Bone & Joint Research
Vol. 3, Issue 3 | Pages 60 - 68
1 Mar 2014
Langton DJ Sidaginamale RP Holland JP Deehan D Joyce TJ Nargol AVF Meek RD Lord JK

Objectives

Wear debris released from bearing surfaces has been shown to provoke negative immune responses in the recipient. Excessive wear has been linked to early failure of prostheses. Analysis using coordinate measuring machines (CMMs) can provide estimates of total volumetric material loss of explanted prostheses and can help to understand device failure. The accuracy of volumetric testing has been debated, with some investigators stating that only protocols involving hundreds of thousands of measurement points are sufficient. We looked to examine this assumption and to apply the findings to the clinical arena.

Methods

We examined the effects on the calculated material loss from a ceramic femoral head when different CMM scanning parameters were used. Calculated wear volumes were compared with gold standard gravimetric tests in a blinded study.


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_I | Pages 29 - 29
1 Jan 2003
Kellett C Ward T Short A Price A Kyberd P Murray D
Full Access

Polyethylene wear can be an important cause of knee replacement failure. Six TKRs in young, active patients with excellent Oxford Knee Scores and Knee Society Scores, mean 76 months post knee replacement and 5 control patients, 2 weeks post TKR, were selected. Each patient had weight bearing stereo radiographs of at 0, 15, 30, 45 and 60 degrees of flexion while standing in a calibration grid. These x-rays were analysed using our Radio Stereometric Analysis (RSA) system. The three-dimensional shape of the TKR (manufacturer’s computer aided design model) was matched to the TKR silhouette on the calibrated stereo radiographs for each angle of flexion. The relative positions of the femoral and tibial components in space were then determined and the linear and volumetric penetration was calculated using Matlab software. The accuracy of the system was found to be 0.3mm (CAD model tolerance 0.25mm). The mean linear wear in the control patients was 0.02mm (range −0.19 to +0.23mm). Average linear penetration in the study group was found to be 0.6 mm at 6 years, giving an overall linear wear rate of 0.1mm/year. Average penetration volume at 76 months was 399mm3. The average volumetric wear rate was 63mm3/year. It is possible to measure volumetric wear in vivo using RSA. Volumetric wear rate was found to be 63mm3 per year. Studies on retrieved normally functioning hip replacements have shown volumetric wear rates of 35mm3 per year. However, clinical outcomes of knee replacements are comparable to those of hip replacements, suggesting that the knee has a more effective mechanism for dealing with polyethylene wear particles


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 33 - 33
1 Aug 2012
Lord J Langton D Nargol A Joyce T
Full Access

Wear debris induced osteolysis is a recognized complication in conventional metal-on-polyethylene hip arthroplasty. One method of achieving wear reduction is through the use of metal-on-metal articulations. One of the latest manifestations of this biomaterial combination is in designs of hip resurfacing which are aimed at younger, more active patients. But, do these metal-on-metal hip resurfacings show low wear when implanted into patients?. Using a Mitutoyo Legex 322 co-ordinate measuring machine (scanning accuracy less than 1 micron) and a bespoke computer program, volumetric wear measurements for retrieved Articular Surface Replacements (ASR, DePuy) metal-on-metal hip resurfacings were undertaken. Measurements were validated against gravimetric calculations for volumetric wear using a sample femoral head that was artificially worn in vitro. At 5mm3, 10mm3, and 15mm3 of material removal, the method was shown to be accurate to within 0.5mm3. Thirty-two femoral heads and twenty-two acetabular cups were measured. Acetabular cups exhibited mean volumetric wear of 29.00mm3 (range 1.35 - 109.72mm3) and a wear rate of 11.02mm3/year (range 0.30 - 63.59mm3/year). Femoral heads exhibited mean wear of 22.41mm3 (range 0.72 - 134.22mm3) and a wear rate of 8.72mm3/year (range 0.21 - 31.91mm3/year). In the 22 cases where both head and cup from the same prosthesis were available, mean total wear rates of 21.66mm3/year (range 0.51 - 95.50mm3/year) were observed. Revision was necessitated by one of five effects; early femoral neck fracture (4 heads), avascular necrosis (AVN) (2 heads, 1 cup), infection (1 head, 1 cup), adverse reaction to metal debris (ARMD) (19 heads, 18 cups) or ARMD fracture (6 heads, 2 cups). Mean paired wear rates for the AVN and infection retrievals were 0.51mm3/year and 3.98mm3/year respectively. In vitro tests typically offer wear rates for metal-on-metal devices in the region of 2-4mm3. Mean paired wear rates for ARMD and ARMD fracture were 17.64mm3/year and 68.5mm3/year respectively, significantly greater than those expected from in vitro tests. In the 4 cases of early fracture, only the heads were revised so a combined wear rate calculation was not possible. The heads exhibited mean wear rate of 8.26mm3/year. These high wear rates are of concern


