Advertisement for orthosearch.org.uk
Results 1 - 20 of 65
Results per page:
The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 14 - 20
1 Jan 2016
Zywiel MG Cherian JJ Banerjee S Cheung AC Wong F Butany J Gilbert C Overgaard C Syed K Jacobs JJ Mont MA

As adverse events related to metal on metal hip arthroplasty have been better understood, there has been increased interest in toxicity related to the high circulating levels of cobalt ions. However, distinguishing true toxicity from benign elevations in cobalt levels can be challenging. The purpose of this review is to examine the use of cobalt alloys in total hip arthroplasty, to review the methods of measuring circulating cobalt levels, to define a level of cobalt which is considered pathological and to review the pathophysiology, risk factors and treatment of cobalt toxicity. To the best of our knowledge, there are 18 published cases where cobalt metal ion toxicity has been attributed to the use of cobalt-chromium alloys in hip arthroplasty. Of these cases, the great majority reported systemic toxic reactions at serum cobalt levels more than 100 μg/L. This review highlights some of the clinical features of cobalt toxicity, with the goal that early awareness may decrease the risk factors for the development of cobalt toxicity and/or reduce its severity. Take home message: Severe adverse events can arise from the release of cobalt from metal-on-metal arthroplasties, and as such, orthopaedic surgeons should not only be aware of the presenting problems, but also have the knowledge to treat appropriately. Cite this article: Bone Joint J 2016;98-B:14–20


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 66 - 66
19 Aug 2024
Terhune EB Sutter EG Balkissoon R Pallante GD Specht L Leikin JB Kwon YM Lewallen DG Gerlinger TL Jacobs JJ
Full Access

Ceramic-on-ceramic (CoC) articulations in total hip arthroplasty (THA) have low wear, but the unique risk of fracture. After revision for CoC fracture, ceramic third bodies can lead to runaway wear of cobalt chrome (CoCr) causing extremely elevated blood cobalt. We present five cases of ceramic liner fractures revised to a CoCr head associated with the rapid development of severe cobalt toxicity. We identified 5 cases of fractured CoC THA treated with revision to CoCr on highly cross-linked polyethylene (HXLPE) – three to conventional bearings and two to modular dual mobility bearings (CoCr acetabular liner, CoCr femoral head, and HXLPE). Mean follow up was 2.5 years after CoCr/HXLPE re-revision. Symptoms of cobalt toxicity occurred at average 9.5 months after revision for ceramic fracture (range 6–12). All patients developed vision and hearing loss, balance difficulties, and peripheral neuropathy. Several had cardiomyopathy, endocrinopathy, and local skin discoloration. Two reported hip pain. Re-revision for cobalt toxicity occurred at an average of 22 months (range 10–36) after revision for ceramic fracture. Average serum cobalt level at re-revision was 991 μg/L (range 734–1302, normal <1 μg/L). All CoCr heads exhibited massive wear with asphericity; deep tissues exhibited prominent metallosis. Treatment consisted of debridement and revision to a ceramic head with HXLPE. Serum cobalt improved to an average of 25 μg/L at final follow up. All patients reported partial improvement in vision and hearing; peripheral neuropathy and balance did not recover. Systemic cobalt toxicity is a rare but devastating complication of ceramic fracture in THA treated with cobalt-alloy bearings. Cobalt alloy bearings should be avoided in this setting. The diagnosis of systemic cobalt toxicity requires a high index of suspicion and was typically delayed following systemic symptoms. Debridement and revision to a ceramic-on-HXLPE leads to improvement but not resolution of cobalt toxicity complications


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 45 - 45
7 Jun 2023
Howard D Manktelow B DeSteiger R Skinner J Ashford R
Full Access

