Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic head
Aims. The aim of the study was to compare two methods of calculating pelvic incidence (PI) and pelvic tilt (PT), either by using the femoral heads or acetabular domes to determine the bicoxofemoral axis, in patients with unilateral or bilateral primary hip osteoarthritis (OA). Methods. PI and PT were measured on standing lateral radiographs of the spine in two groups: 50 patients with unilateral (Group I) and 50 patients with bilateral hip OA (Group II), using the femoral heads or acetabular domes to define the bicoxofemoral axis. Agreement between the methods was determined by intraclass correlation coefficient (ICC) and the standard error of measurement (SEm). The intraobserver reproducibility and interobserver reliability of the two methods were analyzed on 31 radiographs in both groups to calculate ICC and
Introduction. We studied free (= local powder) tobramycin and doxycycline, and controlled release (= local lipid bilayer) doxycycline formulations in a rat model representing a generic joint infection. We . hypothesized. that evidence of infection (quantitative colony forming units (CFU), qualitative
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and
Introduction. Ultra-high molecular weight polyethylene (UHMWPE) can provide local sustained delivery of therapeutics. 1,2. For example, it can deliver analgesics to address post-arthroplasty pain. 2. Given that several analgesics, such as bupivacaine (anesthetic) and tolfenamic acid (NSAID), were shown to possess antibacterial activity against Staphylococci, we hypothesize that analgesic-loaded UHMWPE can also yield antimicrobial effects, preventing the development of periprosthetic joint infections. Methods. Bupivacaine and tolfenamic acid were incorporated into UHMWPE via phase-separated compression molding. Drug release from the prepared samples was measured using high-performance liquid chromatography. Antibacterial studies of the obtained materials were conducted against methicillin-sensitive, and methicillin-resistant S. aureus, as well as S. epidermidis. Time-kill curves were obtained to characterize antimicrobial activity against planktonic bacteria. The dynamics of bacterial adhesion were assessed to characterize antibiofilm activity. Scanning electron microscopy (SEM) was used to visualize adherent bacteria. Anticolonizing activity of the tested materials was characterized using the “daughter cell” method as outlined elsewhere. 3. Cytotoxicity profile of drug-loaded UHMWPEs was evaluated using MG-63 osteoblast cell line. Results. The bupivacaine release rate generally increased with increasing drug loading (e.g. a model knee implant loaded with bupivacaine would release ca. 15–500 mg over 24 hours). While also proportional, drug release from UHMWPE loaded with tolfenamic acid was much lower. The bacterial viability curves showed that bupivacaine-loaded UHMWPE possessed moderate antibacterial activity against planktonic MSSA, MRSA, and S. epidermidis, slowing bacteria proliferation by up to 70%. Bupivacaine-loaded UHMWPE also mitigated biofilm formation and development during the initial culture period.
Introduction. HXLPE acetabular liners were introduced to reduce wear-related complications in THA. However, post-irradiation thermal free radical stabilization can compromise mechanical properties, leave oxidation-prone residual free radicals, or both. Reports of mechanical failure of HXLPE acetabular liner rims raise concerns about thermal free radical stabilization and in vivo oxidization on implant properties. The purpose of this study is to explore the differences in the mechanical, physical and chemical properties of HXLPE acetabular liner rims after extended time in vivo between liners manufactured with different thermal free radical stabilization techniques. Material and methods. Remelted, single annealed and sequentially annealed retrieved HXLPE acetabular liners with in vivo times greater than 4.5 years were obtained from our implant retrieval laboratory. All retrieved liners underwent an identical sanitation and storage protocol. For mechanical testing, a total of 55 explants and 13 control liners were tested. Explant in vivo time ranged from 4.6 – 14.0 years and ex vivo time ranged from 0 – 11.6 years. Rim mechanical properties were tested by microindentation hardness testing using a Micromet II Vickers microhardness tester following ASTM standards. A subset of 16 explants with ex vivo time under one year along with five control liners were assessed for oxidation by FTIR, crystallinity by Raman spectroscopy, and evidence of microcracking by
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
The aim of this study was to perform a cross-cultural adaptation of Oxford Hip Score (OHS) to Indonesian, and to evaluate its psychometric properties. We performed a cross-cultural adaptation of Oxford Hip Score into Indonesian language (OHS-ID) and determined its internal consistency, test-retest reliability, measurement error, floor-ceiling effect, responsiveness, and construct validity by hypotheses testing of its correlation with Harris Hip Score (HHS), vsual analogue scale (VAS), and Short Form-36 (SF-36). Adults (> 17 years old) with chronic hip pain (osteoarthritis or osteonecrosis) were included.Aims
Methods
Introduction. Porous surfaces developed over the past decades have been shown to promote tissue ingrowth. Hydroxyapatite (HA) coatings have been added to these porous coatings in an attempt to further augment bone ingrowth. The development of additive manufacturing techniques has allowed for precision in building these complex porous structures. The effect of supplemental HA coatings on these new surfaces is unclear. The purpose of this study is to evaluate the biological fixation of a novel 3D printed porous implant in a canine model. In addition, we evaluated the effect of different HA coatings on this 3D printed implant. Methods. A canine transcortical model was used to evaluate the performance of three different laser rapid manufacturing (LRM) Ti6Al4V cylindrical implants (5.2 mm diameter, 10mm length): LRM with precipitated hydroxyapatite (P-HA), LRM with plasma sprayed hydroxyapatite (PS-HA), and a hydroxyapatite-free control (No-HA). The implants were 50–60% porous with a mean pore size of 450 μm and have a random interconnected architecture with irregular pore sizes and shapes that are designed based on the structure of cancellous bone. A lateral approach to the femoral diaphysis was used to prepare 5 mm unicortical, perpendicular drill holes in 12 canines. One of each implant type was press-fit into each femur. The femora were harvested at both 4 and 12 weeks post implantation, radiographed and prepared for either mechanical push-out testing to assess the shear strength of the bone-implant interface (left femora, N=6) or for histological processing (right femora, N=6). An un-paired Student's t-test was used to compare statistical significance between the 4 and 12-week results, as well as differences due to implant type; p<0.05 was considered significant. Results. The post-mortem contact radiographs demonstrated substantial condensation of bone around the implants at both 4 and 12 weeks. Bone ingrowth in the canine femora was observed in all implants, with and without HA, at both time periods under backscattered
Introduction. Hip and knee replacements are predictable orthopaedic procedure with excellent clinical outcomes. Discharging or leaking wounds affect length of hospital stay, affect bed planning and predispose to superficial and potentially deep wound infection. Predictable wound healing therefore remains the first hurdle. This trial aims to study the effectiveness of portable disposable incisional negative pressure wound therapy (NPWT) dressings in hip and knee replacements. This trial aims to study the effectiveness of portable disposable incisional negative pressure wound dressings in hip and knee replacements and the impact on wound healing, length of stay and wound complications. Patients/Materials & Methods. Following ethical approval 110 patients each were randomised to ‘Control group’ and ‘Study group’. Patients in control group received traditional dressings and those in study group received an incisional NPWT (PICO) manufactured by Smith & Nephew. Post operatively, state of the wound, level of wound exudate, length of hospital stay and complications were documented. Results. The average length of stay for control group was 4.72 ± 0.69 (Mean ± SEM) and for the study group 3.79±0.19 (mean ±
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
We report two cases of surface deterioration of a zirconia ceramic femoral head associated with phase transformation after total hip arthroplasty. One head was retrieved at revision due to recurrent dislocation after six years and the other because of failure of the locking mechanism of the polyethylene liner after three years. The monoclinic content of the zirconia ceramics rose from 1% to about 30% on the surface of the heads.
We report the long-term radiological results of 58 total hip arthroplasties (THA) using the Charnley offset-bore acetabular socket. Wear was measured at four sites and radiolucent lines and possible migration were recorded. Four cups were retrieved at revision and were examined using light microscopy,
Between March 1990 and May 1991 we performed 85 primary total hip replacements in 74 patients using the Landos Atoll hydroxyapatite (HA)-coated cup and the Corail HA-coated stem. The patients were followed up for a mean of ten years. Of the 85 cups, 26 (31%) have already been revised and a further six are radiologically unstable and awaiting revision. Two femoral stems have been revised for infection without loosening. The retrieved acetabular cups were studied by
Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds. A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.Objectives
Methods
To explore the effect of different types of articulating antibiotic-loaded cement spacers in two-stage revision for chronic hip prosthetic joint infection (PJI). A retrospective cohort study was performed involving 36 chronic PJI patients treated with different types of articulating antibiotic-loaded cement spacers between January 2014 and December 2017. The incidence of complications and the therapeutic effects of different types of antibiotic-loaded articulating cement spacers were compared.Aims
Methods
This study aimed to determine if macrophages can attach and directly affect the oxide layers of 316L stainless steel, titanium alloy (Ti6Al4V), and cobalt-chromium-molybdenum alloy (CoCrMo) by releasing components of these alloys. Murine peritoneal macrophages were cultured and placed on stainless steel, CoCrMo, and Ti6Al4V discs into a 96-well plate. Cells were activated with interferon gamma and lipopolysaccharide. Macrophages on stainless steel discs produced significantly more nitric oxide (NO) compared to their control counterparts after eight to ten days and remained elevated for the duration of the experiment.Aims
Methods
Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene
THAs, and four primary osteoarthritis synovial membranes. The latter
were analyzed to assess the amount of tissue fibrosis that might
have been present at the time of implantation to enable evaluation,
in relation to implantation time, of any subsequent response in
the tissues.Aims
Patients and Methods
The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA) coatings, in a canine model. A canine transcortical model was used to evaluate the characteristics of bone ingrowth of Ti6Al4V cylindrical implants fabricated using laser rapid manufacturing (LRM). At four and 12 weeks post-implantation, we performed histological analysis and mechanical push-out testing on three groups of implants: a HA-free control (LRM), LRM with precipitated HA (LRM-PA), and LRM with plasma-sprayed HA (LRM-PSHA).Aims
Materials and Methods
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods