Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 119 - 119
1 May 2016
Donaldson T Gregorius S Burgett-Moreno M Clarke I
Full Access

This study presents an unusual recurrent case of pigmented villonodular synovitis (PVNS) around a ceramic-on-metal (COM) hip retrieved at 9-years. PVNS literature relates to metal-polyethylene and ceramic-ceramic bearings. Amstutz reported 2 cases with MOM resurfacing and Xiaomei reported PVNS recurring at 14 years with metal-on-polyethylene THA. Friedman reported on PVNS recurrence in a ceramic THA. Ours may be the first reported case of recurrent PVNS of a ceramic-on-metal articulation. This young female patient (now 38-years of age) had a total hip replacement in 2006 for PVNS in her left hip. In her initial work-up, this case was presumed to be a pseudotumor problem, typical of those related to CoCr debris with high metal-ion concentrations. She had an CoCr stem (AML), 36mm Biolox-delta head (Ceramtec), and a Pinnacle acetabular cup with CoCr liner (Ultramet, Depuy J&J). This patient had no concerns regarding subluxation, dislocation or squeaking. Three years ago she complained of mild to moderate groin and thigh pain in her left hip. This worsened in the past year. She noticed increased swelling now with an asymmetry to her right hip. She went to the emergency room in Dec-2014 and was referred to a plastic surgeon. In our consult we reviewed MARS-MRI and CT-scans that demonstrated multiple mass lesions surrounding the hip. Laboratory results presented Co=0.7, Cr=0.3 ESR=38 and Crp=0.3. At revision surgery, the joint fluid was hemorrhagic/bloody with hemosiderin staining the soft tissues. Multiple large 4–5×5cm nodules were present in anterior aspect of the hip as well as multiple nodules surrounding posterior capsule and sciatic nerve. Pathology demonstrated a very cellular matrix with hemosiderin-stained tissue and multiple giant cells, which was judged consistent with PVNS. The trunnion showed no fretting, no contamination and no discoloration. The superior neck showed impingement due to low-inclination cup. There was minimal evidence of metal-debris staining the tissues. There was a large metallic-like stripe across the ceramic head. This is a particularly interesting case and may be the first reported recurrent PVNS around a ceramic-on-metal bearing (COM). Data is scant regarding clinical results of COM bearings and here we have a nine-year result in a young and active female patient. She was believed to have a metalosis-related pseudotumor yet her metal-ion levels were not alarmingly high and there was no particular evidence of implant damage or gross wear products. In addition, the CoCr trunnion appeared pristine. Our work-up continues with analyses of wear and histopath-evidence. This case may demonstrate the need for a broadening of the differential diagnosis when dealing with hip failures


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 494 - 494
1 Dec 2013
Pace F Randelli F Serrao L Banci L
Full Access

Introduction:. We present the mid-term results of our consecutive series of 155 hips treated with ASR XL Acetabular System (ASR-XL) for large-diameter metal-on-metal total hip arthroplasty and with DePuy ASR Hip Resurfacing System (ASR) for hip resurfacing. Methods:. 114 ASR-XL and 41 ASR were implanted between 2004 and 2008 in 145 patients (69 men and 76 women) with a mean age of 57 years. Twenty-one patients (23 hips) resulted lost to follow-up. All patients were recalled and monitored periodically with clinical, hematological and radiological evaluation. Average follow-up of the 155 hips was 72 months (1 to 104). The mean follow-up, excluding revised patients and those lost to follow-up, was up to 89 months. Results:. At latest follow-up we had revised 54 of 155 hips (35%) with complete removal of ASR device. Revision involved 15 hips with ASR (36.5% within ASR group) and 39 hips with ASR-XL (34.2% within ASR-XL group). Main reasons for revision were aseptic loosening in 24 hips (44%), pain or pain associated with elevation of blood metal ions in 9 (17%), elevation of blood metal ions without pain in 9 (17%), deep infection in 4 hips (7%), recurrent dislocation in 1 hip (2%), periprosthetic fractures in 1 hip (2%), unknown because revised elsewhere in 6 (11%). The cumulative survival with revision for any reason as the end-point for ASR and ASR XL were respectively 59.6% and 59.3%. For patients who did not undergone revision, the mean Harris hip score improved to 91 (57 to 100) at five years and the mean satisfaction after the operation was graded 4.4 in a score from 1 to 5. Metal ions plasma concentration analysis was conducted in 83 patients (87 hips). Elevated metal ion levels (>7 μg/l) was found in 39 patients (42 hips, 48%) with average Cr and Co concentrations respectively of 37.3 μg/l and 81.5 μg/l. The remaining 44 patients (45 hips, 52%) had an average Cr and Co concentrations respectively of 1.2 μg/l and 1.9 μg/l. Metal ion levels in revised patients were significantly (p < 0.001) higher than in non-revised patients. MARS-MRI performed in 38 hips with high metal ions revealed pseudotumour formations in 13 hips (34%). No macroscopic evident clinical sign of cobaltism was reported for any patient. Conclusion:. ASR implants survival could have a considerable drop in the mid-term follow-up due to adverse reactions to high blood metal ions concentrations and metal debris. Adverse reactions could include periprosthetic pseudotumour formations and low-detectable periprosthetic osteolysis even in asymptomatic patients