Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 9 - 9
7 Jun 2023
Jenkinson M Meek D MacMillan S Tate R Grant MH Currie S
Full Access

Elevated blood cobalt secondary to metal-on-metal (MoM) hip arthroplasties has been shown to be a risk factor for developing cardiovascular complications including cardiomyopathy. Published case reports document cardiomyopathy in patients with blood cobalt levels as low as 13µg/l (13ppb, 221nmol/l). Clinical studies have found conflicting evidence of cobalt-induced cardiomyopathy in patients with MoM hips. Global longitudinal strain (GLS) is an echocardiography measurement known to be more sensitive than ejection fraction at diagnosing early cardiomyopathies. The extent of cardiovascular injury, as measured by GLS, in patients with elevated blood cobalt levels has not previously been examined.

Sixteen patients with documented blood cobalt ion levels above 13µg/l were identified from a regional arthroplasty database. They were matched with eight patients awaiting hip arthroplasty with no history of cobalt implants. All patients underwent electrocardiogram and echocardiogram assessment for signs of cardiomyopathy including GLS.

Patients with MoM hip arthroplasties had a mean blood cobalt level of 29µg/l (495nmol/l) compared to 0.01µg/l (0.2nmol/l) in the control group. There was no difference or correlation in ejection fraction (EF), left ventricular (LV) end systolic dimension, LV end diastolic dimension, fractional shortening, ventricular wall thickness or E/e’ ratio. However, GLS was significantly reduced in patients with MoM hip arthroplasties compared to those without (−15.2% v −18%, (MoM v control) p= 0.0125). Pearson correlation demonstrated that GLS is significantly correlated with blood cobalt level (r= 0.8742, p=0.0009).

For the first time, this study has demonstrated reduced cardiac function in the presence of normal EF as assessed by GLS in patients with elevated cobalt above 13µg/l. As GLS is a more sensitive measure of systolic function than EF, routine echocardiogram assessment including GLS should be performed in all patients with MoM hip arthroplasties and elevated blood cobalt above 13µg/l. Further work is recommended to assess if these cardiac changes are present in patients with elevated blood cobalt levels below 13µg/l.