Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and recurrent dislocation and three
other patients required closed manipulation for a single dislocation.
One stem subsided more than 5 mm but then stabilised and required
no further intervention. In this series, a modular, tapered, conical stem provided a versatile
reconstruction solution with a low rate of complications. Cite this article:
Aims. The aim of this study was to identify modifiable risk factors associated with mortality in patients requiring revision total hip arthroplasty (THA) for
Short stems have been developed to conserve bone stock, especially in younger populations undergoing a total hip arthroplasty (THA), and showed similar functional outcomes compared to conventional stems. Recent literature suggested that stem length might be an independent risk factor for acute periprosthetic femoral fracture in direct anterior THA (1) or with different short stem designs (2). The purpose of the present study was to compare the incidence of acute periprosthetic femoral fractures, between stems of the same manufacture (Taperloc microplasty vs Taperloc complete), which have the same stem characteristics, except for the stem length which is 35 mm shorter in Taperloc microplasty during posterior and lateral THA. Our institution's arthroplasty database was searched for all primary total hip arthroplasties utilizing short femoral stems performed between August 2016 and August 2023. Pre-operative X-rays for each case were analysed to characterize the proximal femoral geometry, specifically the canal bone ratio (CBR) and canal flare index (CFI). Data analysis was performed to identify risk factors for periprosthetic fractures. For the time period assessed, 2107 femoral stems (Taperloc Microplasty:1727, Taperloc complete: 380) were implanted. Females constituted 53% of the cohort. The average patient age was 70±11 years. The periprosthetic fracture rate was 0.94%, with 20 periprosthetic fractures (Taperloc Microplasty:17, Taperloc complete: 3) excluding 2 greater trochanteric fractures (1 at each group), identified at follow-up of three months. There was no significant difference between the periprosthetic fracture rates between the two stems (0.98% vs 0.79%, p>0.72) The multivariate regression analysis demonstrated that the stem length, CBR, CFI, age, and gender were not risk factors for periprosthetic fracture. The present study demonstrated that both Taperloc Microplasty and Taperloc complete stems had similar rates of periprosthetic fractures, and the stem length was not a risk factor for a periprosthetic fracture during uncemented THA. 1. Tamaki T, Cementless Tapered-Wedge Stem Length Affects the Risk of Periprosthetic Femoral Fractures in Direct Anterior Total Hip Arthroplasty. J Arthroplasty. 2018 Mar;33(3):805-809. 2. Staunton P. Acute
Aims.
We are currently facing an epidemic of periprosthetic
fractures around the hip. They may occur either during surgery or
post-operatively. Although the acetabulum may be involved, the femur
is most commonly affected. We are being presented with new, difficult
fracture patterns around cemented and cementless implants, and we
face the challenge of an elderly population who may have grossly
deficient bone and may struggle to rehabilitate after such injuries.
The correct surgical management of these fractures is challenging.
This article will review the current choices of implants and techniques
available to deal with periprosthetic fractures of the femur. Cite this article: