Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 14 - 14
17 Apr 2023
Bartolo M Newman S Dandridge O Provaggi E Accardi M Dini D Amis A
Full Access

No proven long-term joint-preserving treatment options exist for patients with irreparable meniscal damage. This study aimed to assess gait kinematics and contact pressures of novel fibre-matrix reinforced polyvinyl alcohol-polyethylene glycol (PVA-PEG) hydrogel meniscus implanted ovine stifle joints against intact stifles in a gait simulator.

The gait simulator controlled femoral flexion-extension and applied a 980N axial contact force to the distal end of the tibia, whose movement was guided by the joint natural ligaments (Bartolo; ORS 2021;p1657- LB). Five right stifle joints from sheep aged >2 years were implanted with a PVA-PEG total medial meniscus replacement, fixed to the tibia via transosseous tunnels and interference screws. Implanted stifle joint contact pressures and kinematics in the simulator were recorded and compared to the intact group. Contact pressures on the medial and lateral condyles were measured at 55° flexion using Fujifilm Prescale Low Pressure film inserted under the menisci. 3D kinematics were measured across two 30 second captures using the Optotrak Certus motion-tracking system (Northern Digital Inc.).

Medial peak pressures were not significantly different between the implanted and intact groups (p>0.4), while lateral peak pressures were significantly higher in the implanted group (p<0.01). Implanted stifle joint kinematics in the simulator did not differ significantly from the intact baseline (p>0.01), except for in distraction-compression (p<0.01).

Our findings show that the fibre-matrix reinforced PVA-PEG hydrogel meniscal replacement restored the medial peak contact pressures. Similar to published literature (Fischenich; ABE 2018;46(11):1–12), the lateral peak pressures in the implanted group were higher than the intact. Joint kinematics were similar across groups, with slightly increased internal-external rotation in the implanted group. These findings highlight the effectiveness of the proposed approach and motivate future work on the development of a total meniscal replacement.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 55 - 55
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

The need for gender specific knee arthroplasty is debated. This research aimed to establish whether gender differences in patellar tendon moment arm (PTMA), a composite measure that characterises function of both the patellofemoral and tibiofemoral joints, are a consequence of knee size or other variation.

Methods

PTMA about the instantaneous helical axis was calculated from positional data acquired using optical tracking. First, data post-processing was optimised, comparing four smoothing techniques (raw, Butterworth filtered, generalised cross-validation cubic spline interpolated and combined filtered/interpolated) using a fabricated knee. Then PTMA was measured during open-chain extension for N=24 (11 female) fresh-frozen cadaveric knees, with physiologically based loading and extension rates (420°/s) applied. Gender differences in PTMA were assessed before and after accounting for knee size with epicondylar width.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 27 - 27
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

Unicompartmental (UKA) and bicompartmental (BCA) knee arthroplasty are associated with improved functional outcomes compared to Total Knee Arthroplasty (TKA) in suitable patients, although the reason is poorly understood. The aim of this study was to measure how the different arthroplasties affect knee extensor function.

Methods

Extensor function was measured for sixteen cadaveric knees and then re-tested following the different arthroplasties. Eight knees underwent medial UKA then BCA, then posterior-cruciate retaining TKA, and eight underwent the lateral equivalents then TKA. Extensor efficiency was calculated for ranges of knee flexion associated with common activities of daily living. Data were analyzed with repeated measures analysis of variance (α=0.05).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 83 - 83
1 Aug 2012
Younge A Phillips A Amis A
Full Access

Finite element (FE) modelling has been widely used to create and assess musculoskeletal models. However to achieve a high degree of resolution in describing the structure, significant computational power and time are required. The objective of this study was to introduce a complimentary approach to FE modelling using structural beam theory. This requires far less computational power and models can be analyzed in a fraction of a second, offering quick, intuitive results for engineers and surgeons.

Beam theory was first introduced as a method for analyzing the stresses in long bones in 1917. It was used as the de facto method for several decades. The introduction of FE modelling offered great advances; beam theory calculations were considered laborious and less accurate. However with the advances in computational power so too comes the ability to create modern automated beam theory models.

A study was conducted using the commercially available general structural analysis software Oasys GSA. A synthetic biomechanical femur was CT scanned and the solid model constructed. This model was sectioned into approximately seventy sections in the regions of the shaft and condyles, thirty in the neck and thirty in the head. Line plots of the shape of each of the sections, for both cortical and trabecular parts, were then imported into Oasys GSA. The centroid, area, second moments of area and torsion constant were calculated for each section. The sections were plotted at the position of the cortical centroid and parallel axis theorem was used to plot the trabecular section in the same position. A force representing the hip joint reaction force was applied to a node corresponding to the centre of the femoral head. Muscular forces were applied to stiff radial elements according to those active at the point of peak joint contact force during gait.

Oasys GSA produced instant results showing moment and deflection characteristics of the femur. This data was then used to predict strain plots, which were directly compared to FE results. Initial results compare favourably.

This study has demonstrated an updated fast, efficient and intuitive alternative to finite element modelling.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 82 - 82
1 Aug 2012
Younge A Phillips A Amis A
Full Access

Finite element models of the musculoskeletal system have the possibility of describing the in vivo situation to a greater extent than a single in vitro experimental study ever could. However these models and the assumptions made must be validated before they can be considered truly useful. The object of this study was to validate, using digital image correlation (DIC) and strain gauging, a novel free boundary condition finite element model of the femur.

The femur was treated as a complete musculoskeletal construct without specific fixed restraint acting on the bone. Spring elements with defined force-displacement relationships were used to characterize all muscles and ligaments crossing the hip and knee joints. This model was subjected to a loading condition representing single leg stance. From the developed model muscle, ligament and joint reaction forces were extracted as well as displacement and strain plots. The muscles with the most influence were selected to be represented in the simplified experimental setup.

To validate the finite element model a balanced in vitro experimental set up was designed. The femur was loaded proximally through a construct representative of the pelvis and balanced distally on a construct representing the tibio-femoral joint. Muscles were represented using a cabling system with glued attachments. Strains were recorded using DIC and strain gauging. DIC is an image analysis technique that enables non-contact measurement of strains across surfaces. The resulting strain distributions were compared to the finite element model.

The finite element model produced hip and knee joint reaction forces comparable to in vivo data from instrumented implants. The experimental models produced strain data from both DIC and strain gauging; these were in good agreement with the finite element models. The DIC process was also shown to be a viable method for measuring strain on the surface of the specimen.

In conclusion a novel approach to finite element modeling of the femur was validated, allowing greater confidence for the model to be further developed and used in clinical settings.