Ceramic bearing fractures are rare events, but mandate revision and implantation of new bearings. Revisions using metal heads have been reported to lead to gross volumetric head wear (due to abrasive retained ceramic micro-debris), cobalt toxicity, multi-organ failure and death. Such complications are widely published (50+ reports), yet we know that patients continue to be put at risk. Using data from the NJR and AOANJRR, this study seeks to compare the risk of re-revision and death by revision bearing combination following a ceramic bearing fracture. Data were extracted from the NJR and AOANJRR, identifying revisions for ceramic bearing fracture. Subsequent outcomes of survival, re-revision and death were compared between revision bearing combinations (ceramic-on-ceramic, ceramic-on-polyethylene, and metal-on-polyethylene). 366 cases were available for analysis from the NJR dataset (MoP=34, CoP=112, CoC=221) and 174 from the AOANJRR dataset (MoP=17, CoP=44, CoC=113). The overall incidence rate of adverse outcome (revision or death) was 0.65 for metal heads and 0.23 for ceramic head articulations (p=0.0012) across the whole time period (NJR). Kaplan-Meir survival estimates demonstrate an increased risk of both re-revision and death where a metal head has been used vs a ceramic head following revision for ceramic fracture. There are few decisions in arthroplasty surgery that can lead to serious harm or death for our patients, but revision using a metal head following ceramic bearing fracture is one of them. This study enhances the signal of what is already known but previously only reported as inherently low-level evidence (case reports and small series) due to event rarity. Use of a metal head in revision for ceramic fracture represents an avoidable patient safety issue, which revision guidelines should seek to address


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 31 - 37
1 Jan 2013
Zywiel MG Brandt J Overgaard CB Cheung AC Turgeon TR Syed KA

Symptomatic cobalt toxicity from a failed total hip replacement is a rare but devastating complication. It has been reported following revision of fractured ceramic components, as well as in patients with failed metal-on-metal articulations. Potential clinical findings include fatigue, weakness, hypothyroidism, cardiomyopathy, polycythaemia, visual and hearing impairment, cognitive dysfunction, and neuropathy. We report a case of an otherwise healthy 46-year-old patient, who developed progressively worsening symptoms of cobalt toxicity beginning approximately six months following synovectomy and revision of a fractured ceramic-on-ceramic total hip replacement to a metal-on-polyethylene bearing. The whole blood cobalt levels peaked at 6521 µg/l. The patient died from cobalt-induced cardiomyopathy. Implant retrieval analysis confirmed a loss of 28.3 g mass of the cobalt–chromium femoral head as a result of severe abrasive wear by ceramic particles embedded in the revision polyethylene liner. Autopsy findings were consistent with heavy metal-induced cardiomyopathy. We recommend using new ceramics at revision to minimise the risk of wear-related cobalt toxicity following breakage of ceramic components. Cite this article: Bone Joint J 2013;95-B:31–7


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 7 - 7
1 Jun 2017
Berber R Abdel-Gadir A Palla L Moon J Manisty C Skinner J Hart A
Full Access

Circulating cobalt and chromium from metal-on-metal implants cause rare but fatal autopsy-diagnosed cardiotoxicity. Concern exists that milder cardiotoxicity may be common and under-recognized. Unacceptably high failure rates of metal-on-metal hip implants have prompted regulatory authorities to issue worldwide safety alerts. Despite this, approximately 1 million patients continue to live with metal-on-metal implants, putting them at risk of systemic toxicity. Although blood cobalt and chromium levels are easily measured and track local toxicity, no non-invasive tests for organ deposition exist. We recently demonstrated the utilisation of a T2* protocol (cardiovascular MRI) to detect cobalt and chromium deposition within the liver of a patient with elevated blood cobalt levels (confirmed by liver biopsy tissue analysis and X-ray fluorescence spectroscopy). We sought to detect and constrain the correlation between blood metal ions and a comprehensive panel of established markers of early cardiotoxicity. In addition we applied T2* protocols with the aim of detecting cardiac metal deposition. 90 patients were recruited through RNOH clinics into this prospective single centre blinded study. Patients were divided into 3 age and gender-matched groups according to type of implant and blood metal ion levels as follows: [Group A] Non-metal bearing hip implants; [Group B] Metal-on-metal implants, low blood metal ion levels (<7ppb); and [Group C] Metal-on-metal implants, high blood levels (>7ppb). All underwent detailed cardiovascular phenotyping using cardiac MRI (with T2*, T1 and ECV mapping, in addition to LV size and ejection fraction), advanced echocardiography (LV size and ejection fraction), and cardiac blood biomarker (Troponin and BNP) sampling in the same sitting at the Heart Hospital London. Primary outcomes were pre-specified. See study flow diagram – figure 1. (The study was registered with . clinicaltrials.gov. : NCT02331264). Blood cobalt levels were significantly different between groups (0.17ppb (range 0·10–0·47, SD 0·08) vs. 2·47 (0·72–6·9, SD 1·81) vs. 30·0 (7·54–118.0, SD 29·1) respectively for groups A, B and C). No significant between-group differences were found for LV size, ejection fraction (CMR or echocardiography), LA size, T1, T2*, ECV, BNP or troponin, with all results within normal ranges. There was no relationship between blood cobalt levels and either left ventricular ejection fraction or T2* (r=-0·022 and r=-0·108 respectively). Although small, the study was sufficiently powered to detect, as a minimum, a difference in ejection fraction of 4.8% (Cohen's d effect size 0·8). Using best available technologies, exposure of patients with metal-on-metal hip implants to high (but not extreme) blood cobalt and chromium levels has no detectable effect on the heart. We believe these findings will offer reassurance to one million patients worldwide living with a metal-on-metal hip implant and will support clinicians caring for such patients. For any figures or tables, please contact the authors directly by clicking on ‘Info & Metrics’ above to access author contact details


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 11 - 11
1 Oct 2019
Wyles CC Paradise CR Masters TL Patel R Abdel MP Trousdale RT Sierra RJ
Full Access

Introduction. Adverse local tissue reactions (ALTR) can result in devastating soft tissue and osseous destruction, while potentially increasing the risk of concomitant periprosthetic joint infection (PJI). The aims of this study were to evaluate cobalt (Co) and chromium (Cr) levels generated in simulators from metal-on-polyethylene (MoP) and ceramic-on-polyethylene (CoP) constructs, and determine their impact on native tissues and PJI risk through evaluation of human adipose-derived mesenchymal stem cells (AMSCs) and Staphylococcus epidermidis isolates. Methods. Ten hip simulator constructs were assembled with 36-mm high-offset femoral heads, highly cross-linked polyethylene liners, and titanium stems. Five constructs used CoCr femoral heads and five used ceramic. Constructs were submerged in bovine serum (BS) and run for 1,000,000 cycles. Samples of BS were collected and evaluated for CoCr concentration. Various concentrations of CoCr were chosen for further assessment of cytotoxicity and growth impact on AMSCs and S. epidermidis and compared to inert SiO2. Results. After 1,000,000 cycles, mean MoP and CoP Co concentration was 2264 ng/mL and 0.6 ng/mL, respectively (p<0.001). Mean MoP and CoP Cr concentration was 217 ng/mL and 4.3 ng/mL, respectively (p<0.001). Mean MoP Co:Cr ratio was 10. Co nanoparticles were significantly more toxic to human AMSCs than control SiO2 in a dose-response manner (p<0.001). S. epidermidis growth was not significantly impacted by Co concentrations derived from the simulators. Conclusions. MoP constructs built in ideal conditions generated substantial CoCr debris, highlighting a baseline risk with these implants that may be exacerbated by host factors or imperfect surgical technique. Evaluation of impact on AMSCs suggests that debris levels produced under ideal conditions can be cytotoxic. Additionally, these concentrations did not potentiate or inhibit S. epidermidis growth, suggesting elevated PJI rates with ALTR may be related to other factors potentially associated with tissue necrosis. For any tables or figures, please contact the authors directly


Bone & Joint Open
Vol. 4, Issue 8 | Pages 628 - 635
22 Aug 2023
Hedlundh U Karlsson J Sernert N Haag L Movin T Papadogiannakis N Kartus J

Aims

A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision THAs without infection performed using the same lateral approach.

Methods

A group of eight patients revised due to a PJI with a previous lateral approach was compared with a group of 21 revised THAs without infection, performed using the same approach. The primary variables of the study were the fibril diameter, as seen in transmission electron microscopy (TEM), and the total degeneration score (TDS), as seen under the light microscope. An analysis of bacteriology, classification of infection, and antibiotic treatment was also performed.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 511 - 517
1 May 2023
Petrie MJ Panchani S Al-Einzy M Partridge D Harrison TP Stockley I

Aims

The duration of systemic antibiotic treatment following first-stage revision surgery for periprosthetic joint infection (PJI) after total hip arthroplasty (THA) is contentious. Our philosophy is to perform an aggressive debridement, and to use a high local concentration of targeted antibiotics in cement beads and systemic prophylactic antibiotics alone. The aim of this study was to assess the success of this philosophy in the management of PJI of the hip using our two-stage protocol.