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 267 - 268
1 Jul 2008
ADAM P PHILIPPOT R COUMERT S FARIZON F FESSY M
Full Access

Purpose of the study: The double-mobility concept was introduced for clinical applications for total hip arthroplasty in 1976. The concept preserves joint range of motion while increasing stability. In this study we evaluated the consequences of these advantages in terms of polyethylene wear, measuring wear both on the concave and convex surfaces and volumetrically. Material and methods: Forty polyethylene inserts were explanted and analyzed. Explantation had been performed for mechanical or septic failure after eight years implantation on average. Mean age of patients at implantation was 46 years. After examining the gross aspect of the insert, surface analysis was performed with direct measurement of changes in the curvature using a BHN 706 position sensor for the inner concave surface and lateral projection for the outer convex surface. Estimated measurement error was ±5μm for each method; the manufacturer's tolerance for production of the inserts was 50μm. Volumetric wear was determined by reference to the manufacturer’s data. Student’s t test for paired series was applied. Results: At gross inspection, all inserts had lost the strips originally present on the convex surface; 40% presented visible wear of the retaining ring. Mean annual wear (± standard deviation, SD) obtained with the measuring system was 9±9 μm/yr) for the convex surface and 73 ± 69 μm/yr for the concave surface. Total annual wear, the sum of inner and outer surface wear, was 82±72 μm/yr. The mean volumetric wear was 28±28 mm3/year for the convex surface and 25±23 mm3/year for the concave surface and 53.4±40 mm3/year for total wear. Discussion: Total wear for these 40 double-mobility inserts which had functioned in vivo was not greater than the values reported for the metal-polyethylene bearing with 22.2 mm femoral heads. The double mobility is not associated with greater wear. While there was no significant difference between the wear volume of the convex versus the concave surfaces, the differentials wear were widespread, which can be considered to result from functional differences


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 115 - 115
1 Jan 2016
Knowlton C Wimmer M
Full Access

Introduction. Wear of the UHMWPE tibial component remains a major reason for aseptic loosening and subsequent revision or failure of TKAs [1]. Many retrieval studies measure surface damage patterns as surrogates for the severity of wear, but little is known about how these patterns relate to the volume of material lost. This study (a) examines the wear rate of a cruciate retaining TKA design and (b) relates observed wear patterns to volume loss on the surface. We hypothesize that damage patterns are good predictors for volumetric wear. Methods. 43 revision and 21 postmortem-retrieved MG II (Zimmer Inc.) tibial UHMWPE components were included in this study. Wear scars and damage patterns on the superior articular surfaces were digitized using a video microscope (SmartScope, OGP). Patterns were parsed into four spatially exclusive categories: delamination, polishing, striations and pitting. The surfaces were measured at 100×100µm using a low-incidence laser on a coordinate measuring machine (SmartScope, OGP). Autonomous mathematical reconstruction of the original surface was used [2] to calculate volume changes on the medial and lateral surfaces as an estimate of wear volume [Fig. 1] Total volume loss was calculated within the observed wear scar, and volume loss under each pattern was calculated and normalized to the total volume loss of its insert. Results. Excluding delaminated components, total wear correlated linearly with time in situ (Pearson's r=0.53) with a volumetric wear rate of 13.0±2.9 mm. 3. /year. Total wear area correlated linearly with total wear volume (Pearson's r=0.44), while delaminated area correlated strongly with total wear volume (Pearson's r=0.80). Excluding delaminated components, striated areas correlated more strongly to total volume loss (Pearson's r=0.54) than did total wear scar area (Pearson's r=0.34), while other patterns showed no correlation [Fig. 2]. When present, delaminated areas contributed most to total volume loss in postmortem- and revision-retrievals (58.3% and 38.7% respectively), striations second most (36.2% and 30.7%) and polished areas third most (24.6% and 27.6%), although significant differences were not observed [Fig. 3]. Pitted areas contributed significantly less (heteroscedastic t-test p=0.010) to total wear on postmortem- (1.3±1.7%) than on revision-retrievals (11.3±19.6%), although they were observed with a slightly higher frequency in the postmortem group (81% vs. 76%). Discussion. We found that damage patterns were not reliable surrogates for material volume loss. Other than delamination, area of striated patterns best predicted and contributed most to volume loss. Furthermore, our data suggests that polished and striated patterns in the absence of pitting are markers of a well-functioning UHWMPE TKR. While other groups have reported striations on retrievals [3], this damage pattern remains widely unrecognized in retrieval and knee simulator studies, with the mechanism poorly understood. The higher contribution of pitting to total wear volume in revision-retrieved TKR suggests that the fatigue wear mechanism leading to pitting contributes to the need for the early revision of the implant


