Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 60 - 60
1 May 2017
Alizai M Lipperts M Houben R Heyligers I Grimm B
Full Access

Background

To complement subjective patient-reported outcome measures, objective assessments are needed. Activity is an objective clinical outcome which can be measured with wearable activity monitors (AM). AM's have been validated and used in joint arthroplasty patients to count postures, walking or transfers. However, for demanding patients such as after sports injury, running is an important activity to quantify. A new AM algorithm to distinguish walking from running is trialed in this validation study.

Methods

Test subjects (n=9) performed walking and running bouts of 30s duration on a treadmill at fixed speeds (walking: 3, 4, 5, 7km/h, running: 5, 7, 9, 12, 15km/h) and individually preferred speeds (slow, normal, fast, maximum, walk/run transition). Flat and inclined surfaces (8%, 16%), different footwear (soft, hard, barefoot) and running styles (hind/fore-foot) were tested. An AM (3D accelerometer) was worn on the lateral thigh. Previously validated algorithms to classify all gait as walking were adapted to differentiate running from walking, the main criterium being vertical acceleration peaks exceeding 2g within each subsequent 2s-interval. Independently annotated video observation served as reference.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 77 - 77
2 Jan 2024
Gueorguiev B Varga P
Full Access

Intramedullary nails (IMNs) are the current gold standard for treatment of long bone diaphyseal and selected metaphyseal fractures. Their design has undergone many revisions to improve fixation techniques, conform to the bone shape with appropriate anatomic fit, reduce operative time and radiation exposure, and extend the indication of the same implant for treatment of different fracture types with minimal soft tissue irritation. The IMNs are made or either titanium alloy or stainless steel and work as load-sharing internal splints along the long bone, usually accommodating locking elements – screws and blades, often featuring angular stability and offering different configurations for multiplanar fixation – to secure secondary fracture healing with callus formation in a relative-stability environment. Bone cement augmentation of the locking elements can modulate the construct stiffness, increase the surface area at the bone-implant interface, and prevent cut-through of the locking elements. The functional requirements of IMNs are related to maintaining fracture reduction in terms of length, alignment and rotation to enhance fracture healing. The load distribution during patient's activities is along the entire bone-nail interface, with nail length and anatomic fit being important factors to avoid stress risers


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 137 - 137
2 Jan 2024
Ghaffari A Lauritsen RK Christensen M Thomsen T Mahapatra H Heck R Kold S Rahbek O
Full Access

Smartphones are often equipped with inertial sensors capable of measuring individuals' physical activities. Their role in monitoring the patients' physical activities in telemedicine, however, needs to be explored. The main objective of this study was to explore the correlation between a participant's daily step counts and the daily step counts reported by their smartphone. This prospective observational study was conducted on patients undergoing lower limb orthopedic surgery and a group of non-patients. The data collection period was from 2 weeks before until four weeks after the surgery for the patients and two weeks for the non-patients. The participants' daily steps were recorded by physical activity trackers employed 24/7, and an application recorded the number of daily steps registered by the participants' smartphones. We compared the cross-correlation between the daily steps time-series taken from the smartphones and physical activity trackers in different groups of participants. We also employed mixed modeling to estimate the total number of steps. Overall, 1067 days of data were collected from 21 patients (11 females) and 10 non-patients (6 females). The cross-correlation coefficient between the smartphone and physical activity tracker was 0.70 [0.53–0.83]. The correlation in the non-patients was slightly higher than in the patients (0.74 [0.60–0.90] and 0.69 [0.52–0.81], respectively). Considering the ubiquity, convenience, and practicality of smartphones, the high correlation between the smartphones and the total daily step time-series highlights the potential usefulness of smartphones in detecting the change in the step counts in remote monitoring of the patient's physical activity


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 123 - 123
1 Mar 2021
Jelsma J Schotanus M van Kuijk S Buil I Heyligers I Grimm B
Full Access

