Introduction. Aseptic loosening of the acetabular cup in total hip replacement (THR) remains a major problem. Current diagnostic imaging techniques are ineffective at detecting early loosening, especially for the acetabular component. The aim of this preliminary study was to assess the viability of using a vibration analysis technique to accurately detect acetabular component loosening. Methods. A simplified acetabular model was constructed using a Sawbones foam block into which an acetabular cup was fitted. Different levels of loosening were simulated by the interposition of thin layer of silicon between the acetabular component and the Sawbones block. This included a simulation of a secure (stable) fixation and various combinations of cup zone loosening. A constant amplitude sinusoidal excitation with a sweep range of 100–1500 Hz was used. Output vibration from the model was measured using an accelerometer and an ultrasound probe.
Pedicle screw loosening in posterior instrumentation of thoracolumbar spine occurs up to 60% in osteoporotic patients. These complications may be alleviated using more flexible implant materials and novel designs that could be optimized with reliable computational modeling. This study aimed to develop and validate non-linear homogenized finite element (hFE) simulations to predict pedicle screw toggling. Ten cadaveric vertebral bodies (L1-L5) from two female and three male elderly donors were scanned with high-resolution peripheral quantitative computed tomography (HR-pQCT, Scanco Medical) and instrumented with pedicle screws made of carbon fiber-reinforced polyether-etherketone (CF/PEEK). Sample-specific 3D-printed guides ensured standardized instrumentation, embedding, and loading procedures. The samples were biomechanically tested to failure in a toggling setup using an electrodynamic testing machine (Acumen, MTS) applying a quasi-static cyclic testing protocol of three ramps with exponentially increasing peak (1, 2 and 4 mm) and constant valley displacements. Implant-bone kinematics were assessed with a stereographic 3D motion tracking camera system (Aramis SRX, GOM). hFE models with non-linear, homogenized bone material properties including a strain-based damage criterion were developed based on intact HR-pQCT and instrumented 3D C-arm scans. The experimental loading conditions were imposed, the maximum load per cycle was calculated and compared to the experimental results. HR-pQCT-based bone volume fraction (BV/TV) around the screws was correlated with the experimental peak forces at each displacement level.Introduction
Method
Total joint replacement (TJR) was one of the most revolutionary breakthroughs in joint surgery. The majority studies had shown that most implants could last about 25 years, anyway, there is still variation in the longevity of implants. In US, for all the hip revisions from 2012 to 2017 in the United States, 12.0% of the patients were diagnosed as aseptic loosening. Variable studies have showed that any factor that could cause a systemic or partial bone loss, might be the risk of periprosthetic osteolysis and aseptic loosening. Breast cancer is the most frequent malignancy in women, more than 2.1 million women were newly diagnosed with breast cancer, 626,679 women with breast cancer died in 2018. It's been reported that the mean incidence of THA was 0.29% for medicare population with breast cancer in USA, of which the incidence was 3.46% in Norwegian. However, the effects of breast cancer chemotherapy and hormonotherapy, such as aromatase inhibitors (AI), significantly increased the risk of osteoporosis, and had been proved to become a great threat to hip implants survival. In this case, a 46-year-old female undertook chemotherapy and hormonotherapy of breast cancer 3 years after her primary THA, was diagnosed with aseptic loosening of the hip prosthesis. Her treatment was summarized and analyzed. Breast cancer chemotherapy and hormonotherapy might be a threat to the stability of THA prosthesis. More attention should be paid when a THA paitent occurred with breast cancer. More studies about the effect of breast cancer treatments on skeleton are required.
