Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 13 - 13
1 Dec 2022
Reeves J Spangenberg G Elwell J Stewart B Vanasse T Roche C Faber KJ Langohr GD
Full Access

Shoulder arthroplasty humeral stem design has evolved to accommodate patient anatomy characteristics. As a result, stems are available in numerous shapes, coatings, lengths, sizes, and vary by fixation method. This abundance of stem options creates a surgical paradox of choice. Metrics describing stem stability, including a stem's resistance to subsidence and micromotion, are important factors that should influence stem selection, but have yet to be assessed in response to the diametral (i.e., thickness) sizing of short stem humeral implants.

Eight paired cadaveric humeri (age = 75±15 years) were reconstructed with surgeon selected ‘standard’ sized short-stemmed humeral implants, as well as 2mm ‘oversized’ implants. Stem sizing conditions were randomized to left and right humeral pairs. Following implantation, an anteroposterior radiograph was taken of each stem and the metaphyseal and diaphyseal fill ratios were quantified. Each humerus was then potted in polymethyl methacrylate bone cement and subjected to 2000 cycles of 90º forward flexion loading. At regular intervals during loading, stem subsidence and micromotion were assessed using a validated system of two optical markers attached to the stem and humeral pot (accuracy of <15µm).

The metaphyseal fill ratio did not differ significantly between the oversized and standard stems (0.50±0.06 vs 0.50±0.10; P = 0.997, Power = 0.05); however, the diaphyseal fill ratio did (0.52±0.06 vs 0.45±0.07; P < 0.001, Power = 1.0). Neither fill ratio correlated significantly with stem subsidence or micromotion. Stem subsidence and micromotion were found to plateau following 400 cycles of loading. Oversizing stem thickness prevented implant head-back contact in all but one specimen with the least dense metaphyseal bone, while standard sizing only yielded incomplete head-back contact in the two subjects with the densest bone. Oversized stems subsided significantly less than their standard counterparts (standard: 1.4±0.6mm, oversized: 0.5±0.5mm; P = 0.018, Power = 0.748;), and resulted in slightly more micromotion (standard: 169±59µm, oversized: 187±52µm, P = 0.506, Power = 0.094,).

Short stem diametral sizing (i.e., thickness) has an impact on stem subsidence and micromotion following humeral arthroplasty. In both cases, the resulting three-dimensional stem micromotion exceeded, the 150µm limit suggested for bone ingrowth, although that limit was derived from a uniaxial assessment. Though not statistically significant, the increased stem micromotion associated with stem oversizing may in-part be attributed to over-compacting the cancellous bed during broaching, which creates a denser, potentially smoother, interface, though this influence requires further assessment. The findings of the present investigation highlight the importance of proper short stem diametral sizing, as even a relatively small, 2mm, increase can negatively impact the subsidence and micromotion of the stem-bone construct. Future work should focus on developing tools and methods to support surgeons in what is currently a subjective process of stem selection.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 80 - 80
1 Dec 2022
Reeves J Spangenberg G Elwell J Stewart B Vanasse T Roche C Langohr GD Faber KJ
Full Access

Shoulder arthroplasty is effective at restoring function and relieving pain in patients suffering from glenohumeral arthritis; however, cortex thinning has been significantly associated with larger press-fit stems (fill ratio = 0.57 vs 0.48; P = 0.013)1. Additionally, excessively stiff implant-bone constructs are considered undesirable, as high initial stiffness of rigid fracture fixation implants has been related to premature loosening and an ultimate failure of the implant-bone interface2. Consequently, one objective which has driven the evolution of humeral stem design has been the reduction of stress-shielding induced bone resorption; this in-part has led to the introduction of short stems, which rely on metaphyseal fixation. However, the selection of short stem diametral (i.e., thickness) sizing remains subjective, and its impact on the resulting stem-bone construct stiffness has yet to be quantified.