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 212 - 213
1 May 2011
Ilchmann T Gunzenhauser T Pannhorst S Clauss M
Full Access

Introduction: The biological activity of PE-particles released due to wear is an established risk-factor for osteolysis and loosening after Total Hip Arthroplasty (THA). Cup position and orientation might have an effect on the risk of impingement and wear, thus contribute to the risk of aseptic loosening in the long-term what should be studied. Methods: Between 1984 and 1987 a total of 149 cemented total hips (Müller all-poly cup, Müller straight stem, 32 mm head) have been implanted. All implants had a standardised clinical and radiological follow-up. The pre- and postoperative centre of rotation of the hip and the orientation of the cup were determined. Migration, linear wear and direction of wear were measured twice with standard Methods: and the digital EBRA method. Wear-volume was calculated, taking direction of wear and cup orientation into account. Radiographs were analysed for progressive osteolysis and loosening. Results: 1 patient was lost to follow-up, 47 had died, 7 had been revised before 10 years follow-up. 18 patients had a missing or poor final radiograph, leaving 75 hips for long-term analysis. 41 were in male patients, mean age was 66.2 (+/− 11.0) years, mean follow-up 15.4 (+/−4.1) years. Mean inclination was 40.7° (+/− 7.1), mean anteversion was 14.8° (+/−8.4) And the mean cup positioning was 3.8 mm (+/− 4.3) medial and 5.3 mm (+/− 3.5) cranial. Osteolysis was found in 36 cups, 18 of them have been revised. The average linear wear was 1.1 (+/− 0.9) mm, the average wear volume 798.7 (+/−622.3) mm3, the linear wear rate 0.07 (+/−0.06) mm/year and the volumetric wear rate 54.5 (+/− 43.2) mm3/year. Younger patients had increased linear wear rates (p=0.035). Osteolysis of the cup, cup migration and cup revision were correlated with linear and volumetric wear (all p=< 0.001). There was no correlation of the cup position, inclination and anteversion with osteolysis, loosening and any of the wear parameters. Volume calculation did not provide further information. Conclusion: We found a strong correlation between wear and loosening of PE cups, but cup position and orientation did not affect osteolysis and loosening. Thus in contrast to hard-hard bearings polyethylene is a forgiving bearing surface and improvement of the cup orientation (e.g. due to navigation) will not result in increased cup survival, as long as extreme positioning errors are avoided


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_III | Pages 559 - 559
1 Aug 2008
Kannan V Heaslip R Richards R Sauret V Cobb J
Full Access