Hip resurfacing arthroplasty (HRA) became a popular procedure in the early 90s because of the improved wear characteristic, preserving nature of the procedure and the optimal stability and range of motion. Concerns raised since 2004 when metal ions were seen in blood and urine of patients with a MoM implant. Design of the prosthesis, acetabular component malpositioning, contact-patch-to-rim distance (CPR) and a reduced joint size all seem to play a role in elevated metal ion concentrations. Little is known about the influence of physical activity (PA) on metal ion concentrations. Implant wear is thought to be a function of use and thus of patient activity levels. Wear of polyethylene acetabular bearings was positively correlated with patient's activity in previous studies. It is hypothesized that daily habitual physical activity of patients with a unilateral resurfacing prosthesis, measured by an activity monitor, is associated with habitual physical activity. A prospective, explorative study was conducted. Only patients with a unilateral hip resurfacing prosthesis and a follow-up of 10 ± 1 years were included. Metal ion concentrations were determined using ICP-MS. Habitual physical activity of subjects was measured in daily living using an acceleration-based activity monitor. Outcome consisted of quantitative and qualitative activity parameters. In total, 16 patients were included. 12 males (75%) and 4 females (25%) with a median age at surgery of 55.5 ± 9.7 years [43.0 – 67.9] and median follow-up of 9.9 ± 1.0 years [9.1 – 10.9]. The median cobalt and chromium ion concentrations were 25 ± 13 and 38 ± 28 nmol/L. A significant relationship, when adjusting for age at surgery, BMI, cup size and cup inclination, between sit-stand transfers (p = .034) and high intensity peaks (p = .001) with cobalt ion concentrations were found (linear regression analysis). This study showed that a high number of sit-stand transfers and a high number of high intensity peaks is significantly correlated with high metal ion concentrations, but results should be interpreted with care. For patients it seems save to engage in activities with low intensity peaks like walking or cycling without triggering critical wear or metal ions being able to achieve important general health benefits and quality of life, although the quality (high intensity peaks) of physical activity and behaviour of patients (sit-stand-transfers) seem to influence metal ion concentrations


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 11 - 11
1 Mar 2021
Barcik J Ernst M Balligand M Dlaska CE Drenchev L Todorov S Gueorguiev B Skulev H Zeiter S Epari D Windolf M
Full Access

The course of secondary fracture healing typically consists of four major phases including inflammation, soft and hard callus formation, and bone remodeling. Callus formation is promoted by mechanical stimulation, yet little is known about the healing tissue response to strain stimuli over shorter timeframes on hourly and daily basis. The aim of this study was to explore the hourly, daily and weekly variations in bone healing progression and to analyze the short-term response of the repair tissue to well-controlled mechanical stimulation. A system for continuous monitoring of fracture healing was designed for implantation in sheep tibia. The experimental model was adapted from Tufekci et al. 2018 and consisted of 3 mm transverse osteotomy and 30 mm bone defect resulting in an intermediate mobile bone fragment in the tibial shaft. Whereas the distal and proximal parts of the tibia were fixed with external fixator, the mobile fragment was connected to the proximal part via a second, active fixator. A linear actuator embedded in the active fixator moved the mobile fragment axially, thus stimulating mechanically the tissue in the osteotomy gap via well-controlled displacement being independent from the sheep's functional weightbearing. A load sensor was integrated in the active fixation to measure the force acting in the osteotomy gap. During each stimulation cycle the displacement and force magnitudes were recorded to determine in vivo fracture stiffness. Following approval of the local ethics committee, experiments were conducted on four skeletally mature sheep. Starting from the first day after surgery, the daily stimulation protocols consisted of 1000 loading events equally distributed over 12 hours from 9:00 to 21:00 resulting in a single loading event every 44 seconds. No stimulation was performed overnight. One animal had to be excluded due to inconsistencies in the load sensor data. The onset of tissue stiffening was detected around the eleventh day post-op. However, on a daily basis, the stiffness was not steadily increasing, but instead, an abrupt drop was observed in the beginning of the daily stimulations. Following this initial drop, the stiffness increased until the last stimulation cycle of the day. The continuous measurements enabled resolving the tissue response to strain stimuli over hours and days. The presented data contributes to the understanding of the influence of patient activity on daily variations in tissue stiffness and can serve to optimize rehabilitation protocols post fractures


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 1 - 1
1 Jan 2017
Reeder I Lipperts M Heyligers I Grimm B
Full Access

Introduction: Physical activity is a major outcome in total hip arthroplasty (THA) and discharge criterion. Increasing immediate post-op activity may accelerate discharge, enable fast track surgery and improve general rehabilitation. Preliminary evidence (O'Halloran P.D. et al. 2015) shows that feedback via motivational interviewing can result in clinically meaningful improvements of physical activity. It was the aim of this study to use wearable sensor activity monitors to provide and study the effect of biofeedback on THA patients' activity levels. It was hypothesized that biofeedback would increase in-hospital and post-discharge activity versus controls. Methods: In this pilot study, 18 patients with osteoarthritis receiving elective primary THA followed by a rapid recovery protocol with discharge on day 3 after surgery (day 0) were randomized to the feedback group (n=9, M/F: 4:5, age 63.3 ± 5.9 years, BMI 26.9 ± 5.1) or a non-feedback control group (n=9, M/F: 0:9, age 66.9 ± 5.1 years, BMI 27.1 ± 4.0). Physical activity was measured using a wearable sensor and parameters (Time-on-Feet (ToF), steps, sit-stand-transfers (SST), mean cadence (steps/min)) were calculated using a previously validated algorithms (Matlab). For the in-hospital period data was calculated twice daily (am, ca. 8–13:00h and pm, ca. 13–20:00h) of day 1 (D1) and 2 (D2). The feedback group had parameters reported back twice (morning, lunch) using bar charts comparing visually and numerically their values (without motivational instructions) to a previously measured reference group (n=40, age 71 ±7 years, M:F 16:24) of a conventional discharge protocol (day 4/5). Activity measures continued from discharge (D3) until day 5 (D5) at home. Results: Randomization resulted in matched groups regarding age and BMI, but not gender. The first post-op activity assessment (D1am) was identical between groups. Also thereafter similar values with no significant differences in any parameter were seen, e.g. the time-on-feet at D2PM was 59.2 ±31.7min (feedback) versus 62.9 ±39.2min (controls). Also on the day of discharge and beyond, no effect from the in-hospital feedback was measured. For both groups the course of activity recovery showed a distinct drop on day 4 following a highly active day of discharge (D3). On day 5, activity levels only recovered partially. For both groups, all quantitative activity parameters were significantly higher than the reference values used for feedback. Only cadence as a qualitative measure was the same like reference values. Discussion: Biofeedback using activity values from a body-worn monitor did not increase in-hospital or immediate post-op home activity levels compared to a control group when using the investigated feedback protocol. In general, while the day of discharge steeply boosts patient activity, the day after at home results in an activity drop to near in-patient levels before discharge. In a fast track surgery protocol, it may be of value to avoid this drop via patient education or home physiotherapy. Biofeedback using activity monitors to increase immediate post-op activity for fast track surgery or improved recovery may only be effective when feedback goals are set higher, are personalised or have additional motivational context


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 11 - 11
1 Apr 2017
Grupp T Fritz B Kutzner I Bergmann G Schwiesau J
Full Access

Background. Wear simulation in total knee arthroplasty (TKA) is currently based on the most frequent activity – level walking. A decade ago multi-station knee wear simulators were introduced leading to optimisations of TKA designs, component surface finish and bearing materials. One major limitation is that current wear testing is mainly focused on abrasive-adhesive wear and in vitro testing does not reflect “delamination” as an essential clinical failure mode. The objective of our study was to use a highly demanding daily activities wear simulation to evaluate the delamination risk of polyethylene materials with and without vitamin E stabilisation. Methods. A cruciate retaining fixed bearing TKA design (Columbus CR) with artificially aged polyethylene knee bearings (irradiation 30±2 kGy) blended with and without 0.1% vitamin E was used under medio-lateral load distribution and soft tissue restrain simulation. Daily patient activities with high flexion (2×40% stairs up and down, 10% level walking, 8% chair raising, 2% deep squatting) were applied for 5 million cycles. The specimens were evaluated for gravimetric wear and analysed for abrasive-adhesive and delamination wear modes. Results. The total amount of gliding surface wear was 28.7±1.9 mg for the vitamin E stabilised polyethylene compared to 355.9±119.8 mg for the standard material. The combination of artificial ageing and high demanding knee wear simulation leads to visible signs of delamination in the articulating bearing areas in vitro. Conclusion. To evaluate Vitamin E stabilised polyethylenes in regard to ageing and wear behaviour in vitro, conditions are simulated to create clinical relevant failure modes in the reference material


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 56 - 56
1 May 2017
Jelsma J Senden R Schotanus M Kort N Heyligers I Grimm B
Full Access

Background. Metal-on-metal hip implants can produce adverse tissue reactions to wear debris. Increased metal ion concentrations in the blood are measured as a proxy to wear and the complications it can trigger. Many studies have examined various factors influencing the metal ion concentrations. This is the first study to investigate the effect of physical activity level, as objectively measured in daily life, on blood ion levels, expecting higher concentrations for higher patient activity. Methods. Thirty-three patients (13F/20M, 55.8 ± 6.2 years at surgery) with a unilateral resurfacing hip prosthesis were included. At last follow-up (6.8 ± 1.5 years) cobalt and chromium concentrations in the blood were determined by inductively coupled plasma mass spectrometry. Physical activity was measured during 4 successive days using a 3D-acceleration-based activity monitor. Data was analysed using validated algorithms, producing quantitative and qualitative parameters. Acetabular cup position was measured radiographically. Correlations were tested with Pearson's r'. Results. Wide ranges in cobalt (10-833nmol/l) and chromium (16-592nmol/l) concentrations were measured. No correlations were found between metal ion concentrations and patient characteristics. The mean time of walking per day of was 5475 ± 2730 seconds (≈ 91 minutes) per day, with 6953 ± 3104 steps made during the day. The cadence, the number of steps per minute, was 99.5 ± 7.4. The subjects performed 41 ± 13 sit-stand transfers per day. The number of peak intensity declines when peaks where more intense. No correlations were found between quantitative or qualitative parameters and metal ion concentrations. Conclusions. Higher metal ion concentrations were not correlated with higher activity levels contrary to original expectations. Based on the results the orthopaedic surgeons is able to reassure his or her patients that physical activity in daily living is presumably safe en does not influence metal ion concentrations. Level of Evidence. 2b. Approval. the ethics committee of the Atrium MC, Heerlen approved this study


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 62 - 62
1 May 2017
Lipperts M Senden R Heyligers I Grimm B
Full Access

Background. The goal of total hip arthroplasty (THA) is to reduce pain, restore function but also activity levels for general health benefits or social participation. Thus evaluating THA patient activity can be important for diagnosis, indication, outcome assessment or biofeedback. Methods. Physical activity (PA) of n=100 primary THA patients (age at surgery 63 ±8yrs; 49M/51F; 170 ±8cm, 79.8 ±14.0kg) was measured at 8 ±3yrs follow-up. A small 3D accelerometer was worn for 4 successive days during waking hours at the non-affected lateral upper leg. Data was analysed using validated algorithms (Matlab) producing quantitative (e.g. #steps, #transfers, #walking bouts) and qualitative (e.g. cadence, temporal distribution of events) activity parameters. An age matched healthy control group (n=40, 69 ±8yrs, 22M/18F) served as reference. Results. Daily steps were only 13% lower (n.s) for patients (avg. ±SD: 5989 ±3127) than controls (6890 ±2803). However, the Nr. of walking bouts (187 ±85 vs 223 ±78, −16%) and sit-stand transfers (35 ±14 vs 48 ±15, −27%) were sign. less in patients (p<0.05, Mann-Whitney). Patients showed equal amounts of walking bouts in medium duration (30-60s, 1–5min) but sign. less (−25%) short (<10s, 10–30s) and less (−43%) long events (>5min). This corresponds with sign. less (−32%) short sitting periods (>10min) in patients. Also cadence was sign. lower in patients (93.8 ±11.7 vs 98.9 ±7.3 steps/min). Conclusions. PA varies widely in patients with a substantial proportion (35%) being more active than average controls. Thus, THA must not per se reduce or limit PA. Only 17% of controls and 11% of patients reached the WHO target (10,000 steps/day) suggesting that the THA related drop in activity may inflate the risks for cardiovascular, metabolic or mental disease associated with low activity. Patients avoided short and long activities, both associated with effort. Targeted interventions may address this behaviour