Eight paired cadaveric humeri (age = 75±15 years) were reconstructed with surgeon selected ‘standard’ sized and 2mm ‘oversized’ short-stemmed implants. Standard stem sizing was based on a haptic assessment of stem and broach stability per typical surgical practice. Anteroposterior radiographs were taken, and the metaphyseal and diaphyseal fill ratios were quantified. Each humerus was then potted in polymethyl methacrylate bone cement and subjected to 2000 cycles of compressive loading representing 90º forward flexion to simulate postoperative seating. Following this, a custom 3D printed metal implant adapter was affixed to the stem, which allowed for compressive loading in-line with the stem axis (Fig.1). Each stem was then forced to subside by 5mm at a rate of 1mm/min, from which the compressive stiffness of the stem-bone construct was assessed. The bone-implant construct stiffness was quantified as the slope of the linear portion of the resulting force-displacement curves.

The metaphyseal and diaphyseal fill ratios were 0.50±0.10 and 0.45±0.07 for the standard sized stems and 0.50±0.06 and 0.52±0.06 for the oversized stems, respectively. Neither was found to correlate significantly with the stem-bone construct stiffness measure (metaphysis: P = 0.259, diaphysis: P = 0.529); however, the diaphyseal fill ratio was significantly different between standard and oversized stems (P < 0.001, Power = 1.0). Increasing the stem size by 2mm had a significant impact on the stiffness of the stem-bone construct (P = 0.003, Power = 0.971; Fig.2). Stem oversizing yielded a construct stiffness of −741±243N/mm; more than double that of the standard stems, which was −334±120N/mm.

The fill ratios reported in the present investigation match well with those of a finite element assessment of oversizing short humeral stems3. This work complements that investigation's conclusion, that small reductions in diaphyseal fill ratio may reduce the likelihood of stress shielding, by also demonstrating that oversizing stems by 2mm dramatically increases the stiffness of the resulting implant-bone construct, as stiffer implants have been associated with decreased bone stimulus4 and premature loosening2. The present findings suggest that even a small, 2mm, variation in the thickness of short stem humeral components can have a marked influence on the resulting stiffness of the implant-bone construct. This highlights the need for more objective intraoperative methods for selecting stem size to provide guidelines for appropriate diametral sizing.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 77 - 77
1 Dec 2022
Spangenberg G Langohr GD Faber KJ Reeves J
Full Access

Total shoulder arthroplasty implants have evolved to include more anatomically shaped components that replicate the native state. The geometry of the humeral head is non-spherical, with the sagittal diameter of the base of the head being up to 6% (or 2.1-3.9 mm) larger than the frontal diameter. Despite this, many TSA humeral head implants are spherical, meaning that the diameter must be oversized to achieve complete coverage, resulting in articular overhang, or portions of the resection plane will remain uncovered. It is suspected that implant-bone load transfer between the backside of the humeral head and the cortex on the resection plane may yield better load-transfer characteristics if resection coverage was properly matched without overhang, thereby mitigating proximal stress shielding.

Eight paired cadaveric humeri were prepared for reconstruction with a short stem total shoulder arthroplasty by an orthopaedic surgeon who selected and prepared the anatomic humeral resection plane using a cutting guide and a reciprocating sagittal saw. The humeral head was resected, and the resulting cortical boundary of the resection plane was digitized using a stylus and an optical tracking system with a submillimeter accuracy (Optotrak,NDI,Waterloo,ON). A plane was fit to the trace and the viewpoint was transformed to be perpendicular to the plane. To simulate optimal sizing of both circular and elliptical humeral heads, both circles and ellipses were fit to the filtered traces using the sum of least squares error method. Two extreme scenarios were also investigated: upsizing until 100% total coverage and downsizing until 0% overhang.

