Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 133 - 133
1 Mar 2010
Taylor RM Bernero J Patel A Brodke D Khandkar A
Full Access

Background: Implantable grade silicon nitride (Si3N4), such as MC2 TM from Amedica, is an attractive alternative to the common bearing materials used today for spinal and orthopedic applications. Ceramic bearings used in total hip arthroplasty (THA) consist of a femoral head (ball) articulating inside an acetabular cup (socket); the ball and socket are made of alumina (Al2O3) or Al2O3-based composite materials such as zirconia toughened alumina. Extensive studies on the mechanical, imaging and tribological characteristics of MC2 for spinal implants have been conducted. This paper summarizes some of the results from these studies.

Methods: Si3N4 materials properties were evaluated per ASTM standards C-1161 for flexural strength and E-399 for toughness, using the single edge notched beam method. Strength was also measured after subjecting the ceramic to accelerated aging conditions. Imaging compatibility of the MC2, PEEK, titanium, cobalt-chromium (CoCr) and trabecular metal discs was determined under a variety of imaging methods, including MRI. Wear testing of Si3N4 acetabular cups articulating against Si3N4 and CoCr femoral heads was also conducted. Biocompatibility evaluation of the Si3N4 was conducted per ISO 10993 protocols.

Results: MC2 Si3N4 exhibits improved mechanical properties over modern Al2O3 and Al2O3-based composite THA bearings, with a flexural strength of 920 ± 70 MPa, a Weibull modulus of 19, and a fracture toughness of 10 ± 1 MPa.m1/2. Accelerated aging of Si3N4 did not adversely affect the flexural strength. The imaging comparison study showed that Si3N4 did not exhibit any MRI imaging artifacts similar to PEEK, but unlike the metallic materials used for spinal implants. In wear tests using a hip simulator, Si3N4 acetabular cups produced low volumetric wear when articulating against Si3N4 or CoCr femoral heads. The implantable grade Si3N4 material was also found to be biocompatible.

Conclusions: Si3N4 exhibits a superior combination of properties including strength, toughness, MRI compatibility, and wear resistance. This may allow spinal implants with an important combination of properties than hitherto possible, including a wider range of design options, improved anatomic fit and imaging compatibility. Additionally, spinal vertebral body replacement implants were tested in static and fatigue conditions per the ASTM F 2077 and have been cleared by the FDA. An MC2 Si3N4 based total cervical disc implant has been designed and is undergoing pre-clinical testing. The versatility of the MC2 Si3N4 ceramic may enable its use in total disc replacements owing to its combination of superior wear resistance and imaging compatibility.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 28 - 29
1 Mar 2010
Slosar P Youssef JA Reynolds J Patty CA Brodke D
Full Access

Purpose: Few opportunities exist for physicians to easily obtain immediate insights or solicit direct advice from a respected peer on a complicated case. As a result, physicians end up relying heavily on journals, textbooks, and other dated resources that may prove to be dated or inadequate in addressing unique patient problems. Furthermore, the typical training modules available are limited to a one- or two-day experience. Often times this situation does not provide an adequate understanding of the surgical indications, techniques, and potential complications in the application of new technology. Consequently, many surgeons fail to adopt new technologies due to their lack of understanding of the application of such novel technologies. Recently, Reynolds et al studied the effect of peer collaboration among surgeons at a weekly surgery conference. They found that changes in surgical decision-making occurred in 12.5% of all cases reviewed and in 20% of the complex cases. However, collaboration should not be limited to weekly meetings or annual conferences. The Internet has introduced Collaborative Knowledge Networks (CKNs) that allow communities of surgeons to collaborate virtually on complex patient cases—allowing for rapid, bidirectional information. SpineConnect has surfaced as an example of a CKN that allows orthopedic spine surgeons to discuss surgical decisions.

Method: We studied the use of Internet collaboration among spine surgeons using the SpineConnect website. We also compared the improvement in the adoption rate of novel technologies using technology fellowships versus traditional training modules. These fellowships use both in-person and virtual collaboration through the Spine-Connect platform as a new means of training physicians.

Results: Using the SpineConnect website, surgical decision-making and planning was altered in 34% of active users and 90% agreed that the ability to collaborate with their peers, using a dedicated website, has improved their ability to practice spinal surgery. Continuous communication and dialogue resulting from the technology training fellowships encouraged better understanding of surgical indications, a greater comprehension of surgical technique, and a mechanism to discuss and solve complications. When such fellowships were implemented, we found a 30% improvement in adoption rate over traditional training modules.

Conclusion: The treatment of spine surgery patients has benefited from the utilization of collaborative platforms such as SpineConnect. In spine surgeon training and education, we found that CKNs promoted better surgical decision-making and increased understanding of new technology application.