Wear and loosening are the major causes for long tem failure in Total Hip Replacement (THR). Accurate three dimensional wear analysis of radiographs has its own limitations. We report the results of our clinical study of three dimensional volumetric wear measurements using our custom low radiation risk CT based algorithm and special software. Twenty four patients (32 hips) agreed to take part in our study. The male: female ratio was 1:4. The mean age was 75 years and the mean follow up was 5.4 years. All patients had 28 mm diameter ceramic heads. Of the 32 hips, 17 hips had polyethylene inserts and 15 hips had ceramic inserts. The maximum follow up for the polyethylene and ceramic groups were 12 years and 5.5 years respectively. All the patients were scanned using Somatom Sensation 4 scanner. Using custom software, 3D reconstruction of the components was done and landmark acquisition done on the femoral head, acetabular metal component and the insert. From these landmarks, a dedicated program was used to calculate the centre of the femoral head in relation to the centre of the acetabular component in all three axes and an indirect measurement of wear obtained. Using the axes measurements graphical 3D models of migration of the femoral head component into the acetabular liner were created and volume of wear measured using special software. Accuracy of the method was assessed by measuring the radius of the femoral head since all patients had 28mm diameter heads implanted in them. Assessment of precision of method was done by calculating the level of agreement between two independent observers. In the polyethylene group, there was no significant (< 1mm) wear in x and y axis with time. However there was significant evidence of wear in relation to time in the z axis (max wear = −2.5 mm). In the ceramic group with relatively shorter follow up, there was no evidence of significant wear in all three axes. The mean volume measured in the polyethylene group was 685 mm3 (max = 1629 mm3, min = 132mm3 ). The mean volume measured in the ceramic group was 350mm3 (max = 1045 mm3, min = 139mm3 ). The mean radius of the femoral head measured in both groups was 14.02mm (range =13.8 to 14.4 mm). Accuracy was limited by artifacts particularly in bilateral hip arthroplasties and further in the ceramic group because of the restricted access to the ceramic head for placement of markers. Measurements obtained by two independent observers showed a strong correlation (0.99, p value = 0.001) for the polyethylene group. In the ceramic group the correlation (0.69, p value=0.0126) was not as strong as the polyethylene group. This study has produced a method for three dimensional estimation of wear that can be obtained from low dose CT scans with better accuracy and repeatability (< 0.5 mm) even than to ex vivo studies particularly in polyethylene bearings(wear rate 0.14mm/yr). Noise reduction with appropriate artefact reduction software may further improve the accuracy of this simple and repeatable method


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 78 - 78
1 Dec 2013
Hothi H Cro S Bills P Blunt L Racasan R Blunn GW Skinner JA Hart A
Full Access

Introduction. Metal-on-metal (MOM) total hip arthroplasty using large diameter femoral heads offer clinical advantages however the failure rates of these hips is unacceptably high. Retrieved hips have a wide range of wear rates of their bearing and taper surfaces and there is no agreement regarding the cause of failure. Detailed visual inspection is the first step in the forensic examination of failed hip components and may help explain the mechanisms of failure. The aim of this study was to determine if there was a correlation between the results of detailed inspections and the volumetric wear of the bearing and taper surfaces of retrieved hips. Method. Detailed, non-destructive macroscopic and stereomicroscopic examinations of 89 retrieved MOM hip components were performed by a single experienced examiner using quantitative assessment to document the severity of 10 established damage features:. Light scratches, Moderate scratches, Heavy scratches, Embedded particles, Discolouration, Haziness, Pitting, Visible wear zone, Corrosion, Fretting. Each surface was considered in terms of zones comprising of quadrants (cup, head, and taper) and subquadrants (cup and head), Figure 1. Each zone was scored on a scale of 0 to 3 by determining the percentage of the surface area of the zone that exhibited the feature in question: a score of 0=0%, 1<25%, 25%<2<75%, 3>75%. The sum of the scores of each zone was used for the assessment of each damage feature. The volume of wear at the surfaces of each hip was measured with a Zeiss Prismo coordinate measuring machine (cup and head) and a Talyrond 365 roundness measurement instrument (taper), using previously reported methods. 1, 2. . Simple linear regression models were used to asses the univariable associations between the inspection scores and wear volumes. Multiple linear regression models were subsequently used to asses the simultaneous contribution of the inspection scores, found significant in univariable analyses, on the wear outcome variables. All statistical analysis was performed using Stata/IC version 12.1 (StataCorp, USA) and throughout a p value < 0.05 was considered statistically significant. Results. Visible wear zone, moderate scratches, discolouration and haziness scores were all significantly positively correlated with cup (R. 2. = 70%, 23%, 72% and 33% respectively) and head (R. 2. = 73%, 34%, 67% and 47% respectively) wear volumes. Visible wear zone and discoloration scores were significant predictors in multivariable analysis (p < 0.01) for both surfaces, together explaining 77% and 79% of the variance in the cup and head wear volumes respectively. Corrosion and discoloration scores were significantly positively correlated with taper wear volume (R. 2. = 57% and 53% respectively) and there was a significant interaction between the two damage features (p = 0.01). Discussion. This study demonstrates the importance of detailed visual inspections in retrieval analysis, suggesting that they may help in predicting the severity of bearing and taper surface wear. Future studies will involve assessments of the inter-observer errors of inspections and their relationship with many other variables such as implant design and patient factors