Total resection plane coverage for the fitted ellipses was found to be 98.2±0.6% and fitted circles was 95.9±0.9%Cortical coverage was found to be 79.8 ±8.2% and 60.4±6.9% for ellipses and circles respectively. By switching to an ellipsoid humeral head, a small 2.3±0.3% (P < 0.001) increase in total coverage led to a 19.5±1.3%(P < 0.001) increase in cortical coverage. The overhang for fitted ellipses and circles was 1.7 ±0.7% and 3.8 ±0.8% respectively, defined as a percentage of the total enclosed area that exceeded the bounds of the humerus resections. Using circular heads results in 2.0 ±0.1% (P < 0.001) greater overhang. Upsizing until 100% resection coverage, the ellipse produced 5.4 ±3.5% (P < 0.001) less overhang than the circle. When upsizing the overhang increases less rapidly for the ellipsoid humeral head that the circular one (Figure 1). Full coverage for the head is achieved more rapidly when up-sizing with an ellipsoid head as well. Downsizing until 0% overhang, total coverage and cortical coverage were 7.5 ±2.8% (P < 0.001) and 7.9 ±8.2% (P = 0.01) greater for the ellipse, respectively. Cortical coverage exhibits a crossover point at −2.25% downsizing, where further downsizing led to the circular head providing more cortical coverage.

Reconstruction with ellipsoids can provide greater total resection and cortical coverage than circular humeral heads while avoiding excessive overhang. Elliptical head cortical coverage can be inferior when undersized. These initial findings suggest resection-matched humeral heads may yield benefits worth pursuing in the next generation of TSA implant design.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 139 - 139
1 Jul 2020
Sims L Aibinder W Faber KJ King GJ
Full Access

Essex-Lopresti injuries are often unrecognized acutely with resulting debilitating adverse effects. Persistent axial forearm instability may affect load transmission at both the elbow and wrist, resulting in significant pain. In the setting of both acute and chronic injuries metallic radial head arthroplasty has been advocated, however there is little information regarding their outcome. The purpose of this study was to assess the efficacy of a radial head arthroplasty to address both acute and chronic Essex-Lopresti type injuries.

A retrospective review from 2006 to 2016 identified 11 Essex-Lopresti type injuries at a mean follow-up of 18 months. Five were diagnosed and treated acutely at a mean of 11 days (range, 8 to 19 days) from injury, while 6 were treated in a delayed fashion at a mean of 1.9 years (range, 2.7 months to 6.2 years) from injury with a mean 1.5 (range, 0 to 4) prior procedures. The cohort included 10 males with a mean age was 44.5 years (range, 28 to 71 years). A smooth stem, modular radial head arthroplasty was used in all cases. Outcomes included range of motion and radiographic findings such as ulnar variance, capitellar erosion, implant positioning and implant lucency using a modification of the method described by Gruen. Reoperations, including the need for ulnar shortening osteotomy, were also recorded.

Three patients in each group (55%) reported persistent wrist pain. The mean ulnar variance improved from +5 mm (range, 1.8 to 7 mm) to +3.7 mm (range, 1 to 6.3 mm) at the time of final follow-up or prior to reoperation. Three (50%) patients in the chronic group underwent a staged ulnar shortening osteotomy (USO) to correct residual ulnar positive variance and to manage residual wrist pain. There were no reoperations in the acute group. Following USO, the ulnar variance in those three cases improved further to +3.5, +2.1, and −1.1 mm. No radial head prostheses required removal. Capitellar erosion was noted in five (45%) elbows, and was rated severe in one, moderate in two, and mild in two. Lucency about the radial head prosthesis stem was noted in eight (73%) cases, and rated as severe in 2 (18%), based on Gruen zones.

Treatment of acute and chronic Essex-Lopresti lesions with radial head arthroplasty often results in persistent wrist pain. In the chronic setting, a planned USO was often necessary to restore axial forearm stability after radial head arthroplasty. Essex-Lopresti lesions represent a rare clinical entity that are difficult treat, particularly in the chronic setting. Early recognition and management with a smooth stem modular radial head arthroplasty may provide improved outcomes compared to chronic reconstruction.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 13 - 13
1 Sep 2012
Glennie RA Giles JW Athwal GS Johnson JA Faber KJ
Full Access

Purpose

Glenoid component loosening is a common reason for failed total shoulder arthroplasty. Multiple factors have been suggested as causes for component loosening that may be related to cement technique. The purpose of the study was to compare the load transfer across a polyethylene glenoid bone construct with two different cementing techniques.

