INTRODUCTION.
Introduction.
Explanations for “bearing” noise in ceramic-on-ceramic hips (COC) included stripe-wear formation and loss of lubrication leading to higher friction. However clinical and retrieval studies have clearly documented stripe wear in patients that did not have squeaking. Seldom highlighted has been the risk of metal-on-metal or metal-on-ceramic impingement present in total hip arthroplasty (THA) with metal and ceramic cup designs. The limitation in THA positioning studies has been (i) reliance on 2-dimensional radiographic images and (ii) patients lying supine on the examination table, thus not imaged in squeaking positions. We collected eleven squeaking COC cases for an EOS 3D-imaging functional study. Hip positions were documented in each patient's functional ‘squeaking’ posture using standard and 3-D EOS images for sitting, rising from a chair, hip extension in striding, and single-legged stance. EOS imaging documented for the 1st time that postural dysfunctions with potential impingements were demonstrable for each squeaking case. The 1st major insight in this study came from a female patient who complained of squeaking while walking in flat-soled shoes (Figs. 1a, b). She found that when wearing high-heeled shoes her hip stopped squeaking (Figs. 1c, d). Her lateral EOS view in standing position with heeled shoes revealed that the femoral stem had approximately 3o less hyper-extension compared to flat shoes (Figs. 1b, d, arrows #1,3). The three-dimensional ‘sky-view’ EOS reconstruction of pelvis and femurs (Fig. 2) showed that her femur was also more internally rotated when she wore heels. These subtle shifts in position changed her COC hip from one of squeaking to non-squeaking. A squeaking male patient observed similar postural effects while walking up his boat ramp but not going down the ramp. In both cases, the squeaking was a consequence of cup impinging on a metal femoral neck. Thus the primary cause of squeaking appeared to be hip impingement, i.e. repetitive subluxations that patients generally were not aware of. Another case is representative of situations due to atypical and subtle cup/stem mal-adjustments (Fig. 3); frontal pelvic-tilt, thoracolumbar scoliosis, with 1cm of femur lengthening and a significant increase of offset are observed. Also evident was the femoral-neck retroversion in both standing and sitting.
Increasing numbers and incidence rates of noisy (squeaking, scratching or clicking) ceramic-on-ceramic (CoC) total hip arthroplasties (THA) are being reported. The etiology seems to always involve stripe wear producing a stick-slip effect in the bearing which excites vibrations. As stripe wear is also found in silent CoC bearings, a theory has been developed that the vibrations become audible only via amplification through the vibrating stem. This was supported by showing that the excitation frequency and the resonance frequency of the plain stem are similar [1]. However, stem resonance in-vivo would be influenced by the periprosthetic bone damping and transmitting stem vibrations. Thus, if stem resonance is conditional for noisy COC hips, these should show periprosthetic bone different to silent hips. This study compares stem fit&fill and periprosthetic bone between noisy and silent CoC hips. In a consecutive series of 186 primary CoC hips with identical stems, cups (Stryker ABG-II) and femoral heads (Alumina V40, 28mm) a dedicated patient questionnaire [2] identified 38 noisy hips (incidence rate: 20.4%, squeakers: n=23). Stem fit&fill and cortical wall thickness (CWT, medial and lateral) were measured on post-op AP x-rays according to an established method [3, Fig 1]. Measurements were repeated by a single blinded observer in a control group of silent hips matched for gender, age, stem size and follow-up time (4.6yrs). Fit&fill and CWT were compared between the noisy and silent group at proximal, mid-stem and distal level and on the medial and lateral side.Introduction
Methods