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 111 - 111
1 May 2016
Knowlton C Bhutani P Wimmer M
Full Access

Introduction

Wear of the ultra-high molecular weight polyethylene (UHWMPE) component and the subsequent aseptic loosening remains a primary reason for late revision of total knee replacements (TKRs).[1] While improved measurement techniques have provided more quantitative information on the wear of surgically retrieved inserts, it is not well understood how observed damage patterns translate to volume loss of polyethylene in vivo. The overall purpose of this study is to investigate the relationship of damage patterns and volume loss at the articular surface of total knee replacements. We hypothesize that damage patterns are reliable predictors of volume loss.

Methods

Two different investigators independently analyzed damage patterns and volume loss on 43 revision- and 21 postmortem-retrieved MG II (Zimmer Inc.) tibial UHMWPE components. Areas of damage patterns on the articular surfaces were outlined with a video microscope (SmartScope, OGP) and were separated into four spatially exclusive categories (Fig. 1): delamination, pitting, striations and polishing. Articular surfaces were digitized with a low-incidence laser coordinate measuring machine (SmartScope, OGP). Autonomous reconstruction, a previously described and validated method,[2] calculated volume loss on the medial and lateral sides of each component. To investigate the predictability of volume loss using observed patterns, stepwise linear regression models were rendered in PASW Statistics 18 (SPSS Inc).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 57 - 57
1 May 2016
Rad E Knowlton C Fullam S Lundberg H Laurent M Pourzal R Wimmer M
Full Access

Introduction

Failure of total knee replacements due to the generation of polyethylene wear debris remains a crucial issue in orthopedics. Unlike the hip, it is difficult to accurately determine knee implant wear rates from retrieved components. Several studies have relied on thickness measurements to estimate penetration, but the complicated geometry of contemporary tibial liners poses a challenge to accurately assess wear. In this study we address the question whether linear penetration can serve as a surrogate measure for volumetric material loss.

Methods

Eighty-one retrieved UHMWPE NexGen cruciate-retaining tibial liners (Zimmer, Warsaw, IN) with an average time in situ of 5.27±2.89 years were included in the study. Metrology data for the surfaces of the tibial liners were obtained with a coordinate measuring machine (OGP, Rochester, NY). Using a laser scanner with two micrometer depth accuracy, at least 400,000 measurement points were taken by investigator #1. Areal thickness changes were mapped for the lateral and medial sides with the help of an autonomous mathematical reconstruction algorithm and volume loss was calculated based on wear scar area and local thickness change. Investigator #2, blinded from these results, measured the minimum thickness of the medial and lateral tibial plateau using a dial indicator with a spherical tip radius of 3mm. Twenty-three short term retrievals (3 to 4 per implant size), removed due to infection and without any signs of wear, served as “unused” reference. Linear penetration was then calculated by subtracting the minimum thickness of each plateau from the average thickness of the reference components.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 91 - 91
1 Feb 2017
Baykal D Day J Underwood R
Full Access

In the retrieval analysis of explanted hip joints, the estimation of wear volume and visualization of wear pattern are commonly used to evaluate in-vivo performance. While many studies report wear volumes from explanted hips, it is important to understand the limitations of these estimates including the sources and magnitude of uncertainty of the reported results. This study builds on a previous uncertainty analysis by Carmignato et al. to quantify the magnitude of uncertainty caused by the assumption that the as-manufactured shape of an explanted hip component is a perfect sphere.

Synthetic data sets representing idealized measurements of spheroidal explants (prolate, oblate and pinched) with a nominal diameter of 50 mm were generated. These data sets represent the shape and magnitude of form deviations observed for explanted hip components (Figure 1). Data were simulated for either unworn components or those with a known volume and magnitude of wear simulated to represent 5 µm penetration of a 49.90 mm femoral head into an acetabular cup (Table 1). The volume of wear and wear pattern were estimated using a custom Matlab script developed for analysis of metrology data from explanted hip joints. This script fits a least squares sphere to data points in unworn, as manufactured regions of the surface to estimate the as-manufactured shape of the component. The diameter of the best fit sphere, and wear volume were compared to the known wear depths and volumes from the synthetic datasets.