Up to 20% of patients can remain dissatisfied following TKR. A proportion of TKRs will need early revision with aseptic loosening the most common. The ATTUNE TKR was introduced in 2011 as successor to its predicate design The PFC Sigma (DePuy Synthes, Warsaw, In). However, following reports of early failures of the tibial component there have been ongoing concerns of increased loosening rates with the ATTUNE TKR. In 2017 a redesigned tibial baseplate (S+) was introduced, which included cement pockets and an increased surface roughness to improve cement bonding. Given the concerns of early tibial loosening with the ATTUNE knee system, this study aimed to compare revision rates and those specific to aseptic loosening of the ATTUNE implant in comparison to an established predicate as well as other implant designs used in a high-volume arthroplasty centre. The Attune TKR was introduced to our unit in December 2011. Prior to this we routinely used a predicate design with an excellent long-term track record (PFC Sigma) which remains in use. In addition, other designs were available and used as per surgeon preference. Using a prospectively maintained database, we identified 10,202 patients who underwent primary cemented TKR at our institution between 01/04/2003–31/03/2022 with a minimum of 1 year follow-up (Mean 8.4years, range 1–20years): 1) 2406 with ATTUNE TKR (of which 557 were S+) 2) 4652 with PFC TKR 3) 3154 with other cemented designs. All implants were cemented using high viscosity cement. The primary outcome measures were all-cause revision, revision for aseptic loosening, and revision for tibial loosening. Kaplan-Meier survival analysis and Cox regression models were used to compare the primary outcomes between groups. Matched cohorts were selected from the ATTUNE subsets (original and S+) and PFC groups using the nearest neighbor method for radiographic analysis. Radiographs were assessed to compare the presence of radiolucent lines in the Attune S+, standard Attune, and PFC implants.Abstract
Objective
Methods
For patients who took joint replacement, one of the complications, aseptic joint loosening, could cause a high risk of revision surgery. Studies have shown that MSCs have the ability of homing and differentiating, and also have highly effective immune regulation and anti-inflammatory effects. However, few studies had focused on the stem cells in preventing the occurrence and development of aseptic loosening. In this research, we aimed to clarify whether human umbilical cord mesenchymal stem cells could inhibited the aseptic joint loosening caused by wear particles. A Cranial osteolysis mice model was established on mice to examine the effect of hUC-MSCs on the Titanium particles injection area through micro-CT. The amount of stem cells injected was 2 × 10 5 cells. One week later, the mouse Cranial were obtained for micro-CT scan, and then stained with HE analysis immunohistochemical analysis of TNF-α, CD68, CCL3 and Il-1β. All mice were free of fever and other adverse reactions, and there was no death occurred. Titanium particles caused the osteolysis at the mice cranial, while local injection of hUC-MSCs did inhibit the cranial osteolysis, with a lower BV/TV and a higher porosity. Immunohistochemical results suggested that the expression of TNF-α, CD68, CCL3 and Il-1β in the cranial in Titanium particles mice increased significantly, but was significantly reduced in mice injected with hUC-MSCs. The inhibited CD68 expression indicated that the number of macrophage was lower, which might be a result of the inhibition of CCL3. According to the studies above, HUC-MSCs treatment of mouse cranial osteolysis model can significantly reduce osteolysis, inhibit macrophage recruitment, alleviate inflammatory response, without causing adverse reactions. It may become a promising treatment of aseptic joint loosening.
The patients with a total hip arthroplasty is growing in world manly in Europe and USA, and this solution present a high success at 10years in several orthopaedic registers. The application of total press-fit hip fixation presents the most used solution, but presents some failures associated to the acetabular component fixation, associated to the load transfer and bone loss at long term. The aim of this work is to investigate the influence of different acetabular bone loss in the strain distribution in iliac bone. To evaluate implant fixation, an experimental study was performed using acetabular press-fit component simulating different acetabular bone loss and measuring the strain distribution. The experimental samples developed was based in an iliac bone model of Sawbones supplier and a acetabular component Titanium (Stryker) in a condition press-fit fixation and was implanted according surgical procedure with 45º inclination angle and 20º in the anteversion angle. Were developed five models with same initial bone, one with intact condition simulating the cartilage between bones and four with different bone loss around the acetabular component. These four models representing the evolution of bone support of acetabular components presented in the literature. The evolution of bone loss was imposed with a CAD CAM process in same iliac bone model. The models were instrumented with 5 rosettes in critical region at the cortical bone to measure the strain evolution along the process.Introduction and Objective