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 31 - 31
1 Jan 2017
Morellato K Fink B Grupp T Cristofolini L
Full Access

The stem and the rasp for cemented arthroplasty are typically designed to obtain a cement mantle 2–5 mm thick. However, sometimes a line-to-line cementation is preferred, where the femoral cavity is prepared with the same dimension as the actual stem. There are contrasting reports [1,2] about the suitability of this technique to withstand the long-term fatigue loads. While the theoretical geometry allows no space for the cement, a sort of cement mantle is formed as the cement penetrates in the spongy bone. The scopes of this study were: 1) developing a dedicated in vitro method to test line-to-line cementation; 2) assessing if a short, polished hip stem designed for a standard cementation can be safely cemented line-to-line. In order to perform long-term mechanical in vitro tests, composite bones must be used, as cadaveric bones cannot withstand millions of loading cycles [3]. For this study, the Sawbones Mod. 3406-4 were chosen: they feature an open-cell polyurethane core simulating low-density spongy bone. Post-implantation x-rays confirmed that a relevant cement-bone interdigitation was obtained. Four femurs were prepared with a CoreHip (Aesculap) with regular cement mantle (Regular). Another 4 femurs were rasped to the same rasp size, and implanted with line-to-line cementation with a larger stem (Line-to-line). The implanted femurs were subjected to an accelerated test derived from a validated protocol [3] which replicates the most demanding motor tasks of 24 years of patient activity. Implant elastic micromotions and permanent migrations were measured throughout the test. The implants were then sectioned and treated with dye penetrants to highlight the cement cracks. Elastic and permanent motions did not show any loosening trend, and never exceeded few micrometers. As expected, some damage was visible in the cement mantles after test completion, for both types of implantation (similar to retrieved cement mantles surrounding stable implants [3]. The cement damage was similar in all specimens. No sign of major disruption was visible, neither within the Regular nor in the Line-to-line specimens: in fact, the cracks were limited in length, did not seem to cross the entire mantle thickness, and did not result in any loose cement fragments. The cracks in the line-to-line implants showed the same position and distribution compared to those found in the regular implants, but were slightly longer in some specimens. This in vitro study confirmed the feasibility of simulating line-to-line cementation in vitro. Our results suggest that a stem designed for a regular cement mantle could induce slightly more damage when implanted line-to-line, but no significant trend toward loosening


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_15 | Pages 88 - 88
1 Nov 2018
Griffin MTA Simpson A Hamilton D
Full Access

The first three months following Total Knee Arthroplasty (TKA) provide an early window into a patient's functional outcomes, with the change of function in this time yielding valuable insight. 20 patients due to undergo primary TKA were recruited to the study. Data were recorded at three time points; pre-assessment clinic (PAC) before the operation, 6-weeks-post-operation (6WKs), at 12-weeks-post-operation (12WKs). Functional activity levels were monitored during early post-operative recovery for changes in early functional outcome, and allowed a comparison of metrics at each time point. This included direct functional testing of power output, timed functional performance in clinic, patient reported outcome measures, and multiday activity monitoring devices. Maximal power output symmetry (Power) was similar at 6WKs vs PAC (p = 0.37). At 12WKs, it had increased (p < 0.05). Timed functional performance (Performance) remained similar across all three time points (p = 0.27). Patient reported activities of daily living (ADL) performance significantly increased at 6WKs vs PAC (p < 0.05). At 12WKs, it remained similar (p = 0.10). Patient daily step count significantly decreased at 6WKs vs PAC (p < 0.05). By 12WKs, this had increased to similar levels to PAC (p = 0.30). Within the functional outcome measures, strong post-operative correlations were observed between Power and Performance (r = 0.62), Power and ADL (r = 0.49), and Performance and ADL (r = 0.61). Despite reduced measured step count and similar functional performance, patients report improved ADL at 6WKs. When symmetrical power output and measured step count have improved at 12WKs, patients report similar ADL to that at 6WKs. Multiple measures are required to get a full picture, however this highlights the different aspects measured by different tools


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 717 - 725
1 May 2010
Kamali A Hussain A Li C Pamu J Daniel J Ziaee H Daniel J McMinn DJW

