MAVRIC-SL (MVSL), a multi-spectral magnetic resonance imaging (MRI) sequence [1] used for visualization of tissues near hip arthroplasty, can be accelerated by a short calibration scan [2], permitting an isotropic acquisition in a clinically feasible scan time. Decreasing the repetition time (TR) can further reduce scan time. This study evaluated the feasibility of acquiring isotropic MAVRIC-SL images (MVISO) and reduced TR isotropic MAVRIC-SL images (TRMVISO) of hip arthroplasties to elucidate the added clinical benefit of such acquisitions. Following IRB approval, MVSL, MVISO and TRMVISO images were acquired for 93 hip arthroplasties (84 subjects). Images were reviewed by 2 radiologists to evaluate clinical and image quality features. Comparison of features between scanning sequences was assessed by mixed-effects ordinal logistic regression and odds ratios (OR). Inter/intra-rater agreement was determined using Gwet's agreement coefficient (AC).Introduction
Methods