Vol. 99-B, Issue SUPP_8 | Pages 93 - 93
1 Apr 2017
Karaaslan A Karakaşlı A Ertem F Aycan H
Full Access

Background

Intramedullary nailing is a widely accepted treatment method for femoral fractures. Failure of locking screws is often a threatening complication, particularly on comminuted fractures. For comminuted fractures, the locking nails are load-bearing devices. The load transfer between fractured fragments is made through especially the locking screws for these load bearing situations. Nonunion, malunion, delayed union, shortening, and nail migration are the expected results if early failure of locking screws is present with comminuted fractures. In this study our aim was to compare the bending resistance of titanium and stainless steel locking screws.

Methods

We tested 60 locking screws in six groups (titanium, stainless steel, unthreaded, low threaded and high threaded) in a steel tube that has 30 mm inner diameter, which imitates the lesser trochanter level. We determined the yield points at three-point bending tests that permanent deformation started in the locking screws using an axial compression testing machine.