Methods

The study involved a retrospective review of our prospectively collected database from which we identified all patients who underwent an intended two-stage revision for PJI of the hip. All patients had a diagnosis of PJI according to the major criteria of the Musculoskeletal Infection Society (MSIS) 2013, a minimum five-year follow-up, and were assessed using the MSIS working group outcome-reporting tool. The outcomes were grouped into ‘successful’ or ‘unsuccessful’.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 304 - 311
15 Apr 2024
Galloway R Monnington K Moss R Donaldson J Skinner J McCulloch R

Aims

Young adults undergoing total hip arthroplasty (THA) largely have different indications for surgery, preoperative function, and postoperative goals compared to a standard patient group. The aim of our study was to describe young adult THA preoperative function and quality of life, and to assess postoperative satisfaction and compare this with functional outcome measures.

Methods

A retrospective cohort analysis of young adults (aged < 50 years) undergoing THA between May 2018 and May 2023 in a single tertiary centre was undertaken. Median follow-up was 31 months (12 to 61). Oxford Hip Score (OHS) and focus group-designed questionnaires were distributed. Searches identified 244 cases in 225 patients. Those aged aged under 30 years represented 22.7% of the cohort. Developmental dysplasia of the hip (50; 45.5%) and Perthes’ disease (15; 13.6%) were the commonest indications for THA.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 867 - 874
1 Jul 2022
Ji B Li G Zhang X Xu B Wang Y Chen Y Cao L

Aims

Periprosthetic joint infections (PJIs) with prior multiple failed surgery for reinfection represent a huge challenge for surgeons because of poor vascular supply and biofilm formation. This study aims to determine the results of single-stage revision using intra-articular antibiotic infusion in treating this condition.

Methods

A retrospective analysis included 78 PJI patients (29 hips; 49 knees) who had undergone multiple prior surgical interventions. Our cohort was treated with single-stage revision using a supplementary intra-articular antibiotic infusion. Of these 78 patients, 59 had undergone more than two prior failed debridement and implant retentions, 12 patients had a failed arthroplasty resection, three hips had previously undergone failed two-stage revision, and four had a failed one-stage revision before their single-stage revision. Previous failure was defined as infection recurrence requiring surgical intervention. Besides intravenous pathogen-sensitive agents, an intra-articular infusion of vancomycin, imipenem, or voriconazole was performed postoperatively. The antibiotic solution was soaked into the joint for 24 hours for a mean of 16 days (12 to 21), then extracted before next injection. Recurrence of infection and clinical outcomes were evaluated.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims

A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database.

Methods

Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 847 - 851
1 Jul 2008
Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gibbons CLM Ostlere S Athanasou N Gill HS Murray DW

We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken. All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision. We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 22 - 22
1 Nov 2015
Eyre-Brook A Pearce A Lyle N Stranks G Briant-Evans T
Full Access

Introduction. The prevalence of adverse reactions to metal debris around metal-on-metal (MOM) hip replacements has been reported to range from 7 to 69%. Little has been published on MRI scans with conventional total hip bearing surfaces. This study aimed to establish the prevalence of soft tissue lesions associated with metal-on-polyethylene (MOP) and ceramic-on-ceramic (COC) bearings, compared to MOM prostheses. Patients/Materials & Methods. All Metal Artefact Reduction Sequence (MARS) MRI scans for MOM THRs performed at our unit from January 2009 to present were reviewed, identifying those with contralateral primary MOP or COC THRs included on the scan. These were compared to a previously analysed cohort of 281 MOM THRs. Scans were classified using the Modified Oxford Classification as ‘Normal’, ‘Trochanteric Fluid’, ‘Effusions’ or Adverse Reactions to Debris. Results. 72 MOP and 17 COC prostheses were identified. More abnormalities were seen with scans performed within 12 months of surgery; 7 of 11 having findings consistent with post-operative seromas. These 11 were therefore excluded from further analysis. A higher proportion of normal MRIs were seen in the conventional prostheses group (78% v 34%, p<0.001). Adverse reactions to debris were significantly more common in MOM prostheses (39% v 4%, p<0.001). Effusions were more common with MOM bearings (13% v 4%, p=0.028). There was no difference in the rates of trochanteric fluid (both groups 14%). Discussion. A much higher rate of clinically significant fluid collections were seen with MOM prostheses as a result of increased toxicity of metal debris versus polyethylene or ceramic. There was little difference between MOP and COC scans. Trochanteric fluid collections are more likely to represent bursitis than any problem with the bearing surface. Conclusion. These results provide a useful baseline for comparison of MOM MARS MRI scans, adding to our understanding of the unique failure patterns of MOM prostheses


Bone & Joint Open
Vol. 3, Issue 1 | Pages 61 - 67
18 Jan 2022
van Lingen CP Ettema HB Bosker BH Verheyen CCPM

Aims

Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels.

