Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral
Background and aim. Recent proposals have been introduced to modify
Background and aim. Recent proposals have been introduced to modify
INTRODUCTION. Combining novel diverse population-based software with a clinically-demonstrated implant design is redefining total hip arthroplasty. This contemporary
Introduction. The origins of the uncemented tapered wedge hip
Objectives. The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single
The main reasons of aseptic loosening of the cemented hip stem are three: Bone cement fracture, bone cement debonding, and rupture of cement/bone interface. These are caused by normal/shear stress in the cement mantle. In past studies, there are introduced some optimum design of the hip prosthesis. But all there are not considered enough design objectives. The purpose of this study is to design the optimum stem geometry, which reduces the many stress in the cement mantle at the same time. We reserched the relationship between
Introduction. Resorptive bone remodeling secondary to stress shielding has been a concern associated with cementless total hip arthroplasty (THA). At present, various types of cementless implants are commercially available. The difference in femoral
The use of stem provides consistent component alignment with immediate stable fixation and protects grafted bone by reducing stress on metaphyseal area in revision total knee arthroplasty. One of major concern with use of stems involves stem tip pain in cementless diaphyseal engaging stem. The purpose of this study is to evaluate the effect of
Introduction:. Extensive bone defects of the proximal femur e.g. due to aseptic loosening might require the implantation of megaprostheses. In the literature high loosening rates of such megaprostheses have been reported. However, different fixation methods have been developed to achieve adequate implant stability, which is reflected by differing design characteristics of the commonly used implants. Yet, a biomechanical comparison of these designs has not been reported. The aim of our study was to analyse potential differences in the biomechanical behaviour of three megaprostheses with different designs by measuring the primary rotational stability in vitro. Methods:. Four different stem designs [Group A: Megasystem-C® (Link), Group B: MUTARS®(Implantcast), Group C: GMRS™ (Stryker) and Group D: Segmental System (Zimmer); see Fig. 1] were implanted into 16 Sawbones® after generating a segmental AAOS Typ 2 defect. Using an established method to analyse the rotational stability, a cyclic axial torque of ± 7.0 Nm along the longitudinal stem axis was applied. Micromotions were measured at defined levels of the bone and the implant [Fig. 2]. The calculation of relative micromotions at the bone-implant interface allowed classifying the rotational implant stability. Results:. All four different implants exhibited low micromotions, indicating adequate primary stability. Lowest micromotions for all designs were located near the femoral isthmus [Fig. 3]. The extent of primary stability and the global implant fixation pattern differed considerably and could be related to the different design concepts. Discussion:. Compared to other implant designs, all stems resulted in low relative motions regardless their design. The conical Megasystem-C® stem seems to lock in the proximal isthmus of the femur, whereas the MUTARS® stem seems to have a total fixation. Its hexagonal cross-section might have a good interlocking effect against rotational force application. Similarly, the GMRS™ stem shows a total fixation with little tendency to the distal part. The very rough porous-coated surface seems to generate a comparable fixation method to the hexagonal MUTARS® stem. However, the four longitudinal expansions in the proximal part of the GMRS™ stem might not have such a high rotational stability effect as expected. Compared to the other stems, the Segmental System stem showed very low relative micromotions in the proximal part. This sharp fluted stem seems to engrave itself into the bone. Within this study all stems seemed to achieve an adequate primary rotational stability. We could show that
Aims: The aim of this retrospective study is comparatively analysing cemented versus hydroxyapatite coated cementless þxation. A 10-year survival analysis of 2 patient cohorts operated by the same senior surgeons and with the same
Introduction: The purpose of this paper is to present the results of a prospective study involving one
Purpose of the study: The purpose of this prospective controlled trial was to evaluate the influence of the tibial
Introduction: In total hip arthroplasty (THA) an optimal fit and fill of the stem is essential for stable fixation. Thus femur morphology must be studied during pre-op planning (implant choice, sizing, positioning) or when a new stem is to be designed. Using plain AP x-ray analysis and the definition of a simple two-level parameter (canal flare index, CFI), Noble et al. identified an age related transition of the endosteal canal in AP view from a ‘champagne flute’ to a ‘stove pipe’. This reference data is 2D only, limited to the endosteal geometry and the elderly age range was defined as 60–90yrs so that the number of octogenerians >
80yrs was too low to analyze morphological features of this rapidly growing and critical THA patient population. In this study the endosteal and periosteal femur morphology of subjects >
80yrs was studied using 3D CT analysis. It was the goal to. describe age related changes of the femur morphology in 3D,. to study the influence of gender. to investigate if the results may affect fit &
fill of current stem designs. Methods: High-resolution CT-scans (slice thickness 1mm) were made of 170 consenting volunteers (m/ f=101/69). The old group consisted of 119 subjects ≥80yrs (m/f=65/54, mean age: 84.1yrs [80–105]) and the young group of 51 subjects <