Acetabular cup orientation has been shown to be a factor in edge-loading of a ceramic-on-ceramic THR bearing. Currently all recommended guidelines for cup orientation are defined from static measurements with the patient positioned supine. The objectives of this study are to investigate functional cup orientation and the incidence of edge-loading in ceramic hips using commercially available, dynamic musculoskeletal modelling software that simulates each patient performing activities associated with edge-loading. Eighteen patients with reproducible squeaking in their ceramic-on-ceramic total hip arthroplasties were recruited from a previous study investigating the incidence of noise in large-diameter ceramic bearings. All 18 patients had a Delta Motion acetabular component, with head sizes ranging from 40 – 48mm. All had a reproducible squeak during a deep flexion activity. A control group of thirty-six patients with Delta Motion bearings who had never experienced a squeak were recruited from the silent cohort of the same original study. They were matched to the squeaking group for implant type, acetabular cup orientation, ligament laxity, maximum hip flexion and BMI. All 54 patients were modelled performing two functional activities using the Optimized Ortho Postoperative Kinematics Simulation software. The software uses standard medical imaging to produce a patient-specific rigid body dynamics analysis of the subject performing a sit-to-stand task and a step-up with the contralateral leg, Fig 1. The software calculates the dynamic force at the replaced hip throughout the two activities and plots the bearing contact patch, using a Hertzian contact algorithm, as it traces across the articulating surface, Fig 2. As all the squeaking hips did so during deep flexion, the minimum posterior Contact Patch to Rim Distance (CPRD) can then be determined by calculating the smallest distance between the edge of the contact patch and the true rim of the ceramic liner, Fig 2. A negative posterior CPRD indicates posterior edge-loading.Introduction
Methodology
Polyimide (MP-1, MMATech, Haifa, Israel), is a high performance aerospace thermoplastic used for its lubricity, stability, inertness and radiation resistance. A wear resistant thin robust bearing is needed for total hip arthroplasty (THR). After independent laboratory testing, in 2006, the author used the material as a bearing in two Reflection (Smith and Nephew, USA) hip surgeries. The first, a revision for polyethylene wear, survives with no evidence of wear, noise, new osteolysis or complications related to the MP-1 bearing after 16 yrs. The second donated his asymptomatic MP-1 hip at 6.5yrs for post-mortem examination. There were no osteoclasts, cellular reaction bland in contrast to that of polyethylene. In 2013 a clinical study with ethical committee approval was started using a Biolox Delta (Ceramtec, Germany) head against a polyimide liner in 97 patients. MMATech sold all liners, irradiated: steam 52:45. Sixteen were re-machined in New Zealand. Acetabular shells were Delta PF (LIMA, Italy). The liner locked by taper. The cohort consisted of 46:51 M:F, and ages 43 to 85, mean 65. Ten received cemented stems. For contralateral surgery, a ceramic or polyethylene liner was used. Initial patients were lower demand, later, more active patients, mountain-biking and running. All patients have on-going follow up, including MP-1 liner revision cases. There has been no measurable wear, or osteolysis around the acetabular components using weight-bearing radiographs.
The problem associated with ceramic on ceramic total hip replacement (THR) is audible noise.
Intentionally crosslinked polyethylene has improved the survivorship of total hip replacement and is the current standard bearing material for total hip arthroplasty. Regardless of the manufacturing method and counter-surface, the wear rates have been reduced on the order of 90% compared to historical materials, with a substantial reduction in the occurrence of osteolysis.
Ceramic-on-ceramic bearings provide a solution to the osteolysis seen with traditional metal-on-polyethylene bearings. Sporadic reports of ceramic breakage and squeaking concern some surgeons and this bearing combination can show in vivo signs of edge loading wear which was not predicted from in vitro studies. Taper damage or debris in the taper between the ceramic and metal may lead to breakage of either a ceramic head or insert. Fastidious surgical technique may help to minimise the risk of ceramic breakage.
Hip arthroplasty surgeons have various bearing choices to make on behalf of their patients. We make those choices based on our knowledge of pre-clinical wear testing data and the outcome of clinical and radiological follow-up studies. The initial use of conventional polyethylene revealed limitations in its use in younger patients. Modern highly crosslinked polyethylene is a vastly improved bearing surface that means less wear and its consequences. Despite this, registry data still suggests that loosening, lysis and dislocation are problematic causes of implant failure. The functional success of hip replacement surgery, the ageing population and younger patients requesting arthroplasty means we should predict ongoing issues consequent to wear related events even with the newer polyethylenes. Ceramic-on-ceramic bearings surfaces have a long history of successful clinical use. The benefits of ceramic bearings are its superior wear characteristics, the minimal biological response to the ceramic wear products and the ability of ceramics to be offered in larger head sizes. Its limitations have been reports of fracture and squeaking. Fourth generation ceramic articulations have reduced the fracture incidence.
The leading cause for total hip arthroplasty (THA) revision remains aseptic loosening due to bearing wear. The younger and more active patients currently undergoing arthroplasty present unprecedented demands on THA-bearings. Ceramic-on-ceramic (CoC) bearings have consistently shown the lowest wear rates. The recent advances, especially in alumina CoC bearings, have solved many past problems and produced preferable results in vitro. Alumina ceramics are extremely hard, scratch resistant, biocompatible, offer a low coefficient of friction, superior lubrication and lower wear rates in comparison to other bearings in THA. The major disadvantage of ceramics used to be fracture. The new generation of alumina ceramics, has reduced the risk of ball fracture to 0.03–0.05%. The risk for liner fracture is even lower. Assuming an impingement-free component implantation, CoC bearings have major advantages over other bearing combinations. Due to the superior hardness, CoC bearings produce less third body wear and are virtually impervious to damage from instruments during the implantation process. A complication specific to CoC bearings is squeaking.
Ceramic-on-ceramic bearings provide a solution to the osteolysis seen with traditional metal-on-polyethylene bearings. Sporadic reports of ceramic breakage and squeaking concern some surgeons and this bearing combination can show in vivo signs of edge loading wear which was not predicted from in vitro studies. Taper damage or debris in the taper between the ceramic and metal may lead to breakage of either a ceramic head or insert. Fastidious surgical technique may help to minimise the risk of ceramic breakage.
Malorientation of the acetabular cup in Total Hip replacement (THR) may contribute to premature failure of the joint through instability (impingement, subluxation or dislocation), runaway wear in metal-metal bearings when the edge of the contact patch encroaches on the edge of the bearing surface, squeaking of ceramic-ceramic bearings and excess wear of polyethylene bearing surfaces leading to osteolysis. However as component malorientation often only occurs in functional positions it has been difficult to demonstrate and often is unremarkable on standard (usually supine) pelvic radiographs. The effects of spinal pathology as well as hip pathology can cause large rotations of the pelvis in the sagittal plane, again usually not recognized on standard pelvic views. While Posterior pelvic rotation with sitting increases the functional arc of the hip and is protective of a THR in regards to both edge loading and risk of dislocation, conversely Anterior rotation with sitting is potentially hazardous. We developed a protocol using three functional positions – standing, supine and flexed seated (posture at “seat-off” from a standard chair). Lateral radiographs were used to define the pelvic tilt in the standing and flexed seated positions. Pelvic tilt was defined as the angle between a vertical reference line and the anterior pelvic plane. Supine pelvic tilt was measured from computed tomography. Proprietary software (Optimized Ortho, Sydney) based on Rigid Body Dynamics then modelled the patients’ dynamics through their functional range producing a patient-specific simulation which also calculates the magnitude and direction of the dynamic force at the hip and traces the contact area between prosthetic head/liner onto a polar plot of the articulating surface. Given prosthesis specific information edge-loading can then be predicted based on the measured distance of the edge of the contact patch to the edge of the acetabular bearing. Results and conclusions. The position of the pelvis in the sagittal plane changes significantly between functional activities. The extent of change is specific to each patient. Spinal pathology can be an insidious “driver” of pelvic rotation, in some cases causing sagittal plane spinal imbalance or changes in orientation of previously well oriented acetabular components.