Methods

All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.


Bone & Joint Research
Vol. 10, Issue 10 | Pages 693 - 703
1 Oct 2021
Wang X Wang D Xia P Cheng K Wang Q Wang X Lin Q Song J Chen A Li X

Aims

To evaluate the effect of ultrasound-targeted simvastatin-loaded microbubble destruction (UTMDSV) for alleviation of the progression of osteoarthritis (OA) in rabbits through modulation of the peroxisome proliferator-activated receptor (PPARγ).

Methods

In vitro, OA chondrocytes were treated with ultrasound (US), US-targeted microbubble destruction (UTMD), simvastatin (SV), and UTMDSV on alternate days for four weeks. Chondrocytes were also treated with PPARγ inhibitor, PPARγ inhibitor+ UTMDSV, and UTMDSV. The cholesterol efflux rate and triglyceride levels were measured using an assay kit and oil red O staining, respectively. In vivo, the OA rabbits were treated with a single intra-articular injection of UTMD, SV, and UTMDSV every seven days for four weeks. Cartilage histopathology was assessed by safranin-O staining and the Mankin score. Total cholesterol (TC) and high-density lipoprotein-cholesterol (HDL-C) in rabbit knee synovial fluid were detected by enzyme-marker assay. Aggrecan, collagen II, and PPARγ expression levels were analyzed by Western blotting (WB).


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 192 - 199
1 Mar 2000
Engelbrecht E von Foerster G Delling G

Glass ionomer cement (Ionocem) was developed for use in bone surgery and is reported to be notably biocompatible. Between 1991 and 1994 we performed revision operations for aseptic loosening of arthroplasties of the hip on 45 patients using this material in its granulate form (Ionogran) mixed with homologous bone as a bone substitute. Of these 45 patients, 42 were followed up for a mean of 42 months. Early reloosening of the acetabular component has occurred in ten after a mean of 30 months. Histological examination showed large deposits of aluminium in the adjacent connective tissue and bone. Osteoblastic function and bone mineralisation were clearly inhibited. The serum levels of aluminium were also increased. The toxic damage at the bone interface caused by high local levels of aluminium must be seen as an important factor in the high rate of early reloosening. Our findings cast doubt on the biocompatibility of this material and we do not recommend continuation of its further use in orthopaedic surgery


Bone & Joint Research
Vol. 10, Issue 8 | Pages 498 - 513
3 Aug 2021
Liu Z Lu C Shen P Chou S Shih C Chen J Tien YC

Aims

Interleukin (IL)-1β is one of the major pathogenic regulators during the pathological development of intervertebral disc degeneration (IDD). However, effective treatment options for IDD are limited. Suramin is used to treat African sleeping sickness. This study aimed to investigate the pharmacological effects of suramin on mitigating IDD and to characterize the underlying mechanism.

Methods

Porcine nucleus pulposus (NP) cells were treated with vehicle, 10 ng/ml IL-1β, 10 μM suramin, or 10 μM suramin plus IL-1β. The expression levels of catabolic and anabolic proteins, proinflammatory cytokines, mitogen-activated protein kinase (MAPK), and nuclear factor (NF)-κB-related signalling molecules were assessed by Western blotting, quantitative real-time polymerase chain reaction (qRT-PCR), and immunofluorescence analysis. Flow cytometry was applied to detect apoptotic cells. The ex vivo effects of suramin were examined using IDD organ culture and differentiation was analyzed by Safranin O-Fast green and Alcian blue staining.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims

We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice.

Methods

Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1035 - 1041
1 Sep 2019
Markel DC Bou-Akl T Rossi MD Pizzimenti N Wu B Ren W

Aims

The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components.

Patients and Methods

A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.