The results showed that the Matlab script estimated a wear volume of up to 1.4 mm3 for an unworn cup with a radial deviation of 10 µm. The maximum error of 13.3 mm3 was for a pinched cup with wear at the pole. The complete results are shown in Table 2.

In some cases with aspherical form deviations, the least squares sphere fitted to the synthetic data was displaced in the Z direction with respect to the origin of the spheroid and the radius of the least squares sphere was outside the range of the principal radii of the spheroid. For instance, in case 5, the center was shifted 22 µm vertically from the mathematical center.

The results from this study show that the magnitude of uncertainty due to form deviations on wear volume varies depending on the shape and magnitude of the form deviations and in some cases was greater than 10 mm3. A further important finding is that in some instances, the diameter and center of the least squares sphere fitted to the unworn regions may not be consistent with the mathematical radius and center of the synthetic data. This may have important implications for the “reverse engineering” of the as-manufactured dimensions from worn explanted hip joints.

Please contact authors directly for the figure:

Figure 1 Graphical depiction of a) synthetic data set, b) deviation map of a hemispherical acetabular cup with simulated wear, c) deviation map of a prolate spheroid with simulated wear at rim with color bar set to ±5 microns, d) deviation map of pinched ellipsoid with simulated wear at 45 degrees from pole.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims

We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.

Patients and Methods

We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.


Bone & Joint Research
Vol. 2, Issue 5 | Pages 84 - 95
1 May 2013
Sidaginamale RP Joyce TJ Lord JK Jefferson R Blain PG Nargol AVF Langton DJ

Objectives

The aims of this piece of work were to: 1) record the background concentrations of blood chromium (Cr) and cobalt (Co) concentrations in a large group of subjects; 2) to compare blood/serum Cr and Co concentrations with retrieved metal-on-metal (MoM) hip resurfacings; 3) to examine the distribution of Co and Cr in the serum and whole blood of patients with MoM hip arthroplasties; and 4) to further understand the partitioning of metal ions between the serum and whole blood fractions.

Methods

A total of 3042 blood samples donated to the local transfusion centre were analysed to record Co and Cr concentrations. Also, 91 hip resurfacing devices from patients who had given pre-revision blood/serum samples for metal ion analysis underwent volumetric wear assessment using a coordinate measuring machine. Linear regression analysis was carried out and receiver operating characteristic curves were constructed to assess the reliability of metal ions to identify abnormally wearing implants. The relationship between serum and whole blood concentrations of Cr and Co in 1048 patients was analysed using Bland-Altman charts. This relationship was further investigated in an in vitro study during which human blood was spiked with trivalent and hexavalent Cr, the serum then separated and the fractions analysed.


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims. Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. Methods. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear. Results. Overall survival rate for all-cause revision was 94.5% (95% confidence interval (CI) 81.7% to 97.2%) at a mean follow-up of 12.8 years (10.9 to 18.7). Three patients were revised, none for liner wear, fracture, or dissociation. A total of 22 patients were included in the radiological analysis (mean follow-up 9.9 years (7.5 to 13.7)). Mean linear liner wear was 0.085 mm (95% CI -0.086 to 0.257) and the volumetric wear rate was 11.097 mm. 3. /year (95% CI -6.5 to 28.7). Conclusion. Using HXLPE liners with 36 mm heads in 52 mm acetabular components or smaller is safe, with excellent survival and low rates of linear and volumetric wear at medium-term follow-up. Patients did not require revision surgery for liner complications such as fracture, dissociation, or wear. Our results suggest that the advantages of using larger heads outweigh the potential risks of using thin HXLPE liners. Cite this article: Bone Joint J 2023;105-B(1):29–34


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 38 - 43
1 Mar 2024
Buckner BC Urban ND Cahoy KM Lyden ER Deans CF Garvin KL

Aims. Oxidized zirconium (OxZi) and highly cross-linked polyethylene (HXLPE) were developed to minimize wear and risk of osteolysis in total hip arthroplasty (THA). However, retrieval studies have shown that scratched femoral heads may lead to runaway wear, and few reports of long-term results have been published. The purpose of this investigation is to report minimum ten-year wear rates and clinical outcomes of THA with OxZi femoral heads on HXLPE, and to compare them with a retrospective control group of cobalt chrome (CoCr) or ceramic heads on HXLPE. Methods. From 2003 to 2006, 108 THAs were performed on 96 patients using an OxZi head with a HXLPE liner with minimum ten-year follow-up. Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up (mean 13.3 years). Linear and volumetric liner wear was measured on radiographs of 85 hips with a minimum ten-year follow-up (mean 14.5 years). This was compared to a retrospective control group of 45 THAs using ceramic or CoCr heads from October 1999 to February 2005, with a minimum of ten years’ follow-up. Results. Average HHS improved from 50.8 to 91.9 and 51.0 to 89.8 in the OxZi group and control group, respectively (p = 0.644), with no osteolysis in either group. Linear and volumetric wear rates in the OxZi group averaged 0.03 mm/year and 3.46 mm. 3. /year, respectively. There was no statistically significant difference in HHS scores, nor in linear or volumetric wear rate between the groups, and no revision for any indication. Conclusion. The radiological and clinical outcomes, and survivorship of THA with OxZi femoral heads and HXLPE liners, were excellent, and comparable to CoCr or ceramic heads at minimum ten-year follow-up. Wear rates are below what would be expected for development of osteolysis. OxZi-HXLPE is a durable bearing couple with excellent long-term outcomes. Cite this article: Bone Joint J 2024;106-B(3 Supple A):38–43


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims. We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. Methods. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes. Results. At two years, 44 patients in each group were available for RSA assessment. The median total two-year proximal head penetration was -0.02 mm (IQR -0.09 to 0.07; p = 0.548) for the largest possible head and -0.01 mm (IQR -0.07 to 0.10; p = 0.525) for 32 mm heads. Their difference was not statistically significant (p = 0.323). Neither group demonstrated a bedding-in period. The median steady-state volumetric wear rates were 6.1 mm. 3. /year (IQR -59 to 57) and 3.5 mm. 3. /year (-21 to 34) respectively, and did not differ between the groups (p = 0.848). There were no statistically significant differences in periacetabular radiolucencies or patient-reported outcomes. Conclusion. The use of the largest possible metal head did not increase vitamin E-doped cross-linked polyethylene wear compared with 32 mm heads at two years. Linear wear was negligible and volumetric wear rates were very low in both head size groups. There was a tendency towards higher values of volumetric wear in large heads that warrants longer-term evaluation before any definite conclusions about the association between head size and wear can be drawn. Cite this article: Bone Joint J 2021;103-B(7):1206–1214


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims. A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA). Patients and Methods. Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear. Results. Linear wear occurred at 0.015 mm/year (supine) and 0.220 mm/year (standing). Volumetric wear occurred at 10.3 mm. 3. /year (supine) and 39.3 mm. 3. /year (standing). Wear occurred primarily on the medial side of the joint. Weightbearing imaging greatly improved the reliability of measurement. Clinical precision of volumetric wear was 34 mm. 3. No significant associations were found between patient demographics or function scores and measured wear. Conclusion. In vivo volumetric wear of TKAs can be assessed at short-term follow-up using MBRSA. Cite this article: Bone Joint J 2019;101-B:1348–1355


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 129 - 137
1 Jun 2020
Knowlton CB Lundberg HJ Wimmer MA Jacobs JJ

Aims. A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients. Methods. In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component. Results. Volumetric wear rates on eight non-delaminated components measured 15.9 mm. 3. /year (standard error (SE) ± 7.7) on the total part, 11.4 mm. 3. /year (SE ± 6.4) on the medial side and 4.4 (SE ± 2.6) mm. 3. /year on the lateral side. Volumetric wear rates modelled from patient gait mechanics predicted 16.4 mm. 3. /year (SE 2.4) on the total part, 11.7 mm. 3. /year (SE 2.1) on the medial side and 4.7 mm. 3. /year (SE 0.4) on the lateral side. Measured and modelled wear volumes correlated significantly on the total part (p = 0.017) and the medial side (p = 0.012) but not on the lateral side (p = 0.154). Conclusion. In the absence of delamination, patient-specific knee mechanics during gait directly affect wear of the tibial component in TKA. Cite this article: Bone Joint J 2020;102-B(6 Supple A):129–137


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 32 - 32
1 Dec 2022
Fransen B Bengoa F Neufeld M Sheridan G Garbuz D Howard L
Full Access