Hip simulators have been used for ten years to determine the tribological performance of large-head metal-on-metal devices using traditional test conditions. However, the hip simulator protocols were originally developed to test metal-on-polyethylene devices. We have used patient activity data to develop a more physiologically relevant test protocol for metal-on-metal devices. This includes stop/start motion, a more appropriate walking frequency, and alternating kinetic and kinematic profiles. There has been considerable discussion about the effect of heat treatments on the wear of metal-on-metal cobalt chromium molybdenum (CoCrMo) devices. Clinical studies have shown a higher rate of wear, levels of metal ions and rates of failure for the heat-treated metal compared to the as-cast metal CoCrMo devices. However, hip simulator studies in vitro under traditional testing conditions have thus far not been able to demonstrate a difference between the wear performance of these implants. Using a physiologically relevant test protocol, we have shown that heat treatment of metal-on-metal CoCrMo devices adversely affects their wear performance and generates significantly higher wear rates and levels of metal ions than in as-cast metal implants


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_31 | Pages 6 - 6
1 Aug 2013
Boyd A Soon V Sapare S McAllister J Deakin A Sarungi M
Full Access

Patient reported outcome measures (PROMs) are important for assessing the results of lower limb arthroplasty. Unrealistic or uneducated expectations may have a significant negative impact on PROMs even when surgery is technically successful. This study's aim was to quantify pre-operative expectations of Scottish patients undergoing total hip and knee replacement (THR/TKR). 100 THR and 100 TKR patients completed validated questionnaires (from the Hospital for Special Surgery) prior to their operation after receiving standard pre-operative information (booklet, DVD, consultations). Each patient rated expectations from very important to not having the expectation. A total score was calculated using a numerical scale for the grading of each expectation. Univariate regression analysis was used to investigate the relationship between demographics and expectation score. The THR cohort had mean age 66.2 (SD 10.5), 53% female, mean BMI 29.0 (SD 5.1) and mean Oxford score 44 (SD 7). The TKR cohort had mean age 67.6 (SD 8.5), 59% female, mean BMI 32.8 (SD 5.8) and mean Oxford score 44 (SD 8). 100% THR and 96% TKR patients had 10 or more expectations of their operation. All expected pain relief. Other improvements expected were: walking for 100% THA and 99% TKA patients; daily activities for 100% THAs and 96% TKAs; recreational activities for 96% THAs and 93% TKAs; sexual activity for 66% THAs and 59% TKAs; psychological well-being for 98% THAs and 91% TKAs. Regression analysis showed increasing age lowered expectations in both THR (p=0.025) and TKR (p=0.031) patients but that gender, BMI and Oxford score were not significantly related to expectations. This study highlights that patients expect far more than pain relief and improved post-operative mobility from their operation. It is important to discuss and manage these expectations with patients prior to surgery. By doing so, patient satisfaction and PROMs should further improve


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 333 - 333
1 Jul 2014
Senden R Heyligers I Grimm B
Full Access