Materials and Methods
Surgical treatment of fragility sacrum fractures with percutaneous sacroiliac (SI) screw fixation is associated with high failure rates in terms of screw loosening, cut-through and turn-out. The latter is a common cause for complications, being detected in up to 20% of the patients. The aim of this study was to develop a new screw-in-screw concept and prototype implant for fragility sacrum fracture fixation and test it biomechanically versus transsacral and SI screw fixations. Twenty-seven artificial pelves with discontinued symphysis and a vertical osteotomy in zone 1 after Denis were assigned to three groups (n = 9) for implantation of their right sites with either an SI screw, the new screw-in-screw implant, or a transsacral screw. All specimens were biomechanically tested to failure in upright position with the right ilium constrained. Validated setup and test protocol were used for complex axial and torsional loading, applied through the S1 vertebral body. Interfragmentary movements were captured via optical motion tracking. Screw motions in the bone were evaluated by means of triggered anteroposterior X-rays. Interfragmentary movements and implant motions in terms of pull-out, cut-through, tilt, and turn-out were significantly higher for SI screw fixation compared to both transsacral screw and screw-in-screw fixations. In addition, transsacral screw and screw-in-screw fixations revealed similar construct stability. Moreover, screw-in-screw fixation successfully prevented turn-out of the implant, that remained at 0° rotation around the nominal screw axis unexceptionally during testing. From biomechanical perspective, fragility sacrum fracture fixation with the new screw-in-screw implant prototype provides higher stability than with the use of one SI screw, being able to successfully prevent turn-out. Moreover, it combines the higher stability of transsacral screw fixation with the less risky operational procedure of SI screw fixation and can be considered as their alternative treatment option.
Considerable evidence exists that aseptic loosening is initiated by wear particles that recruit macrophages and stimulate their production of pro-inflammatory cytokines. The cytokines primarily act indirectly by inducing production of RANKL, which stimulates osteoclast differentiation, osteolysis, and inflammatory bone loss. There is also considerable evidence that activation of macrophage Toll-like Receptors (TLRs) contributes to this cascade of events. It is however controversial whether bacterially-derived immunostimulatory molecules known as Pathogen-Associated Molecular Patterns (PAMPs) can contribute to aseptic loosening by stimulating their cognate TLRs on macrophages. Priming and subsequent activation of the NLRP3 inflammasome is essential for macrophage production of mature, active IL-1β in response to wear particles. We recently confirmed that wear particles can activate pre primed NLRP3 inflammasomes in the absence of PAMPs. Thus, activation of the NLRP3 inflammasome is the only macrophage-based event in the aseptic loosening cascade that we have found to date is independent of PAMPs. In contrast, priming of the NLRP3 inflammasome by wear particles requires PAMPs as well as their cognate TLRs. These results add to the growing body of evidence that bacterially-derived PAMPs can contribute to aseptic loosening.
Mechanisms underlying implant failure remain incompletely described, though the presence of macrophage-mediated inflammatory reactions is well documented. Hypoxia has a critical role in many diseases and is known to be interdependent with inflammation. Metals used for joint replacements have also been reported to provoke hypoxia-like conditions. In view of this, we aim to investigate hypoxia-associated factors in aseptic loosening and osteoarthritis with a focus on macrophages. Western blotting, calorimetric assay, haematoxylin-eosin staining, immunohistochemistry, double-immunofluorescence and transmission electron microscopy were performed on capsular tissue obtained from patients undergoing primary implantation of a total hip replacement for osteoarthritis and from patients undergoing revision surgery for aseptic loosening to investigate the presence of hypoxia-associated factors.Background
Methods
Inflammation and chemokines play a pivotal role in aseptic loosening (AL) and prosthetic joint infection (PJI) of total joint arthroplasty (TJA). Recently, the Duffy antigen receptor for chemokines (DARC) on erythrocytes was identified as a potent chemokine receptor able to bind and carry without deactivating a wide range of CXC and CC chemokines from circulation to tissues. The role of DARC and its functional polymorphism (SNP) influencing the number of the DARC molecules on the erythrocytes in AL/PJI has not been studied yet. We genotyped functional polymorphism in the DARC gene (rs12075) using MassArray technology (Agena Bioscience) in 354 patients with TJA (hip and knee arthroplasties). Patients were further subdivided into those with a complication (AL, n = 110; PJI, n = 126) and a control group without complications for at least 10 years (n = 118). Statistics was performed by Plink 1.07 and relative entropy.Background