80yrs (m/f=36/15, mean age 67.8yrs [39–79]). After thresholding the bone boundaries in Mimics V12 (Materialise, B), the endos-teal and periosteal coordinates were analyzed for width, wall thickness, surface areas and various CFI’s relating dimensions at 20mm above LT and at a distal level (e.g. 60mm below LT, isthmus): Surface CFI (3D-CFI), frontal and lateral CFI based on the AP and ML projections (2D-CFI) and flaring in each of the four directions (1DCFI). Results: The surface CFI was sign. lower in subjects ≥80yrs (5.08 ±1.23) than in subjects <
80yrs (6.61 ±1.72, p<
0.0001). This difference was sign. larger in females than in males (−32% vs. −17%), an observation valid with reference to any distal level. Equivalent age differences were found in both the frontal and lateral 2D-CFI as well as the medial, lateral and anterior 1D-CFI with changes in the anterior direction (−26.3%) being most dominant. In addition wall thickness was sign. reduced in the very elderly. E.g. at 20mm above LT, the medial wall measured 10.40mm at <
80yrs and 7.61 at ≥80yrs, a reduction of −27% (p<
0.001). In females (−35%) this difference was sign. larger (males: −23%, p<
0.001) even when corrected for height. Discussion: The age driven transition of proximal femur morphology continues in the octogenarian population. This transition is not limited to two discrete levels in the AP plane as previously reported but it is a continuous 3D phenomenon with high directional asymmetry. In addition, this transition also affects the wall thickness and the periosteal shape. Furthermore a strong gender effect was identified with aging females showing increasingly and asymmetrically less flaring and thinner walls. An age and gender specific THA
One of the most important characteristic of the developmental dysplastic hip (DDH) is high anteversion in femoral neck. Neck-shaft angle is also understood to be higher (i.e. coxa-valga) in DDH femora. From this understanding many DDH intended
There is increasing impetus to use rapid recovery care pathways when treating patients undergoing total hip arthroplasty (THA). The direct anterior (DA) approach is a muscle sparing technique that is believed to support these new pathways. Implants designed for these approaches are available in both collared and collarless variations and understanding the impact each has is important for providing the best treatment to patients. This study aims to examine the role of implant design on implant fixation and patient recovery.Background
Purpose/Aim of Study
There have been many attempts to reduce the risk of femoral component loosening. Using a tapered stem having a highly polished stem surface results in stem stabilization subsequent to debonding and stem-cement taper-lock and is consistent with force-closed fixation design. In this study, we assessed the subsidence of two different polished triple tapered stems and two different cements in primary THA.Introduction
Purpose
Increasing pressure to use rapid recovery care pathways when treating patients undergoing total hip arthroplasty (THA) is evident in current health care systems for numerous reasons. Patient autonomy and health care economics has challenged the ability of THA implants to maintain functional integrity before achieving bony union. Although collared stems have been shown to provide improved axial stability, it is unclear if this stability correlates with activity levels or results in improved early function to patients compared to collarless stems. This study aims to examine the role of implant design on patient activity and implant fixation. The early follow-up period was examined as the majority of variation between implants is expected during this time-frame. Patients (n=100) with unilateral hip OA who were undergoing primary THA surgery were recruited pre-operatively to participate in this prospective randomized controlled trial. All patients were randomized to receive either a collared (n=50) or collarless (n=50) cementless femoral stem. Patients will be seen at nine appointments (pre-operative, < 2 4 hours post-operation, two-, four-, six-weeks, three-, six-months, one-, and two-years). Patients completed an instrumented timed up-and-go (TUG) test using wearable sensors at each visit, excluding the day of their surgery. Participants logged their steps using Fitbit activity trackers and a seven-day average prior to each visit was recorded. Patients also underwent supine radiostereometric analysis (RSA) imaging < 2 4 hours post-operation prior to leaving the hospital, and at all follow-up appointments. Nineteen collared stem patients and 20 collarless stem patients have been assessed. There were no demographic differences between groups. From < 2 4 hours to two weeks the collared implant subsided 0.90 ± 1.20 mm and the collarless implant subsided 3.32 ± 3.10 mm (p=0.014). From two weeks to three months the collared implant subsided 0.65 ± 1.54 mm and the collarless implant subsided 0.45 ± 0.52 mm (p=0.673). Subsidence following two weeks was lower than prior to two weeks in the collarless group (p=0.02) but not different in the collared group. Step count was reduced at two weeks compared to pre-operatively by 4078 ± 2959 steps for collared patients and 4282 ± 3187 steps for collarless patients (p=0.872). Step count increased from two weeks to three months by 6652 ± 4822 steps for collared patients and 4557 ± 2636 steps for collarless patients (p=0.289). TUG test time was increased at two weeks compared to pre-operatively by 4.71 ± 5.13 s for collared patients and 6.54 ± 10.18 s for collarless patients (p=0.551). TUG test time decreased from two weeks to three months by 7.21 ± 5.56 s for collared patients and 8.38 ± 7.20 s for collarless patients (p=0.685). There was no correlation between subsidence and step count or TUG test time. Collared implants subsided less in the first two weeks compared to collarless implants but subsequent subsidence after two weeks was not significantly different. The presence of a collar on the stem did not affect patient activity and function and these factors were not correlated to subsidence, suggesting that initial fixation is instead primarily related to implant design.
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article: