Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_12 | Pages 18 - 18
10 Jun 2024
Haston S Langton D Townshend D Bhalekar R Joyce T
Full Access

Despite advancements, revision rates following total ankle replacement (TAR) are high in comparison to other total joint replacements. This explant analysis study aimed to investigate whether there was appreciable metal particulate debris release from various contemporary TARs by describing patterns of material loss. Twenty-eight explanted TARs (9 designs: 3 fixed and 6 mobile bearing), revised for any reason, were studied. The articulating surfaces of the metal tibial and talar components as well as the polyethylene insert were assessed for damage features using light microscopy. Based on the results of the microscopic analysis, scanning electron microscopy with energy dispersive X-ray spectroscopy was performed to determine the composition of embedded debris identified, as well as non-contacting 3D profilometry. Pitting, indicative of material loss, was identified on the articulating surfaces of 54% of tibial components and 96% of talar components. Bearing constraint was not found to be a factor, with similar proportions of fixed and mobile bearing metal components showing pitting. More cobalt-chromium than titanium alloy tibial components exhibited pitting (63% versus 20%). Significantly higher average surface roughness (Sa) values were measured for pitted areas in comparison to unpitted areas of these metal components (p<0.05). Additionally, metallic embedded debris (cobalt-chromium likely due to pitting of the tibial and talar components or titanium likely from loss of their porous coatings) was identified in 18% of polyethylene inserts. The presence of hard 3. rd. body particles was also indicated by macroscopically visible sliding plane scratching, identified on 79% of talar components. This explant analysis study demonstrates that metal debris is released from the articulating surfaces and the coatings of various contemporary TARs, both fixed and mobile bearing. These findings suggest that metal debris release in TARs may be an under-recognised issue that should be considered in the study of painful or failed TAR moving forwards


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 907 - 914
1 Jul 2009
Koivu H Kohonen I Sipola E Alanen K Vahlberg T Tiusanen H

Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities > 10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1366 - 1371
1 Oct 2013
Muir D Aoina J Hong T Mason R

We performed a retrospective review of a consecutive series of 178 Mobility total ankle replacements (TARs) performed by three surgeons between January 2004 and June 2009, and analysed radiological parameters and clinical outcomes in a subgroup of 129 patients. The mean follow-up was 4 years (2 to 6.3). A total of ten revision procedures (5.6%) were undertaken. The mean Ankle Osteoarthritis Scale (AOS) pain score was 17 (0 to 88) and 86% of patients were clinically improved at follow-up. However, 18 patients (18 TARs, 14%) had a poor outcome with an AOS pain score of > 30. A worse outcome was associated with a pre-operative diagnosis of post-traumatic degenerative arthritis. However, no pre- or post-operative radiological parameters were significantly associated with a poor outcome. Of the patients with persistent pain, eight had predominantly medial-sided pain. Thirty TARs (29%) had a radiolucency in at least one zone.

The outcome of the Mobility TAR at a mean of four years is satisfactory in > 85% of patients. However, there is a significant incidence of persistent pain, particularly on the medial side, for which we were unable to establish a cause.

Cite this article: Bone Joint J 2013;95-B:1366–71.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 61 - 65
1 Jan 2010
Morgan SS Brooke B Harris NJ

We present the outcomes in 38 consecutive patients who had total ankle replacement using the Ankle Evolution System with a minimum follow-up of four years. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) score and regular standardised anteroposterior and lateral weight-bearing radiographs were obtained. Patient satisfaction and complications were recorded and the survival of the implants was demonstrated by the Kaplan-Meier method.

The mean follow-up was for 57.8 months (48 to 80). The cumulative survival rate at six years was 94.7% (95% confidence interval 80.3 to 98.7). The mean total AOFAS score was 88.1 (53 to 100). The mean score for pain was 35.8 (20 to 40). Ten patients presented with edge-loading of whom nine had corrective surgery. Two ankles were revised, one to an arthrodesis and the other to replace the tibial component. Nine patients showed radiological evidence of osteolysis. They had minimal non-progressive symptoms and further surgery was not undertaken. Nevertheless, the concerns about osteolysis led to the implant being withdrawn by the manufacturer.

The medium-term results of the ankle evolution system ankle replacement are satisfactory with high patient satisfaction, but the rate of osteolysis is of some concern. The long-term benefit of this procedure has yet to be determined.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 69 - 74
1 Jan 2009
Wood PLR Sutton C Mishra V Suneja R

We describe the results of a randomised, prospective study of 200 ankle replacements carried out between March 2000 and July 2003 at a single centre to compare the Buechel-Pappas (BP) and the Scandinavian Total Ankle Replacement (STAR) implant with a minimum follow-up of 36 months. The two prostheses were similar in design consisting of three components with a meniscal polyethylene bearing which was highly congruent on its planar tibial surface and on its curved talar surface. However, the designs were markedly different with respect to the geometry of the articular surface of the talus and its overall shape.

A total of 16 ankles (18%) was revised, of which 12 were from the BP group and four of the STAR group. The six-year survivorship of the BP design was 79% (95% confidence interval (CI) 63.4 to 88.5 and of the STAR 95% (95% CI 87.2 to 98.1). The difference did not reach statistical significance (p = 0.09). However, varus or valgus deformity before surgery did have a significant effect) (p = 0.02) on survivorship in both groups, with the likelihood of revision being directly proportional to the size of the angular deformity. Our findings support previous studies which suggested that total ankle replacement should be undertaken with extreme caution in the presence of marked varus or valgus deformity.