 - 103
1 Apr 2017
Kocialkowski C Peach C
Full Access

Background

Functional outcomes of conservatively managed, valgus impacted proximal humeral fractures are poor. Operative fixation can improve results but can be technically challenging. We prospectively assessed outcomes following fixation of valgus unstable proximal humeral fractures with a novel hybrid fixed angle blade plate, at a minimum of 12 months follow up.

Methods

A new hybrid fixed angle blade plate design (Fx Plate, Exactech, Gainesville, USA) was used by the senior author to treat patients with displaced and potentially valgus unstable proximal humeral fractures, at our institution between November 2012 and March 2014. The study was approved by the research and development department. Outcomes were prospectively assessed using the Oxford and Constant shoulder scores and quality of life was assessed using the SF-36 questionnaire. Radiographs were taken at regular intervals to assess fracture healing.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 36 - 36
4 Apr 2023
Pastor T Zderic I van Knegsel K Link B Beeres F Migliorini F Babst R Nebelung S Ganse B Schöneberg C Gueorguiev B Knobe M
Full Access

Proximal humeral shaft fractures are commonly treated with long straight locking plates endangering the radial nerve distally. The aim of this study was to investigate the biomechanical competence in a human cadaveric bone model of 90°-helical PHILOS plates versus conventional straight PHILOS plates in proximal third comminuted humeral shaft fractures.

Eight pairs of humeral cadaveric humeri were instrumented using either a long 90°-helical plate (group1) or a straight long PHILOS plate (group2). An unstable proximal humeral shaft fracture was simulated by means of an osteotomy maintaining a gap of 5cm. All specimens were tested under quasi-static loading in axial compression, internal and external rotation as well as bending in 4 directions. Subsequently, progressively increasing internal rotational loading until failure was applied and interfragmentary movements were monitored by means of optical motion tracking.

Flexion/extension deformation (°) in group1 was (2.00±1.77) and (0.88±1.12) in group2, p=0.003. Varus/valgus deformation (°) was (6.14±1.58) in group1 and (6.16±0.73) in group2, p=0.976. Shear (mm) and displacement (°) under torsional load were (1.40±0.63 and 8.96±0.46) in group1 and (1.12±0.61 and 9.02±0.48) in group2, p≥0.390. However, during cyclic testing shear and torsional displacements and torsion were both significantly higher in group 1, p≤0.038. Cycles to catastrophic failure were (9960±1967) in group1 and (9234±1566) in group2, p=0.24.

Although 90°-helical plating was associated with improved resistance against varus/valgus deformation, it demonstrated lower resistance to flexion/extension and internal rotation as well as higher flexion/extension, torsional and shear movements compared to straight plates. From a biomechanical perspective, 90°-helical plates performed inferior compared to straight plates and alternative helical plate designs with lower twist should be investigated in future paired cadaveric studies.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 37 - 37
4 Apr 2023
Pastor T Zderic I van Knegsel K Richards G Gueorguiev B Knobe M
Full Access

Proximal humeral shaft fractures are commonly treated with long straight plates or intramedullary nails. Helical plates might overcome the downsides of these techniques as they are able to avoid the radial nerve distally. The aim of this study was to investigate in an artificial bone model: (1) the biomechanical competence of different plate designs and (2) to compare them against the alternative treatment option of intramedullary nails.

Twenty-four artificial humeri were assigned in 4 groups and instrumented as follows: group1 (straight 10-hole-PHILOS), group2 (MULTILOCK-nail), group3 (45°-helical-PHILOS) and group4 (90°-helical-PHILOS). An unstable proximal humeral shaft fracture was simulated. Specimens were tested under quasi-static loading in axial compression, internal/external rotation and bending in 4 directions monitored by optical motion tracking.

Axial displacement (mm) was significantly lower in group2 (0.1±0.1) compared to all other groups (1: 3.7±0.6; 3: 3.8±0.8; 4: 3.5±0.4), p<0.001. Varus stiffness in group2 (0.8±0.1) was significantly higher compared to groups1+3, p≤0.013 (1: 0.7±0.1; 3: 0.7±0.1; 4: 0.8±0.1). Varus bending (°) was significantly lower in group2 compared to all other groups (p<0.001) and group4 to group1, p=0.022. Flexion stiffness in group1 was significantly higher compared to groups2+4 (p≤0,03) and group4 to group1, p≤0,029 (1: 0.8±0.1; 2: 0.7±0.1; 3: 0.7±0.1; 4: 0.6±0.1). Flexion bending (°) in group4 was higher compared to all other groups (p≤0.024) and lower in group2 compared to groups1+4, p≤0.024. Torsional stiffness remained non significantly different, p≥0.086. Torsional deformation in group2 was significantly higher compared to all other groups, p≤0.017. Shear displacement remained non significantly different, p≥0.112.

From a biomechanical perspective, helical plating with 45° and 90° may be considered as a valid alternative fixation technique to standard straight plating of proximal third humeral fractures. Intramedullary nails demonstrated higher axial and bending stiffness as well as lower fracture gap movements during axial loading compared to all plate designs. However, despite similar torsional stiffness they were associated with higher torsional movements during internal/external rotation as compared to all investigated plate designs.