Summary. Physical activity monitoring using a single accelerometer works reliably in clinical practice and is of added value as clinical outcome tool, as it provides objective and more precise information about a patient's activity compared to currently used questionnaires. Introduction. Standard clinical outcome tools do not comply with the new generation of patients who are younger and more active. To capture the high functional demands of these patients, current outcome scales have been optimised (e.g. New-Knee Society Score: New-KSS), new outcome scales have been developed (e.g. Knee disability and Osteoarthritis Outcome score: KOOS). Also objective measurement tools (e.g. activity monitors) have become increasingly popular. This study evaluates the pre- and postoperative TKA status of patients using such optimised and new outcome tools. Patients and Methods. Physical activity of 18 preoperative (68 ± 6yrs) and 16 postoperative (72 ± 7yrs, follow up range 9–10 yrs) TKA patients was assessed using a most recommended patient reported questionnaire, SQUASH (high value=more active) and a body-fixed 3D-accelerometer based activity monitor (AM). The AM was worn for 4–7 successive days at the lateral side of the nonaffected upper leg. Activity parameters (e.g. # steps, # transfers, # walking bouts of short/long duration, cadence) were derived using validated algorithms. Function was measured using patient reported questionnaires: KOOS-PS (range 0–100=worse) and New-KSS (higher score=better). Independent t-test, Mann-Whitney test and Pearson's r were used to compare groups and to investigate correlations (p<0.05). Results. All AMs and questionnaires were returned (response rate 100%) and showed similar or higher completion rates for the AM (100%) than questionnaires (range 82–100%). At 10yrs follow-up, the function of TKA patients was significantly improved with regards to preoperative showing lower KOOS-PS and higher New-KSS subscores. Also physical activity was higher at 10yrs showing significantly more steps/day (+39%). Other activity parameters like walking bouts and transfers were also higher in the postoperative group (resp.8% and 21%), but not significantly. Most walking bouts were short taking 10–30s (pre: 63%, post: 59% of the bouts) and consisting of 10–15 steps (pre: 78%, post: 75% of the bouts). The only correlation found between activity and functional outcomes was between AM data (amount steps, sitting, walking time) and the New-KSS Function walking & standing subscale (r-range 0.37–0.54). No correlation was found between AM data and SQUASH. Moderate to high correlations were found between functional outcomes (KOOS-PS vs. New-KSS, r-range −0.56 – −0.81). Discussion/Conclusion. At 10yr follow up, TKA patients continue to have higher functional scores and also maintain higher activity levels than preoperative, as is mainly shown in steps/day. The fact that only the New-KSS Function Walking & Standing subscale correlated with AM data indicates that function and activity are two widely independent outcome dimensions. This suggests that patients are active largely independent of their functional limitations (e.g. high activity, worse KOOS). The lack of correlation between objectively measured and patient reported physical activity indicates patients are less reliable in estimating their actual activity. The correlations between functional outcomes indicate redundancy. AM meets and exceeds response and completion rates of questionnaires. AM seems to be more objective, precise and sensitive to measure physical activity than questionnaires and adds a largely independent outcome dimension to clinical assessments


Bone & Joint Research
Vol. 6, Issue 11 | Pages 631 - 639
1 Nov 2017
Blyth MJG Anthony I Rowe P Banger MS MacLean A Jones B

Objectives

This study reports on a secondary exploratory analysis of the early clinical outcomes of a randomised clinical trial comparing robotic arm-assisted unicompartmental knee arthroplasty (UKA) for medial compartment osteoarthritis of the knee with manual UKA performed using traditional surgical jigs. This follows reporting of the primary outcomes of implant accuracy and gait analysis that showed significant advantages in the robotic arm-assisted group.

Methods

A total of 139 patients were recruited from a single centre. Patients were randomised to receive either a manual UKA implanted with the aid of traditional surgical jigs, or a UKA implanted with the aid of a tactile guided robotic arm-assisted system. Outcome measures included the American Knee Society Score (AKSS), Oxford Knee Score (OKS), Forgotten Joint Score, Hospital Anxiety Depression Scale, University of California at Los Angeles (UCLA) activity scale, Short Form-12, Pain Catastrophising Scale, somatic disease (Primary Care Evaluation of Mental Disorders Score), Pain visual analogue scale, analgesic use, patient satisfaction, complications relating to surgery, 90-day pain diaries and the requirement for revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 261 - 266
1 Feb 2005
Földhazy Z Arndt A Milgrom C Finestone A Ekenman I

Strains applied to bone can stimulate its development and adaptation. High strains and rates of strain are thought to be osteogenic, but the specific dose response relationship is not known. In vivo human strain measurements have been performed in the tibia to try to identify optimal bone strengthening exercises for this bone, but no measurements have been performed in the distal radial metaphysis, the most frequent site of osteoporotic fractures. Using a strain gauged bone staple, in vivo dorsal metaphyseal radial strains and rates of strain were measured in ten female patients during activities of daily living, standard exercises and falls on extended hands. Push-ups and falling resulted in the largest compression strains (median 1345 to 3146 με, equivalent to a 0.1345% to 0.3146% length change) and falling exercises in the largest strain rates (18 582 to 45 954 με/s). On the basis of their high strain and/or strain rates these or variations of these exercises may be appropriate for distal radial metaphyseal bone strengthening


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_VIII | Pages 48 - 48
1 Mar 2012
Beaulé PE
Full Access

The renewed interest in the clinically proven low wear of the metal-on-metal bearing combined with the capacity of inserting a thin walled cementless acetabular component has fostered the reintroduction of hip resurfacing. As in other forms of conservative hip surgery, i.e. pelvic osteotomies and impingement surgery, patient selection will help minimize complications and the need for early reoperation. Patient Selection and Hip Resurfacing. Although hip resurfacing was initially plagued with high failure rates, the introduction of metal on metal bearings as well as hybrid fixation has shown excellent survivorships of 97 to 99% at 4 to 5 years follow-up. However, it is important to critically look at the initial published results. In all of these series there was some form of patient selection. For example, in the Daniel and associates publications, only patients with osteoarthritis with an age less than 55 were included with 79% of patients being male. Treacy and associates stated that: “the operation was offered to men under the age of 65 years and women under the age of 60 years, with normal bone stock judged by plain radiographs and an expectation that they would return to an active lifestyle, including some sports”. However in the materials and methods, although the mean age is 52 years, the range is from 17 to 76 years including some patients with rheumatoid arthritis as well as osteonecrosis. Obviously, some form of patient selection is needed; but how one integrates them is where the Surface Arthroplasty Risk Index (SARI) is useful. With a maximum score of 6, points are assigned accordingly: femoral head cyst >1cm: 2 points; patient weight <82kg: 2 points; previous hip surgery: 1 point; UCLA Activity level >6: 1 point. A SARI score >3 represented a 4 fold increase risk in early failure or adverse radiological changes and with a survivorship of 89% at four years. The SARI index also proved to be relevant in assessing the outcome of the all cemented McMinn resurfacing implant (Corin¯, Circentester, England) at a mean follow-up of 8.7 years. Hips which had failed or with evidence of radiographic failure on the femoral side had a significantly higher SARI score than the remaining hips, 3.9 versus 1.9. Finally, one must consider the underlying diagnosis when evaluating a patient for hip resurfacing. In cases of dysplasia, acetabular deficiencies combined with the inability of inserting screws through the acetabular component may make initial implant stability unpredictable. This deformity in combination with a significant leg length discrepancy or valgus femoral neck could compromise the functional results of surface arthroplasty, and in those situations a stem type total hip replacement may provide a superior functional outcome. In respect to other diagnoses (osteonecrosis, inflammatory arthritis), initial analyses have not demonstrated any particular diagnostic group at greater risk of earlier failure. The only reservation we have is in patients with compromised renal function since metal ions generated from the metal-on-metal bearing are excreted through the urine and the lack of clearance of these ions may lead to excessive levels in the blood. Surgical Technique. Because resurfacing has not been within the training curriculum of orthopaedic surgeons for the last 2 decades, there will most likely be a learning curve in the integration of this implant within clinical practice. This data was confirmed for hip resurfacing when looking at the Canadian Academic Experience where in the first 50 cases of five arthroplasty surgeons only a 3.2% failure rate was noted of which 1.6% were due to neck fracture. Femoral neck fracture can occur because of significant varus positioning as well as osteonecrosis of the femoral head due to either disruption of the blood supply or over cement penetration. Finally, abnormal wear patterns leading to severe soft tissue reactions are being increasingly recognized and are related to either impingement or vertically placed acetabular components. Although impingement has long been recognized after total hip arthroplasty to limit range of motion and in extreme cases to hip instability, the risk after hip resurfacing may be greater since the femoral head-neck unit is preserved. Beaulé and associates have reported that 56% of hips treated by hip resurfacing have an abnormal offset ratio pre-operatively, with the two main diagnostic groups presenting deficient head-neck offset being osteonecrosis and osteoarthritis both of which have been associated with femoroacetabular impingement in the pre arthritic state. Conclusion. Although patients with a high activity level are likely to put their hip arthroplasties at risk for earlier failure, limiting a patient's activity because of fear of revision with a stem type hip arthroplasty has been shown to negatively impact the quality of life at long term follow-up. Thus hip resurfacing arthroplasty plays a significant role in the treatment of hip arthritis by permitting a return to full activities or what the patient perceives as his/her full capacities to do so, permitting them to enjoy a better quality of life without fearing a major hip revision