Methods
Aseptic loosening is a major cause of failure of total hip arthroplasty. The adverse tissue response to prosthetic wear particles, with activation of cytokine and prostanoid production, contributes to bone loss around the implants. We have investigated the possibility that inducible nitric oxide synthase (iNOS) and cyclo-oxygenase-2 (COX-2) are expressed in macrophages in the pseudomembrane at the bone-implant interface, thereby contributing to the periprosthetic bone resorption. We also assessed whether peroxynitrite, a nitric oxide (NO)-derived oxidant associated with cellular injury, is generated in the membrane. Enzymatic activity of iNOS was measured using the arginine-citrulline assay technique and prostaglandin E2 (PGE2), as an indicator of COX-2 activity, was measured using an enzyme immunoassay. Cellular immunoreactivity for iNOS, nitrotyrosine (a marker of peroxynitrite-induced cellular injury) and COX-2 was assessed by quantitative peroxidase immunocytochemistry while immunofluorescence methods were used for subsequent co-localisation studies with CD68+ macrophages. The presence of calcium-independent iNOS activity and PGE2 production was confirmed in the homogenized interface membrane. Immunocytochemistry showed that periprosthetic CD68+ wear-debris-laden macrophages were the most prominent cell type immunoreactive for iNOS, nitrotyrosine and COX-2. Other periprosthetic inflammatory and resident cell types were also found to immunolocalise nitrotyrosine thereby suggesting peroxynitrite-induced protein nitrosylation and cellular damage not only in NO-producing CD68+ macrophages, but also in their neighbouring cells. These data indicate that both iNOS and COX-2 are expressed by CD68+ macrophages in the interface membrane and peroxynitrite-induced cellular damage is evident in such tissue. If high-output NO and peroxynitrite generation were to cause macrophage cell death, this would result in the release of phagocytosed wear debris into the extracellular matrix. A detrimental cycle of events would then be established with further phagocytosis by newly-recruited inflammatory cells and subsequent NO, peroxynitrite and prostanoid synthesis. Since both NO and have been implicated in the induction and PGE2 maintenance of chronic inflammation with resulting loss of bone, and peroxynitrite in the pathogenesis of disease states, they may be central to the pathogenesis of aseptic loosening.
We analysed impaction bone grafting used together with cemented or uncemented fixation in acetabular revision surgery. The overall risk for re-revision did not differ between the cemented and uncemented group. However, aseptic loosening was more common in the cemented group. Several surgical techniques address bone defects in cup revision surgery. Bone impaction grafting, introduced more than thirty years ago, is a biologically and mechanically appealing method. The primary aim of this study was to evaluate the effect of bone impaction grafting when used with uncemented and cemented fixation in cup revision surgery. Uncemented cups resting on more than 50% host bone were used as controls.Summary Statement
Background
This study investigates the use of porous biphasic ceramics as graft extenders in impaction grafting of the femur during revision hip surgery. Impaction grafting of the femur was performed in four groups of sheep. Group one received pure allograft, group two 50% allograft and 50% BoneSave, group three 50% allograft and 50% BoneSave type 2 and group four 10% allograft and 90% BoneSave as the graft material. Function was assessed using an index of pre- and post-operative peak vertical ground reaction force ratios. Changes in bone mineral density were measured by dual energy X ray absorptiometry (DEXA) scanning.
We analysed revised Mathys isoelastic polyacetal femoral stems with stainless-steel heads and polyethylene acetabular cups from eight patients in order to differentiate various types of particle of wear debris.
Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons.Objectives
Methods
The Capital Hip implant was a Charnley-based system which included a flanged and a roundback stem, both of which were available in stainless steel and titanium. The system was withdrawn from the market because of its inferior performance. However, all four of the designs did not produce poor rates of survival. Using a simulated-based, finite-element analysis, we have analysed the Capital Hip system. Our aim was to investigate whether our simulation was able to detect differences which could account for the varying survival between the Capital Hip designs, thereby further validating the simulation. We created finite-element models of reconstructions with the flanged and roundback Capital Hips. A loading history was applied representing normal walking and stair-climbing, while we monitored the formation of fatigue cracks in the cement. Corresponding to the clinical findings, our simulation was able to detect the negative effects of the titanium material and the flanged design in the Capital Hip system. Although improvements could be made by including the effect of the roughness of the surface of the stem, our study increased the value of the model as a predictive tool for determining failure of an implant.