With the introduction of highly crosslinked polyethylene (HXLPE) in total hip arthroplasty (THA), orthopaedic surgeons have moved towards using larger femoral heads at the cost of thinner liners to decrease the risk of instability. Several short and mid-term studies have shown minimal liner wear with the use HXLPE liners, but the safety of using thinner HXPLE liners to maximize femoral head size remains uncertain and concerns that this may lead to premature failure exist. Our objective was to analyze the outcomes for primary THA done with HXLPE liners in patients who have a 36-mm head or larger and a cup of 52-mm or smaller, with a minimum of 10-year follow-up. Additionally, linear and volumetric wear rates of the HXLPE were evaluated in those with a minimum of seven-year follow-up. We hypothesized that there would be minimal wear and good clinical outcome. Between 2000 and 2010, we retrospectively identified 55 patients that underwent a primary THA performed in a high-volume single tertiary referral center using HXLPE liners with 36-mm or larger heads in cups with an outer diameter of or 52-mm or smaller. Patient characteristics, implant details including liner thickness, death, complications, and all cause revisions were recorded. Patients that had a minimum radiographic follow-up of seven years were assessed radiographically for linear and volumetric wear. Wear was calculated using ROMAN, a validated open-source software by two independent researchers on anteroposterior X-rays of the pelvis. A total of 55 patients were identified and included, with a mean age of 74.8 (range 38.67 - 95.9) years and a mean BMI of 28.98 (range 18.87 - 63-68). Fifty-one (94.4%) of patients were female. Twenty-six (47.7%) patients died during the follow-up period. Three patients were revised, none for liner wear, fracture or dissociation. Twenty-two patients had a radiographic follow-up of minimum seven years (mean 9.9 years, min-max 7.5 –13.7) and were included in the long-term radiographic analysis. Liner thickness was 5.5 mm at 45 degrees in all cases but one, who had a liner thickness of 4.7mm, and all patients had a cobalt-chrome head. Cup sizes were 52mm (n=15, 68%) and 50mm (n=7, 32%). Mean linear liner wear was 0.0470 mm/year (range 0 - 0.2628 mm) and mean volumetric wear was 127.69 mm3/year (range 0 - 721.23 mm3/year). Using HXLPE liners with 36-mm heads or bigger in 52-mm cups or smaller is safe, with low rates of linear and volumetric wear in the mid to long-term follow-up. Patients did not require revision surgery for liner complications, including liner fracture, dissociation, or wear. Our results suggest that the advantages of using larger heads should outweigh the potential risks of using thin HXLPE liners


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_20 | Pages 3 - 3
12 Dec 2024
Bhalekar R Wells S Nargol M Nargol A Waller S Wildberg L Sidaginamale R Langton DJ
Full Access

Objectives. Unicompartmental knee replacement (UKR) is a conservative option for degenerative disease, with mobile (Oxford UKR) and fixed bearing (Physica ZUK) the most commonly used devices. The primary reasons for revising UKRs include disease progression (36.9%), loosening (31.7%), and pain (7.5%). Loosening typically occurs due to osteolysis caused by wear particles from the polyethylene (PE) inserts. There is limited published literature which has quantified volumetric material loss from the PE inserts of cemented fixed-bearing UKRs. This study aimed to quantify bearing wear and backside deformation of these PE components. Design and Methods. At our national retrieval centre, we measured changes volumetric bearing wear and backside deformation of 31 explanted fixed-bearing Physica ZUK UKR PE inserts using a peer-reviewed methodology based on coordinate measuring machine analysis. These explants had been revised for any indication [Females (19) and Males (12)]. The location of the wear scars was identified and mapped. We compared the volumetric wear from the bearing surface with contemporary total knee replacement (TKR) inserts. Results. The median (IQR) total volumetric wear of the UKR PEs [96(50-152) mm. 3. ] was significantly higher (p<0.001) than contemporary TKR PEs [40 (24-83) mm. 3. ]. The median (IQR) total volumetric wear of the PEs in males [148(122-200) mm. 3. ] was significantly higher(p=0.016) than in females [56(61-119) mm. 3. ]. The wear scars were situated in the anterior third in 61% of cases, with 26% located posteriorly. Examination of the PE backside surfaces exhibited damage in the area directly inferior to the bearing surface wear scar in the majority of cases. Conclusion. In conclusion, this study highlights significant PE wear of Physica ZUK UKR inserts, with higher wear rates compared to contemporary TKR inserts, particularly in males