Method

Eight cadaveric specimens underwent polyethylene glenoid component implantation. Four had cement around the pegs only (CPEG) and four had cement across the entire back (CBACK) of the implant including around the pegs. Step loading was performed with a pneumatic actuator and a non-conforming humeral head construct capable of applying loads at various angles. Strain gauges were placed at the superior and inferior poles of the glenoid and position trackers were applied to the superior and inferior aspects polyethylene component. Micro CT data were obtained before and after the loading protocol.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 153 - 153
1 Sep 2012
Faber KJ Pike JM Grewal R Athwal GS King GJ
Full Access

Purpose

Limited information is available regarding the functional outcomes of radial head fractures managed with open reduction and internal fixation (ORIF). The purpose of this study was to determine the functional outcomes of radial head fractures treated with ORIF.

Method

Fifty-two patients, with a mean age of 4412 years, who were treated with radial head ORIF were evaluated at a mean of 4.42.4 years. Thirty were isolated radial head fractures (Group A), 13 (Group B) were associated with a complex fracture-dislocation (terrible triad variants), and 5 (Group C) were associated with a proximal ulnar fracture (Monteggia/trans-olecranon variants). Fourty-four were partial articular fractures and 8 were complete articular fractures. Outcomes were assessed with physical and radiographic examination, and validated self-reported questionnaires.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 155 - 155
1 Sep 2012
Elkinson I Giles JW Faber KJ Boons HW Ferreira LM Johnson JA Athwal GS
Full Access

Purpose

The remplissage procedure may be performed as an adjunct to Bankart repair to address an engaging Hill-Sachs defect. Clinically, it has been reported that the remplissage procedure improves joint stability but that it may also restrict shoulder range of motion. The purpose of this biomechanical study was to examine the effects of the remplissage procedure on shoulder motion and stability. We hypothesized that the remplissage procedure would improve stability and prevent engagement but may have a deleterious effect on motion.

Method

Eight cadaveric forequarters were mounted on a custom biomechanical testing apparatus which applied simulated loads independently to the rotator cuff muscles and to the anterior, middle and posterior deltoid. The testing conditions included: intact shoulder, Bankart defect, Bankart repair, 2 Hill-Sachs defects (15%, 30%) with and without remplissage. Joint range of motion and translation were recorded with an optical tracking system. Outcomes measured were internal-external rotation range of motion in adduction and 90 combined abduction, extension range of motion and stability, quantified in terms of joint stiffness and engagement, in abduction.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 18 - 18
1 Sep 2012
Elkinson I Giles JW Faber KJ Boons HW Ferreira LM Johnson JA Athwal GS
Full Access

Purpose

The remplissage technique of insetting the infraspinatus tendon and posterior joint capsule into an engaging Hill-Sachs lesion has gained in popularity. However, a standardized technique for suture anchor and suture placement has not been defined for this novel procedure. The purpose of this biomechanical study was to compare three remplissage techniques by evaluating their effects on joint stiffness and motion.

Method

Cadaveric forequarters (n=7) were mounted on a custom active biomechanical shoulder simulator. Three randomly ordered techniques were conducted: T1- anchors in the valley of the defect, T2- anchors in the rim of the humeral head; T3- anchors in the valley with medial suture placement. The testing conditions included: intact, Bankart, Bankart repair, and 15% & 30% HS lesions with repairs (T1, T2, T3). Outcome measures including internal-external range of motion and stability were recorded. Stability was quantified in terms of glenohumeral joint stiffness against an externally applied anterior force of 70N.